Annexure D, Legal Metrology (General) Rule, 2011: ANNEXURE D

Legal Metrology (General) Rule, 2011 - Annexure D

Text of annexure D

ANNEXURE D

ANNEXURE D CLINICAL TEST OF RESPONSE TIME (Mandatory) 1. Non-predicting clinical electrical thermometers The minimum time for achieving thermal equilibrium at each appropriate body site shall be determined on the basis of testing at least ten persons. 2. Predicting (calculating) clinlcal electrical thermometers (1) The difference between the displayed calculated temperature and the corresponding measured temperature at thermal equilibrium of a calculating (predicting) thermometer shall be determined on the basis of testing at least 100 persons. The predicted temperature of each person at an appropriate body site shall be determined by the method specified by the manufacturer. After the predicted Indication, the thermometer shall remain at the site for measuring and indicating the actual temperature of Its sensor. The total time allowed shall be sufficient to attain thermal equlllbrlum. The difference in the first and second Indicated temperatures for 95% of the persons tested shall not be more than 0.2°C. (2) If an oral (subllngual) test has been carried out, the minimum number of persons required for rectal measurement shall be twenty. PART VII MANOMETERS OF INSTRUMENTS FOR MEASURING ARTERIAL BLOOD PRESSURE (SPHYGMOMANOMETARS) PARTVII•A NON-INVASIVE MECHANICAL SPHYGMOMANOMETERS 1, Scope This Part specifies general performance, efficiency and mechanical and electrical safety requirements, Including test methods for type approval, for non-invasive mechanical sphygmomanometers and their accessories which, by means of an inflatable cuff, are used for the non· invasive measurement of arterial blood pressure. The application of the cuff is not limited to a particular extremity of the human body (e.g. the upper arm). Within the scope of this specification are sphygmomanometers with a mechanical pressure sensing element and display, used In conjunction with a stethoscope or other manual methods for detecting Korotkoff sounds and for cuff Inflation. Note : Luer locks shall not be usedwith these devices. 2. Terminology (1) Bladder: The Inflatable component of the cuff is called bladder. (2) 1'ressure in a blood vessel: It refers to pressure In the arterial system of the body. (3) Cuff: It is that component of the sphygmomanometer, comprising a bladder and a sleeve, which is wrapped around the limb of the patient. (4) Diastolic blood pressure (value): The minimum value of the arterial blood pressure as a result of relaxation of the systemic ventricle Is called diastolic blood pressure. Note: Because of hydrostatic effects, this value should be measured with the cuff at the heart level. (S) Meon arterial blood pressure (value): It Is the value of the integral of one cycle of the blood pressure curve divided by the time of one heart beat period. Note: Because of hydrostatic effects, this value should be measured with the cuff at the heart level. (6) Non-invasive blood pressure measurement: Indirect measurement of the arterial blood pressure without arterial puncture is called non· invasive blood pressure measurement. - __J_?--? ---- . 592 THE GAZETTE OF INDIA: EXTRAORDINARY [PARTII-SEC. 3(i)] -··--··-· .. ·-- t (7) Pneumatic-system : A system·that:·:includes .all pressurized and pressure-controlling .. parts .such :as ·cuff, ·tubing, connectors, valves,' transducer .and .purnp is called pneumatic system. (8) Sleeve: It is essentially· Inelastic part of the cuff that encloses the bladder. (9) Sphygmomanometer: It is an Instrument .used for the non-Invasive measurement oHhe arterial :blood ·pressure. {10) Systolic-blood pressure (value): The maximum ·value of the arterial blood pressure as a result of the contraction of the systemic ventricle is called.systolic blood pressure. Note: Because of .hydrostatic effects, this value should be meesuredwlth the cuff at the heart level. (11) MechanlC'lilsphygmomanometer: . The. sphygmomanometer which uses either a mercury or an aneroid manometer or another mechanical measuring device for the non-Invasive measurement of the arterial blood pressure by means of an inflatable cuff, is called mechanical sphygmomanometer. (12) Auscultatory method: It is that technique whereby sounds (known as Korotkoff sounds) are heard over an occluded artery as the occluding pressure is slowly released, the appearance of sounds coinciding with the systolic blood pressure and the disappearance of sounds with the diastolic blood pressure in adults. In children under age of 13, "k4" {I.e. 4th phase Korotkoff sound) may be appropriate. (13) Deflation valve : The valve used for controller exhaust of the pneumatic system during measurement. (14) Rapid exhaust valve: The valve used for rapidly exhausting the pneumatic system. (15) lamper proofing : It is that means of preventing the user from gaining easy access to the measuring· mechanism of the device. 3. Description of the category of Instrument The basic components of a sphygmomanometer are a cuff and bladder that can be wrapped around a patient's limb, a manual system for applying and releasing pressure to the bladder, and a means of measuring and displaying the Instantaneous pressure in the bladder. Mechanical sphygmomanometers, which use either mercury or an aneroid manometer or another mechanical measuring device for the non­ invasive measurement of the arterial blood pressure by means of an inflatable cuff. · Notes : Components of these devices are manometer, cuff, valve for deflation (often In combination with rapid exhaust valve), hand pump or electro-mechanical pump and connection hoses. These devices may also contain electro-mechanical components for pressure control. 4. Units of measurement The blood pressure shall be indicated either In kilo-Pascal (kPa) or in millimetres of mercury(mmHg). s. Metrologlcal requirements · (1) Maximum permissible errors of the cuff pressure Indication {I) Under ambient conditions For any set ofconditions within the ambient temperature range of 15°C to 25°C and the relative humidity range of 20% to 85%, both for increasing and for decreasing pressure, the maximum permissible error for the measurement of the cuff pressure at any point of the scale range shall be ± 0.4 kPa (± 3 mmHg) In case of verifying the first time and ± 0.5 kPa {± 4 mmHg) for sphygmomanometers in use. Testing shall be carried out In accordance with paragraph 1 of Annexure A. (II) Under storage conditions The sphygmomanometer shall maintain the maximum permissible error requirements specified in this paragraph S(l){i) after storage for 24 h at a temperature of -20°C and for 24h at a temperature of 70°C and a relative humidity of85% (non-condensing). Testing shall be carried out in accordance with paragraph 3 of Annexure A. {Ill) Under varying temperature conditions For the ambient temperature range of 10°C to40°Cand the relative humidityof BS% (non­ condensing), the difference of the cuff pressure Indication of the sphygmomanometer shall not exceed ± 0.4 kPa (± 3 mmHg). Testing shall be carried out in accordance with paragraph 2 of Annexure A. 6. Technical requirements (1) Technical requirements for the cuff and bladder The cuff shall contain a bladder. For ["lfll II-? 3(i)] 593 reusable cuffs the manufacturer shall indicate the method for cleaning in the accompanying documents. · Note : The optimum bladder size Is one with dimensions such that its width is 40% of the limb circumference at the midpoint of the cuff application, and its length is at least 80%, preferably 100%, of the limb circumference at the midpoint of cuff application. Use of the wrong size can affect the accuracy of the measurement. (2) Technical requirements for the pneumatic system (i) ?Ir leakilge Air leakage shall not exceed a pressure drop of 0.5 kPa/min (4 mmHg/min). Testing shall be carried out In accordance with paragraph 4 of Annexure A. (ii) Pressure reduction rate Manually operated deflation valves shall be capable of adjustment to a deflation rate from 0.3 kPa/s to 0.4 kPa/s (2 mmHg/s to 3 mmHg/s). Manually operated deflation valves shall be easily adjusted to these values. Deflation valves shall be tested in accordance with paragraph 5 of Annexure A. (iii) Rapid exhaust During the rapid exhaust of the pneumatic system, with the valve fully opened, the time for the pressure reduction from 35 kPa to 2 kPa (260 mmHg to 15 mmHg) shall not exceed 1 Os. Testing shall be carried out In accordance with paragraph 6 of Annexure A. (3) Technlcal requirements for the pressure Indicating devices · (i) Nomlnal range and measuring range I . The nominal range shall be equal to the measuring range. _ The nominal range for the cuff gauge pressure shall extend from O kPa to at least 35 kPa (O mmHg to at least 260 mm Hg). (ii) Analogue indication (a) Scale The sea le shall be designed and arranged so that the measuring values can be read clearly and are easlly recognized. Testing shall be carried out by visual Inspection. (b) First scale mark The graduation shall begin with the first scale mark at O kPa (0 mmHg). Testing shall be carried out by visual inspection. (c) Scale interval The scale interval shall be- 0.2 kPa for a scale graduated in kPa; or 2 mmHg for a scale graduated in mmHg. Each fifth · scale mark shall be indicated by greater length and each tenth scale mark shall be numbered. An example of a scale In mmHg is given In Agure 1. Testing shall be carried out by visual inspection. · (d) Scale spacing and thickness of the scale marks The distance between adjacent scale marks shall be not less than LO mm. The thickness of the scale marks shall not exceed 20% of the smallest scale spacing. All scale marks shall be of equal thickness. Testing shall be carried out in accordance with paraqraph 7 of Annexure A. (Figure 1 Exampleof an aneroid manometer scale division In mmHg without a tolerance zone at zero) - -· _L_ ----? 594 THE GAZETTE OF INDIA: EXTRAORDINARY [PARTII-SEC. 3(i)] (4) Additional technical requirements for mercury manometers (i) Internal diameter of the tube containing mercury The nominal internal diameter of the mercury tube shall be at least 3.5 mm. The tolerance on diameter shall not exceed ± 0.2mm. Testing shall be carried out in accordance with paragraph 8 of Annexure A. (ii) Portable devices A portable device shall be provided with an adjusting or locking mechanism to secure it In the specified position of use. Testing shall be carried out by visual inspectlon. (iii) Devices to prevent mercury from being spilled during use and transport A device shall be placed in the tube to prevent mercury from being spilled during use and transport (for example: stopping device, locking device, etc.). This device shall be such that when the pressure in the systemdrops rapidly from 27 kPa to O kPa (from 200 mmHg to O mmHg), the time taken for the mercury column to fall from 2? kPa to 5 kPa (from 200 mmHg to 40 mmHg) shall not exceed 1.5 s. This time is known as the "exhaust time". Testing shall be carried out In accordance with paragraphs 9 and 10 of Annexure A. (iv) Quality of the mercury (a) The mercury shall have a purity of not less than 99.99% according to the declaration of the supplier of the mercury. (b) The mercury shall exhibit a clean meniscus and shall not contain air bubbles. (v) Graduation of the mercury tube Graduations shall be permanently marked on the tube containing mercury. If, numbered at each fifth scale mark, the numbering shall be alternately on the right and left-hand side of, and adjacent to, the tube. Testing shall be carried out by visual 'Inspection. (5rAdditlonal technical requirements for aneroid manometers · (I) Scale mark at zero If a tolerance zone Is shown at zero it shall notexceed ± 0.4 kPa (± 3 mmHg) and shall be clearly marked. A scale mark at zero shall be indicated. Note : Graduations within the tolerance zone are optional. Testing shall be carried out by visual Inspection. (II) Zero The movement of the elastic sensing . element including the pointer shall not be obstructed within 0.8 kPa (6 mmHg) below zero. Neither the dial nor the pointer shall be adjustable by the user. Testing shall be carried out by visual Inspection. (iii) Pointer The pointer shall cover between 1/3 and 2/3 of the length of the shortest scale mark of the scale. A.t the place of indication It · shall be not thicker than the scale mark. The distance between the pointer and the dial shall not exceed 2 mm. Testing shall be carried out by visual Inspection. (Iv) Hyster/sis error The hysterisis error throughout the pressure range shall be within the range O kPa to 0.5 kPII (0 mmHg to 4 mmHg). Testing shall be carried out in accordance with paragraph 11 of Annexure A. (v) Construction and materials The construction of the aneroid manometer and the material for the elastic sensing elements shall ensure an adequate stability of the measurement. The elastic sensing elements shall be aged with respect to pressure and temperature. After 10,000 alternating pressure cycles the change In the pressure indication of the aneroid manometer shall be not more than 0.4 kPa (3 mmHg) throughout the pressure range. Testing shall be carried out in accordance with paragraph 12 of Annexure A. (6) Safety reqmrements (i) Resistance to vibration and shock The sphygmomanometer shall comply with the requirements of paragraph S(l)(i). . ['WT II-- 3(i)] 595 (ii) Mechanical safety It shall be possible to abort the blood pressure measurement at any time by activating the manual rapid exhaust valve, which shall be easily accessible. (iii) Tamper proofing Tamper proofing of the manometer shall be achieved by requiring the use of a tool or breaking a seal. Testing shall be carried out by visual, inspection. 7. Metrologlcal controls (1) Model approval At least three samples of a new type of sphygmomanometer shall be tested. The tests to verify conformity to metrological and technical requirements shall be carried out according to Annexure A. (2) Verification (i) Initial verification At initial verification the requirements of paragraphs S(l)(i), 6(2)(i) and 6(4)(1v)shall be fulfilled. Testing shall be carried out according to paragraphs 1, 4 and 11 of Annexure A. (ii) Subsequent verification Each instrument of an approved type of sphygmomanometer shall be verified every 2 years or after repair. Requirement of paragraphs S(l)(i) and 6(2)(1) shall be fulfilled and tests must be carried out according to paragraph 1 of Annexure A. (3) Seaiing (i) Control marks will be put on lead seals for which corresponding punched screws shall be attached whenever necessary. These seals shall prevent, without destruction of the control marks: in the case of mercury manometers: the separation of reservoir and scale; in the case of all other manometers: the opening of the casing. (ii) If the construction of the Instrument guarantees security against any interference, the metrological control marks or the security marks may be attached In form of labels. (Iii) All seals shall be accessible without using a tool. (4) Marking of the ctevlce The device shall be marked with the following information:- name and/or trademark of manufacturer; serial number and year of fabrication; measuring range and measuring unit; model approval number; centre of the bladder, Indicating the correct position for the cuff over the artery; and marking on the cuff indicating the limb circumference for which it is appropriate In paragraph 6(1). · The following additional markings are required for mercury manometers:-- lndlcation of the internal nominal diameter and the tolerance of the tube containing mercury in paragraph 6( 4 )(i). (5) Manufacturer's information (i) The manufacturer's Instruction manual shall contain the following lnformation::- (a) explanation of the operating procedures which are Important for correct application (such as the selection of the appropriate cuff size, positioning of the cuff. and adjustment of the pressure reduction rate); (b) a warning to users of equipment intended for use in environments employing lntravascularfluid systems not to connect the output of the blood pressure measuring device to such systems as · air might inadvertently be pumped into a blood vessel if, for example, Luer locks were used; (c) methods for cleaning reusable cuffs; (d) nature and frequency of the maintenance required to ensure that the device operates correctly and safely at all times; it Is recommended that the performance should be checked at least every 2 years and after maintenance and repair, by re· verifying at least the requirements in paragraphs S(l)(i), 6(2)(i) (testing 596 'THE, GAZEl:TE iQF IND!l<\ : EXTRAORDINARY [PARTII-SEC. 3(i)] c,,at leastat7 kPa (SOmmHg) and 27 'kPa (200 mmHg). and .paraqraph :6{4}(1v); {e) Internal· nominal diameter and tolerance .. of .the . tube 'Containing mercury; and (f) :' detailed· instructions .for the safe ··handeling of'mercu,;y. ,ANNEXUREA TEST PROCEDURES: (Mandatory) 1, Method of test for.the maximum,permlsslble errors of the cliff pressure'indicatlon (1) ApparlJtuS (i) rigid metal.vessel with.acapadty of 500 ml ±5 %; (ii) calibrated •reference manometer with an uncertainty less than 0,1 kPa (0.8 mmHg); (iii) pressure .generator, e.g. ball pump (hand pump ),with a deflation valve; (iv) T-piece,connectors and hoses. (2) Procetlure Replace the cuff .with the vessel. Connect the calibrated reference manometer by means of a T-piece connector and hoses to the pneumatic system !OOml '(seeFiqure 2). After disabling the electro-mechanical .pump (if fitted), connect the pressure generator into the pressure system by means of another T-piece .connectcr, Carry out the test In pressure steps of not rnore than 7 kPa (SO mmHg) between a kPa (O mmHg) and the maximum pressure of the scale ·range. (3) Expression of results Express the results as the differences between the indicated pressure of the manometer of the device to be tested and the corresponding readings of the reference manometer. 2. Method of test for the Influence of temperature on cuff pressure Indication (1) Appsretus (I) apparatus as specified In paragraph 1(1) of this Annexure plus (II} a climatic chamber. (2) Procedure Replace the cuff with the vessel. Connect the calibrated reference manometer by means of a T-plece connector to the pneumatic system (see Figure 3). After disabling the electromechanical pump (if fitted) connect the additional pressure generator into the pneumatic system by means of another T-piece connector. 1 · - Reference manometer 3 - Metal vessel 2 - Manometer of the device to be tested 4 - Pressure generator Figure 2. MelJSurement system for determining the limits of error of the cuff pressure indiClltion ['1f'T II-'? 3(i)] 597 1 - Reference manometer 3 - Manometer of the device to be tested 2 - Climatic chamber 4 - Metal vessel rigid metal cylinder of an appropriate size; pressure generator, e.g. ball pump (hand pump) with a deflation valve; time measuring device. (2) Procedure Wrap the cuff around the cylinder. Note : Electro-mechanical pumps which are part of the device may be used for the test. carry out the test over the whole measuring range at least five equally spaced pressure steps 20°C ambient temperature, 85% relative humidity (non-condensing); 40°C ambient temperature, 85% relative humidity (non-condensing). 3. Method of test for the maximum permissible error after storage , (1) Apparatus • apparatus as specified in paragraph 1(1) of this Annexure. (2) Procedure Replace the cuff with the vessel. Connect the calibrated reference manometer by means of a T-piece connector to the pneumatic system Express the results as the differences between the indicated pressure of the manometer of the device to be tested and the corresponding readings of the reference manometer at the relevant temperature value. carry out the test of the cuff pressure Indication as described in paragraph 1(2) of this Annexure for each of the combinations of temperature and humidity mentioned above. (3) Expression of results 5 - Pressure generator Figure 3. Measurement system for determining the Influence of temperature For each of the following combinations of (see Figure 3). After disabling the electro-mechanical temperature and humidity, condition the device for pump (If fitted), connect the additional pressure at least 3 h in the climatic chamber to allow the generator Into the pneumatic system by means of device to reach steady conditions:- another T·plece connector. 10°c ambient temperature, 85% relative Store the instrument under test for 24 h at a humidity (non-condensing); temperature of-20°C and subsequently for 24 hours at a temperature of 70°C and a relative humidity of 85% (non-condensing). Note: This is one test and not two separate tests. carry out the test in pressure steps of not more than 7 kPa (50 mmHg) between O kPa (0 mmHg) and the maximum pressure of the scale range. (3) Expression of results Express the results as the differences between the indicated pressure of the manometer of the device to be tested and the corresponding readings of the reference manometer. 4. Method of test for air leakage of the pneumatic system (1) Apf)8ratus _j __ ?-- 598 'THE· .. GAZETTE OF JNDIA: EXTRAORDINARY [PARTII-SEC. 3(i)] (e.g.) 7 kPa (50 mmHg), 13 kPa (lOO·mmHg),·20 kPa (150 mmHg);.27 'kPa (200 mmHg) and· 34 kPa (250 mmHg). Test the air leakage over a. period of 5 min and determine-the measured vatue.rrom this. (3) Expression of results Express the air)eakage,as·the .rate of the pressure Joss per minute. 5. Method of test .for pressure.reduction •rate for deflation valves {1}Apparatus · T-piece'.connector; · calibrated reference manometer with signal output and an uncertainty less than 0.1 kPa (0.8 mmHg); · artificial limbs [see Notes-under paragraph 5(2) of this;Annexure]. · recording .unlt. (2) Procedure Measure the-pressure reduction rate either on human limbs or ,artifrc:lal limbs. Note 1 : 'The.intention is to use artificial limbs, but as these are still under consideration, measurements performed with human volunteers are acceptable. Note 2 : It is Intended that the properties of - the artificial limbs reflect some elastic properties of human limbs. Because cuff deflation rate may be Influenced by the way that a cuff Is applied, the cuff should be applied and removed for each of at least ten repeated measurements, on at least two different limb sizes. These two limb sizes should be equal to the upper and lower limits of limb circumferences for which a particular size of cuff is recommended to be used. A resetting of the deflation valve is permitted during the test. Connect the calibrated reference manometer to the cuff by means of a T-piece connector. Connect the output of the calibrated reference manometer to the recording unit. Plot the pressure reduction in the form of a pressure curve as a function of time. (3) Expression of results Determine the rate of pressure reduction by graphical evaluation (by drawing tangents) at the pressure values of 8.0 kPa (60 mmHg), 16.0 kPa (120 mmHg) and 24.0 kPa (180 mmHg). The pressure reduction rate is the mean value calculated separately for these three pressure values and for the various limb circumferences. 6. Method of test for the rapid exhaust Naive (1) Apparatus rigid metal vessel, with a capacity of 500 ml±5 %; calibrated reference manometer, with an uncertainty less than 0.1 kPa (0.8 mmHg); · 'r-piece connector; · time measuring device. (2) Procedure Carry out the test with the vessel in place of the cuff. Connect the calibrated reference manometer by means of a T-piece to the pneumatic system. Inflate to the maximum pressure and open the rapid exhaust valve. (3) EXpression of results Measure the time between the pressure values specified In paragraph 6(2)(111). 7. Method of test for the thickness of the scale marks and the scale spacing (1) Apparatus · scaled magnifying lens or similar device. (2) Procedure Determine the thickness of the scale marks and the scale spacing using the scaled magnifying lens. 8. Method of test for the internal diameter of the mercury tube (1) Apparatus · limit plug gauges or similar devices, with a tolerance less than 0.05 mm. (2) Procedure Test the nominal internal diameter of the tube at each end by using the limit plug gauge. 9, Method oftest for security against mercury losses (1) Apparatus collecting vessel of an adequate size; calibrated reference manometer, with an uncertainty less than 0.1 kPa (0.8 mmHg); T -piece connector; • pressure generator, e.g. ball pump (hand pump) with a deflation valve. (2) Procedure Place the sphygmomanometer to be tested In the collectingvessel. Connect the pressure generator and a T-plece connector attached to a calibrated reference manometer directly to the hose leading to ['WI II-? 3(i)] 599 the mercury reservoir. Use the pressure generator to raise the pressure in the manometer to 13.3 kPa (100 mmHg) greater than the maximum indicated scale reading on the test manometer. Maintain this pressure for 5s and then release the pressure in the system. Check that no mercury has spilled. r 10. Method of test for the influence of the mercury stopping device {1] Apparatus · time measuring device, e.g. a stopwatch or an electronic timing device; · pressure generator, e.g. ball pump (hand pump) with a deflation valve. (2) Procedure Connect the pressure generator directly to the hose leading to the mercury reservoir, I.e. without connecting a cuff. When a gauge pressure of more than 27 kPa (200 mmHg) has been reached, occlude the tube and remove the pressure generator. After removing the occlusion from the tube, measure the time taken for the mercury column to fall from the 27 kPa (200 mm Hg) mark to the 5 kPa (40 mmHg) mark. Check that the exhaust time does not exceed 1.5 s. 11. Method of test for the hysterlsls error of the aneroid manometer (1) Apparatus (i) rigid metal vessel, with a capacity of 500 ml±So/o; (ii) calibrated reference manometer, with an uncertainty less than 0.1 kPa (0.8 mmHg); (iii) pressure generator, e.g. ball pump (hand pump) with a deflation valve; (iv) T-piece connectors. (2) Procedure Replace the cuff with the vessel. Connect the calibrated reference manometer by means of a T-piece connector to the pneumatic system. After disabling the electro-mechanical pump (If fitted) connect the additional pressure generator Into the pneumatic system by means of another T·piece connecto? · Test the device with increasing pressure steps of not more than 7 kPa (50 mmHg) to the scale maximum, at which point hold the pressure for 5 minutes and then decrease it by_the same steps. Disconnect the calibrated reference manometer during the S minutes at maximum pressure. (3) Expression of results· Express the results as the difference between the Indicated values on the manometer at the same test pressure steps when increasing the pressure and when decreasing the pressure. 12. Method of test for the construction (1) Apparatus alternating pressure generator, which generates a sinusoidal pressure variation between 3 kPa and 30 kPa (20 mmHg and 220 mm Hg) at a maximum rate of 60 cycles per minute. · (2) Procedure . Carry out the procedure specified in paragraph 1 of this Annexure. Connect the aneroid manometer directly to the alternating pressure generator and perform 10,000 alternating pressure cycles. One hour after the stress test carry out the procedure as specified in paragraph 1 of this Annexure at the same test pressure levels as before the stress test. (J) Expression of results Express the results as the differences between the indicated values on the manometer at the same test pressure steps before and after the stress test. PARTVII·B NON•INVASIVE AUTOMATED SPHYGMOMANOMETERS 1.Scope This specification gives general performance, efficiency and mechanical and electrical safety requirements, including test methods for type approval, for non-invasive electronic or automated sphygmomanometers and their accessories which, by means of an inflatable cuff; are used for the non· invasive measurement of arterial blood pressure. This specification only applies to devices measuring at the upper arm, the wrist or the thigh. Note : Luer locks sha II not be used with these devices [see paragraphs 6(11)(iii) and 7(5)]. 2. Terminology (1) Bladder: It Is the Inflatable component of the cuff. (2J Pressure in a blood vessel : It refers to pressure in the arterial system of the body. · (3) Cuff: It Is that component of the sphygmomanometer, _J? --- - 600 THE GAZE'JffE OF 1INDIA : EXTRAORDINARY [PARTII-SEC. 3(i)] comprising a bladder and a sleeve, which is wrapped around the limb.of the·patient. (4) Diastolicbloodpressurefva/ue).: The minimum value ·of the .arterlat blood pressure as a result of relaxation: of\the -systernlc ventricle. Note: Because of hydrostatic effects,this value should be measured. with the cuff at the.heart level. (S) Mean stteristbkxd.pressure (value) : It is the value.of..the· integral of.one cycle of the blood pressure curv.edlvided:by.thetime·.of.oneheart beat period. Note: Because of:?ydrostatic effects, this value should be measured withthe cuff at the heart level. (6) Non-Invasive blood pressure measurement: Indirect measurement of the arterial blood pressure without ,arterial puncture is called non­ lnvasiyeblood pressure ,measurement. (7) Pneumatic,system c- A system that tncludes all pressurized and pressure-controlling parts such as cuff, tubing, connectors, valves, transducer and pump Is called pneumatic system. (8) Sleeve: It is essentially Inelastic part of the cuff that encloses the bladder. (9) Sphygmomanometer: It is an instrument used for the non-invasive measurement of the arterial blood pressure. (10) Systolic blood pressure (value): The maximum value of the arterial blood pressure as a result of the contraction of the systemic ventricle is called systolic blood pressure. Note: Because of hydrostatic effects, this value shou Id be measured with the cuff at the heart level. (11) Electro-mechanical blood pressure measuring system : It is that system that consists of: (i) at least one cuff, which is connected to the pneumatic system; (ii) at least one electro-mechanical transducer to measure cuff pressure; (iii) at least one measured value display; and (iv) if needed, srgnal inputs and outputs. (12) Electro-mechanical pressure transducer : It is that component that transforms pressure signals into electrical signals. (13) Oscillometrlc method : It is that method, wherein a cuff Is placed on the limb and the pressure in the cuff Is Increased until the blood flow in the artery is interrupted and then the pressure in the cuff is slowly reduced. Note: During the inflation and deflation of the cuff small pressure changes (oscillations) occur In the cuff as a result of the arterial blood pressure pulses. These oscillations, which first increase and then decrease, are detected and stored together with the corresponding cuff pressure values in the measurement system. With these stored values the systolic, diastolic and mean arterial blood pressure values can be mathematically derived using an appropriate algorithm. It is possible to carry out the measurement during the inflation phase. (14) Zero setting : The procedure that corrects a deviation of the pressure reading to O kPa (0 mmHg) at atmospheric pressure (gauge pressure: O kPa (0 mmHg)] is called zero setting. (15) Patient simulator : Device for simulating the oscillometric cuff pulses and/or auscultatory sounds during inflation and deflation. Note : This device is not used for testing accuracy but is required in assessing stability of performance. ' (16) Auscultatory method : It Is that technique whereby sounds (known as Korotkoff sounds) are heard over an occluded artery as the occluding pressure Is slowly released, the appearance of sounds coinciding with the systolic blood pressure and the disappearance of sounds with the diastolic blood pressure. In children under the age of 13, "k4" (i.e. 4th phase Korotkoff sound) may be appropriate. (17) Self-linearizing deflation valve : The valve used for controlled linearizing exhaust of the pneumatic system during measurement. 3. Description of the category of instrument The basic components of a sphygmomanometer are a cuff and bladder that can be wrapped around a patient's limb, a system for applying and releasing pressure to the bladder, and a means of measuring and displaying the instantaneous pressure In the bladder. 4, Units of measurement The blood pressure shall be indicated either in kilo-pascals (kPa) or in millimetres of mercury (mmHg). 5, Metrologlcal requirements (1) Maximum permissible errors of the cuff pressure Indication ['WT n-= 3(i)J For any set of conditions within the ambient temperature range of 15°C to 25°C and the relative humidity range of 20% to 85%, both for increasing and for decreasing pressure, the maximum permissible error for the measurement of the cuff pressure at any point of the scale range shall be ± 0.4 kPa (± 3 mmHg) in case of verifying the first time and ± 0.5 kPa ( + · 4 mm Hg) for sphygmomanometers in use. Testing shall be carried out in accordance with para 2 of Annexure A. (2) Maximum permissible errors of the overall system as measured by clinical tests" The following maximum permissible errors shall apply for the overall system:- maximum mean error of measurement : ± 0.7 kPa ± 5 mmHg); maximum experimental standard deviation : 1.1 kPa (8 mm Hg). (3) Environmental performance (i) Storage: Blood pressure measuring systems shall maintain the requirements specified after storage for 24 h at a temperature of -5°C and for 24 hours at a temperature of S0°C and a relative humidity of 85% (non­ condensing). Testing shall be carried out at environmental conditions In accordance with paragraph 2 of Annexure A after the test sample has been placed for 24 h at a temperature of -5°C and immediately afterwards for 24 hours at a temperature of 50°C In a climatic chamber. Note : Integrated multi-parameter monitors may contain components which may be damaged during storage. The general temperature range as stated in para 3 of Annexure A has therefore been reduced. (ii) Temperature, relative humidity: For the ambient temperature range of 10°c to 40°C and a relative humidity of 85% (non-condensing), the difference of the cuff pressure Indication of the sphyg­ momanometer shall not exceed + 0.4 kPa (±3 mmHg). - Testing shall be carried out in accordance with paragraphs 2 and 11 of Annexure A. The signal processing for the determination of the blood pressure values shall not be Influenced within the range of temperature and relative humidity. . For any set of conditions all the deviations between l/7// Gif/-lS'/ ... ·-··· --- 601 the reference pressure and the indicating cuff pressure of the instrument must be less than or equal to the maximum permissible error. 6. Technical requirements (1) General : Equipment, or parts thereof, using materials or having forms of construction different from those detailed In this Recommendation shall be accepted If it can be demonstrated that an equivalent degree of safety and performance ls.obtained. (2) Technical requirements for the cuff and bladder: The cuff shall contain a bladder. For reusable cuffs the manufacturer shall indicate the method for cleaning In the accompanying documents. Note : The optimum bladder size is one with dimensions such that its width is 40% of the limb circumference at the midpoint of the cuff application and its length is at least 80%, preferably 100% of the limb circumference at the midpoint of cuff application. Use of the wrong size can affect the accuracy of the measurement. . (3) Technical requirements for the display: The display shall be designed and arranged so that the Information including measuring values can be read and easily recognized. Testing shall be carried out by visual inspection; If abbreviations are used on the display they shall be as follows:- "S" or "SYS" : systolic blood pressure (value); "D" or "DIA" : diastolic .blood pressure (value); "M" or"MAP": mean arterial blood pressure (value). Single letter abbreviations shall be positioned in such a way to avoid confusion with SI units. (4) Effect of voltage variations of the power source (i) Internal electrical power source (a) Changes of the voltage within the working range determined according to para 4( 1) of Annexure A shall not influence the cuff pressure reading and the result of the blood pressure measurement. (b) Outside this working range no cuff pressure reading and no result of the blood pressure measurement shall be displayed . ,t 602 THE· .GAZETTE ,OF INDIA : EXTRAORDINARY [PARTII-SEC. 3(i)] Testing shall .be .carried out in accordance with paragraphs 4(1) and 5(1) of AnnexureA. (ii) External-electrical power source (a) ·changes of the voltage within the , working range specified by the , manufacturer shall not Influence the :cuff pressure reading and the result .. of.the blood pressure measurement. (b) .Incorrect ·values resulting from voltage variations outside the limits ·given In paragraph 6(4)(11)(a)shall •not:be displayed. Testing shall be carried out according to para 4(4) (alternating current) and para 4(5) (direct current) of Annexure A. Note: In the case of any malfunction of the equipment, deflation to below 2 kPa (15 mmHg) must be guaranteed within 180s in the case of adult patients and to.below 0.7 kPa (5 mmHg) within 90s in the case of neonatal/Infant patients. (5) Pneumatic system (i) Air leakage: Air leakage shall not exceed a pressure drop of 0.8 kPa/mln (6 mmHg/min), Testing shall be carried out In accordance with paragraph 6 of Annexure A. (Ii) Pressure reducing system for devices using the auscultatory method: The pressure reducing system for manually operated and automated deflation valves shall be capable of maintaining a deflation rate of 0.3 kPa/s to 0.4 kPa/s (2 mmHg/s to 3 mmHg/s) within the target range of systolic and diastolic blood pressure. For devices which control .the pressure reduction as a function of the pulse rate, a deflation rate of 0.3 kPa/pulse to 0.4 kPa/ pulse (2 mmHg/pulse to 3 mmHg/pulse) shall be maintained. Note: Manually operated deflation valves should be easily adjustable to these values. Testing shall be carried out In accordance with paragraph 7 of Annexure 7. (iii) Rapid exhaust: During the rapid exhaust of the pneumatic system, with the valve fully opened, the time for the pressure reduction from 35 kPa to 2 kPa (260 mmHg to 15 mmHg) shall not exceed 10 seconds. For blood pressure measuring systems having the capability to measure in a neonatal/in-fant mode, the time for the pressure reduction from 20 kPa to 0.7 kPa (150 mmHg to 5 mmHg) during the rapid exhaust of the pneumatic system with the valve fully opened shall not exceed Ss. Testing shall be carried out In accordance with paragraph 8 of Annexure A. (Iv) Zero setting: Blood pressure measuring systems shall be capable of automatic zero setting. The zero setting shall be carried out at appropriate Intervals, at least starting after switching on the device. At the moment of the zero setting a gauge pressure ofO kPa (0 mmHg) shall exist and be displayed thereafter. Devices performing zero setting only Immediatelyafter switching on, shall switch off autornatically when the drift of the pressure transducer and the analog signal processing exceeds 0.1 kPa (1 mmHg). Testing shall be carried out In accordance with para 9 and paragraph 10 of Annexure A. (6) Electromagnetic compatibility Either: electrical and/or electro-magnetic Interferences shall not lead to degradations In the cuff pressure Indication or in the result of the blood pressure measurement; or If elettrical and/or electro-magnetic Interferences lead to an abnormality, the abnormality shall be clearly indicated and It shall be possible to restore normal operation within 30 s after cessation of the electro-magnetic disturbance. (7) Stability of the cuff pressure indication The change in the cuff pressure indication shall not" be more than 0.4 kPa (3 mm Hg) throughout the pressure range after 10,000 simulated measurement cycles. Testing shall ·be carried out in accordance with paragraph 12 of Annexure A. (8) Pressure indicating device (i) Nominal range and measuring range: The nominal range for the cuff pressure measurement shall be specified by the manufacturer. The measuring and indication ranges of the cuff pressure shall be equal to the nominal range. Values _of blood [ 'fT'T Jl-<911'6 3( i)] - .:,,_- ,, ... , -- 603 pressure measurement results outside the nominal range of cuff pressure shall be dearly indicated as out of range. Testing shall be carried out by visual inspection. (ii) Digital indication : The digital scale interval shall be 0.1 kPa (1 mmHg). If the measured value of a parameter is to be indicated on more than one display, all the displays shall indicate the same numerical value. Measured numerical values on the display(s), and the symbols defining the units of measurement shall be arranged in such a way so as to avoid misinterpretation. Numbers and characters should be clearly legible. Testing shall be carried out by visual Inspection. (9) Signal input and output ports The construction of the signal input and output ports (excluding Internal interfaces, e.g. microphone signal input) relevant to the non-invasive blood pressure measurement shall ensure that Incorrectly fitted or defective accessories shall not result in erroneous indication of cuff pressure or erroneous indication of blood pressure. Testing shall be carried out in accordance with paragraph 13 of Annexure A. (10) Alarms If alarms are used they shall be of at least medium priority. (11) Safety (i) Cuff pressure: It shall be possible to abort any blood · pressure measurement at any time by single key operation and this shall lead to a rapid exhaust. Testing shall be carried out in accordance with paragraph 14 of Annexure A. (ii) Unauthorlred access: All controls which affect accuracy shall be sealed against unauthorized access. Testing shall be carried out by visual Inspection. (iii) Tubingconnectors: Users of equipment intended for use in environments employing intravascular fluid systems shall take all necessary precautions to avoid connecting-the output of the blood pressure measuring device to such systems as air might inadvertently be pumped Into a blood vessel if, for example, Luer locks were used. {Iv) E/ectrieal safety: Electronic or automated sphygmomano· meters shall comply with the relevant natlona I safety regulations. (v) Resistance to vibration and shock: After testing, the device shall comply with the requirements of paragraph 5(1). 7. Metrologlcal controls (1) Model approval At least three samples of a new type of sphygmomanometer shall be tested. The tests to verify conformity to metrologlcal and technical requirements shall be carried out according to Annexure A. (2) Verifieation (I) Inltlal verification At Initial verification the requirements of paragraphs 5(1) and 6(5)(i) shall be fulfilled. · Testing shall be carried out according to paraqraphs 2 and 6 of Annexure A. (II) Subsequent verification Each instrument of an approved type of sphygmomanometer shall be verified every 2 years or after repair. Requirement of paragraphs 5(1) and 6(5)(1)shall be fulfilled and tests must be carried out according to paragraphs 2 and 6 of Annexure A. (3) Sealing (I) Control marks will be put on lead seals for which corresponding punched screws shall be attached whenever necessary. These seals shall prevent, without destruction of the control marks: (a) In the case of patient-monitors In which the sphygmomanometer Is one pa rt of a system: the manipulation of the metrologlcally relevant parts for measuring blood pressure; (b) in the case of all other manometers: the opening of the' casing. 604 THE .GAZETifE .OF 'INDIA : EXTRAORDINARY [PARTII-SEc. 3(i)] (ii) If tnecconstructlon of the .lnstrument guarantees .securttv against any interference; the metrological control marks or the.security.marks may be·attached in the form ciflabels. (iii) All seals shalt be accessible without using a tool. (4) MarkingoNhedevice The devke.shall'be.rnarked with the following information:- (i) namerend/cr-tracemerk of manufacturer; (ii) seriaL number :and year of fabrication; (iii) measuring :range and measuring unit; (iv) model approval number; (v) centre:ofthe bladder. indicating the correct position .forthe cuff over the artery; and (vi) marking om the cuff indicating ttie limb drcumference for which it Is appropriate. (5) Manufacturer's information (i) The manufacturer's instruction manual shall contain the following information: (a) explanation of the operating procedures which are important for correct application (such as the selection of the appropriate cuff size, positioning of the cuff and adjustment of the pressure reduction rate); (b) a warning to users of equipment Intended for use in environments employing intravascular fluid systems not to connect the output of the blood pressure measuring device to . such systems as air might Inadvertently be pumped into a blood vessel if, for example, Luer locks were used; (c) methods for cleaning reusable cuffs; (d) nature and frequency of the maintenance to ensure that the device operates properly and safely at all times; it is recommended that the performance should be checked at least every 2 years and after maintenance and repair, by re­ verifying at least the requirements in paragraphs 5(1) and 6(5)(1) (testing at \east at 7 kPa (SOmmHg) and 27 kPa (200 mmHg)); (e) a reference method for clinical tests carried out according to Annexure C · or an equivalent method; (f) a list of all components belonging to the pressure measuring systems, including (g) a description of the operating principles of the blood pressure measuring device; (h) remarks on the environment or operational factors which affect the performance ( e.g. electro- magnetic fields, arrhythmia); (I) specification of the single input/ output ports(s); (J) specification of the rated voltage, If applicable; (k) specification of the intended power source, if applicable; (I) nominal range for the result of the blood pressure measurement; (m) warm up time, If applicable; (n) description of the meaning .of the "out of range signal" and (o) description of the alarms, If applicable.

Previous and next

All provisions of Legal Metrology (General) Rule, 2011

Questions people ask

Is annexure D of Legal Metrology (General) Rule, 2011 still in force?

No repeal of Legal Metrology (General) Rule, 2011 is recorded.