Appendix A, Legal Metrology (General) Rule, 2011: APPENDIX A
Legal Metrology (General) Rule, 2011 - Appendix A
Text of appendix A
APPENDIX A
APPENDIX A
TEST FOR HARDNESS OF MAXIMUM
INDICATING DEVICE
1. Apparatus
(1) Centrifuge-A centrifuge with radial arms
·
15 cm each. Each arm shall be provided with a pocket
for keeping one or more clinical thermometers. Speed
of the centrifuge shaU be either fixed or adjustable
to 600 rpm.
2. Procedure
(1) Place the thermometers for some time in
water at a temperature anywhere between 42 and
43°c.
(2) Then put them In the pockets of all the
centrifuge, bulb facing outwards, that is away from
the axis of the centrifuge while In rotation.
(3) Let the centrifuge work steadily at its correct
speed for at least 2 minutes. Then stop it.
(4) Take the thermometers out of the pockets
and observe the mercury column.
(5) Thermometers shall be taken as having
satisfied the requirement of this test, if the mercury
rests below or at 35°C mark.
Important-It is necessary that the room
temperature does not exceed 34.5°C during the test.
APPENOIXB
TEST FOR ACCURACY
1. Apparatus
(1) Comparator Bat?Elther screw type or
bubbler type, as prescribed In JS 6274-1971 Method
of calibrating liquid-in-glass thermometers, filled with
water.
(2) Reference Thermometers-Calibrated
reference thermometers [see Schedule Mark 8 and
9 of the IS 4825-1982 specification of liquid-in-glass
solid-stem reference thermometers (first revision)].
2. Procedure ,
(1) After preliminary check of the clinical
thermometers, carry out the accuracy test In
accordance with 6.2 ofIS 6274-1971 at 37.6 ± 0.5°C,
39.0 ± O.S°C and 41 ± 0.5°C respectively.
(2) The thermometers shall be considered to
have passed this test if the accuracy so determined
lies within limits prescribed in paragraph 6(2).
APPENOIXC
TEST OF TIME OF RESPONSE
1. Apparatus
Same as in paragraphs 1(1) and 1(2) of
Appendix B provided that some automatic
arrangement is made to switch off stirring at the
instant the bulb of thermometers under test touch
the water surface and to switch it again immediately
after the thermometers are taken out of the water.
2, Procedure
(1) Set the comparator oath at a temperature
of 41 ± 0.5°C. Note the temperature on the reference
thermometer.
[ 'WT II - "!!!11'6 3 ( i) ] 583
(2) Shake down the clinical thermometers under
test to maximum extent.
Note: Centrifuge mentioned under para 1(1)
of Annexure A may be used for this purpose, if
required.
(3) Plunge the clinical thermometers Into the
bath upto 35°C mark. Take out suddenly after 7.5 to
8.0 seconds wipe and dry. Take their readings.
(4) Thermometers sha II be considered to have
passed this test, If they Indicate temperature within
± 0.1 °C of the comparator bath temperature noted
in paragraph 2( 1) of this Appendix.
Important-It Is necessary that the room
temperature does not exceed 34.5°C during the test.
·
PARTC
CLINICAL ELECTRICAL THERMOMETERS
WITH MAXIMUM DEVICE
1. Scope
(1) This Part specifies the metrological and
technical requirements for clinical electrical
thermometers with a maximum device. Such
instruments are designed to measure human or
animal body temperature. They indicate a maximum
temperature measured after a steady state is reached
or predicted after a time specific to the design of the
instrument. Until the maximum temperature is
indicated, the actual temperature may be indicated
by the thermometer.
(2) The measuring range of clinical
thermometer shall be either of 35°C to 42°C or 35°C
to 43°C. Two accuracy classes, class I and class II,
are covered by this specification.
(3) This specification appliesto battery-powered
instruments which provide a digital Indication of
temperature.
(4) Oinical electrical thermometers designed
to measure skin temperature are not covered by this
specification.
(5) This specification does not exclude the use
of any contract device based on other measurement
principles that meets equivalent performance
standards in determining maximum body
temperature at specified time intervals.
2. Terminology
(1) A clinical electrical thermometer, as covered
by this specification Is a contact thermometer
comprising a temperature probe and an indicating
unit, and that Is designed to measure human or
animal body temperature.
(2) A temperature probe is the component of a
thermometer of which part is applied to a body cavity
or tissue with which It establishes thermal equilibrium.
It comprises a temperature sensor with associated
parts includ;ng coverage, seals, Inner leads, and
connecting plug, where appropriate.
Notes : 1. A body or tissue may be the mouth
(sublingual}, rectum, or armpit.
2. The part of the probe In contact with a body
cavity or tissue is called the 'applied part'.
3. An Indicating unit Is the component of a
thermometer that process the output signal of the
temperature sensor and displays the measured
temperature.
4. A maximum device Is the component of a
thermometer that monitors over a specified time the
temperature measured by a probe in contact with a
body cavity or tissue, at which it indicates the
maximum temperature and maintains the indication
until reset by the user.
5. Predicting clinical electrical thermometer
calculates the maximum temperature of a probe In
contact with a body cavity or tissue, without waiting
for thermal equilibrium to occur, by using heat transfer
data and a mathematical algorithm.
3. Description of the instrument
(1) A complete thermometer consists of a
temperature probe connected to any indicating unit.
The Instrument may be of the following type:..:.
-Indicating unit that is compatible with
the characteristic response of a probe.
4. Metrologlcal requirements
(1) Unit of measurement-measuring range-scale
interval
·
(a) The unit of temperature shall be the degree
Celsius 0C.
(b) The measurement range shall be a
minimum of 35°C to 43°C or 35°C to 42°C.
Greater measuring ranges may be sub·
divided into partial ranges; however, the
range 35°C to 42°C shall be continuous.
(c) The scale interval or digital increment shall
not exceed-
, 0.01 °C for class I thermometers,
• 0 .1 °C for class II thermometers.
(2) Maximum permissible errors
(a) The maximum permissible error under
reference conditions for the temperature
range 32.0°C to 42.0°C for the two accuracy
classes covered shall be as follows:-
·--- ---- ---------
_J
_
584 .THE tG'AZETTE ,OF 1JND1A: EXTRAORDINARY [PARTII-SEC. J(i)]
Maximum·permissible errors'(range:32;0°c
to 42:ooq
(b) outsice.the temperature range 32°C to
42°C,.the maximum errors.shall be at twice
the vatues.specltled .m ;paragraph 4(2)(a).
(3) Reference conditions
The reference. conditions for the requirements
of paragraph 4?),shall be:
ambiennemperature of 27°C ± s•c
relative·humldlty to 50% ± 20%
the instrument operating within the
specified range of battery voltage (specified
power:sup,plyconditions).
(4) Time response
The :thermometer shall be submitted by
,
the manufacturer to a testing laboratory
to determine Its time response. The test
shall be based on any analysis of a
significant sample of human subjects.
The difference between the displayed
calculated temperature and the
corresponding measured temperature at
thermal equilibrium of calculating
(predicting)thermometer shall not exceed
0.2°c.
5. Technical requirements
(1) tempersture probe
(i) For an Interchangeable probe of the
resistance type, the manufacturer shall
specify the maximum power that may be
supplied to the probe by an Indicating unit;
this power shall not cause energy
dissipation (FR) giving rise to any increase
in temperature by more than 0.02°C when
immersed In a reference water bath
maintained at a temperature range of
36.9°C to 37.1°C.
Note :(1) For a description of the reference
water bath, see Annexure A.
(2) A test of this requirement is only
applicable to inter-changeable probes
submitted for pattern approval without a
specific Indicating unit. When a probe is
submitted with an associated Indicating
unit, the requirement in 5(2)(7) applies.
Accuracy
dass
Class I
Class II
Complete -Indicating ;Temperature
thermometer unit probe
± o.1s•c ±.o.os•c ± O.l'C
± 0.2!C + 0.1 •c :.± 0.1°C
(ii) The thermal stability of the probe, after
exposure for 100 hours at 80°C ± 2°C or
300 hours at 55°C ± 2°c, shall be such
that the requirement for maximum
permissible errors specified in paragraph
4(2) Is met.
(Ill) The electrical insulation of the probe shall
be sufficient to prevent a change In
Indicated temperatures greater than
± 0.02°C when the probe is immersed In
an electrically conducting liquid having an
Insulation resistance of 80 MW at 20°C. This
Insulation includes that between the Inner
lead wires, that between the wires and the
surface of the probe, and that encasing and
protecting connections and transitions.
(Iv) The locations of the sensors in the probe
shall be such that, when the probe Is
Immersed to depths greater than SO mm
from its tip in a reference water bath at a
temperature does not vary by more than
o.os•c from that Indicated at a depth of
SO mm.
(v) The probe shall be strong enough to
withstand mechanical stresses expected
under normal conditions of the use.
(vi} If the probe is inter-changeable, it shall be
fitted with either a plug-in or quick dis·
connectable electrical connector. The
contact resistance of the connector or the
Insulation resistance between the circuits
of the' connector or to ground shall not
cause a variation In the indicated
temperature greater than 0.02°C.
Note : The connectors may not be required to be
water resistant.
(vii) The probe shall meet the requirements for
maximum permissible errors specified In
paragraph 4(2) when the applied part has
been subjected to the cleaning In and
disinfecting procedure specified by the
manufacturer.
Notes: (1) For small compact thermometers this
applies to the complete instrument.
(2) The materials of the probe that come
into contact with the body should be
selected for compatibllity with body tissue.
(viii) The output signal of the probe shall not
vary by more than ± O.OS°C when the
temperature of the cable connecting It to
an Indicating unit varies by 20°C.
['WT 11-¥16 3(i)]
(2) Indicating unit
(i) When connected to a resistance-type
temperature probe, the indicating unit shall
provide an energizing potential sufficiently
low so that energy dissipation ( 1' R) In the
probe shall not cause an increase in
indicated temperature of over 0.01 •c when
the probe Is immersed in a reference water
bath at a temperature wi.thin the specified
measuring range.
(ii) The Indicating unit shall not Indicate a
temperature when connected to a battery
charger.
(iii) The digital display of temperature shall be
at least 4 mm In height or it shall be optically
magnified so as to appear at least 4 mm in
height.
(iv) The indicating unit shall provide a clear
indicating or warning signal when the
measured temperature is outside the
specified measuring range.
(v) The Indicating unit shall include a self
checking device that meets the
requirements of the paragraph 4(2). This
device, which may be manual or automatic,
shall input a predetermined electrical signal.
Failure shall be clearly indicated.
Note : This device checks only the operation of
the indicating unit and does not ensure that
a temperature measurement Is correct. It
provides a means of detecting a faulty
operation caused by a defective component
or other disturbance.
(vi) In the case of a predicting thermometer,
the indicating unit shall provide a means
of displaying the measured temperature
after reaching the thermal equilibrium.
(3) Complete thermometer
Note : The reference thermometer Is that
indicated (either before the test, or before and after
the test, as appropriate) by the thermometer probe
immersed in the reference water bath according to
the paragraph 1 of Annexure A, the temperature
being held constant within the working range of the
thermometer.
·
(I) The thermometer shall provide a clear
indication or warning signals when the
battery voltage Is outside the specified limits
and it shall meet the requirement specified
in paragraph 4(2) when the voltage Is within
these limits.
u t o r;r/!l-1 lf 7
585
(ii) The indicated temperature shall not vary
by more than ± 0 .1 °C from the reference
temperature when the temperature of the
thermometer casing varies from 10°C to
40°C.
(Iii) The indicated temperature shall not vary
more than ± 0.1 °C from the reference
temperature after a thermal shock resulting
from an abrupt change In temperature from
-s-c tc-so-c,
(Iv) The indicated temperature shall not vary
by more than ± O.1 •c from the reference
temperature after storage for 24 hours at·
20°c ± 2°c and at 60°C ± z-c,
(v) The indicated temperature shall not vary
by more than ± O.1 •c from the reference
temperature after storage at a relative
· humidity of 91 % to 95% at a temperature
constant within± 2°C in the range 20°C to
az-c.
(vi) The indicated temperature shall not vary
by more than ± 0.3°C from the reference
.
temperature during exposure to
electromagnetic field having a frequency
between 150 kHz and 500 MHz and a field
strength of lOV/m.
(vllf The indicated temperature shall not vary
·
by more than ± 0.1 °C from the reference
temperature after fall on to a hard surface
from a height of tm from three different
orientations.
(viii) Small and compact complete thermometers
shall be water resistant.
6. Practical Instructions
(1) Manufacturers shall provide an operating
manual, or Instructions, including the following
information:-
• description of appropriate uses and means
of applications,
• Identification of the specified temperature
measuring range of the complete
thermometer taking into account, If
applicable, the specified measuring ranges
of both the interchangeable probes and the
indicating unit,
Instructions and precautions for deaning
and disinfecting the complete thermometer
or the inter-changeable probes,
Identification of components and suitable
Inter-changeable parts such as probes and
586 THE GAZETTE OF INDIA : EXTRAORDINARY [PARTII-SEC. 3(i)]
battertes.vtncludlnq nominal voltage,; if
applicable,
• minimum time. for
..
achlevlnq .thermel
equilibrium,
• description of transition fromthe·predicted
temperature,measuring -rnede Into •the
actual .temperature,measuring mode,
instruction for the self,checklng device,
• information on the.correct ·environmental
conditions .of use, .storaqe, and transport
of the:thermometer.
(2) Specit1c·1nformation should be provided by
the manufacturer, on request, regarding possible sub
standard performance if used under the following
conditions:-
• outside 'the ,presc.ribed environmental
temperature .and humidity range,
• after an accidental mechanical shock.
7. Metrologlcal controls
(1) Pattem .evsiuetion
(i) Manufacturers shall provide the following
information 1-
locatlon of sensor from tip of probe,
description and principles of
measurements of complete
thermometer,
description of electrical principlesand
of any necessary equipments
provided,
description of test for self-checking
device,
specified working range for battery,
nominal and specified temperature
measuring ranges,
nominal values of calibrations data
for type of temperature probe, as
applicable,
precautions for cleaning and
disinfecting complete thermometer
or temperature probes, as
appropriate, including test results as
specified in paragraph 3 of Annexure
B,
Indicating on instrument if a
displayed value is calculated,
test results,
results of clinical test of time
responses [paragraph 4(4) and
Annex OJ,
operating manual and/or instructions
(see clause 6).
(ii) Thermometers shall be subjected to the
following tests:-
Note : Requirements for the reference water bath
and the test for maximum permissible
errors are provided In Annexure A. The
performance requirements for the
Instrument and Its major components are
provided In paragraphs 4 and 5. Where
· appropriate, an additional description of
required tests is provided In Annexure B.
Probe
maximum permissible errors (4(2)
and paragraph 2(2)(1) of Annexure
A]
long-term thermal stability [5(1)(11)]
electrical insulation and water
resistance [5(1)(111)and paragraph
2 of Annexure BJ.
location of sensor [S(l)(iv)]
mechanical strength [5{1)(v)]
electrical contact resistance of
connector [5{1 )(vi)]
cleaning and disinfecting [5( 1 )(vii)
and paragraph 3 of Annexure BJ
stability with changes in temperature
of cables [S(l)(viii]
Indicating unit
maximum permissible errors (4(2)
and Annexure A]
power provided to probe [5(2)(1)and
paragraph 1 of Annexure BJ
indication when connected to battery
charger [5(2)(ii)]
indication If the thermometer· Is
outside the specified measuring
range [5(2)(1v)]
self-checking device [5(2)(v)]
display of predicting thermometer
[5(2)(vi)]
Complete thermometer
maximum permissible errors (4(2)
and Annexure A]
low battery Indication [5(3)(1) and
paragraph 4 of Annexure BJ
ambient temperature [5{3)(ii) and
paragraph 6 of Annexure BJ
['IJT'Tn-= 3(ilJ
··;'""" •. ,,
.
..:.., .. ··" ,. ,··?--?·--
587
thermal shock [5(3)(iil) and.
paragraph 6 of Annexure BJ
storage temperatures [5(3)(1v)J
humidity [5(3)(v) and paragraph 7
of Annexure BJ
electromagnetic radiation inter
ference [5(3)(vi) and paragraph 8 of
Annexure BJ
mechanical shock. [5(3)(vli) and
.
paragraph 9 of Annexure BJ
water resistance of small compact
thermometers [5(3)(vlll) and
Annexure CJ
Instructions and precautions for
cleaning and disinfecting the
complete thermometer or the inter
changeable probes,
Identification of components and
suitable inter-changeable parts such
as probes and batteries, Including
nominal voltage, if applicable,
minimum time for achieving thermal
equilibrium by an indicating energy
dissipation requirements [S(l)(a)
and paragraph 1 of Annexure BJ
(iii) A report of the results of tests required in
paragraphs 7(1)(ii) and 7(1)(111)shall be
prepared.
(2) Marks and labels
(i) Manufacturers shall provide a space for
marks and labels.
·
(ii) Manufacturers shall affix on the thermo
meter or Indicating unit, if separate, the
following marks or labels:-
name and address of manufacturer
or supplier, and/or trademark,
model or type designation, and serial
or lot number,
temperature values or Indications
given by the self-check.Ingdevice,
indication of the orientation or
position in sue, where appropriate,
indication if a displayed value is
calculated.
(iii) Inter-changeable temperature probes shall
bear the following marks or labels::-
name and address of manufacture
and/or trademark,
type designation,
serial or lot number, or relevant
manufacturing production data.
(iv) A single-use temperature probe shall be
sealed In a package on which the
information specified In paragraph 7(2)(iii)
and the measuring range shall be indicated.
In addition, sufficient space on the package
shall be provided for the application of
official approval marks. It shall be clear If
the package has been opened and the
Instructions shall stipulate that the user only
opens the package Immediately before use.
(v) The testing laboratory shall permit the
application in a conspicuous place, of the
following:-
pattern approval mark or label, on
each complete thermometer or
indicating unit and associated
temperature probe(s).
indication of. the specified
temperature measuring range If the
.
total range of the thermometer Is
greater.
(3) Verfflcatlon
(I) The laboratory shall examine the
Information provided by manufacturers as
specified in paragraph 6.
(II) The laboratory shall examine the
Instrument's pattern approval certificate
mark(s) or label(s).
(Ill) The laboratory shall cany out any of the
tests indicated in paragraph 7(1)(ii) that
may be critical for the designated
application of the instrument.
Note: The tests indicated In paragraph 2 of
Annexure A may be sufficient for
verification.
(Iv) The laboratory shall provide a verified
Instrument with a mark or label.
(v) The water resistance of small and compact
complete thermometers shall be examined
by means of the procedure described In
paragraph 2 of Annexure C.
(vi) The laboratory shall indicate the period of