Appendix C, Legal Metrology (General) Rule, 2011: APPENDIXC

Legal Metrology (General) Rule, 2011 - Appendix C

Text of appendix C

APPENDIXC

APPENDIXC TEST FOR ACCURACY 1. Apparatus (1) Comparator Bath-Either screw type or bubbler type, as prescribed In IS 6274-1971 Method of calibratingliquid-in-glass thermometers, filled with water. · (2) Reference Thermometers-Calibrated reference thermometers (see schedule mark ISTI of IS 4825 : 1968) 2. Procedure (1) After preliminary check of the clinical thermometers, carry out the accuracy test In accordance with 6.2 of IS 6274-1971 at 37.0 ± 0.5°C, 39.0 ± 0.5°C and 41.0°C ± O.S°C respectively. (2) The thermometers shall be considered to have passed this test if the accuracy so determined lies within limits prescribed In paragraph 6(2)(ii). APPEND/XO DITTRMINATION OF THE MEAN DEPRESSION OF ZERO OF THERMOMETERS 1. It Is not possible to determine the depression of zero of dlnlcal thermometers (mercury-in-glass, with the maximum indicating device) covered by this specification. Therefore, special test thermometers (paragraph 2) shall be manufactured from the glass being examined in order to conduct the necessary measurements. 2. The test thermometers must meet the following requirements- (1) Scale Range At least from • 3.o•c to + 3.D°C. (2) Scale Interval 0.02°c, 0.05°C or 0.1 °C (3) The scale spacing must be at least 1.0 mm. (4) The expansion chamber must be large enough to allow the thermometers to be heated to 400°C without damage. (5) The thermometers must be properly stablllzed by the manufacturer and must meet the requirements of the stabilization test (see paragraph 3). ' . : ·t··«·· [ '1T'1 II -lsfUs 3 ( i) ] 579 3. The proper stabilization of each test thermometer must be tested in accordance with the following provisions:- (1) The thermometer is heated In a test bath (liquid bath or metal block type oven) from ambient temperature up to 350°C ± 10°c and keep at this temperature for at least five minutes. It is then cooled to 50°C in the test bath, which decreases in temperature by 10 to 15°C/h. (2) When the thermometer has reached a temperature of 50°C, it is removed from the test bath and its 0°C correction value (K,) is determined with the help of a zero point (0°C) Ice bath, which consists of a Dewar flask filled with finely crushed ice covered over with water. The water used to make the Ice and the water in which ice is submerged must be pure. Its electrical conductivity must not exceed 10-3 S.ml at 20°c. The Ice must be carefully tamped so that there are no air bubbles In the ice­ water mixture. It must be compacted as much as possible both prior to measurement and periodically during measurement. It is recommended that a water purifier, a refrigerator with ice trays and an ice crusher be employed for preparing the ice-water mixture. (3) The thermometer is then heated a second time to 350°C ± 10°c in the test bath and kept at this temperature for 24 hours. It is then cooled to 50°C, as before {paragraph 3(1)}. (4) When the thermometer has reached a temperature of 50°C, it is removed from · the test bath and its 0°C correction value (K,) Is determined once more. (5) K, must not differ from K, by more than 0.15°C. Thermometers which do not meet , this requirement must not be used to determine the depression of zero. 4. The mean depression of zero Is determined in accordance with the following provislons:- (1) At least three test thermometers must be used. They must be manufactured from the glass being tested, must have met the requirements of the stabilization test (paragraph 3), and not have been heated above the ambient temperature once value K,has been determined [paragraph 3(4)]. (2) Eac::h of these thermometers must be tested at least three times la accordance with the provisions of [paragraphs 4(2)(1), 4(2)(11) and 4(2)(iii) below]. (I) The thermometer is kept in a test bath at 100°C ± 1.0°C for 30 minutes. It is then removed from the bath and allowed to cool in air. While it is cooling to ambient temperature, its bulb must not come into contact with other objects. (II) The 0°C correction of the thermometer is determined not later than 15 minutes after the thermometer has been removed from the test bath. The correction value obtained Is designated by the symbol K,. (Ill) The thermometer Is then kept for one week at a temperature between 20°c and 25°C. At the end of the week the 0°C correction Is determined. This correction value Is designated K,. The procedures described in paragraphs 4(2)(1) and 4(2)(11)are then repeated, and a o•c correction value, designated I<,,Is obtained. (iv) The procedures described In paragraph 4(2)(iii) are repeated to obtain a series of n differences K2• K,,K, - K,;,K1,- K,..,. These are the values of the thermometer's depression of zero from the first, second and n the series of measurements, respectively. (v) When n series of measurements have been made with m test thermometers, the following expression is obtained for the mean depression of zero of these thermometers. _. l ?((k(i)_k(i))+(k(i)_k(i))+(k(i)_k(i) )l mn ? r l 3 4 ' 2n 2n+l 1 ,., which must not exceed 0.07°C {paragraph 7(2)(11)} In accordance with the provisions of paragraphs 4(1) and 4(2), the conditions m 3 and n 3 must be . met for m and n, and the standard deviation ofthe · mean depression of zero determined in'accordance with the aforementioned provisions, must not exceed ± 0.01°c. 580. · 'THE-:GAZETTE ,OF i!NDIA: EXTRAORDINARY [PARTll-SEc. 3(i)] (vi) ,If: a ·more accurate value for ·the . mean.depresson of zero is required, ,. atleast five.series of measurements on .. ·at least' five: test thermometers must be carried·out. /PARTS CLINICAi; THERMOMETER-ENCLOSED SCALE TYPE Scope : This::Part specifies the requirements and methods ottesrfor,enclosed:scalectype clinical thermometershevmqe.rnadrnum indicating device. 1, Terminology (1) For the purpose of this standard the definitions given In.rs 2627°1979 Glossary of terms relating to liquid'in,glass thermometers as revised . from time to time,:.ln addition to the following, shall apply. (2) Glass:tu/Je9-Sheath tubes, capillary tubes and bulb tubes. 2. Type The thermometers shali be of the enclosed­ scale rnercury-ln-qlass type. 3. Temperature Scale The thermometers shall have a translucent paper or plastic .material strip duly graduated in degrees Celsius (°C) and shall have a range from 35 to 42°c or 35 to 43°C. 4. Immersion Thermometers shall be calibrated for total vertical/horizontal Immersion. 5 .. Requirements (1) Patternr-There shall be two patterns of bulb, namely, oral and rectal as follows: (c) Oral-for thermometers for use in mouth, and (d) Rectal-for thermometers for use in rectum. Note l : Oral thermometers may also be used in armpit or groin. Note 2: Rectal thermometers may also be used in the mouth, armpit or groin after proper disinfection (2) Materials (I) Glass tubing-The thermometers shall be made from glass tubing conforming to IS 4529-1968 Specification for glass tubes for medical thermometers . · (II) Thermometric liquid-The thermometric liquid shall be pure, dry mercury. (Ill) Strip-The strip bearing the scale shall be of a translucent or plastic material suitable for the temperature to be read. (3) Construction (I) The thermometer sheath shall be In alignment with the bulb and it shall be round or oval. (a) The free end of the sheath shall be finished smooth, preferably hemispherical In shape as shown In Figure 2 JS (a) The maximum indicting device shall pass the hardness test prescribed in Appendix A when an acceleration of FIGURE-2 CLINICAL THERMOMETER : ENCLOSED SCALE TYPE (ii) Maximum Indicating device (constriction)- 600 m/sec' is applied at the closed The thermometers shall have a maximum end of the bulb for a period of at indicatingdevices (constriction) constructed least 2 minutes. In the capillary, between the mercury bulb (Ill) Bulb-The bulb of the oral thermometers and the scale in order to prevent the shall be cylindrical with a smooth rounded mercury In the stem from returning to the end and shall be in alignment with the bulb on cooling. · sheath. The bulb of the rectal thermometers shall be pear-shaped. (a) The diameter of bulb and the Joint of bulb with sheath shall no where ['WT II-? 3(i)] 581 exceed that of the sheath. The joint shall be smooth and regular. (b) The bulb shall be hemispherical at the tip, (iv) Strip bearing the scale-It shall be placed tightly against the capillary tube Inside the sheath and shall be firmly and securely fastened at the top of the thermometer In such a way that it can freely expand In length. The fixing shall not obscure the scale. Note : A suitable method of fixing is by fusing a glass tube -or rod to the sheath and to the upper end of the strip bearing the scale. (v) CapilliJry tube-The capillary tube shall be smooth and uniform. (4) Dimension!l-The dimensional and scale requirements for clinical thermometers, enclosed scale shall be as given in Table 2. TABLE 2 DlMENSIONAL AND SCALE REQUIREMENTSFOR CUNICAL THERMOMETERS,ENCLOSED SCALE SI. No. Characteristic ( 1) (2) (i) Scale range, °C (ii) Smallest scale division, °C (iii) Over all length, mm, Max (iv) Scale length, mm (v) External diameter of sheath, mm (round or oval), max (vi) External diameter of capillary, mm (vii) External diameter of the bulb, mm (viii) Bulb length, mm, max (ix) Thickness of strip, mm (x) Distance above the constriction upto the 35°C mark, mm, Min (xi) Distance from the highest graduation line to top of sheath, mm, Min (5) Graduation and Figuring (i) The thermometer scale shall be sub-divided in 0.1 °C as shown in Figure SSR. (ii) The graduation lines shall be of black pigment, marked permanently on the strip. They shall be equally spaced and at right angles to the axis of the thermometer. (iii) All graduation lines shall be of equal and uniform thickness of not more than 0.2 mm. (d) Every tenth graduation line shall be a long line (about 2 mm) which shall be figured. (e) There shall be a medium line (about 1.5 mm) midway between two consecutive long lines, and Requirements (3) 35 to 42 or 35 to 43 0.1 120 45 to 65 12 2± 0.2 4 ± 0.5 21 0.6 ± 0.1 10 8 (f) There shall be four short lines (about 1 mm) equally spaced between consecutive medium and long lines. (iv) When the thermometer Is held in a vertical position and viewed from the front, the left­ hand ends of the graduation lines shall lie on an Imaginary vertical fine (see Figure SSR). (v) The figures shall be placed to the right of the axis in such a way that extension of the fine to which they refer, would bisect them. The figures shall be placed parallel to the axis of the thermometer. · (vi) A distinguishing mark (say, an arrow) shall be marked at 37.0°C mark to indicate the normal temperature. -·- _ _L . - ?---'------ 582 THEWAZETTE ·OF ,INDIA: EXTRAORDINARY [PART!I-SEC, 3(i)] (vii) All graduation lines and'figures shall be clearly Nisible. 6. Performance requirements (1) Appearance-',When heatedto 42°C or so the mercury column.shall look like a-strip of uniform width with clarity arid brJghtnessthroughout. Note: This test-takes care of allvlsual defects, for example.' twist, in _ the -,glass, aberrations, _ cloudiness, devitrification, etc. (2) Agei(lg and Accuracy-'Clinical thermometers.shall be .adequately aged to ensure that their accuracy as :measured :by the method prescribedin Appendil(B.shall'be±0,1 °C for at least one year after the first:officialverification. (3) Time.of•Response-'--The time of response of clinical thermometers, as checked by the method prescribed in/Appendix C shall not exceed 8.0 seconds. When the temperature is rising very slowly at a uniform rate, the Jumping of the meniscus of the thermometric liquld,does mot exceed one-half of the smallest scale division. 7. Marking and Packing (1) Marking-f:ach clinical thermometer shall be legibly and indelib1ymarked on its stem with the following:- . (a) The letter °C near the top of the scales; (b) The word Oral or rectal as the case may be; (c) The manufacturer's name or his recognized trade-mark, If any; (d) A code number to trace the batch of manufacture; and (e) A recognized mark of the verifying authority showing the year of initial verification; (f) The Standard Mark of the Bureau of Indian Standards. · (2) Packin?llnlcal thermometers shall be securely packed, Individuallyand collectively,in any manner acceptable to the purchaser so as to minimize the risk of damage In handling, transport and storage.

Previous and next

All provisions of Legal Metrology (General) Rule, 2011

Questions people ask

Is appendix C of Legal Metrology (General) Rule, 2011 still in force?

No repeal of Legal Metrology (General) Rule, 2011 is recorded.