Annexure C, Legal Metrology (General) Rule, 2011: ANNEXUREC
Legal Metrology (General) Rule, 2011 - Annexure C
Text of annexure C
ANNEXUREC
ANNEXUREC
TEST OF WATER RESISTANCE OF COMPLETE
THERMOMETERS
(Mandatory)
1. Pattern approval
(1) A total of 10 samples shall be tested.
(2) The battery casing shall be opened and
closed several times before _the tests if the
thermometer Is equipped with replaceable batteries.
(3) The thermometer shall be totally immersed
in an equivalent physiological solution (9.Sg NaCl per
litre of distilled water) to a depth of 15 cm and at
temperatures of 50°C and 20°C for the following
periods of time and in the sequence indicated :-
1 hour at 50°C ± 2°c
l hour at 20°C ± 2°c
24 hours at 50°C ± 2°C
24 hours at 20°C ± 2°c
(4) The Indicated values shall be measured at
two or more temperatures near the lower and upper
limit of the measuring range before the first
Immersion and after the second and last immersion.
The thermometers shall have reached equilibrium
with room temperature before recording the indicated
values. After the last immersion, the thermometers
shall be stored for 14 days in air at room temperature
before taking the last measurement.
(5) The test may be discontinued if it is obvious
that water has penetrated into the casing of a
thermometer.
(6) The thermometer pattern shall be declared
to be water resistant if, for nine out of ten
thermometers, the difference in indicated
['1TITu-= J(i)J 591
temperatures for any individual thermometer is less
than-
• 0.04°C for thermometers with a minimum
digital increment of 0.01 °C (class!),
• 0.1°C for thermometers with a minimum
digital Increment of 0.1 °C (class II).
2, Verification
(1) The thermometers shall be totally immersed
in an equivalent physiological solution at a
temperature of so•c ± z-c to a depth of 15 cm for
one hour, after which they shall be immersed for
another hour under same conditions but at a
temperature of 20°C ± 2°C. Before the first
immersion and after the second immersion, the
indicated values shall .be measured at two
temperatures.
(2) A thermometer shall be accepted if the
performance requirements specified In paragraph
· 1(6) are met.