Appendix A, Legal Metrology (General) Rule, 2011: APPENDIX A
Legal Metrology (General) Rule, 2011 - Appendix A
Text of appendix A
APPENDIX A
APPENDIX A
TEST FOR HARDNESS-OF MAXIMUM INDICATING
DEVICE
1. Apparatus
(1) Centrifuge-A centrifuge with radial arms
15 cm each. Each arm shall be provided with a pocket
for keeping one or more clinical thermometers. Speed
of the centrifuge shall be either fixed or adjustable
to 600ppm.
2. Procedure
(1) Place the thermometers for sometime in
water at a temperature anywhere between 42°C -
43°C. Then put them In the pockets of all the
centrifuge, bulb facing outwards, that Is away from
the axis of the centrifuge while in rotation.
Let the centrifuge work steadily at Its correct
speed for at least 2 minutes. Then stop It.
Take the thermometers out of the pockets and
observe the mercury column.
Thermometers shall be taken as having satisfied
the requirement of this test, if the mercury rests
below or at 35°C mark.
Important-It is necessary that the room
temperature does not exceed 34.5°C during the test.
APPEND/XS
TEST FOR PERMANENCY OF MARKING
1. Procedure
(1) Place the thermometers in a S per cent (m/
v) solution of phenol in water maintained at 37 ±
0.5°C for a period ot 20 minutes.
·
·
(2) Wipe the thermometers dry with a piece of
·soft cloth and examine.
(3) The thermometers shall be considered to
'have passed the test if, after this treatment, the
marking does not peel off anywhere.
·