Annexure C, Legal Metrology (General) Rule, 2011: ANNEXUREC

Legal Metrology (General) Rule, 2011 - Annexure C

Text of annexure C

ANNEXUREC

ANNEXUREC TEST OF WATER RESISTANCE OF COMPLETE THERMOMETERS (Mandatory) 1. Pattern approval (1) A total of 10 samples shall be tested. (2) The battery casing shall be opened and closed several times before _the tests if the thermometer Is equipped with replaceable batteries. (3) The thermometer shall be totally immersed in an equivalent physiological solution (9.Sg NaCl per litre of distilled water) to a depth of 15 cm and at temperatures of 50°C and 20°C for the following periods of time and in the sequence indicated :- 1 hour at 50°C ± 2°c l hour at 20°C ± 2°c 24 hours at 50°C ± 2°C 24 hours at 20°C ± 2°c (4) The Indicated values shall be measured at two or more temperatures near the lower and upper limit of the measuring range before the first Immersion and after the second and last immersion. The thermometers shall have reached equilibrium with room temperature before recording the indicated values. After the last immersion, the thermometers shall be stored for 14 days in air at room temperature before taking the last measurement. (5) The test may be discontinued if it is obvious that water has penetrated into the casing of a thermometer. (6) The thermometer pattern shall be declared to be water resistant if, for nine out of ten thermometers, the difference in indicated ['1TITu-= J(i)J 591 temperatures for any individual thermometer is less than- • 0.04°C for thermometers with a minimum digital increment of 0.01 °C (class!), • 0.1°C for thermometers with a minimum digital Increment of 0.1 °C (class II). 2, Verification (1) The thermometers shall be totally immersed in an equivalent physiological solution at a temperature of so•c ± z-c to a depth of 15 cm for one hour, after which they shall be immersed for another hour under same conditions but at a temperature of 20°C ± 2°C. Before the first immersion and after the second immersion, the indicated values shall .be measured at two temperatures. (2) A thermometer shall be accepted if the performance requirements specified In paragraph · 1(6) are met.

Previous and next

All provisions of Legal Metrology (General) Rule, 2011

Questions people ask

Is annexure C of Legal Metrology (General) Rule, 2011 still in force?

No repeal of Legal Metrology (General) Rule, 2011 is recorded.