Appendix C, Legal Metrology (General) Rule, 2011: APPENDIXC
Legal Metrology (General) Rule, 2011 - Appendix C
Text of appendix C
APPENDIXC
APPENDIXC
TEST FOR ACCURACY
1. Apparatus
(1) Comparator Bath-Either screw type or
bubbler type, as prescribed In IS 6274-1971 Method
of calibratingliquid-in-glass thermometers, filled with
water.
·
(2) Reference Thermometers-Calibrated
reference thermometers (see schedule mark ISTI of
IS 4825 : 1968)
2. Procedure
(1) After preliminary check of the clinical
thermometers, carry out the accuracy test In
accordance with 6.2 of IS 6274-1971 at 37.0 ±
0.5°C, 39.0 ± 0.5°C and 41.0°C ± O.S°C respectively.
(2) The thermometers shall be considered to
have passed this test if the accuracy so determined
lies within limits prescribed In paragraph 6(2)(ii).
APPEND/XO
DITTRMINATION OF THE MEAN DEPRESSION OF
ZERO OF THERMOMETERS
1. It Is not possible to determine the depression
of zero of dlnlcal thermometers (mercury-in-glass,
with the maximum indicating device) covered by this
specification. Therefore, special test thermometers
(paragraph 2) shall be manufactured from the glass
being examined in order to conduct the necessary
measurements.
2. The test thermometers must meet the
following requirements-
(1) Scale Range
At least from • 3.o•c to + 3.D°C.
(2) Scale Interval
0.02°c, 0.05°C or 0.1 °C
(3) The scale spacing must be at least 1.0 mm.
(4) The expansion chamber must be large
enough to allow the thermometers to be
heated to 400°C without damage.
(5) The thermometers must be properly
stablllzed by the manufacturer and must
meet the requirements of the stabilization
test (see paragraph 3).
' . : ·t··«··
[ '1T'1 II -lsfUs 3 ( i) ] 579
3. The proper stabilization of each test
thermometer must be tested in accordance with the
following provisions:-
(1) The thermometer is heated In a test bath
(liquid bath or metal block type oven) from
ambient temperature up to 350°C ± 10°c
and keep at this temperature for at least
five minutes. It is then cooled to 50°C in
the test bath, which decreases in
temperature by 10 to 15°C/h.
(2) When the thermometer has reached a
temperature of 50°C, it is removed from
the test bath and its 0°C correction value
(K,) is determined with the help of a zero
point (0°C) Ice bath, which consists of a
Dewar flask filled with finely crushed ice
covered over with water. The water used
to make the Ice and the water in which ice
is submerged must be pure. Its electrical
conductivity must not exceed 10-3 S.ml
at 20°c. The Ice must be carefully tamped
so that there are no air bubbles In the ice
water mixture. It must be compacted as
much as possible both prior to
measurement and periodically during
measurement.
It is recommended that a water purifier, a
refrigerator with ice trays and an ice crusher
be employed for preparing the ice-water
mixture.
(3) The thermometer is then heated a second
time to 350°C ± 10°c in the test bath and
kept at this temperature for 24 hours. It is
then cooled to 50°C, as before {paragraph
3(1)}.
(4) When the thermometer has reached a
temperature of 50°C, it is removed from ·
the test bath and its 0°C correction value
(K,) Is determined once more.
(5) K, must not differ from K, by more than
0.15°C. Thermometers which do not meet
,
this requirement must not be used to
determine the depression of zero.
4. The mean depression of zero Is determined
in accordance with the following provislons:-
(1) At least three test thermometers must be
used. They must be manufactured from the
glass being tested, must have met the
requirements of the stabilization test
(paragraph 3), and not have been heated
above the ambient temperature once value
K,has been determined [paragraph 3(4)].
(2) Eac::h of these thermometers must be tested
at least three times la accordance with the
provisions of [paragraphs 4(2)(1), 4(2)(11)
and 4(2)(iii) below].
(I) The thermometer is kept in a test
bath at 100°C ± 1.0°C for 30
minutes. It is then removed from
the bath and allowed to cool in air.
While it is cooling to ambient
temperature, its bulb must not come
into contact with other objects.
(II) The 0°C correction of the
thermometer is determined not later
than 15 minutes after the
thermometer has been removed
from the test bath. The correction
value obtained Is designated by the
symbol K,.
(Ill) The thermometer Is then kept for one
week at a temperature between
20°c and 25°C. At the end of the
week the 0°C correction Is
determined. This correction value Is
designated K,. The procedures
described in paragraphs 4(2)(1) and
4(2)(11)are then repeated, and a o•c
correction value, designated I<,,Is
obtained.
(iv) The procedures described In
paragraph 4(2)(iii) are repeated to
obtain a series of n differences K2•
K,,K, - K,;,K1,- K,..,. These are the
values of the thermometer's
depression of zero from the first,
second and n the series of
measurements, respectively.
(v) When n series of measurements
have been made with m test
thermometers, the following
expression is obtained for the mean
depression of zero of these
thermometers.
_.
l ?((k(i)_k(i))+(k(i)_k(i))+(k(i)_k(i) )l
mn ? r l 3 4 ' 2n 2n+l 1
,.,
which must not exceed 0.07°C {paragraph
7(2)(11)}
In accordance with the provisions of paragraphs
4(1) and 4(2), the conditions m 3 and n 3 must be
.
met for m and n, and the standard deviation ofthe ·
mean depression of zero determined in'accordance
with the aforementioned provisions, must not exceed
± 0.01°c.
580. · 'THE-:GAZETTE ,OF i!NDIA: EXTRAORDINARY [PARTll-SEc. 3(i)]
(vi) ,If: a ·more accurate value for ·the
. mean.depresson of zero is required,
,. atleast five.series of measurements
on
..
·at least' five: test thermometers
must be carried·out.
/PARTS
CLINICAi; THERMOMETER-ENCLOSED
SCALE TYPE
Scope : This::Part specifies the requirements
and methods ottesrfor,enclosed:scalectype clinical
thermometershevmqe.rnadrnum indicating device.
1, Terminology
(1) For the purpose of this standard the
definitions given In.rs 2627°1979 Glossary of terms
relating to liquid'in,glass thermometers as revised
.
from time to time,:.ln addition to the following, shall
apply.
(2) Glass:tu/Je9-Sheath tubes, capillary tubes
and bulb tubes.
2. Type
The thermometers shali be of the enclosed
scale rnercury-ln-qlass type.
3. Temperature Scale
The thermometers shall have a translucent
paper or plastic .material strip duly graduated in
degrees Celsius (°C) and shall have a range from 35
to 42°c or 35 to 43°C.
4. Immersion
Thermometers shall be calibrated for total
vertical/horizontal Immersion.
5
.. Requirements
(1) Patternr-There shall be two patterns of
bulb, namely, oral and rectal as follows:
(c) Oral-for thermometers for use in mouth,
and
(d) Rectal-for thermometers for use in rectum.
Note l : Oral thermometers may also be used
in armpit or groin.
Note 2: Rectal thermometers may also be used
in the mouth, armpit or groin after proper disinfection
(2) Materials
(I) Glass tubing-The thermometers shall be
made from glass tubing conforming to IS
4529-1968 Specification for glass tubes for
medical thermometers
.
·
(II) Thermometric liquid-The thermometric
liquid shall be pure, dry mercury.
(Ill) Strip-The strip bearing the scale shall be
of a translucent or plastic material suitable
for the temperature to be read.
(3) Construction
(I) The thermometer sheath shall be In
alignment with the bulb and it shall be
round or oval.
(a) The free end of the sheath shall be
finished smooth, preferably
hemispherical In shape as shown In
Figure 2
JS
(a) The maximum indicting device shall
pass the hardness test prescribed in
Appendix A when an acceleration of
FIGURE-2 CLINICAL THERMOMETER : ENCLOSED SCALE TYPE
(ii) Maximum Indicating device (constriction)- 600 m/sec' is applied at the closed
The thermometers shall have a maximum end of the bulb for a period of at
indicatingdevices (constriction) constructed least 2 minutes.
In the capillary, between the mercury bulb (Ill) Bulb-The bulb of the oral thermometers
and the scale in order to prevent the shall be cylindrical with a smooth rounded
mercury In the stem from returning to the end and shall be in alignment with the
bulb on cooling.
·
sheath. The bulb of the rectal
thermometers shall be pear-shaped.
(a) The diameter of bulb and the Joint
of bulb with sheath shall no where
['WT II-? 3(i)] 581
exceed that of the sheath. The joint
shall be smooth and regular.
(b) The bulb shall be hemispherical at
the tip,
(iv) Strip bearing the scale-It shall be placed
tightly against the capillary tube Inside the
sheath and shall be firmly and securely
fastened at the top of the thermometer In
such a way that it can freely expand In
length. The fixing shall not obscure the
scale.
Note : A suitable method of fixing is by
fusing a glass tube -or rod to the sheath
and to the upper end of the strip bearing
the scale.
(v) CapilliJry tube-The capillary tube shall be
smooth and uniform.
(4) Dimension!l-The dimensional and scale
requirements for clinical thermometers, enclosed
scale shall be as given in Table 2.
TABLE 2
DlMENSIONAL AND SCALE REQUIREMENTSFOR CUNICAL THERMOMETERS,ENCLOSED SCALE
SI. No. Characteristic
( 1) (2)
(i) Scale range, °C
(ii) Smallest scale division, °C
(iii) Over all length, mm, Max
(iv) Scale length, mm
(v) External diameter of sheath, mm
(round or oval), max
(vi) External diameter of capillary, mm
(vii) External diameter of the bulb, mm
(viii) Bulb length, mm, max
(ix) Thickness of strip, mm
(x) Distance above the constriction
upto the 35°C mark, mm, Min
(xi) Distance from the highest
graduation line to top of sheath,
mm, Min
(5) Graduation and Figuring
(i) The thermometer scale shall be sub-divided
in 0.1 °C as shown in Figure SSR.
(ii) The graduation lines shall be of black
pigment, marked permanently on the strip.
They shall be equally spaced and at right
angles to the axis of the thermometer.
(iii) All graduation lines shall be of equal and
uniform thickness of not more than 0.2 mm.
(d) Every tenth graduation line shall be
a long line (about 2 mm) which shall
be figured.
(e) There shall be a medium line (about
1.5 mm) midway between two
consecutive long lines, and
Requirements
(3)
35 to 42 or
35 to 43
0.1
120
45 to 65
12
2± 0.2
4 ± 0.5
21
0.6 ± 0.1
10
8
(f) There shall be four short lines (about
1 mm) equally spaced between
consecutive medium and long lines.
(iv) When the thermometer Is held in a vertical
position and viewed from the front, the left
hand ends of the graduation lines shall lie
on an Imaginary vertical fine (see Figure
SSR).
(v) The figures shall be placed to the right of
the axis in such a way that extension of
the fine to which they refer, would bisect
them. The figures shall be placed parallel
to the axis of the thermometer. ·
(vi) A distinguishing mark (say, an arrow) shall
be marked at 37.0°C mark to indicate the
normal temperature.
-·- _
_L
. -
?---'------
582 THEWAZETTE ·OF ,INDIA: EXTRAORDINARY [PART!I-SEC, 3(i)]
(vii) All graduation lines and'figures shall be
clearly Nisible.
6. Performance requirements
(1) Appearance-',When heatedto 42°C or so
the mercury column.shall look like a-strip of uniform
width with clarity arid brJghtnessthroughout.
Note: This test-takes care of allvlsual defects,
for example.' twist, in
_
the -,glass, aberrations,
_
cloudiness, devitrification, etc.
(2) Agei(lg and Accuracy-'Clinical
thermometers.shall be .adequately aged to ensure
that their accuracy as :measured :by the method
prescribedin Appendil(B.shall'be±0,1 °C for at least
one year after the first:officialverification.
(3) Time.of•Response-'--The time of response
of clinical thermometers, as checked by the method
prescribed in/Appendix C shall not exceed 8.0
seconds.
When the temperature is rising very slowly at
a uniform rate, the Jumping of the meniscus of the
thermometric liquld,does mot exceed one-half of the
smallest scale division.
7. Marking and Packing
(1) Marking-f:ach clinical thermometer shall
be legibly and indelib1ymarked on its stem with the
following:-
.
(a) The letter °C near the top of the scales;
(b) The word Oral or rectal as the case may
be;
(c) The manufacturer's name or his recognized
trade-mark, If any;
(d) A code number to trace the batch of
manufacture; and
(e) A recognized mark of the verifying authority
showing the year of initial verification;
(f) The Standard Mark of the Bureau of Indian
Standards.
·
(2) Packin?llnlcal thermometers shall be
securely packed, Individuallyand collectively,in any
manner acceptable to the purchaser so as to minimize
the risk of damage In handling, transport and storage.
Previous and next