Annexure A, Legal Metrology (General) Rule, 2011: ANNEXUREA
Legal Metrology (General) Rule, 2011 - Annexure A
Text of annexure A
ANNEXUREA
ANNEXUREA
TEST PROCEDURES
(Mandatory)
1. General
For dig Ital i?dicationsan uncertainty of 0.1 kPa
(1 mmHg) shall allowed in any displayed value,
because the display system cannot indicate a change
of less than one unit.
2. Method oftest for the maximum permissible
errors of .the cuff pressure Indication
Requirements in paragraph 5(1) shall apply.
(1) Apparatus
(a) rigid metal vessel with a capacity of 500 ml
:I: 5%;
(b) calibrated reference manometer with an
uncertainty less than 0.1 kPa (0.8 mmHg);
(c) pressure generator, e.g. ball pump (hand
pump) with a deflation valve;
(d) T-plece connectors and hoses.
(2) Procedure:
Replace the cuff with the vessel. Connect the
calibrated reference manometer by means of a
T-piece connector and hoses to the pneumatic circuit
(see Figure 1). After disabling the electro-mechanical
['WI 11-Wl& J(i)]
mlll1<f]•
Q
JOO"''
© 0
605
1 - Reference manometer 2 - Device to be tested
·
3 - Metal vessel 4 - Pressure generator
Figure 1: Measurement system for determining the limits of error of the cuff pressure indication
pump (if fitted), connect the additional pressure
generator into the pressure system by means of
another T-piece connector. Carry out the test in
pressure steps of not more than 7 kPa (50 mmHg)
between O kPa (0 mmHg) and the maximum pressure
of the scale range.
(3) Expression of results:
Express the results as the differences between
the indicated pressure of the manometer of the device
to be tested and the corresponding readings of the
reference manometer.
1 - Reference manometer
3 - Device to be tested
2 • Climatic chamber
4 • Metal vessel 5 • Pressure generator
Figure 2: Measurement system for determining the Influence of temperature
f,,f?{f c;.z:/t1-J52
606 THE GAZETTE· OF INDIA-: 'EXTRAORDINARY [PARTII-SEC. 3(i)]
3. Method of test for the influence of
temperature on cuff pressure indication
(1) Apparatus
apparatus as specified in paragraph 2(1)
of Annexure A;
climatic chamber.
(2) Procedure:
Replace the. cuff with the vessel. Connect the
calibrated reference manometer ·by means of a T
piece connector to the pneumatic system,(see Figure
2). After disablinq.the electro-mechanlcal pump (if
fitted), connect the additional pressure generator into
the pneumatic system by:means-oLanother T-piece
connector.
For each of the following combinations of
temperature and· humidity, condition the device for
at least 3 h in ·the ,climatic chamber to allow the
device to reach steady conditions:
(i) 10°C ·ambient temperature, 85% relative
humidity (non-condensing);
(ii) 20°C ,ambient temperature, 85% relative
humidity (non-condensing);
(iii) 40°C ambient temperature, 85% relative
humidity (non-condensing).
carry out the test of the cuff pressure Indication
as described in paragraph 2(2) of Annexure A for
each of the combinations of temperature and
humidity mentioned above.
(3) Expression of results:
Express the results as the differences between
the indicated pressure of the manometer of the device
to be tested and the corresponding Indications of
the reference manometer at the relevant temperature
value.
4. Test methods for the effect of voltage
variations of the power source on the cuff
pressure indication
(1) Internal electrical power source
(i) Apparatus
adjustable direct current voltage
supply;
voltmeter with an uncertainty of less
than 0.5% of the measured value;
callbrated reference manometer with
an uncertainty of less than 0.1 kPa
(0.8 mmHg).
(ii) Procedure:
Replace the internal electrical power source
of the blood pressure measuring system
with a DC voltage supply having an
Impedance which Is equivalent to the
impedance of the Internal electrical power
source specified by the manufacturer.
Measure the variation In applied DC voltage
supply with a voltmeter. Test the blood
pressure measuring system by altering the
DC voltage supply in steps of 0.1 V and
determine the lowest voltage limit at which
the cuff pressure reading is still displayed.
Carry out the test with the maximum
permissible impedance of the internal
electrical power source.
Carry out the test according to the
procedure specified In paragraph 2 of
Annexure A at the lowest voltage limit
Increased by 0.1 V and also at the nominal
voltage.
(Ill) Expression of results:
Express the results as the difference
between the cuff pressure indication of the
blood pressure measuring system to be
tested and that of the reference manometer
at the lowest voltage limit increased by
0.1 V and at nominal voltage.
(2) Extern(!!electrical power source-alternating
current
(I} Apparatus
(a) adjustable alternating current
voltage supply;
(b) voltmeter with an uncertainty of less
than 0.5 % of the measured value;
(c) calibrated reference manometer with
an uncertainty of less than 0.1 kPa
(0.8 mmHg).
(II) Procedure :
Connect the blood pressure measuring
system to the adjustable alternating current
voltage supply. Measure the variation in AC
voltage supply with the voltmeter.
Carry out the test according to the
procedure specified in paragraph 2 of this
Annexure.
['lT'T II-? 3(i)] 607
the maximum rated voltage,
declared by the manufacturer,
increased by 10%;
the mean value of the maximum and
minimum rated voltage, declared by
the manufacturer;
the minimum rated voltage, declared
by the manufacturer, decreased by
10%,
(iii) Expression of results:
Express the results as the difference
between the cuff pressure indication of the
blood pressure measuring system to be
tested and that of the reference
manometer.
(3) External electrical power source-direct
current
(i) Apparatus:
Use the apparatus listed In paragraph
4(1)(1) of this Annexure.
(ii) Procedure:
Connect the blood pressure measuring
system to the DC voltage supply. Control
the DC voltage supply by reference to a
voltmeter.
Carry out the test according to the
procedure specified in paragraph 2 of
Annexure A at-
(a) the maximum rated voltage,
declared by the manufacturer,
Increased by 10%;
(b) the mean value of the maximum and
minimum rated voltage, declared by
the manufacturer;
(c) the minimum rated voltage, declared
by the manufacturer, decreased by
10%.
(iii) Expression of results :
Express the results as the difference
between the cuff pressure indication of the
'blood pressure measuring system to be
tested and that of the reference
manometer.
(4) Voltage variations of the external electrical
power source· alternating current
(I) Apparatus:
Use the apparatus· listed in paragraph
4(2)(i) of this Annexure.
(ii) Procedure:
Connect the blood pressure measuring
system to the AC voltage supply. Measure
the variation in the AC voltage supply with
the voltmeter. Test the blood pressure
measuring system by altering the DC
voltage supply in steps of SV and determine
the lowest voltage limit at which the cuff
pressure reading Is displayed.
Carry out the test according to the
procedure specified In paragraph 2 of this
Annexure at the lowest voltage limit
increased by 5V and also at the rated
voltage.
(iii) Expression of results:
Express the results as the difference
between the cuff pressure Indication of the
blood pressure measuring system to be
tested and that of the reference manometer
at rated voltage and the lowest voltage limit
Increased by 5V.
(5) Voltage variations of the external electrical
power source • direct current
(I) Apparatus:
Use the apparatus listed in paragraph
4(1)(i) of this Annexure.
(ii) Procedure:
Connect the blood pressure measuring
system to the DC voltage supply. Measure
the variation in the DC voltage supply with
the voltmeter.
Test the blood pressure measuring system
by altering the DC voltage supply in steps
of O.1 V and determine the lowest voltage
limit at which the cuff pressure reading Is
displayed.
Carry out the test according to the
procedure specified in paragraph 2 of
Amiexure A at the lowest voltage limit
Increased by 0.1 V and also at the rated
voltage.
(Iii) /:xpression of results:
Express the results as the difference
between the cuff pressure indication of the
blood pressure measuring system to be
608 THE 'GAZETTE OF INDIA: 'EXTRAORDINARY [PARTII-SEC. 3(i)]
tested.endthetotthe reference manometer
at rated voltcige and at. the lowest voltage
limit increased.by. O,LV.
5. Test methods· for the ·effect .of .voltage
variations of the power source on.the.resuttor
the blood pressure measurement
(1) Internal electrical power source
(i) Apparatus
adjustable direct .cuirent voltage
supply;
'
voltmeter -wlth 'an ·uncertainty less
tthan .0.5% .otthe measured value;
. patient simulator [see paragraph
'2(1?)] ,for the auscultatory and/or
,osclllometric method, having
raddltfonal deviations originatingfrom
the 'Simulator of not more than 0.27
.kPa,(2 mmHg) for the mean value of
.the ,measurements and generating
signals for blood pressure values of
.approximately
-.systolic: 16 kPa (120 mmHg);
- diastolic: 11 kPa (80 mmHg);
- pulse rate: 70 min· 1 ·80mln· 1.
(ii) Procedure:
Replace the Internal electrical power source
of the blood pressure measuring system
by a DC voltage supply having an
impedance which is equivalent to the
impedance of the internal electrical power
source specified by the manufacturer.
Devices Intended to be used with consumer
batteries shall be tested with an Impedance
of less than 1112.
·
Control the DC voltage supply by reference
to the voltmeter.
Connect the blood pressure measuring
system to the patient simulator. Carry out
the test at the maximum permissible
impedance of the internal electrical power
source.
Carry out 20 simulated blood pressure
measurements at the lowest voltage limit
as determined in paragraph 4(1)(ii) of
Annexure A increased by 0.1 V and at
no'minal voltage.
(ill) Expression of results:
Determine the mean value (systolic and
diastolic values separately) of the 20
consecutive readings taken at each voltage
level.
(2) External electrical power source •
altemating current
(I) Apparatus:
adjustable alternating current
voltage supply;
voltmeter with an uncertainty less
than 0.5% of the measured value;
patient simulator as described In
paragraph 5(1 )(1)of Annexure A.
(II) Procedure:
Connect the blood pressure measuring
system to the AC voltage supply.
Control the AC voltage supply by reference
to the voltmeter. Connect the blood
pressure measuring to the simulator.
Carry out 20 simulated blood pressure
measurements each at:
the maximum rated voltage,
declared by the manufacturer,
increased by 10%;
'the mean value of the maximum and
minimum rated voltage, declared by
the, manufacturer;
the minimum rated voltage, declared
by the manufacturer, decreased by
10%.
( iii) Expression of results :
Determine the mean value (systolic and
diastolic values separately) of the 20
consecutive readings taken at each voltage
level.
(J) External electrical power source - direct
current
(I) Apparatus
(a) adjustable direct current voltage
supply;
(b) voltmeter with an uncertainty less
than 0.5 % of the measured value;
--.s
·------ --·-'
['1T'TII = 3(i)]
•.. -->-----
609
(c) patient simulator as described in
paragraph S(l)(i) of Annexure A.
(ii) Procedure:
Connect the blood pressure measuring
system to the DC voltage supply.
Control the DC voltage supply by reference
to the voltmeter. Connect the blood
pressure measuring system to the
simulator.
Carry out 20 simulated blood pressure
measurements each at:
(a) the maximum rated voltage,
declared by the manufacturer,
increased by 10%;
(b) the mean value of the maximum and
minimum rated voltage, declared by
the manufacturer;
(c) the minimum rated voltage declared
.
by the manufacturer, decreased by
10%.
(ill) Expression of results:
Detenmine the mean value (systolic and
diastolic values separately) of the 20
consecutive readings taken at each voltage
level.
6, Method of test for air leakage of the
pneumatic system
( 1) Apparatus:
6(1)(i) rigid metal cylinder of an
appropriate size;
6(1)(ii) pressure generator, e.g. ball pump
(hand pump) with deflation valve;
6(1)(iii) stopwatch.
(2) Procedure
If because of technical reasons, the test as
described in this sub-clause cannot be performed,
use an alternative test procedure specified by the
manufacturer.
Carry out the test at constant temperature in
the range 15°C to 25°C.
Before beginning the test, allow the blood
pressure measuring system to reach working
temperature.
Wrap the cuff around the cylinder (see 6.2) such
that, for devices measuring et the upper arm and
the thigh, the circumference of the applied cuff does
not exceed that of the cylinder by more than 7%.
Note 1 : Electro-mechanical pumps which are
a part of the system may be used for the test. Valves
which are permanently opened may be disconnected
for the test.
Note 2 : For this test no calibrated reference
manometer is required because the cuff pressure
.
display of the unit under test can be used when the
error of the cuff pressure Indication is taken Into
account. The advantage of this test Is that the unit
under test Is In its original configuration. Additional
connections can increase the leakage.
Carry out the test over the whole measuring
range at least five equally spaced pressure steps [e.g.
7 kPa(SOmmHg), 13 kPa (100 mmHg), 20 kPa (150
mmHg), 27 kPa (200 mmHg) and 34 kPa (250
mmHg)J. Because the thermodynamic equilibrium is
influenced by decreasing or Increasing the pressure
when changing to the next pressure step, wait at
least 60 s before reading the values. Test the air
leakage over a period of 5 minutes and determine
the measured value from this.
_
(3) Expression of results:
Express the air leakage as the rate of pressure
loss per minute.
7. Method of test for the pressure reduction
rate
(1) Apparatus
(I) T-piece connectors;
(ii) calibrated reference manometer with
signal output port and an uncertainty
less than 0.1 kPa (0.8 mmHg);
(iii) artificial or human limbs;
(iv) recording unit.
(2) Procedure
Measure the pressure reduction rate either on
human subjects or artificial limbs.
Note 1 :The intention Is to use artificial limbs,
but as these are still under consideration,
measurements perfonmed with human volunteers are
acceptable.
Note 2: Two limb sizes should be used, being
equal to the upper and lower limits of limb
Lf 7 t.t c;.r/11- I? 3
..
·- - --- ---------
---- ---?-.
610 THE GAZETTE OF INDIA : EXTRAORDINARY [PART11--SEC. J(i)]
circumferences with which-a particular size of cuff is
recommended for use.
Note 3 : It is intended that the characteristics
of the artificial limbs reflect some elastic
characteristics of human limbs.
Because the cuff deflation rate may be
influenced by the way that a cuff is applied, apply
and remove the cuff for each of at least ten repeated
measurements on at least two different limb sizes.
The deflation may be reset.
Connect the calibrated reference manometer
to the cuff by means of a T-piece. Connect the output
part of the calibrated reference manometer to the ·
recording unit.
(3) Expression of results
Determinethe rate of pressure reduction (e.g.
by graphical evaluation and drawing tangents) at the
pressure values 8 kPa (60 mmHg), 16 kPa (120
mmHg) and 24 .kPa (180 mmHg). Calculate the
pressure reduction rate as the mean value calculated
separately for the pressure values 8 kPa (60 mmHg),
16 kPa (120 mmHg) and 24 kPa (180 mmHg) and
for the various limb circumferences.
If the pressure reduction rates are dependent
on the pulse, record the pulse rate. In this case,
express the result as pressure reduction rate per
pulse.
8. Method or test for the rapid exhaust valve
(1) Apparatus
two rigid vessels with capacities of
100 ml ± 5% and 500 ml ± 5%,
respectively;
calibrated reference manometer with.
an uncertainty less than 0.1 kPa (0.8
mmHg);
T-plece connector;
stopwatch.
(2) Procedure ·
Carry out the test with the 500 ml vessel in
place of the cuff. For blood pressure
measuring systems having the capability of
measuring In a neonatal/infant mode and
for devices measuring at the wrist, carry
out the test with the 100 ml vessel In place
of the cuff.
Connect the calibrated reference
manometer by means of a T-piece to the
pneumatic system.
Inflate at least to the maximum pressure
given in paragraph 6(5}(iii),wait 60 seconds
and activate the rapid exhaust valve.
Measure the time between the pressure
values specified in paragraph 6(5)(iii) using
the stopwatch.
(3) Expression of results
Express the results as the measured
exhaust times.
9. Test method for the zero setting
(1} Apparatus
(I} rigid vessel with a capacity of 500
ml ± 5%;
(II) calibrated reference manometer with
an uncertainty less than 0.1 kPa (0.8
mmHg};
(Ill} electro-mechanical pressure/suction
.
pump;
(Iv) pressure generator, e.g. ball pump
(hand pump) with deflation valve;
(v) T-piece connectors;
(vi) hoses.
(2)
·
Procedure and evaluation
If, because of technical reasons, the test
as described in this sub·clause cannot be
performed, use an alternative test
procedure specified by the manufacturer.
To test the function of the zero setting,
apply a pressure of+ 0.8 kPa ( + 6 mmHg)
and subsequently - 0.8 kPa (- 6 mmHg) to
the pneumatic system and initiate a zero
· setting of the device. Ensure that all
displayed pressure values have a systematic
error of - 0.8 kPa (- 6 mmHg) and + 0.8
kPa (+ 6 mmHg), respectively.
Before beginning the test, allow the blood
pressure measuring system to reach
working temperature .
Set up the blood pressure measuring
system to be tested as follows:-
['ll'T II-? 3(i)] 6ll
replace the cuff with the 500 ml
vessel;
insert the calibrated reference
manometer into the pneumatic
system by means of a T-piece
connector;
insert the pressure/suction pump
into the pneumatic system by means
of a T-plece connector;
Insert the pressure generator into the
pneumatic system by means of a T
plece connector.
Note: If convenient, one adjustable pump may
be used in place of the pressure/suction pump and
pressure generator to generate the pressures.
Proceed in the following way:-
(a) Initiate a zero setting as described by the
manufacturer. Set the blood pressure
measuring system to the service mode, if
available. Raise the pressure to 13 kPa (100
mmHg) immediately afterwards and record
the displayed value.
(b) Generate a constant gauge pressure of
+ 0.8 kPa ( + 6 mmHg) in the pneumatic
system by using the pressure/suction pump
at the moment of zero setting. During this
period close the deflation valve of the device
under test or close the hose to It, e.g. by
pinching the hose tightly. Set the blood
pressure measuring system to the service
mode, If available. Raise the pressure to
13 kPa. (100 mmHg) Immediately
afterwards. The zero setting Is operating
correctly If the displayed value decreases
by 0.8 kPa (6 mmHg) compared to the value ·
taken In (a).
(c) Repeat (b) with a constant gauge pressure
of - 0.8 kPa (- 6 mmHg) In the pneumatic
system. Set the blood pressure measuring
system to the service mode, if available.
Raise the pressure to 13 kPa.(100 mmHg)
immediately afterwards. The zero setting
is operating correctly If the displayed value
increases by 0.8 kPa (6 mmHg) compared
to the value taken in (a).
10. Test method for the drift of the cuff
·
pressure indication
(1) General
This test applies for devices performing zero
setting only immediately after switching on.
(2) Apparatus
rigid vessel with a capacity of 500 ml ±
5%;
calibrated reference manometer with an
uncertainty less than 0.1 kPa (0.8 mmHg);
stopwatch;
· T-plece connectors;
· patient simulator as described In paragraph
·
S(l)(i) of this Annexure.
(3) Procedure and evaluation
Replace the cuff with the 500 ml vessel. Insert
the calibrated reference manometer and the patient
simulator Into the pneumatic circuit by means of T
piece connectors.
Before beginning the test, allow the blood
pressure measuring system to reach operating
temperature as described in the Instructions for use,
Test the stability of the cuff pressure indication
after the zero setting at a pressure value of 7 kPa
(50 mmHg) according to the procedure specified In
paragraph 2 of this Annexure.
Under the same environmental conditions
determine the time (t.) until the change of the cuff
pressure Indication exceeds 0.1 kPa (1 mmHg).
Switch off the device and switch on afterwards.
Perform one blood pressure measurement and wait
until the device has switched off automatically.
Determine the time (t,) between switching on and
automatically switching off. The time (t,) shall be
less than or equal to the time (t.),
11. Test method for the stability of the blood
pressure determination. (influence of
temperature and humidity)
( 1) Apparatus
(I) patient simulator as described In
paragraph 5(1)(1) of this Annexure.
(II) climatic chamber, capable of adjustment
to an accuracy of 1 °C for the
temperature and 5% for the relative
humidity.
(2) Procedure
Carry out the testing of the signal processing
by means of the patient simulator. For each of the
---
. --- - ---
_L
612
.
THE GAZETTE OF. INDIA : EXTRAORDINARY [PARTJI-SEC. 3(i)]
followingcombinations of temperature and humidity,
place the blood pressure measuring system for at
least 3 h in the climatic chamber to allow the system
to reach steady conditions:
(i) 10°C ambient temperature, 85% relative
humidity(non-condensing);
(ii) 20°C ambient temperature, 85% relative
humidity (non-condensing);
(iii) 40°C ambient temperature, 85% relative
humidity (non-condensing).
For each combination of temperature and
humidity, take 20 consecutive readings of the blood
pressure measuring system under test.
Place the blood pressure measuring system in
the climatic chamber for at least 3 hours. At each
combination of temperature and humidity switch on
the blood pressure measuring system before starting
the test. Wait until the warm up time (described In
the instructions'for use) has elapsed, carry out the
measurement (20 consecutive readings) and switch
off the blood pressure measuring system afterwards.
(3) Expression of results
Determine.the mean value (systolic and diastolic
values separately) of the 20 consecutive readings
taken at each combination of temperature and
humidity.
Note: Because the testing of the influence of
temperature and humidity for the signal processing
cannot be separated from the temperature/humidity
effect on the pressure transducer and the deviations
originating from the simulator, both contributions
should be taken into account for the evaluation of
the test.
12, Test methods for the stablllty of cuff
pressure indication following prolonged usage
(1) Procedure
Carry out the test according to the procedure
specified in paragraph 2 of this Annexure prior to
prolonged usage.
Perform 10,000 simulated measurement cycles
with the complete blood pressure measure-ment
system at which at least the following cuff pressure
values shall be reached:
· adult mode: 20 kPa (150 mmHg);
· neonatal/Infant mode: 10 kPa (75 mmHg).
Note l: For devices which measure with the
auscultatory and oscillometric method this test should
be carried out for both modes.
Note 2 : For devices which measure In both
modes (adult and neonatal/infant) the test should
be carried out In both modes.
(2) Expression of results
Express the result as the difference between
the cuff pressure indication before and after 10,000
simulated blood pressure measurement cycles at the
same test pressure and under the same
environmental conditions.
13. Test methods for the effect of external
voltages and abnormal connections to the
signal Input/output ports
(1) Apparatus
rigid vessel with a capacity of 500 ml
:I: 5%;
calibrated reference manometer with an
uncertainty less than 0.1 kPa (0.8 mmHg);
T-plece connectors;
pressure generator, e.g. ball pump (hand
pump) with deflation valve.
(2) Procedure
Replace the cuff with the 500 ml vessel, insert
the calibrated reference manometer into the
pneumatic system by means of a T·piece and proceed
as follows:-
(i) Raise the pressure to 13 kPa (100 mmHg)
and record the displayed value.
(ii) Repeat (i) whilst short circuiting all contacts
of the signal input/output ports belonging
to the non-invasive blood pressure
measuring system.
(Ill) Repeat (i) whilst applying the maximum
voltage specified by the manufacturer to
each contact belonging to the non-invasive
blood pressuremeasuring system.
(3) Evaluation
Compare the indicated value under (i) with the
indicated values under (ii) and (Iii).
14, Test method for the cuff pressure deflation
following an aborted measurement
(1) Apparatus
(I) calibrated reference manometer with
an uncertainty less than 0.1 kPa (0.8
mmHg);
(JI) T·piece connectors.
(2) Procedure and evaluation
Insert the calibrated reference manometer into
the pneumatic system by means of a T·piece.
Start a blood pressure measurement. Abort the
measurement during inflation. Start an-other
measurement and abort it during the pressure
reduction. If interval measurements are possible
repeat the test in this mode.
Check by visual inspection whether the rapid
exhaust is activated.
['11'1 n-= J(i)J 613
PART VIII
TAXIMETERS
l. Terminology
(a) 'Taximeter'-A measuring instrument which
totalizes continuously and indicates at any moment
of the journey the charges payable by the passenger
of a public vehicle as function of the distance travelled
and, below a certain speed, of the length of time
occupied, independent of supplementary charges,
according to the authorised tariffs.
(b) Basicdtstence tariff-The tariff for distance
corresponding to all the Intervals except the initial
interval.
·
(c) Basic time tariff-The tariff for time
correspondingto all the Intervals except the Initial interval.
(d) Cleared-A taximeter is cleared when no
indication of fare Is shown and when all parts are in
the positions In which they are designed to be, when
the vehicle on which the taximeter is Installed Is not
engaged by a passenger.
(e) Reading face-The side of a taximeter upon
which the indications of interest to the passenger
are indicated.
(f) Fare-That portion of the charge for the
hire of a vehicle that Is automatically calculated by a
taximeter through the operation of the distance or
time mechanism,
(g) Flag-A device by which the operating
conditions of a taximeter is controlled.
(h) Initial distance or time inteNal-The Interval
corresponding to the Initial money drop.
(i) Money drop-An increment of fare Indication.
(j) Initial money drop-The Initial charges
appearing on the reading face of the taximeter at
the time when it Is hired by passenger.
(k) Distance of time inteNals-The intervals
corresponding to money drops following the initial
money drop.
(I) Constant 'k' of the taxkneter-The constant
'k' of a taximeter Is a characteristic quantity showing
the type and number of signals which the Instrument
must receive in order to indicate correctly a covered
distance of 1 km.
This constant 'k' Is expressed-
(i) in 'revolutions per indicated kilometre' (rev/
km) If the information relating to the
distance covered by the vehicle Is Introduced
into the taximeter in the form of a number
of revolutions of its main shaft (drive shaft
at entry point to the instrument).
(ii) in 'impulse per indicated kilometre' (imp/
km) if this information is introduced in
the form of electrical signals.
According to the construction of the instrument
the constant 'k' may be fixed or may be adjustable
by fixed amounts.
(m) Characteristic coemcient 'w' of the vehicle-
The characteristic coefficient 'w' of a vehicle is a
quantity Indicating the type and number of signals
Intended to drive the taximeter which appear at the
component provided for this purpose, for a distance
travelled of 1 km.
This coefficient 'w' is expressed-
(!) In 'revolutions per kilometre travelled'
(rev/km.); or
(II) In 'impulse per kilometre travelled' (Imp/
km),
depending on whether the Information relating to
the distance travelled by the vehicle appears In the
form of a number of revolutions of the component
driving the taximeter or in the form of electrical
signals,
This coefficient varies as a function of several
factors, principallythe wear and pressure of the tyres,
the load carried by the vehicle, the conditions under
which the vehicle makes a journey. It shall be
measured under the standard test conditions for the
vehicle.
(n) Adapting device-A special device which
allows the values of'k' and 'w' to be adjusted in such
a way that maximum penmisslble error laid down In
paragraph S(c) shall not· be exceeded.
2. General
(a) The following units of measurements shall
be used for taximeters:-
(!) the metre or kilometre, for distance
(11) the second, minute or hour for time.
(b) The fare for the journey shall be expressed
In the legal monetary units.
·
3. Technical characteristics
(A) General constructional features
(a) The taximeter shall be robust and well·
constructed. The functional parts of the
taximeter shall be made of materials
which guarantee adequate strength and
stability.
(b) The casing of the taximeter and that of
the adapting device as well as the covers
of the transmission devices shall be so
made that the essential parts of the
mechanism are out of reach from outside
and are protected against dust and
humidity.
--- ·------?,-- ?. -----
_L
--- ------
614 THE GAZETTE OF INDIA: .EXTRAORDINARY [PARTJI-SEC. 3(i)]
Access to devices for. adjustment shall be
made impossible without damage to the
sealing arrangements.
(c) In the case of electrontc'texlrneters.fbe
electronic devices which calculate the
charge payable .shall .operate without
failure;:and.at any time it shall.be possible
to checktheir correct operation.
This may· be achieved :by means of a
special -controt .proqramme which Is
automatk.or manually activated.
Any fault -ldentifled 'by this control
programme shall be clearly Indicated.
(B) Measuring device-Calculating device
(a) Except when the taximeter ls being
cleared, Indications offare shall be clearly
visible:at:all times.
(b) The taximeter shall be so designed that it
calculates and indicates the fare for the
journey solely on the basis of:
{i) the distance travelled when the
vehicle moves at a speed higher than
the changeover speed; or
(ii) the period of time when the vehicle
moves at a speed less than the
changeover speed.
The changeover speed Is attained
by dividing the time tariff by the
distance tariff and may vary
according to the variation In these
tariffs.
(c) The distance drive shall be made through
the medium of the wheels, and the reverse
motion shall not cause a reduction in the
fare or distance shown.
(d) The time drive shall be obtained by a
mechanical or electronic movement of the
dock work, which can be activated only
by operating the mechanism of the
taximeter.
(e) If the working of the clockwork
mechanism Is operated by manual
winding, It shall work for at least 8 hours
without rewinding or for at least two hours
if rewinding Is necessary at each manual
operation before it is set in operation.
The electronic clockwork, shall be capable
of functioning at any time.
(f) During the distance drive, the first
increment offare indication (money drop)
shall occur after travelling the Initial
distance. The subsequent money drops
shall correspond to equal distance
between each of them.
{g) During the time drlve, the first increment
of fare indication (money drop) shall occur
after the initial time interval. The
subsequent money drops shall correspond
to equal time intervals between each of
them.
(h) Without change of drive, the ratio
between the lnltlal distance and the
subsequent distances shall be the same
as the ratio between the Initial time and
subsequent time Intervals.
(i) An adapting device, situated inside or
outside the instrument case, shall allow
the adaptation of the taximeter constant
to the characteristic coefficient of the ·
vehicle on which It Is mounted, with an
accuracy such that the maximum
permissible errors laid down in paragraph
S(c) shall not be exceeded.
U) The taximeter shall be so designed as to
facilitate necessary adjustment of the
calculating device for making it conform
to the changes In the tariff.
(k) If the number of tariffs provided on the
Instrument is greater than the number of
tariffs in force, the taximeter shall, In the
superfluous positions, calculate and
Indicate a fare based on one of the
authorised tariffs.
(C) Control mechanism
(a) The mechanism of the taximeter shall be
capable of being set In motion after having
been engaged by a single control
mechanism.
For the electronic taximeter this
mechanism may consist of various push
buttons and switches for special
operations.
The mechanism of the taximeter shall be
capable of being set in motion in one of
the positions indicated In clauses (b), (c)
and (d).
(b) FREE position (for hire)
In this position-
(1) there shall be no indication of the
fare to be paid or, this indication shall
be equal to zero or to a value of the
initial money drop but in the latter
case, the indication shall be covered
·
by a shutter;
['WT ll-e,1ls 3(i)] 615
(ii) the distance drive and the time drive
shall not operate the device which
Indicates the fare to be paid;
(iii) the totalizer indicating the total
.
distance travelled shall remain
turned off;
(iv) the window through which possible
extras are seen shall be blank or
Indicate "Zero".
(c) "WORKING" position (hired)-In this
position, the time and distance drives and
the extras Indicator, if any, shall be
engaged,
(d) "TO PAY" position-In this position, which
shows the final total fare due from the
passenger for the journey excluding any
extras, the time drive shall be
disconnected and the distance drive shall
remain connected to the authorised tariff.
In case of electronic taximeters it is
permitted, by operating a special button
in the TO PAY position, to add any possible
extras to the fare, and to indicate on the
indicator, the total fare payable by the
passenger. When this button Is released
the two amounts shall be indicated
separately.
(e) The control mechanism, shall be so
designed that starting from FREE position,
the taximeter can be set successively in
WORKING position and TO PAY position.
(f) The operation of the control mechanism
is subject to the following restrictions:-
(i) Starting from the WORKING position, it
shall not be possible to put the taximeter
back in the "FREE" position without going
through the "TO PAY" position.
(ii) Starting from the TO PAY position, It shall
not be possible to put the taximeter in
the WORKING position without going
through the FREE position.
(D) Indicating device
(a) The "reading face" of the taximeter shall
be so designed that the indications of
interest to the passenger can be easily
read by him.
·
(b) The fare to be paid, excluding possible
extras, shall be ascertained from the fare
indicator by simple reading of an Indication
in aligned figures having a minimum
height of 10 mm.
When electronic indicating elements are
used it shall be possible to check the
operation of the indication.
(c) As soon as the instrument is put Into
operation from the FREE position by
operating the control mechanism, the
shutter, if any, covering the Indication of
fare to be paid, shall retract and a fixed
amount corresponding to the initial money
drop shall appear.
The fare indicator shall thereafter advance
by successive steps of a constant
· monetary value, as soon as the amount
of the initial money drop has been used
up.
(d) · The taximeter shall be provided with a
device indicating at any moment on the
reading face the engaged working
position.
(e) The taximeter shall have a means for
illuminating the readings appearing on the
reading face and it shall be possible to
replace the light bulbs without opening
the sea led parts of the meter.
In the case of self-luminous indications
no additional illumination is required If
readability of the indications of interest
to the passenger Is ensured.
(E) Optional additional devices
The taximeter may in addition be provided with
supplementary devices, such as-
(a) an Indicator of extras Independent of the
fare indicator mentioned in clause 3(D)(b)
and automatically returning to zero In the
FREE position. In the TO PAY position It Is
permitted to add the extras to the fare by
operating a push button;
(b) totalizers which give, In aligned figures
having a minimum height of 4 mm,
Indications of-
(1) the total distance travelled by the
vehicle;
(Ii) the total distance travelled on hire;
(111)the total number of "engagement";
(iv) the total number of money drop.
The totalizers may be also non-erasable
electronic memories which can be recalled
and indicated in the FREE position of the
taximeter. The data In the memory shall
be protected for at least 72 hours after a
power supply failure;
(c) a device detecting the presence of a
'
passenger (pressure sensor):
?- -· --· --- --- ---
----?
j_>·--
-- ---------------
616 THE GAZETTE OF INDIA: EXTRAORDINARY [PARTII-SEC. 3(i)]
6. Protection and guarantee seals
(a) The under mentioned part of the
taximeter shall be so constructed that they
can be sealed with .lead or by marks of
protection and guarantee:
{i) the case enclosing the Internal
mechanism of the taximeter;
(II) the case of adjusting device (if this
device is outside the case of the
taximeter);
· (iii) the covers of electrical or mechanical
devices forming the connection
between the entrance of the
taximeter and the corresponding
part provided on the vehicle for
alignment with the instrument,
including the detachable parts of the
adjusting device.
(b) A plate of approved size and pattern shall
be attached to the taximeter gear box or
to the taximeter Itself In such a manner
that It cannot be removed without either
.
removing the seals affixed by the testing
.
authority or opening the taximeter or
taximeter gear box. The plate shall show
In raised or sunken words and figures, the
type of cab on which the taximeter Is to
be used.
These seals shall be such that all access
to protected components particularlythe
components of adjustment is impossible
without damaging the seal.
7. standard test conditions for the vehicle
(a) The load carried by the vehicle
corresponds to the weight of four adult
persons including the driver.
(b) The tyres are inflated to the pressure
prescribed by the vehicle manufacturer
and are in good condition (e.g.,
conforming to the road safety rules).·
(c) The vehicle is moving, under its own
power, on level ground in a straight line,
_
at a speed less than 40 km/h.
8. Test methods
To determine compliance with distance
tolerances a distance test of the taximeter shall be
conducted utilizing anyone of the following test
methods:-
(a) Road test method-A road test consists
of driving the vehicles over a precisely
measured road course.
(b) Fifth wheel test method-This test
consists of driving the vehicle over any
reasonable road course and determining
the distance actually travelled through the
use of a mechanism known as "fifth
wheel" that is attached to the vehicle and
that independently measures and
indicates the distance.
(i)
,
/
for the Initial distance, 2 per cent of
the true value. However, for initial
distances less than 1000 metres, the
dispersion may be up to 20 metres;
(ii) for distances succeeding the initial
distance 2 per cent of the true value.
(b) During drive by time the maximum
dispersion of the indications shall not
exceed-
(i) for the initial time : 3 per cent of
the true value however, for Initial
times less than 10 minutes, this error
may be up to 20.seconds,
(ii} for times succeeding the Initial time:
3 per cent of the true value.
(c) The adaption shall be carried out under
the standard test conditions such that the
constant 'K' of the taximeter differs by less
than 1 per cent from the characteristic
number 'W' of the vehicle on which it is
mounted.
4. Inscription
(a) Every taximeter shall bear the following
indicatlons:-
(i) name and address of the
manufacturer or his trade mark.
(ii) the constant 'k' in rev/km. or Imp/km.
(b) Every taximeter shall have places to
permit:
Affixatlon of marks of initial and periodical
verifications.
(c) In the vicinity of the windows of all
indicating devices the meanings of the
values shown shall be given clearly, legibly
and unambiguously.
(d) The name or the symbol of the monetary
unit shall be alongside the fare Indication
for the Journey and the indication of extras
to be paid.
5. Maximum permissible errors
(a) During drive by distance the maximum
dispersion of the indications shall not
exceed:
[ "l1'T II - "@!US" 3 ( i ) ] 617
(c) Simul11ted road test method_;._This test
consists of determining the distance
travelled by computation from rolling
circumference and wheel turn data.
9. Test procedure
(a) The distance test of a taximeter, whether
(a), (b) or (c) shall include at least
duplicate runs of sufficient length to cover
at least the third money drop or one km.,
whichever Is greater, and shall be at a
speed approximating the average speed
travelled by the vehicle in normal service.
(b) If the taximeter is equipped with a
mechanism through which charges are
made for time intervals, a test shall be
conducted to determine whether there is
interference between the time and
distance mechanism. During this test, the
vehicle Is operated at a speed of 3 or 4
km./h faster than changeover speed.
PART IX
COMPRESSED GASEOUS FUEL (CNG)
MEASURING SYSTEMS FOR VEHICLES
HEADING-A
1. Scope
( 1) These specifications lay down the
metrological and technical requirements applicable
to compressed gaseous fuel measuring systems for
vehicles. It also specifies the requirements for the
approval of constituents elements of the measuring
systems (meter, etc.), Measuring systems for liquid
petroleum gas are not in the scope of this
specification as the fluid is at liquid state.
(2) In general, the measuring systems that are
covered by this specification are intended for the
refuelling of roadside motor vehicles, small boats,
aircraft and for trains. '
In principle, these specifications apply to all
measuring systems fitted with a meter as defined in
paragraph 1, sub-paragraph (1) (continuous
measurement), whatever be the measuring principle
of the meters or their application.
TERMINOLOGY
1. Measuring system and its constituents
(1) Meter
An instrument intended to measure
continuously, memorizes and display the quantity of
gas passing through the measurement transducer
at metering conditions.
Note: A meter Includes at least a transducer, a
calculator (includingadjustment or correction devices,
if present) and an indicating device.
(2) neesaremen: transducer
A part of the meter which transforms the flow
of the gas to be· measured Into signals which are
passed to the calculator. !t may be autonomous or
use an external power source.
Note: For the purposes of these specifications,
the measurement transducer includes the flow or
quantity sensor.
(3) C11lc11/ator
A part of the meter that receives the output
signals from the trensduceris) and, possibly, from
associated measuring instruments, transforms them
and, If appropriate, stores In memory the results until
they are used. In addition, the calculator may be
capable of communicating both ways with peripheral
equipment. ·
(4) Indicating device
A part of the meter, which displays continuously
the measurement results.
Note; A printing device which provides an
indication at the end of the measurement is not an
indicating device.
(5) Ancillary device
A device intended to perform a particular
function, directly involved In elaborating, transmitting
or-•displayingmeasurement results.
Main ancillary devices are:
(a) zero setting devices,
(b) repeating indicating device,
(c) printing device,
(d) memory device,
(e) price indicating device,
(f) totalizing indicating device, pre-setting
device, and
(g) self-service device.
(6) Additional device
A part or a device, other than an ancillary
device, required to ensure correct measurement or
intended to facilitate the measuring operations, or
which could In any way affect the measurement.
Main additional devices are:
(a) filter,
(b) device used for the transfer point,
(c) anti-swirl device,
(d) branches or bypasses,
(e) valves, and
(f) hoses.
618 THE GAZETTE OF INDIA: EXTRAORDINARY [PARTII-SEC. 3(i)]
(7) Measuring system
A system which comprises the meter Itself.and
all the ancillary devices.and additional·devlces.
(8) Compressed gaseous fuel measuring systems for
vehicles
A measuring systern intended for the-refuelling
of motor vehicles with compressed gaseous fuel.
(9) Pre-setting device
A device which permits ·the selection of the
quantity to be measured and ,which ·automatically
stops the flow .of .the gas at the end of the
measurement of the selected-quantity.
Note: The.pre-set quantity may be the mass
or the related pnce.to pay.
(10) Adjustment device
A device incorporated ,in the meter, that only
allows shifting of the .error curve generally parallel
to itself, with a wlew to -bringing errors within the
maximum permissible errors.
(11) Associated measuring instruments
Instruments connected to the calculator or the
correction device, for measuring certain quantities
which are characteristics of the gas, with a view to
making corrections.
{12) Correction device
A device connected to or incorporated in the
meter for automatically correcting the mass, by taking
into account the flow rate or the characteristics of
the gas to be measured (viscosity, temperature,
pressure) and the pre-established calibration curves
or both.
{13) Transfer point
A point at which the gas is defined as being
delivered.
2. Self.service measuring systems
(1) Self-service arrangement
An arrangement that allows the customer to
use a measuring system for the purpose of obtaining
gas for his own purchase.
(2) Self-service device
A specific device that is part of a self-service
arrangement and which allows one or more
measuring systems to perform in this self-service
arrangement.
Note: The self-service device Includes all the
elements and constituents that are mandatory so
that a measuring system performs In a self-service
arrangement.
The arrangement is made of a self-service
device and connected measuring systems.
(3) Attended service mode
An operating mode of a self-service
arrangement in which the supplier is present and
controls the authorization for the delivery.
(4) Unattended service mode
An operating mode of a self-service
arrangement In which the self-service arrangement
controls the authorization for a delivery, based on
an action of the customer.
Note: In unattended service mode, the end of
the measurement operation is the end of the
registration (printing and/or memorizing) of
information concerning the measurement operation.
(SJ Pre-payment
A type of payment in attended or unattended
service mode requiring payment for a quantity of
gas before the delivery commences.
(6) Attended post-payment (or post-payment)
A type of payment in attended service mode
requiring payment for the delivery but before the.
customer leaves the site of the delivery.
(7) Unattended post-payment (or delayed payment)
A type of payment in unattended service mode
in which payment for the delivered quantity Is
required after the delivery, but in which the
transaction Is not settled when the customer leaves
the site, following an implicit agreement with the
supplier.
(8) Authorization of a measuring system
An operation that brings the measuring system
into a condition suitable for the commencement of
the delivery.
3. Metro/ogical characteristics
(1) Primary Indication
An Indication (displayed, printed or memorized)
which Is subject to legal metrology control.
Note: Indications other than primary
indications are commonly referred to as secondary
indications.
(2) Absolute error of measurement
The result of a measurement minus the
(conventional) true value of the measurand.
(3) Relative error
The absolute error of measurement divided by
the (conventional) true value of the measurand.
[ '11'1 II-? 3 (i) J 619
(4) Maximum permlsslble errors
The extreme values permitted by the
specification for an error.
(5) Minimum measured quantity of a measuring
system
The smallest mass of gas for which the
measurement Is metrologically acceptable for that
system.
Note: In measuring systems intended to
deliver, this smallest mass is also referred to as the
minimum delivery.
(6) Minimum specified mass deviation
The absolute value of the maximum permissible
error for the minimum measured quantity of a
measuring system.
(7) Minimum specified price deviation
The price to pay corresponding to the minimum
specified mass deviation.
(8) Repeatability error
For the purposes of this specification, the
difference between the largest and the smallest
results of successive measurements of the same
quantity carried out under the same conditions.
(9) Intrinsic error
The error of a measuring system determined
under reference conditions.
(10) Initial intrinsic error
The intrinsic error of a measuring system as
determined prior to all performance tests.
(11) Fault
The difference between the error of Indication
and the (initial) intrinsic error of a measuring system.
(12) Significantfault
For the mass, a fault means magnitude of which
is greater than the larger of these two values :-
(ii) one-tenth of the magnitude of the ·
maximum permissible error for the
measuring system and for the measured
mass,
( iii) the minimum specified mass deviation.
For the price to pay, the price corresponding to
the significant fault for the mass.
Note: No fault is allowed for the unit price.
The following are not considered to be
significant faults :-
(v) faults arising from simultaneous and
mutually Independent causes In the
measuring instrument Itself or in its
· checking facilities;
(vi) transltorv faults being momentary
variations in the Indication, which cannot
be interpreted, memorized or transmitted
as a measurement result.
(13) Durabl/ity
The capability of the measuring system to keep
its performance characteristics over a period of use,
for electron le purposes.
4. Tests and test conditions
(1) Influence quantity
A quantity which is not the subject of the
measurement but which Influences the value of the
rneasurand or the indication of the measuring system.
(2) Influence factor
An influence quantity having a value within the
rated operating conditions of the measuring system,
as specified in these specifications.
(3) Disturbance
An influence quantity having a value outside
the specified rated operating conditions of the
measuring system.
Note ; An influence quantity is a disturbance If
for that Influence quantity the rated operating
conditions are not specified.
(4) Rated operating conditions
Conditions of use, giving the range of values of
influence quantities for which the metrological
characteristics are intended to be within the
maximum permissible errors.
(5) Reference conditions
A set of specified values of influence factors
fixed to ensure valid inter comparison of results of
measurements.
(6) Performance test
A test Intended to verify whether the measuring
system under test (EUT) is capable of accomplishing
its intended functions.
(7) Endurance test
A test intended to verify whether the meter or
the measuring system is able to maintain Its
performance characteristics over a period of use.
(B}Bank
A test reservoir or a set of test reservoirs
manifolded together which forms part of a multl
segment gas storage system. The segments operate
at different pressure levels from one another in
refuelling systems fitted with or using a sequential
control device.
Note; Testing by using banks generate
transient flow rates.
-------·----
--·--·-
--?--
- --
-------------
620 THE GAZETTE ,OF ·1ND1A-: >EXTRAORDINARY [PARTII SEC. 3(i)]
{9) Sequential control device
A device whlch·allows·switching from a bank to
another one. This device mayJie induded.ln a measuring
system or may be partof the refuelling,statlon.
5. Electronlcor electrical equipment
(1) Electronic device
A device employing electronic·sub-assemblles
and performing:a.speclfic function.
Note: Electronic .devices, .as defined above,
may be cornplete.measurmq .systems .or .pert of
measuring systems.
(2) Electronlc sub,essembly
A part of .an electronic device, employing
electronic components and having a .recognizable
function of its own.
(3) Electronic component
The smallestphysical entity which uses electron
or hole conduction In semi-conductors, gases, or in
a vacuum.
(4) Checking facility
A facility Which .is Incorporated In a measuring
system and which enables significant faults to be
detected and acted upon.
Note: The checking of a transmission device
alms at verifying that all the information which is
transmitted is fully received by the receiving
equipment.
(5) Automatic checking facility
A checking facility operating without the
intervention of an operator.
(6) Permanent automatic checking faclllty (type P)
An automatic checking facility operating during
the entire measurement operation.
(7) Intermittent automatic checking facility (tupe I)
An automatic checking facility operating at least once,
either at the beginning or at the end of each
measurement operation.
(8) Power supply device
A device which provides the electronic devices
with the required electrical energy, using one or
several sources of e.c, or d.c.
HEADING-B
J., General requirements
(1) Constituents of a measuring system
A meter itself is not a measuring system. A
measuring system includes at least-
(a) a meter,
(b) a transfer point,
(c) a gas circuit with particular characteristics
which must be taken into account.
The measuring system may be provided with
other ancillary and additional devices [paragraph 2
sub-paragraph (2)).
·
If several meters Intended for separate
measuring operations have common elements
calculator filter, etc.) each meter Is considered to
form, with the common elements, a measuring
system.
A measuring system shall Include only one meter.
(2) Ancillary and additional devices
(I) Ancillary devices may be a part of the
calculator or of the meter, or may be
peripheral equipment, connected through
an Interface to the calculator. These
devices shall bear a legend which is clearly
vlslble to the user to indicate that they
are not controlled when they display a
measurement result visible to the user.
Such a legend shall be present on each
print-out likely to be made available to
the customer.
(11) By definition, additional devices likely to
be Installed in a measuring system shall
not corrupt the metrological behaviour of
the measuring apparatus.
(3) Field of operation
(I} The field of operation of a measuring
system is determined by the following
characteristics:-
minimum measured quantity,
measuring range limited by the
'minimum flow rate, Qmin and the
maximum now rate, Qmax
'
maximum pressure of the gas,
Pmax,
minimum pressure of the gas, Pmln
'
if appropriate, nature and
characteristics of the gases to be
measured,
maximum temperature of the gas,
Tmax,
minimum temperature of the gas,
Tmin•
environment class
The environmental class may be
different according to devices of the
measuring system, provided each
device is used according to its own
environmental class. In particular
this is applicable to some parts of a
self-service device which can be
used at different temperatures than
the rest of the measuring system.
['ll'T J[-"1.§1116 3(i)] 621
(ii) The minimum measured quantity of
a measuring system shall have the
form 1 x 10n, 2 x 10n or 5 x 10n
authorised units of mass, where n
Is a positive or negative whole
number, or zero.
The minimum measured quantity
shall satisfy the conditions of use of
the measuring system. Except in.
exceptional cases, the measuring
system shall not be used for
measuring quantities, less than this
minimum measured quantity.
Measuring systems having a maxi·
mum flow rate not greater than 30 kg/
min shall have a minimum measured
quantity not exceeding 2 kg.
Measuring systems having a
maximum flow rate larger than or
equal to 30 kg/min but not greater
than 100 kg/min shall have a
minimum measured quantity not
exceeding 5 kg.
(iii) The measuring range shall satisfy
the conditions of use of the
measuring system; the latter shall
be designed so that the flow rate is
between the minimum flow rate and
the maximum flow rate, except at
the beginningand at the end of the
measurement or during inter
ruptions.
In normal conditions of use, the
measuring system shall not be
capable of delivering quantities at
flow rates smaller than the minimum
flow rate. The measuring range of a
measuring system shall be within the
measuring range of each of its
elements.
The ratio between the maximum
flow rate and the minimum flow rate
shall be at least fifteen.
(iv) A measuring system shall exclusively
be used for measuring gas having
characteristics within Its field of
operation, as specified In the model
approval certificate. The field of
operation of a measuring system
shall be within the fields of
measurement of each of Its
constituent elements, In particular
the meter.
(4) Indications
(i) Measuring systems shall be provided with
an indicating device gMng the mass of
gas measured. If the system is fitted with
a price indicating device, then-
1(4)(i)(a) indications of unit price and
price to be paid are related only to mass.
1(4)(i)(b) these indications are displayed
only when displaying the mass.
(11) Mass shall only be indicated in kilogram.
The symbol or the name of the unit shall
appear in the immediate vicinity of the
Indication.
(Ill) A measuring system may have several
devices indicating the same quantity. Each
shall meet the requirements of this
specification if subject to control. The scale
Intervals of the various indications shall
be the same.
(Iv) For any measured quantity relating to the
same measurement, the indications
provided by various devices shall not
deviate one from another.
(v) The use of the same indicating device for
the indications of several measuring
systems (which then have a common
Indicating device) is authorized provided
that it is impossible to use any two of these
measuring systems simultaneously, and
that the measuring system providing the
Indication is clearly identified.
(vi) The scale interval shall be in the form 1 x
10n, 2 x 10n or 5 x 10n authorized units
of mass, where n is a positive or negative
whole number, or zero.
The scale interval shall be equal to or
smaller than half the minimum specified
mass deviation.
However non-significant scale intervals
should be avoided. Thus does not apply
to price indications.
(vii) When relevant, the provisions relating to
mass indications apply also to price
Indications by analogy, and to secondary
Indications of other quantities as well.
(S) Suitability of additional devices
(I) Measuring systems shall incorporate a
transfer point. This transfer point Is
located down stream of the meter.
(II) No means shall be provided by which any
measured gas .can be diverted down
stream of the meter during the filling
operation.
(Ill) 1\No or more delivery transfer points may
----?=------=-----=-=--=::..:..:..:....=------=-----=-=---=---=-- ----- --?- - -
622 THE GAZETTE OF INDIA: EXTRAORDINARY [PARTII SEC. 3(i)]
be permanently installed and operated
simultaneously or alternately provided so
that any diversion of gas to other than
the Intended receiving receptacle(s)
cannot be readily accomplished or is
readily apparent. Such means Include, for
example, physical barriers, vlslble valves
or Indications that make it clear which
transfer points are in operation, and
explanatory signs, if necessary.
(iv) When only onetransfer point can be used
during a. delivery, and after the transfer
.
point has been replaced, the next delivery
shall be Inhibited until the indicating
device· has been reset to zero.
When two or more transfer points can be
used simultaneously or alternately, and
after.the utilized transfer points have been
replaced, the next delivery shall be,
inhibited ,until the indicating device has
been reset to zero. Moreover, by design,
the provisions of paragraph 2 sub
paragraph (5) clause (iii) shall be fulfilled.
(v) .
The system shall be designed In order to
ensure that the measured quantity is
delivered. In particular if the hose down
stream of the meter is likely to be de
pressurtzed between two deliveries,this shall
lead for instance to systematic correction or
re-pressurizing before counting.
Whatever is the operating principle, in
particular what constitutes the hose or the
transfer point, in the worst measuring
conditions, the mass which Is measured
but not delivered shall be smaller than or
equal to half the minimum specified mass
deviation.
Note:The purpose of this provision is not
to allow a systematic deviation. This shall
be verified by design examination, tests
or calculation.
(vi) If there Is a risk that the supply conditions
can overload Qmax of the meter, a flow
limiting device shall be provided. It shall
be possible to seal it.
(vii) A pressure gauge shall be permanently
installed on the measuring system In order
to check Pmax and Pmin·
2. Metrologlca/ requirements for measuring
systems and meters
(1) Maximum permissible errors and other
metrological characteristics
(i) Without prejudice to paragraph 2 sub
paragraph (1) clause (iii) the maximum
permissible relative errors on mass
indications, positive or negative, at model
approval are,
- ± 1% of the measured quantity for the
meter alone, and
- ± 1.5% of the measured quantity for
the complete measuring system.
(ii) The maximum permissible relative errors
on mass indications, positive or negative,
.
at initial verification or subsequent
verifications and when these verifications
are performed under rated operating
conditions on-site of use, are ± 2% of
the measured quantity for the complete
measuring system.
(Iii) The maximum permissible errors
applicable to the minimum measured
quantity are twice the corresponding
values stated in paragraph 2, sub
paragraph (1), clause (i).
So the minimum specified mass deviation
(Eminlfor the measuring system is given
by the formula:
Emin = 3 x Mmin/ 100
Where Mmin is the minimum measured
quantity having the form specified In
paragraph 1, sub-paragraph (3), clause
(II}.
Note: The minimum specified mass
deviation is an absolute maximum
perrnlsslbleerror.
(Iv} Whatever the,measured quantity may be,
the magnitude of the maximum
permissible error (expressed as an
absolute error) for the complete system
Is never less than the minimum specified
mass deviation.
(v) For any quantity equal to or greater than
1000 scale intervals, the repeatability error
of the meter shall not be greater than ±
0.6%.
(vi} Within their field of operation, meters shall
present a magnitude of the difference
between the initial intrinsic error and the
error after the endurance test equal to or
less than ± 1%.
The requirement on repeatability applies
after the endurance test.
(2) Conditions for applying maximum permissible
errors
(I) Maximum permissible errors apply for all
gas, all temperatures and all pressures of
['WI n-= 3(i)J 623
the gas, and all flow rates for which the
system or the meter is intended to be
approved.
A measuring system or a meter shall be
capable of fulfilling all requirements
without adjustment or modification during
the relevant evaluation procedure.
(ii) When stated in the model approval
certificate, a verification of a measuring
system or of a meter intended to measure
gas may be carried out with air (or with
another fluid). In this case and if
necessary, the pattern approval certificate
provides a smaller range or a shin for
maximum permissible errors, so that
maximum permissible errors are fulfilled
fo?gas.
3. Requirements for meters and ancillary
devices of a measuring system
The meter and ancillary devices of a measuring
system shall meet the following requirements,
whether or not they are subject to a separate model
approval:-
(1) Meter
(i) Field of operation
The field of operation of a meter is
determined at least by the following
characteristlcs:-
minimum measured quantity,
measuring range limited by the minimum
flow rate, Qmin, and the maximum flow
rate, Qmax,
maximum pressure of the gas, Pmax,
if appropriate, nature and characteristics
of the gases to be measured,
maximum temperature of the gas, T max
'
minimum temperature of the gas T min.
The temperature range shall be + 100C to
+ 40°c.
(ii) Metrological requirements
See paragraph 2 under Headlng-B.
(iii) Connections between the now sensor and
the Indicating device
The connections between the flow sensor
and the indicating device shall be reliable
and, for electronic devices, durable.
(iv) Adjustment device
Meters may be provided with an
adjustment device which permits
modification of the ratio between the
indicated mass and the actual mass of gas
passing through the meter, by a simple
command.
·
·
When this adjustment device modifies this
ratio in a discontinuous manner, the
consecutive values of the ratio shall not
differ by more than 0.001.
Adjustment by means of a by pass of the
meter is prohibited.
(v) Correction device
Meters may be fitted with correction
devices; such devices are always
considered as an integ rat part of the
meter. The whole of the requirements
which apply to the meter, in particular the
maximum permissible errors specified In
paragraph 3 sub-paragraph (1) clause (Iii)
of Heading-B, are therefore applicable
to the corrected mass.
In normal operation, non-corrected mass
shall not be displayed.
The aim of a correction device is to reduce
the errors as dose to zero as possible.
Note: The use of this device for adjusting
the errors of a meter to values other than
as close as practical to zero is forbidden,
even when these values are within the
maximum permissible errors.
The correction device shall not allow the
correction of a pre-estimated drift In
relation to times or mass, for example,
The accuracy of associated measuring
Instruments shall be good enough to
permit that the requirements on the meter
be met as specified In paragraph 2, sub
paragraph ( 3) of Heading-B.
Associated measuring Instruments shall be
fitted with checking devices, as specified
In paragraph 4, sub-paragraph (3) clause
(vi) of Heading-B.
(2) IndlC8tlng device
(I) Measuring systems shall be equipped with
digitalindicatingdevices. The decimal sign
shall appear distinctly.
(II) The continuous display of mass during the
period of measurement is mandatory.
(iii) The height for the figures of the indicating
device shall be equal to or greater than
10 millimetre.
I
?··.l----·--- ----------------
624 THE GAZETTE OF INDIA : EXTRAORDINARY [PARTII SEC. 3(i)]
(3) Zero setting device
(i) Measuring systems shall be equipped with
a device for resetting the mass indicating
device to zero.
(a) The zero setting device shall not
permit any alteration of the
measurement result shown by the
mass Indicating device (other than
by making the result disappear and
displaying zeros).
(b) Once the zeroing operation has
begun It shall be impossible for the
mass Indicating device to show a
result different from that of the
measurement which has Just been
made, until the zeroing operation
has been completed.
The measuring system shall not be
capable of being reset to zero during
measurement.
(ii) If the system also includes a price
indicating device, this indicating device
shall be fitted with a zero resetting device.
The zero setting devices of the price
indicating device and of the mass
indicatingdevice shall be designed In such
a way that zeroing either indicating device
automatically involves zeroing the other.
(iii) If the measuring system Is fitted with a
printing device, any printing operation
shall not be possible in the course of a
measurement and further delivery shall
only be possible efter a reset: to zero has
been performed. However, the printing
operation shall not change the quantity
indicated on· the indicating device.
(iv) If the measuring system is designed so
that registration of mass could occur
without any effective flow rate, a device
shall register this apparent flow rate and
compensate the measurement result for
it.
(4) Price indicating device
(i) A mass indicating device. may be
complemented with a price Indicating
device which displays both the unit price
and the price to be paid.
The monetary unit used or its symbol shall
appear illl the immediate vicinity of the
indk:al:ll:m.
(ii) The selected unit price shall be displayed
by an !Indicating device before the start
of the measurement to unit price shall be
adjustable; changing the unit price may
be carried out either directly on the
measuring system or through peripheral
equipment.
The indicated unit price at the start of a
measurement operation shall be valid for
the whole transaction. A new unit price
shall only be effective at the moment a
new measurement operation may start.
A time of at least five seconds shall elapse
between indicating a new unit price and
before the next measurement operation
can start, if the unit price is set from
peripheral equipment.
(iii) Only rounded errors pertaining to the least
significant digit of the price to be paid are
authorized.
(5) Printing device
(I) The mass printed shall be expressed in
multiples or sub-multiples of kilogram for
the indication of mass.
The figures, the unit used or its symbol
and the decimal sign, if any, shall be
printed on the ticket by the device.
(ii) The printing device may also print
Information identifying the measurement
such as: sequence number, date,
Identification of the measuring system,
type of gas, etc. If the printing device is
conrlected to more than one measuring
system, 1.tmust print the identification of
the relevant system.
(Ill) If a printing device allows repetition of
the printing before a new delivery has
started, copies shall be clearly marked as
such, for example by printing "duplicate".
(Iv) The printing device shall print, in addition
to the measured quantity, either the
corresponding price and the unit price.
(v) The printing devices are also subject to
the requirements In paragraph 4, sub·
paragraph (3), clause (v) of Heading-B.
(6) Memory device
(I) Measuring systems may be fitted with a
memory device to store measurement
results until their use or to keep a trace
of commercial transactions, providing
proof in case of a dispute. Devices used
to read stored. information are considered
as included in the memory devices.
['WT II-? 3(i)]
(ii) The medium on which data are 'stored
must have sufficient permanency to
ensure that the data are not corrupted
under normal storage conditions. There
shall be sufficient data memory storage
for any particular application and for
roadside measuring systems data storage
for at least one month corresponding to
normal use shall be available.
(iii} When the storage is full, it Is permitted to
delete memorized data when both the
following conditions are met:-
(a) data are deleted in the same order
as the recording order and the rules
established for the particular
application are respected,
(b) deletion Is carried out after a special
manual operation.
(iv) Memorization shall be such that it is
impossible In normal use to modify stored
values.
(v} Memory devices shall be fitted with
checking facillties according to paragraph
4, sub-paragraph (3), clause (v) of
Heading-B. The aim of the checking
facility Is to ensure that stored data
correspond to the data provided by the
calculator and that restored data
correspond to stored data.
(7) Pre-setting device
(i) The selected quantity Is pre-set by
operating a digital device which indicates
that quantity. The pre-set quantity shall
be indicated before the start of the
measurement.
(ii) Where it Is possible to view simultaneously
the figures of the display device of the
pre-setting. device and those of the mass
indicating device, the former shall be
clearly distinguishable from the latter.
(iii) Indication of the selected quantity may,
during measurement, either remain
unaltered or return progressively to zero.
(iv) The difference found under normal
operating conditions, between the pre-set
quantity and the quantity shown by the
mass Indicating device at the end of the
measurement operation, shall not exceed
the minimum specified mass deviation.
{v) The pre-set quantities and the quantities
shown by the mass indicating device shall
625
be expressed in the same unit. This unit
or Its symbol shall be marked on the pre
setting device.
(vi) The scale interval of the pre-setting device
shall be equal to the scale interval of the
Indicating device.
(vii) Pre-setting devices may incorporate a
device to permit the flow of gas to be
stopped quickly when necessary.
(viii} Measuring systems with a price indicating
device may also be fitted with a price pre
?tting device which stops the flow of the
gas when the quantity delivered
corresponds to the pre-set price. The
requirements in dauses (i) to (vii) of sub
paragraph (7) of paragraph 3 of Heading
s apply by analogy.
(8) Calculator
(I) The maximum permissible errors, positive·
or negative, on the gas quantity
Indications applicable for the calculators
when they are checked separately shall
be 0.05 per cent of the true value.
(ii) All factors necessary for the elaboration
of indications such as unit price,
calculation table, correction polynomial,
etc. shall be present in the calculator at
the beginning of the measurement
operation.
(Iii) The calculator may be provided with
Interfaces permitting the coupling of
peripheral equipment. When these
Interfaces are used, the instrument shall
continue to function correctly and Its
metrological functions shall not be capable
of being affected.
4. Technical requirements for electronic
devices·
(1) General requirements
(I) Electronic measuring systems shall be
designed and manufactured such that
their errors do not exceed the maximum
permissible errors as defined in paragraph
2 sub-paragraph (1) under rated operating
conditions.
(11) Electronic measuring systems shall be
designed and manufactured such that,
when they are exposed to the
disturbances:
either (a) significant faults do not
occur,
626 THE GAZETTE OF INDIA : EXTRAORDINARY [PARTII SEC. 3(i)]
..
-or (b) significant faults are detected
and acted upon by means of
checking facilities.
This provision may apply separately to:
each individual cause of significant
fault and/or,
each part of the measuring system.
.
(iii) The requirements in paragraph 4, sub·
paragraph (1), clause (i) and paragraph
4, sub-paragraph (1), clause (Ii) of
Headlng-B shall be met durably. For this
purpose, electronic measuring systems
shall be provided with the checking
facilities specified in paragraph 4, sub·
paragraph (3) of Heading-B.
(iv) Electronic measuring systems shall be
such that the delay time between the
measurement value and the
corresponding indicated value shall not
exceed 500 milli second.
(v) A model of a measuring system is
presumed to comply with the
requirements in sub-paragraph (1) of
paragraph 4 of this Heading If It passes
the examination and tests specified in
paragraph 7, sub-paragraph (1) clause (ix)
sub-clauses (a) and (b) of Headlng-B.
(2) Power supply device
(I) A measuring system shall be provided with
an emergency power supply device
allowing
either (a) to safeguard all measur
ing functions during a
failure of the principal
power supply,
or (b) that data contained at
the moment of a failure
leading to stopping the
flow are saved and
displayable on an
indicating device for
sufficient time to permit
the conclusion of the
current transaction.
The absolute value of the maximum
permissible error for the Indicated
mass, In the second case, is
Increased by 5 per cent of the
minimum measured quantity.
(ii) In case of a failure leading to stopping
the slow, measuring systems shall be such
that the minimum duration of operation
of the ,display shall be either
continuously and automatically at
least 15 minutes following
immediately the failure of the
principal electrical supply, or
a total of at least 5 minutes In one
or several periods controlled
manually during one hour following
Immediately the failure.
Note : During model approval the instrument
has to be supplied with electric power normally for
12 hours which preceded the test. Before this supply
the battery (If provided) may be uncharged.
In addition, measuring systems shall be
designed so that an interrupted delivery cannot be
continued after the power supply device has been
re-established if the power failure has lasted more
than 15 seconds.
(3) Checking facilities
(I) Action of checking facilltles
The detection by the checking facilities of
significant faults shall result in the
following actions:-
(a) automatic correction of the fault, or
(b) stopping only the faulty device,
when the measuring system without
that device continues to comply with
the regulations, or
(c) stopping the flow.
'
(11) Checking facilities for the measurement
transducer
The objective of these checking facilities
Is to verify the presence of the transducer,
Its correct operation and the correctness
of data transmission.
These checking facilities shall be of type
P and the checking shall occur at time
Intervals not exceeding the duration of the
measurement of an amount of gas equal
to the minimum specified mass deviation.
It shall be possible during model approval
and initial verification to check that these
checking facilities function correctly-
(a) by disconnecting the transducer, or
(b) by interrupting one of the sensor's
pulse generators, or
(c) by interrupting the electrical supply
of the transducer.
[ 'WT II - <51"6 3 ( i) ] 627
(iii) Checking facilities for the calculator
The objective of these checking facilities
is to verify that the calculator system
functions correctly and to ensure the
validity of the calculations made.
There are no special means required for
indicating that these checking facilities
function correctly.
(a) The checking of the functioning of
the calculation system shall be of
type P or I. In the latter case, the
checking shall occur at least every
five minutes in the course of a
delivery but at least once during a
delivery.
The objective of the checking is to
verify that-
(a· 1) the values of all permanently
memorized instructions and data are
correct, by such means as
summing up all instruction and
data codes and comparing the
sums with a fixed value,
line and column parity bits,
cyclic redundancy check,
double Independent storage of
data,
storage of data In "safe coding';
for example protected by check
sum, line and column parity bits.
(a·2)all procedures of internal transfer
and storage of data relevant to the
measurement result are performed
correctly, by such means as
write-read routine,
conversion and re-conversion of
codes,
use of "safe coding" (checksum,
parity bit),
double storage.
(b) The checking of the validity of
calculations shall be of type P. This
consists of checking the correct
value of all data related to the
measurement whenever these data
are Internally stored or transmitted
to peripheral equipment through an
interface; this check may be carried
out by such means as parity bit,
check sum or double storage. In
- -
----
----
addition, the calculation system shall
be provided with a means of
controlling the continuity of the
·
calculation program.
(Iv) Checking facility for the indicating device
The objective of this checking facilityis to
verify that the primary indications are
dis played and that they correspond to the
data provided by the calculator. In
addition, it aims at verifying the presence
of the indicating devices, when they are
removable. The control may be performed
according to either the first possibilityIn
paragraph 4, sub-paragraph (3), clause
(Iv) sub-clause (b) or the second
possibility in paragraph 4, sub-paragraph
(3), clause (iv) sub-clause (c) of
Heading-B.
(a) It shall be possible during verification to
detenrnine that the checking facilityof the
Indicating device Is working, either
by disconnecting all or part of the
indicating device,
or by an action which simulates a
failure in the display, such as using
a test button.
(b) The first possibility is to control
automatically the complete indicating
device. The checking facility of the
Indicating device is of type P; however, It
may be of type I if a primary indication Is
provided by another device of the
measuring system, or if the indication may
be easily determined from other primary
Indications (for example, in the case of
the presence of a price indicating device
It Is possible to determine the price to pay
from the mass and the unit price),
Means may include, for example
for indicating devices using
incandescent filaments or light
emitting diode, measuring the
current in the filaments,
for indicating devices using
fluorescent tubes, measuring the
grid voltage,
for indicating devices using electro
magnetic shutters, checking the
impact of each shutter,
for indicating devices using
multiplexed liquid crystals, output
checking of the control voltage of
__
l
... ----
--
628 THE GAZETTE OF INDIA : EXTRAORDINARY [PARTII-SEC. 3(i}]
segment lines and of common
electrodes, so as to detect any
disconnection or short-circuit
between control circuits.
(c) The second possibility is on the one hand
to check automatically the electronic
circuits used for the Indicating device
except the driving circuits of the display
itself and on the other hand to check the
display.
The automatic checking facility of the
electronic circuits used for the indicating
device Is of type P; however, It may be of
type I If a primary indication Is provided
by another device of the measuring
system, or if the indication may be easily
determined from other primary Indications
(for example, in the case of the presence
of a price lndlcatt ng device It Is possible
to determine the price to pay from the
mass and the unit price).
The checking facility of the display shall
provide visual checking of the entire
display which shall meet the following
descrlption:-
displaylng all the elements·("eights"
test If appropriate)
blanking all the elements ("blank"
test)
displaying "zeros"
Each step of the sequence shall last
at least 0.75 second.
This visual checking facility shall be
of type I but it is not mandatory for
a malfunction to result in the actions
described in paragraph 4, sub
paragraph (3), clause (i) of
Headlng-B.
(v) Checking facilities for ancillary devices
An ancillary device (repeating device,
printing device, self-service device,
memory device, etc.) with primary
indications shall include a checkingfacility
of type I or P. The object of this checking
faciUty Is to verify the presence of the
anci!larv device, when it Is a necessary
device, and to verify the correct
transmission of data from the calculator
to the .ancillary device.
In par.tlctilar, the .checking of a printing
de\iiae aims atensuring that the printing
controls correspond to the data.
transmitted by the calculator. At least the
following shall be checked:-
{a) presence of paper, and
{b) the electronic control circuits (except
the driving circuits of the printing
mechanism Itself).
It shall be possible during model
approval and other verifications to
check that the checking facilityof the
printing device Is functioning by an
action simulating a printing fault,
such as using a test button.
Where the action of the checking
facility is a warning, this shall be
given on or by the ancillary device
, concerned.
(vi) Checking facillties for the associated
measuring instruments
Associated measuring Instruments shall
Include a checking facility of type P. The
aim of this checking facility is to ensure
that the signal given by these associated
Instruments is Inside a pre-determined
measuring range.
Examples:
four wire transmission for resistive
sensors,
frequency filters for density meters,
control of the driving current for
4-20 mA pressure sensors.
5. Technltal requirements, for measuring
systems with self-service a"angement
(1) General requirements
(I) Where the self-service device serves two
or more measuring systems, each
measuring system shall be provided with
a measuring system Identification number
that shall be accompanied by a primary
Indication provided by the self-service
device.
{II) Indication of Information that is not
·
subject to rnetroloqlcal control is allowed,
provided that it cannot be confused with
metrological information.
(iii) The control device of the self-service
device should be capable of indicating the
status of the measuring system (e.g.
running, authorized or unauthorized) that
are connected to the self-service device
and in the case of multiple modes of
service and/or type of payment also that
particular status of the measuring system.
·· t-T···•?' .,.,.,,,. -?-, , "" ,-....,? -c··•-?---..- -..-- -
['ll'T II-? 3(i)] 629
· (iv) A change of the type of payment and/or
mode of operation shall not be effective
before the end of the current
measurement operation.
(v) The self-service arrangement, Including
provisions related to clearly defined
methods of operation, shall be such that
at least one primary indication for the
benefit of the customer must be available
at least up to the settlement of the
transaction to enable the delivered quantity
and the price to pay to be checked.
(vi) In the case of a self-service arrangement
that totallzes the delivered mass for
different registered customers over the
course of time, the minimum measured
quantity Is not affected by the scale
interval used for such tantalizations.
(2) Attended service mode
· If the measuring system indicating device
provides the only primary indication, provisions shall
be made to inform die customer that die next
authorization of a particular measuring system can
only be given by the supplier after settlement of the
current transaction.
(i) Attended post-payment
(a) Where the self-service arrangement
includes a device that provides an
additional primary indication (additional
to those of the indicating device of the
measuring system}, it shall consist of at
least one Installation for the reproduction
of the mass and/or the price Indicated by
the measuring system indicating device,
consisting of
a printing device for the Issue of a
receipt to the customer, or
an Indicating device for the benefit
of the supplier together with a
display for the benefit of the
customer.
(b) For self-service devices with temporary
storage (temporary storage mode} of
measurement data of measuring system
the following requirements apply:
(b·l}temporary storage of measurement
data shall be restricted to one
delivery for each measuring system,
(b-2}the primary indication shall be
accompanied by a clear mark
representing the sequence. For
example, the number 1 or 2 or the
letter A or B,
(b-s) when a primaryIndication of the self
service device Is out of service, die ·
self-service arrangement may
continue Its operation provided that
it no longer uses any temporary
storage, and that the measuring
system indicatingdevice remains the
primary indication.
(c) Where the mandatory primary indication
for the benefit of the customer is provided
by a device in the form of a separate
constructional unit and this unit becomes
uncoupled, or If the checking facilities
detect a malfunction, the temporary
storage mode shall be prohibited and the
measuring system Indicating device
remains the primary Indication.
(Ii) Pre-payment in attended service made
(a) The requirements of paragraph 3,
sub-paragraph (7) of Heading-Bare
applicable.
(b) A printed or hand-written receipt of
the pre-paid amount shall be
provided.
(3) Unattended service mode
(I) General
(a) The self-service arrangement shall
provide .
additional primary
indications by means of
a printing device for the Issue
of a receipt to the customer, and
a device (printing or memory
device) on which measurement
data are registered for the
benefit of the supplier.
(b) When the printing devices or
memory device, as required In
paragraph 5, sub-paragraph (3),
clause (i), sub-clause (a) are not
able to provide any indication or
become unserviceable, the customer
shall be clearly warned by automatic
means before the operation
commences.
Passing from attended to
unattended service mode shall not
be possible before correct operation
of the arrangement is concluded as
feasible by the checking facilities,
including compliance with the above
provision.
630 THE GAZETTE OF INDIA : EXTRAORDINARY [PARTII-SEC. 3(i)]
(c) Where the self-service arrangement
Is used by registered customers, the
provisions of sub-clauses (a) and (b)
of clause (i) of sub-paragraph (3) of
paragraph 5 do not apply to
measurements related to such
customers. An additional Individual
mass totalizer is considered to
,
provide a primary indication.
(d) Micro-processors, which upon
disturbance or interference Influence
the measurement operation, shall be
equipped with means for controlling
the continuity of the processor
program and for ensuring the
discontinuation of the current delivery
when the continuity of the processor
program is no longer ensured.
The next effective acceptance of
notes, cards or other. equivalent
mode of payment shall only take
place If the continuity of the
processor program is re-established.
(e) When a power supply failure occurs,
the delivery data shall be
memorized. The requirements of
paragraph 4, sub-paragraph (2),
clause (ii) of Heading-B shall apply.
(ii) Delayed payment
The printed and/or memorized Indications
as mentioned in paragraph 6 sub
paragraph (3) clause (i) of Headlng-B shall
contain sufficient information for further
checking and at least, the measured
quantity, the price to pay and information
to identify the particular transaction (e.g.,
the measuring system number, location,
date, time).
(iii) Pre-payment in unattended service mode
(a) Following the termination of each
delivery,the printed and/or memorized
lndi,:atlons as intended in paragraph
5, sub-paragraph (3), clause (i) of
Heading·B shall be made available,
dearly Indicatingthe amount which has
been pre-paid and the price
corresponding to the gas obtained.
These printed and/or memorized
Indications may be divided Into two
parts as follows:-
(a) one part provided prior to the
delivery on which the pre-paid
amount is shown and
recognizable as such,
(b) one part provided followingthe
termination of delivery, provided
that it is clear from the information
provided on both parts that they
are related to the same delivery.
(b) The requirements of paragraph 3
sub-paragraph (7) are applicable.
6. Markings and seal/ng
(1) Marking
(I) Each measuring system, component or
sub-system for which model approval has
been granted shall bear, placed together
legibly and indelibly either on the dial of
the indicating device or on a special data
plate, the following information:-
(a) model approval sign,
(b) manufacturer's identification mark or
trademark,
(c) designation selected by the
manufacturer, if appropriate,
(d) serial number and year of
manufacture,
(e) characteristics as defined In
paragraph 1, sub-paragraph (3),
clause (i) and paragraph 3, sub
paragraph (1), clause (i) of
Headinq-B.
(f) where relevant, the maximum
allowed speed for the sequential
control device (the tested one).
,
Note:The indicated characteristics
should be the actual characteristics
of use, if they are known when the
plate is affixed. When they are not
known, the indicated characteristics
are those allowed by the model
approval certificate.
However, the minimum and the
maximum temperatures of the gas
shall appear on the data plate only
when they d!ffor from -10 degree
Centigrade Md +40 degree
Centigrade respectively.
The minimum measured quantity of
the measuring system shall In all
cases be clearly visible on the dial
of any indicating device visible to the
user during the measurement.
When a measuring system can be
transported without being
dismantled, the markings required
for each component may also be
combined on a single plate.
['ll'T 11-? 3(i)] 631
(ii) Any information, markings or diagrams
spedfled shall be clearlyvisible on the dial of
the indicatingdevice or within proximityto it.
The markings on the dial of the Indicating
device of a meter forming a part of a
measuring system shall not contravene those
on the data plate of the measuring system.
(2) Sealing devices and stamping plate:
(i) General
Sealing Is preferably carried out by means
of lead seals. However, other types of
sealing are permitted on fragile Instruments
or when these seals provide sufficient
integrity, electronic seals for Instance.
The seals shall, in all cases, be easily
accessible.
·
Sealing should be provided on all parts of
the measuring system which cannot be
materially protected in any other way
against operations liable to affect the
measurement accuracy.
Sealing devices shall prohibit changing of
any parameter that participate In the
determination of measurement results
(parameters for correction and conversion
in partlculer).
A plate, referred to as the stamping plate,
aimed at receiving the control marks, shall
be sealed or permanently attached on a
support of the measuring system. It may
be combined with the data plate of the
measuring system referred to In paragraph
6, sub-paragraph (1) of Headlng-6.
(ii) Electronic sealing devices
(a) When access to parameters that
participate In the determination of results
of measurement is not protected by
mechanical sealing devices, the
protection shall fulfil the following
provlsions:-
(a-1) access shall only be allowed to
authorized people, by means of a special
device (hard key, etc.). Only a code is
not considered as fulfillingthis provision;
(a-2) It shall be possible for at least the
last hundred interventions to be
memorized; the record shall include the
date with time and characteristic
elements Identifying the authorized
person making the intervention [see (a)
above] and the value of the parameters
changed the traceability of these
interventions shall be assured for at least
two years, if it is not over-written on the
occasion of a further intervention; If
deletion of a previous Intervention must
occur to permit a new record, the oldest
record shall be deleted.
(b) For measuring systems with parts which
may be disconnected one from another by
the user and which are lnten:hangeable,
the following provisions shall be fuifilled:-
(b-I) it shall not be possible to access
parameters that participate in the
determination of results of
measurements through dlscon
nected points unless the provisions
in paragraph 6, sub paragraph (2),
clause (ii), sub clause (a) of
Heading-Bare fulfilled;
(b·2) interposing any device which may
influence the accuracy shall be
prevented by means of electronic
and data processing securities or, If
not possible, by mechanical means.
(c) For measuring systems with parts which may
be disconnected one from another by the
user and which are not interchangeable, the
provisions in paragraph 6, sub-paragraph
(2), clause (ii) sub-clause (b) of Headlng·B
apply. Moreover, these measuring systems
shall be provided with devices which do not
allow them to operate if the various parts
are not associated according to the
manufacturer's configuration.
Note: Disconnections which are not
allowed to the user may be prevented, for
example by means of a device that prevents
any measurement after disconnecting and
reconnecting.
7. Metrofogfcal control
When a test is conducted, the expanded
uncertainty on the determination of errors on
indications of mass shall be less than one-fifth of the
rnaxlmum permissible or tolerance applicable for that
test on model approval and one-third of the maximum
permissible error applicable for that test on other
verifications.
However, this provision may not be fulfilled for
tests at the minimum measured quantity or twice
this value.
Note: The expanded uncertainty includes
components of uncertainties that are in relation to
the instrument to be verified, in particular its scale
interval and, If applicable, the periodic variation.
However the repeatability error of the meter or device
to be verified shall not be included in the uncertainty.
--------------- -- ------- ---
J
_
632 THE GAZETTE OF INDIA: EXTRAORDINARY [PARTII-SEC. 3(i)]
(l)Model approval
(i) General
Measuring systems subject to legal
metrology control shall be subjected to
model approval.
The test aims at verifying that the measuring
system complies with the provisions of paragraph 2
sub-paragraph (1). clauses (ii) and (Ill) within the
field of operation. The test will be conducted as per
the procedure outlined in Annexure.
·