Annexure D, Legal Metrology (General) Rule, 2011: ANNEXURE D
Legal Metrology (General) Rule, 2011 - Annexure D
Text of annexure D
ANNEXURE D
ANNEXURE D
CLINICAL TEST OF RESPONSE TIME
(Mandatory)
1. Non-predicting clinical electrical
thermometers
The minimum time for achieving thermal
equilibrium at each appropriate body site shall be
determined on the basis of testing at least ten
persons.
2. Predicting (calculating) clinlcal electrical
thermometers
(1) The difference between the displayed
calculated temperature and the corresponding
measured temperature at thermal equilibrium of a
calculating (predicting) thermometer shall be
determined on the basis of testing at least 100
persons. The predicted temperature of each person
at an appropriate body site shall be determined by
the method specified by the manufacturer. After the
predicted Indication, the thermometer shall remain
at the site for measuring and indicating the actual
temperature of Its sensor. The total time allowed
shall be sufficient to attain thermal equlllbrlum. The
difference in the first and second Indicated
temperatures for 95% of the persons tested shall
not be more than 0.2°C.
(2) If an oral (subllngual) test has been carried
out, the minimum number of persons required for
rectal measurement shall be twenty.
PART VII
MANOMETERS OF INSTRUMENTS FOR
MEASURING ARTERIAL BLOOD PRESSURE
(SPHYGMOMANOMETARS)
PARTVII•A
NON-INVASIVE MECHANICAL
SPHYGMOMANOMETERS
1, Scope
This Part specifies general performance,
efficiency and mechanical and electrical safety
requirements, Including test methods for type
approval, for non-invasive mechanical
sphygmomanometers and their accessories which,
by means of an inflatable cuff, are used for the non·
invasive measurement of arterial blood pressure. The
application of the cuff is not limited to a particular
extremity of the human body (e.g. the upper arm).
Within the scope of this specification are
sphygmomanometers with a mechanical pressure
sensing element and display, used In conjunction with
a stethoscope or other manual methods for detecting
Korotkoff sounds and for cuff Inflation.
Note : Luer locks shall not be usedwith these
devices.
2. Terminology
(1) Bladder:
The Inflatable component of the cuff is called
bladder.
(2) 1'ressure in a blood vessel:
It refers to pressure In the arterial system of
the body.
(3) Cuff:
It is that component of the sphygmomanometer,
comprising a bladder and a sleeve, which is wrapped
around the limb of the patient.
(4) Diastolic blood pressure (value):
The minimum value of the arterial blood
pressure as a result of relaxation of the systemic
ventricle Is called diastolic blood pressure.
Note: Because of hydrostatic effects, this value
should be measured with the cuff at the heart level.
(S) Meon arterial blood pressure (value):
It Is the value of the integral of one cycle of the
blood pressure curve divided by the time of one heart
beat period.
Note: Because of hydrostatic effects, this value
should be measured with the cuff at the heart level.
(6) Non-invasive blood pressure measurement:
Indirect measurement of the arterial blood
pressure without arterial puncture is called non·
invasive blood pressure measurement.
-
__J_?--?
----
.
592 THE GAZETTE OF INDIA: EXTRAORDINARY [PARTII-SEC. 3(i)]
-··--··-·
.. ·--
t
(7) Pneumatic-system :
A system·that:·:includes .all pressurized and
pressure-controlling
..
parts .such :as ·cuff, ·tubing,
connectors, valves,' transducer .and .purnp is called
pneumatic system.
(8) Sleeve:
It is essentially· Inelastic part of the cuff that
encloses the bladder.
(9) Sphygmomanometer:
It is an Instrument .used for the non-Invasive
measurement oHhe arterial :blood ·pressure.
{10) Systolic-blood pressure (value):
The maximum ·value of the arterial blood
pressure as a result of the contraction of the systemic
ventricle is called.systolic blood pressure.
Note: Because of .hydrostatic effects, this value
should be meesuredwlth the cuff at the heart level.
(11) MechanlC'lilsphygmomanometer:
.
The. sphygmomanometer which uses either a
mercury or an aneroid manometer or another
mechanical measuring device for the non-Invasive
measurement of the arterial blood pressure by means
of an inflatable cuff, is called mechanical
sphygmomanometer.
(12) Auscultatory method:
It is that technique whereby sounds (known as
Korotkoff sounds) are heard over an occluded artery
as the occluding pressure is slowly released, the
appearance of sounds coinciding with the systolic
blood pressure and the disappearance of sounds with
the diastolic blood pressure in adults. In children
under age of 13, "k4" {I.e. 4th phase Korotkoff sound)
may be appropriate.
(13) Deflation valve :
The valve used for controller exhaust of the
pneumatic system during measurement.
(14) Rapid exhaust valve:
The valve used for rapidly exhausting the
pneumatic system.
(15) lamper proofing :
It is that means of preventing the user from
gaining easy access to the measuring· mechanism of
the device.
3. Description of the category of Instrument
The basic components of a sphygmomanometer
are a cuff and bladder that can be wrapped around
a patient's limb, a manual system for applying and
releasing pressure to the bladder, and a means of
measuring and displaying the Instantaneous pressure
in the bladder. Mechanical sphygmomanometers,
which use either mercury or an aneroid manometer
or another mechanical measuring device for the non
invasive measurement of the arterial blood pressure
by means of an inflatable cuff.
·
Notes : Components of these devices are
manometer, cuff, valve for deflation (often In
combination with rapid exhaust valve), hand pump
or electro-mechanical pump and connection hoses.
These devices may also contain electro-mechanical
components for pressure control.
4. Units of measurement
The blood pressure shall be indicated either In
kilo-Pascal (kPa) or in millimetres of mercury(mmHg).
s. Metrologlcal requirements
· (1) Maximum permissible errors of the cuff
pressure Indication
{I) Under ambient conditions
For any set ofconditions within the ambient
temperature range of 15°C to 25°C and the
relative humidity range of 20% to 85%,
both for increasing and for decreasing
pressure, the maximum permissible error
for the measurement of the cuff pressure
at any point of the scale range shall be ±
0.4 kPa (± 3 mmHg) In case of verifying
the first time and ± 0.5 kPa {± 4 mmHg)
for sphygmomanometers in use. Testing
shall be carried out In accordance with
paragraph 1 of Annexure A.
(II) Under storage conditions
The sphygmomanometer shall maintain the
maximum permissible error requirements
specified in this paragraph S(l){i) after
storage for 24 h at a temperature of -20°C
and for 24h at a temperature of 70°C and a
relative humidity of85% (non-condensing).
Testing shall be carried out in accordance
with paragraph 3 of Annexure A.
{Ill) Under varying temperature conditions
For the ambient temperature range of 10°C
to40°Cand the relative humidityof BS% (non
condensing), the difference of the cuff pressure
Indication of the sphygmomanometer shall not
exceed ± 0.4 kPa (± 3 mmHg). Testing shall
be carried out in accordance with paragraph
2 of Annexure A.
6. Technical requirements
(1) Technical requirements for the cuff and
bladder
The cuff shall contain a bladder. For
["lfll II-? 3(i)] 593
reusable cuffs the manufacturer shall
indicate the method for cleaning in the
accompanying documents.
·
Note : The optimum bladder size Is one with
dimensions such that its width is 40% of the limb
circumference at the midpoint of the cuff application,
and its length is at least 80%, preferably 100%, of
the limb circumference at the midpoint of cuff
application. Use of the wrong size can affect the
accuracy of the measurement.
(2) Technical requirements for the pneumatic
system
(i) ?Ir leakilge
Air leakage shall not exceed a pressure drop
of 0.5 kPa/min (4 mmHg/min).
Testing shall be carried out In accordance
with paragraph 4 of Annexure A.
(ii) Pressure reduction rate
Manually operated deflation valves shall be
capable of adjustment to a deflation rate
from 0.3 kPa/s to 0.4 kPa/s (2 mmHg/s to
3 mmHg/s). Manually operated deflation
valves shall be easily adjusted to these
values. Deflation valves shall be tested
in accordance with paragraph 5 of
Annexure A.
(iii) Rapid exhaust
During the rapid exhaust of the pneumatic
system, with the valve fully opened, the
time for the pressure reduction from 35 kPa
to 2 kPa (260 mmHg to 15 mmHg) shall
not exceed 1 Os.
Testing shall be carried out In accordance
with paragraph 6 of Annexure A.
(3) Technlcal requirements for the pressure
Indicating devices
·
(i) Nomlnal range and measuring range
I .
The nominal range shall be equal to the
measuring range.
_
The nominal range for the cuff gauge
pressure shall extend from O kPa to at least
35 kPa (O mmHg to at least 260 mm Hg).
(ii) Analogue indication
(a) Scale
The sea le shall be designed and arranged
so that the measuring values can be read
clearly and are easlly recognized. Testing
shall be carried out by visual Inspection.
(b) First scale mark
The graduation shall begin with the first
scale mark at O kPa (0 mmHg).
Testing shall be carried out by visual
inspection.
(c) Scale interval
The scale interval shall be-
0.2 kPa for a scale graduated in kPa;
or
2 mmHg for a scale graduated in
mmHg. Each fifth · scale mark shall
be indicated by greater length and
each tenth scale mark shall be
numbered. An example of a scale In
mmHg is given In Agure 1. Testing
shall be carried out by visual
inspection.
·
(d) Scale spacing and thickness of the scale
marks
The distance between adjacent scale marks
shall be not less than LO mm. The thickness
of the scale marks shall not exceed 20% of
the smallest scale spacing.
All scale marks shall be of equal thickness.
Testing shall be carried out in accordance
with paraqraph 7 of Annexure A.
(Figure 1 Exampleof an aneroid manometer scale division In mmHg without a tolerance zone at zero)
- -·
_L_
----?
594 THE GAZETTE OF INDIA: EXTRAORDINARY [PARTII-SEC. 3(i)]
(4) Additional technical requirements for
mercury manometers
(i) Internal diameter of the tube containing
mercury
The nominal internal diameter of the
mercury tube shall be at least 3.5 mm. The
tolerance on diameter shall not exceed ±
0.2mm.
Testing shall be carried out in accordance
with paragraph 8 of Annexure A.
(ii) Portable devices
A portable device shall be provided with
an adjusting or locking mechanism to
secure it In the specified position of use.
Testing shall be carried out by visual
inspectlon.
(iii) Devices to prevent mercury from being
spilled during use and transport
A device shall be placed in the tube to
prevent mercury from being spilled during
use and transport (for example: stopping
device, locking device, etc.). This device
shall be such that when the pressure in the
systemdrops rapidly from 27 kPa to O kPa
(from 200 mmHg to O mmHg), the time
taken for the mercury column to fall from
2? kPa to 5 kPa (from 200 mmHg to 40
mmHg) shall not exceed 1.5 s. This time is
known as the "exhaust time".
Testing shall be carried out In accordance
with paragraphs 9 and 10 of Annexure A.
(iv) Quality of the mercury
(a) The mercury shall have a purity of
not less than 99.99% according to
the declaration of the supplier of the
mercury.
(b) The mercury shall exhibit a clean
meniscus and shall not contain air
bubbles.
(v) Graduation of the mercury tube
Graduations shall be permanently marked
on the tube containing mercury.
If, numbered at each fifth scale mark, the
numbering shall be alternately on the right
and left-hand side of, and adjacent to, the
tube.
Testing shall be carried out by visual
'Inspection.
(5rAdditlonal technical requirements for
aneroid manometers
·
(I) Scale mark at zero
If a tolerance zone Is shown at zero it shall
notexceed ± 0.4 kPa (± 3 mmHg) and shall
be clearly marked.
A scale mark at zero shall be indicated.
Note : Graduations within the tolerance
zone are optional.
Testing shall be carried out by visual
Inspection.
(II) Zero
The movement of the elastic sensing
.
element including the pointer shall not be
obstructed within 0.8 kPa (6 mmHg) below
zero.
Neither the dial nor the pointer shall be
adjustable by the user. Testing shall be
carried out by visual Inspection.
(iii) Pointer
The pointer shall cover between 1/3 and
2/3 of the length of the shortest scale mark
of the scale. A.t the place of indication It
·
shall be not thicker than the scale mark.
The distance between the pointer and the
dial shall not exceed 2 mm. Testing shall
be carried out by visual Inspection.
(Iv) Hyster/sis error
The hysterisis error throughout the pressure
range shall be within the range O kPa to
0.5 kPII (0 mmHg to 4 mmHg). Testing shall
be carried out in accordance with paragraph
11 of Annexure A.
(v) Construction and materials
The construction of the aneroid manometer
and the material for the elastic sensing
elements shall ensure an adequate stability
of the measurement. The elastic sensing
elements shall be aged with respect to
pressure and temperature. After 10,000
alternating pressure cycles the change In
the pressure indication of the aneroid
manometer shall be not more than 0.4 kPa
(3 mmHg) throughout the pressure range.
Testing shall be carried out in accordance
with paragraph 12 of Annexure A.
(6) Safety reqmrements
(i) Resistance to vibration and shock
The sphygmomanometer shall comply
with the requirements of paragraph
S(l)(i).
.
['WT II-- 3(i)] 595
(ii) Mechanical safety
It shall be possible to abort the blood
pressure measurement at any time by
activating the manual rapid exhaust
valve, which shall be easily accessible.
(iii) Tamper proofing
Tamper proofing of the manometer shall
be achieved by requiring the use of a
tool or breaking a seal.
Testing shall be carried out by visual,
inspection.
7. Metrologlcal controls
(1) Model approval
At least three samples of a new type of
sphygmomanometer shall be tested. The tests to
verify conformity to metrological and technical
requirements shall be carried out according to
Annexure A.
(2) Verification
(i) Initial verification
At initial verification the requirements of
paragraphs S(l)(i), 6(2)(i) and 6(4)(1v)shall
be fulfilled.
Testing shall be carried out according to
paragraphs 1, 4 and 11 of Annexure A.
(ii) Subsequent verification
Each instrument of an approved type of
sphygmomanometer shall be verified every
2 years or after repair. Requirement of
paragraphs S(l)(i) and 6(2)(1) shall be
fulfilled and tests must be carried out
according to paragraph 1 of Annexure A.
(3) Seaiing
(i) Control marks will be put on lead seals for
which corresponding punched screws shall
be attached whenever necessary. These
seals shall prevent, without destruction of
the control marks:
in the case of mercury manometers: the
separation of reservoir and scale;
in the case of all other manometers: the
opening of the casing.
(ii) If the construction of the Instrument
guarantees security against any
interference, the metrological control marks
or the security marks may be attached In
form of labels.
(Iii) All seals shall be accessible without using
a tool.
(4) Marking of the ctevlce
The device shall be marked with the
following information:-
name and/or trademark of manufacturer;
serial number and year of fabrication;
measuring range and measuring unit;
model approval number;
centre of the bladder, Indicating the correct
position for the cuff over the artery; and
marking on the cuff indicating the limb
circumference for which it is appropriate In
paragraph 6(1).
· The following additional markings are
required for mercury manometers:--
lndlcation of the internal nominal diameter
and the tolerance of the tube containing
mercury in paragraph 6( 4 )(i).
(5) Manufacturer's information
(i) The manufacturer's Instruction manual shall
contain the following lnformation::-
(a) explanation of the operating
procedures which are Important for
correct application (such as the
selection of the appropriate cuff size,
positioning of the cuff. and
adjustment of the pressure reduction
rate);
(b) a warning to users of equipment
intended for use in environments
employing lntravascularfluid systems
not to connect the output of the
blood pressure measuring device to
such systems as
·
air might
inadvertently be pumped into a blood
vessel if, for example, Luer locks
were used;
(c) methods for cleaning reusable cuffs;
(d) nature and frequency of the
maintenance required to ensure that
the device operates correctly and
safely at all times; it Is recommended
that the performance should be
checked at least every 2 years and
after maintenance and repair, by re·
verifying at least the requirements
in paragraphs S(l)(i), 6(2)(i) (testing
596 'THE, GAZEl:TE iQF IND!l<\ : EXTRAORDINARY [PARTII-SEC. 3(i)]
c,,at leastat7 kPa (SOmmHg) and 27
'kPa (200 mmHg). and .paraqraph
:6{4}(1v);
{e) Internal· nominal diameter and
tolerance
..
of .the
.
tube 'Containing
mercury; and
(f) :' detailed· instructions .for the safe
··handeling of'mercu,;y.
,ANNEXUREA
TEST PROCEDURES: (Mandatory)
1, Method of test for.the maximum,permlsslble
errors of the cliff pressure'indicatlon
(1) ApparlJtuS
(i) rigid metal.vessel with.acapadty of 500 ml
±5 %;
(ii) calibrated •reference manometer with an
uncertainty less than 0,1 kPa (0.8 mmHg);
(iii) pressure .generator, e.g. ball pump (hand
pump ),with a deflation valve;
(iv) T-piece,connectors and hoses.
(2) Procetlure
Replace the cuff .with the vessel. Connect the
calibrated reference manometer by means of a
T-piece connector and hoses to the pneumatic system
!OOml
'(seeFiqure 2). After disabling the electro-mechanical
.pump (if fitted), connect the pressure generator into
the pressure system by means of another T-piece
.connectcr, Carry out the test In pressure steps of not
rnore than 7 kPa (SO mmHg) between a kPa (O
mmHg) and the maximum pressure of the scale
·range.
(3) Expression of results
Express the results as the differences between
the indicated pressure of the manometer of the device
to be tested and the corresponding readings of the
reference manometer.
2. Method of test for the Influence of
temperature on cuff pressure Indication
(1) Appsretus
(I) apparatus as specified In paragraph 1(1)
of this Annexure plus
(II} a climatic chamber.
(2) Procedure
Replace the cuff with the vessel. Connect the
calibrated reference manometer by means of a
T-plece connector to the pneumatic system (see
Figure 3). After disabling the electromechanical pump
(if fitted) connect the additional pressure generator
into the pneumatic system by means of another
T-piece connector.
1 ·
- Reference manometer
3 - Metal vessel
2 - Manometer of the device to be tested
4 - Pressure generator
Figure 2. MelJSurement system for determining the limits of error of the cuff pressure indiClltion
['1f'T II-'? 3(i)] 597
1 - Reference manometer
3 - Manometer of the device to be tested
2 - Climatic chamber
4 - Metal vessel
rigid metal cylinder of an appropriate size;
pressure generator, e.g. ball pump (hand
pump) with a deflation valve;
time measuring device.
(2) Procedure
Wrap the cuff around the cylinder.
Note : Electro-mechanical pumps which are
part of the device may be used for the test.
carry out the test over the whole measuring
range at least five equally spaced pressure steps
20°C ambient temperature, 85% relative
humidity (non-condensing);
40°C ambient temperature, 85% relative
humidity (non-condensing).
3. Method of test for the maximum permissible
error after storage ,
(1) Apparatus
• apparatus as specified in paragraph 1(1)
of this Annexure.
(2) Procedure
Replace the cuff with the vessel. Connect the
calibrated reference manometer by means of a
T-piece connector to the pneumatic system
Express the results as the differences between
the indicated pressure of the manometer of the device
to be tested and the corresponding readings of the
reference manometer at the relevant temperature
value.
carry out the test of the cuff pressure Indication
as described in paragraph 1(2) of this Annexure for
each of the combinations of temperature and
humidity mentioned above.
(3) Expression of results
5 - Pressure generator
Figure 3. Measurement system for determining the Influence of temperature
For each of the following combinations of (see Figure 3). After disabling the electro-mechanical
temperature and humidity, condition the device for pump (If fitted), connect the additional pressure
at least 3 h in the climatic chamber to allow the generator Into the pneumatic system by means of
device to reach steady conditions:- another T·plece connector.
10°c ambient temperature, 85% relative Store the instrument under test for 24 h at a
humidity (non-condensing); temperature of-20°C and subsequently for 24 hours
at a temperature of 70°C and a relative humidity of
85% (non-condensing).
Note: This is one test and not two separate
tests.
carry out the test in pressure steps of not more
than 7 kPa (50 mmHg) between O kPa (0 mmHg)
and the maximum pressure of the scale range.
(3) Expression of results
Express the results as the differences between
the indicated pressure of the manometer of the device
to be tested and the corresponding readings of the
reference manometer.
4. Method of test for air leakage of the
pneumatic system
(1) Apf)8ratus
_j
__ ?--
598 'THE·
..
GAZETTE OF JNDIA: EXTRAORDINARY [PARTII-SEC. 3(i)]
(e.g.) 7 kPa (50 mmHg), 13 kPa (lOO·mmHg),·20
kPa (150 mmHg);.27 'kPa (200 mmHg) and· 34 kPa
(250 mmHg). Test the air leakage over a. period of 5
min and determine-the measured vatue.rrom this.
(3) Expression of results
Express the air)eakage,as·the .rate of the
pressure Joss per minute.
5. Method of test .for pressure.reduction •rate
for deflation valves
{1}Apparatus
· T-piece'.connector;
· calibrated reference manometer with signal
output and an uncertainty less than 0.1 kPa
(0.8 mmHg);
· artificial limbs [see Notes-under paragraph
5(2) of this;Annexure].
· recording .unlt.
(2) Procedure
Measure the-pressure reduction rate either on
human limbs or ,artifrc:lal limbs.
Note 1 : 'The.intention is to use artificial limbs,
but as these are still under
consideration, measurements
performed with human volunteers
are acceptable.
Note 2 : It is Intended that the properties of
-
the artificial limbs reflect some
elastic properties of human limbs.
Because cuff deflation rate may be Influenced
by the way that a cuff Is applied, the cuff should be
applied and removed for each of at least ten repeated
measurements, on at least two different limb sizes.
These two limb sizes should be equal to the upper
and lower limits of limb circumferences for which a
particular size of cuff is recommended to be used. A
resetting of the deflation valve is permitted during
the test.
Connect the calibrated reference manometer
to the cuff by means of a T-piece connector. Connect
the output of the calibrated reference manometer to
the recording unit.
Plot the pressure reduction in the form of a
pressure curve as a function of time.
(3) Expression of results
Determine the rate of pressure reduction by
graphical evaluation (by drawing tangents) at the
pressure values of 8.0 kPa (60 mmHg), 16.0 kPa (120
mmHg) and 24.0 kPa (180 mmHg). The pressure
reduction rate is the mean value calculated separately
for these three pressure values and for the various
limb circumferences.
6. Method of test for the rapid exhaust
Naive
(1) Apparatus
rigid metal vessel, with a capacity of 500
ml±5 %;
calibrated reference manometer, with an
uncertainty less than 0.1 kPa (0.8 mmHg);
· 'r-piece connector;
· time measuring device.
(2) Procedure
Carry out the test with the vessel in place of
the cuff. Connect the calibrated reference manometer
by means of a T-piece to the pneumatic system.
Inflate to the maximum pressure and open the rapid
exhaust valve.
(3) EXpression of results
Measure the time between the pressure values
specified In paragraph 6(2)(111).
7. Method of test for the thickness of the scale
marks and the scale spacing
(1) Apparatus
· scaled magnifying lens or similar device.
(2) Procedure
Determine the thickness of the scale marks and
the scale spacing using the scaled magnifying lens.
8. Method of test for the internal diameter of
the mercury tube
(1) Apparatus
· limit plug gauges or similar devices, with a
tolerance less than 0.05 mm.
(2) Procedure
Test the nominal internal diameter of the tube
at each end by using the limit plug gauge.
9, Method oftest for security against mercury
losses
(1) Apparatus
collecting vessel of an adequate size;
calibrated reference manometer, with an
uncertainty less than 0.1 kPa (0.8 mmHg);
T -piece connector;
• pressure generator, e.g. ball pump (hand
pump) with a deflation valve.
(2) Procedure
Place the sphygmomanometer to be tested In
the collectingvessel. Connect the pressure generator
and a T-plece connector attached to a calibrated
reference manometer directly to the hose leading to
['WI II-? 3(i)] 599
the mercury reservoir. Use the pressure generator to
raise the pressure in the manometer to 13.3 kPa (100
mmHg) greater than the maximum indicated scale
reading on the test manometer. Maintain this pressure
for 5s and then release the pressure in the system.
Check that no mercury has spilled.
r
10. Method of test for the influence of the
mercury stopping device
{1] Apparatus
· time measuring device, e.g. a stopwatch
or an electronic timing device;
· pressure generator, e.g. ball pump (hand
pump) with a deflation valve.
(2) Procedure
Connect the pressure generator directly to the
hose leading to the mercury reservoir, I.e. without
connecting a cuff. When a gauge pressure of more
than 27 kPa (200 mmHg) has been reached, occlude
the tube and remove the pressure generator.
After removing the occlusion from the tube,
measure the time taken for the mercury column to
fall from the 27 kPa (200 mm Hg) mark to the 5 kPa
(40 mmHg) mark.
Check that the exhaust time does not exceed
1.5 s.
11. Method of test for the hysterlsls error of
the aneroid manometer
(1) Apparatus
(i) rigid metal vessel, with a capacity of 500
ml±So/o;
(ii) calibrated reference manometer, with an
uncertainty less than 0.1 kPa (0.8 mmHg);
(iii) pressure generator, e.g. ball pump (hand
pump) with a deflation valve;
(iv) T-piece connectors.
(2) Procedure
Replace the cuff with the vessel. Connect the
calibrated reference manometer by means of a
T-piece connector to the pneumatic system. After
disabling the electro-mechanical pump (If fitted)
connect the additional pressure generator Into the
pneumatic system by means of another T·piece
connecto?
·
Test the device with increasing pressure steps
of not more than 7 kPa (50 mmHg) to the scale
maximum, at which point hold the pressure for
5 minutes and then decrease it by_the same steps.
Disconnect the calibrated reference manometer
during the S minutes at maximum pressure.
(3) Expression of results·
Express the results as the difference between
the Indicated values on the manometer at the same
test pressure steps when increasing the pressure and
when decreasing the pressure.
12. Method of test for the construction
(1) Apparatus
alternating pressure generator, which
generates a sinusoidal pressure variation
between 3 kPa and 30 kPa (20 mmHg and
220 mm Hg) at a maximum rate of 60 cycles
per minute.
·
(2) Procedure
.
Carry out the procedure specified in paragraph
1 of this Annexure.
Connect the aneroid manometer directly to the
alternating pressure generator and perform 10,000
alternating pressure cycles.
One hour after the stress test carry out the
procedure as specified in paragraph 1 of this
Annexure at the same test pressure levels as before
the stress test.
(J) Expression of results
Express the results as the differences between
the indicated values on the manometer at the same
test pressure steps before and after the stress test.
PARTVII·B
NON•INVASIVE AUTOMATED
SPHYGMOMANOMETERS
1.Scope
This specification gives general performance,
efficiency and mechanical and electrical safety
requirements, including test methods for type
approval, for non-invasive electronic or automated
sphygmomanometers and their accessories which,
by means of an inflatable cuff; are used for the non·
invasive measurement of arterial blood pressure.
This specification only applies to devices
measuring at the upper arm, the wrist or the thigh.
Note : Luer locks sha II not be used with these
devices [see paragraphs 6(11)(iii) and 7(5)].
2. Terminology
(1) Bladder:
It Is the Inflatable component of the cuff.
(2J Pressure in a blood vessel :
It refers to pressure in the arterial system of
the body.
·
(3) Cuff:
It Is that component of the sphygmomanometer,
_J?
--- -
600 THE GAZE'JffE OF 1INDIA : EXTRAORDINARY [PARTII-SEC. 3(i)]
comprising a bladder and a sleeve, which is wrapped
around the limb.of the·patient.
(4) Diastolicbloodpressurefva/ue).:
The minimum value ·of the .arterlat blood
pressure as a result of relaxation: of\the -systernlc
ventricle.
Note: Because of hydrostatic effects,this value
should be measured. with the cuff at the.heart level.
(S) Mean stteristbkxd.pressure (value) :
It is the value.of..the· integral of.one cycle of the
blood pressure curv.edlvided:by.thetime·.of.oneheart
beat period.
Note: Because of:?ydrostatic effects, this value
should be measured withthe cuff at the heart level.
(6) Non-Invasive blood pressure measurement:
Indirect measurement of the arterial blood
pressure without ,arterial puncture is called non
lnvasiyeblood pressure ,measurement.
(7) Pneumatic,system c-
A system that tncludes all pressurized and
pressure-controlling parts such as cuff, tubing,
connectors, valves, transducer and pump Is called
pneumatic system.
(8) Sleeve:
It is essentially Inelastic part of the cuff that
encloses the bladder.
(9) Sphygmomanometer:
It is an instrument used for the non-invasive
measurement of the arterial blood pressure.
(10) Systolic blood pressure (value):
The maximum value of the arterial blood
pressure as a result of the contraction of the systemic
ventricle is called systolic blood pressure.
Note: Because of hydrostatic effects, this value
shou Id be measured with the cuff at the heart level.
(11) Electro-mechanical blood pressure
measuring system :
It is that system that consists of:
(i) at least one cuff, which is connected to the
pneumatic system;
(ii) at least one electro-mechanical transducer
to measure cuff pressure;
(iii) at least one measured value display; and
(iv) if needed, srgnal inputs and outputs.
(12) Electro-mechanical pressure transducer :
It is that component that transforms pressure
signals into electrical signals.
(13) Oscillometrlc method :
It is that method, wherein a cuff Is placed on
the limb and the pressure in the cuff Is Increased
until the blood flow in the artery is interrupted and
then the pressure in the cuff is slowly reduced.
Note: During the inflation and deflation of the
cuff small pressure changes (oscillations) occur In
the cuff as a result of the arterial blood pressure
pulses. These oscillations, which first increase and
then decrease, are detected and stored together with
the corresponding cuff pressure values in the
measurement system. With these stored values the
systolic, diastolic and mean arterial blood pressure
values can be mathematically derived using an
appropriate algorithm. It is possible to carry out the
measurement during the inflation phase.
(14) Zero setting :
The procedure that corrects a deviation of the
pressure reading to O kPa (0 mmHg) at atmospheric
pressure (gauge pressure: O kPa (0 mmHg)] is called
zero setting.
(15) Patient simulator :
Device for simulating the oscillometric cuff
pulses and/or auscultatory sounds during inflation
and deflation.
Note : This device is not used for testing
accuracy but is required in assessing stability of
performance.
'
(16) Auscultatory method :
It Is that technique whereby sounds (known as
Korotkoff sounds) are heard over an occluded artery
as the occluding pressure Is slowly released, the
appearance of sounds coinciding with the systolic
blood pressure and the disappearance of sounds with
the diastolic blood pressure. In children under the
age of 13, "k4" (i.e. 4th phase Korotkoff sound) may
be appropriate.
(17) Self-linearizing deflation valve :
The valve used for controlled linearizing exhaust
of the pneumatic system during measurement.
3. Description of the category of instrument
The basic components of a sphygmomanometer
are a cuff and bladder that can be wrapped around
a patient's limb, a system for applying and releasing
pressure to the bladder, and a means of measuring
and displaying the instantaneous pressure In the
bladder.
4, Units of measurement
The blood pressure shall be indicated either in
kilo-pascals (kPa) or in millimetres of mercury
(mmHg).
5, Metrologlcal requirements
(1) Maximum permissible errors of the cuff
pressure Indication
['WT n-= 3(i)J
For any set of conditions within the ambient
temperature range of 15°C to 25°C and the relative
humidity range of 20% to 85%, both for increasing
and for decreasing pressure, the maximum
permissible error for the measurement of the cuff
pressure at any point of the scale range shall be ±
0.4 kPa (± 3 mmHg) in case of verifying the first
time and ± 0.5 kPa ( +
·
4 mm Hg) for
sphygmomanometers in use.
Testing shall be carried out in accordance with
para 2 of Annexure A.
(2) Maximum permissible errors of the overall
system as measured by clinical tests"
The following maximum permissible errors shall
apply for the overall system:-
maximum mean error of measurement : ±
0.7 kPa ± 5 mmHg);
maximum experimental standard deviation
: 1.1 kPa (8 mm Hg).
(3) Environmental performance
(i) Storage:
Blood pressure measuring systems shall
maintain the requirements specified after
storage for 24 h at a temperature of -5°C
and for 24 hours at a temperature of S0°C
and a relative humidity of 85% (non
condensing).
Testing shall be carried out at
environmental conditions In accordance
with paragraph 2 of Annexure A after the
test sample has been placed for 24 h at a
temperature of -5°C and immediately
afterwards for 24 hours at a temperature
of 50°C In a climatic chamber.
Note : Integrated multi-parameter
monitors may contain components which
may be damaged during storage. The
general temperature range as stated in para
3 of Annexure A has therefore been
reduced.
(ii) Temperature, relative humidity:
For the ambient temperature range of 10°c
to 40°C and a relative humidity of 85%
(non-condensing), the difference of the cuff
pressure Indication of the sphyg
momanometer shall not exceed + 0.4 kPa
(±3 mmHg).
-
Testing shall be carried out in accordance with
paragraphs 2 and 11 of Annexure A.
The signal processing for the determination of
the blood pressure values shall not be Influenced
within the range of temperature and relative humidity.
.
For any set of conditions all the deviations between
l/7// Gif/-lS'/
... ·-···
---
601
the reference pressure and the indicating cuff
pressure of the instrument must be less than or equal
to the maximum permissible error.
6. Technical requirements
(1) General :
Equipment, or parts thereof, using materials or
having forms of construction different from those
detailed In this Recommendation shall be accepted
If it can be demonstrated that an equivalent degree
of safety and performance ls.obtained.
(2) Technical requirements for the cuff and
bladder:
The cuff shall contain a bladder. For reusable
cuffs the manufacturer shall indicate the method for
cleaning In the accompanying documents.
Note : The optimum bladder size is one with
dimensions such that its width is 40% of the limb
circumference at the midpoint of the cuff application
and its length is at least 80%, preferably 100% of
the limb circumference at the midpoint of cuff
application. Use of the wrong size can affect the
accuracy of the measurement.
.
(3) Technical requirements for the display:
The display shall be designed and arranged so
that the Information including measuring values can
be read and easily recognized.
Testing shall be carried out by visual inspection;
If abbreviations are used on the display they
shall be as follows:-
"S" or "SYS" : systolic blood pressure
(value);
"D" or "DIA" : diastolic .blood pressure
(value);
"M" or"MAP": mean arterial blood pressure
(value).
Single letter abbreviations shall be positioned
in such a way to avoid confusion with SI units.
(4) Effect of voltage variations of the power
source
(i) Internal electrical power source
(a) Changes of the voltage within the
working range determined according
to para 4( 1) of Annexure A shall not
influence the cuff pressure reading
and the result of the blood pressure
measurement.
(b) Outside this working range no cuff
pressure reading and no result of the
blood pressure measurement shall
be displayed .
,t
602 THE· .GAZETTE ,OF INDIA : EXTRAORDINARY [PARTII-SEC. 3(i)]
Testing shall .be .carried out in accordance with
paragraphs 4(1) and 5(1) of AnnexureA.
(ii) External-electrical power source
(a) ·changes of the voltage within the
, working range specified by the
, manufacturer shall not Influence the
:cuff pressure reading and the result
..
of.the blood pressure measurement.
(b) .Incorrect ·values resulting from
voltage variations outside the limits
·given In paragraph 6(4)(11)(a)shall
•not:be displayed.
Testing shall be carried out according to para
4(4) (alternating current) and para 4(5) (direct
current) of Annexure A.
Note: In the case of any malfunction of the
equipment, deflation to below 2 kPa (15 mmHg) must
be guaranteed within 180s in the case of adult
patients and to.below 0.7 kPa (5 mmHg) within 90s
in the case of neonatal/Infant patients.
(5) Pneumatic system
(i) Air leakage:
Air leakage shall not exceed a pressure drop
of 0.8 kPa/mln (6 mmHg/min),
Testing shall be carried out In accordance
with paragraph 6 of Annexure A.
(Ii) Pressure reducing system for devices using
the auscultatory method:
The pressure reducing system for manually
operated and automated deflation valves
shall be capable of maintaining a deflation
rate of 0.3 kPa/s to 0.4 kPa/s (2 mmHg/s
to 3 mmHg/s) within the target range of
systolic and diastolic blood pressure. For
devices which control .the pressure
reduction as a function of the pulse rate, a
deflation rate of 0.3 kPa/pulse to 0.4 kPa/
pulse (2 mmHg/pulse to 3 mmHg/pulse)
shall be maintained.
Note: Manually operated deflation valves
should be easily adjustable to these values. Testing
shall be carried out In accordance with paragraph 7
of Annexure 7.
(iii) Rapid exhaust:
During the rapid exhaust of the pneumatic
system, with the valve fully opened, the
time for the pressure reduction from 35 kPa
to 2 kPa (260 mmHg to 15 mmHg) shall
not exceed 10 seconds.
For blood pressure measuring systems
having the capability to measure in a
neonatal/in-fant mode, the time for the
pressure reduction from 20 kPa to 0.7 kPa
(150 mmHg to 5 mmHg) during the rapid
exhaust of the pneumatic system with the
valve fully opened shall not exceed Ss.
Testing shall be carried out In accordance with
paragraph 8 of Annexure A.
(Iv) Zero setting:
Blood pressure measuring systems shall be
capable of automatic zero setting. The zero
setting shall be carried out at appropriate
Intervals, at least starting after switching
on the device. At the moment of the zero
setting a gauge pressure ofO kPa (0 mmHg)
shall exist and be displayed thereafter.
Devices performing zero setting only
Immediatelyafter switching on, shall switch
off autornatically when the drift of the
pressure transducer and the analog signal
processing exceeds 0.1 kPa (1 mmHg).
Testing shall be carried out In accordance with
para 9 and paragraph 10 of Annexure A.
(6) Electromagnetic compatibility
Either:
electrical and/or electro-magnetic
Interferences shall not lead to degradations
In the cuff pressure Indication or in the
result of the blood pressure measurement;
or
If elettrical and/or electro-magnetic
Interferences lead to an abnormality, the
abnormality shall be clearly indicated and
It shall be possible to restore normal
operation within 30 s after cessation of the
electro-magnetic disturbance.
(7) Stability of the cuff pressure indication
The change in the cuff pressure indication shall
not" be more than 0.4 kPa (3 mm Hg) throughout the
pressure range after 10,000 simulated measurement
cycles.
Testing shall ·be carried out in accordance with
paragraph 12 of Annexure A.
(8) Pressure indicating device
(i) Nominal range and measuring range:
The nominal range for the cuff pressure
measurement shall be specified by the
manufacturer. The measuring and indication
ranges of the cuff pressure shall be equal
to the nominal range. Values _of blood
[ 'fT'T Jl-<911'6 3( i)]
-
.:,,_-
,, ... , --
603
pressure measurement results outside the
nominal range of cuff pressure shall be
dearly indicated as out of range.
Testing shall be carried out by visual inspection.
(ii) Digital indication :
The digital scale interval shall be 0.1 kPa
(1 mmHg). If the measured value of a
parameter is to be indicated on more than
one display, all the displays shall indicate
the same numerical value. Measured
numerical values on the display(s), and the
symbols defining the units of measurement
shall be arranged in such a way so as to
avoid misinterpretation. Numbers and
characters should be clearly legible.
Testing shall be carried out by visual Inspection.
(9) Signal input and output ports
The construction of the signal input and output
ports (excluding Internal interfaces, e.g. microphone
signal input) relevant to the non-invasive blood
pressure measurement shall ensure that Incorrectly
fitted or defective accessories shall not result in
erroneous indication of cuff pressure or erroneous
indication of blood pressure.
Testing shall be carried out in accordance with
paragraph 13 of Annexure A.
(10) Alarms
If alarms are used they shall be of at least
medium priority.
(11) Safety
(i) Cuff pressure:
It shall be possible to abort any blood
·
pressure measurement at any time by
single key operation and this shall lead to
a rapid exhaust.
Testing shall be carried out in accordance with
paragraph 14 of Annexure A.
(ii) Unauthorlred access:
All controls which affect accuracy shall be
sealed against unauthorized access.
Testing shall be carried out by visual Inspection.
(iii) Tubingconnectors:
Users of equipment intended for use in
environments employing intravascular fluid
systems shall take all necessary precautions
to avoid connecting-the output of the blood
pressure measuring device to such systems
as air might inadvertently be pumped Into
a blood vessel if, for example, Luer locks
were used.
{Iv) E/ectrieal safety:
Electronic or automated sphygmomano·
meters shall comply with the relevant
natlona I safety regulations.
(v) Resistance to vibration and shock:
After testing, the device shall comply with
the requirements of paragraph 5(1).
7. Metrologlcal controls
(1) Model approval
At least three samples of a new type of
sphygmomanometer shall be tested.
The tests to verify conformity to metrologlcal
and technical requirements shall be carried out
according to Annexure A.
(2) Verifieation
(I) Inltlal verification
At Initial verification the requirements of
paragraphs 5(1) and 6(5)(i) shall be
fulfilled.
·
Testing shall be carried out according to
paraqraphs 2 and 6 of Annexure A.
(II) Subsequent verification
Each instrument of an approved type of
sphygmomanometer shall be verified every
2 years or after repair. Requirement of
paragraphs 5(1) and 6(5)(1)shall be fulfilled
and tests must be carried out according to
paragraphs 2 and 6 of Annexure A.
(3) Sealing
(I) Control marks will be put on lead seals for
which corresponding punched screws shall
be attached whenever necessary. These
seals shall prevent, without destruction of
the control marks:
(a) In the case of patient-monitors In
which the sphygmomanometer Is
one pa rt of a system: the
manipulation of the metrologlcally
relevant parts for measuring blood
pressure;
(b) in the case of all other manometers:
the opening of the' casing.
604 THE .GAZETifE .OF 'INDIA : EXTRAORDINARY [PARTII-SEc. 3(i)]
(ii) If tnecconstructlon of the .lnstrument
guarantees .securttv against any
interference; the metrological control marks
or the.security.marks may be·attached in
the form ciflabels.
(iii) All seals shalt be accessible without using
a tool.
(4) MarkingoNhedevice
The devke.shall'be.rnarked with the following
information:-
(i) namerend/cr-tracemerk of manufacturer;
(ii) seriaL number :and year of fabrication;
(iii) measuring :range and measuring unit;
(iv) model approval number;
(v) centre:ofthe bladder. indicating the correct
position .forthe cuff over the artery; and
(vi) marking om the cuff indicating ttie limb
drcumference for which it Is appropriate.
(5) Manufacturer's information
(i) The manufacturer's instruction manual shall
contain the following information:
(a) explanation of the operating
procedures which are important for
correct application (such as the
selection of the appropriate cuff size,
positioning of the cuff and
adjustment of the pressure reduction
rate);
(b) a warning to users of equipment
Intended for use in environments
employing intravascular fluid systems
not to connect the output of the
blood pressure measuring device to
.
such systems as air might
Inadvertently be pumped into a blood
vessel if, for example, Luer locks
were used;
(c) methods for cleaning reusable cuffs;
(d) nature and frequency of the
maintenance to ensure that the
device operates properly and safely
at all times; it is recommended that
the performance should be checked
at least every 2 years and after
maintenance and repair, by re
verifying at least the requirements
in paragraphs 5(1) and 6(5)(1)
(testing at \east at 7 kPa (SOmmHg)
and 27 kPa (200 mmHg));
(e) a reference method for clinical tests
carried out according to Annexure C
·
or an equivalent method;
(f) a list of all components belonging to
the pressure measuring systems,
including
(g) a description of the operating
principles of the blood pressure
measuring device;
(h) remarks on the environment or
operational factors which affect the
performance ( e.g. electro- magnetic
fields, arrhythmia);
(I) specification of the single input/
output ports(s);
(J) specification of the rated voltage, If
applicable;
(k) specification of the intended power
source, if applicable;
(I) nominal range for the result of the
blood pressure measurement;
(m) warm up time, If applicable;
(n) description of the meaning .of the
"out of range signal" and
(o) description of the alarms, If
applicable.