Annexure B, Legal Metrology (General) Rule, 2011: ANNEXUREB

Legal Metrology (General) Rule, 2011 - Annexure B

Text of annexure B

ANNEXUREB

ANNEXUREB BRIEF DESCRIPTION OF INSTRUMENT PERFORMANCE TESTS (Mandatory) 1, Energy dissipation of a resistance-type Interchangeable probe ( 1) The probe shall be placed in a reference water bath as specified In paragraph 1(1) of Annexure A at a temperature of 37°C ± 1 °C. Measurements shall be carried out at three or more DC currents with the highest power being 2mW. For each applied current, the voltage and current shall be measured. (2) The equivalent resistance values shall be calculated and then converted to temperature values using the manufacturer's characteristic (resistance versus temperature) table for the probe type. A linear (least-squares fit) curve of temperature as a function of applied power shall be drawn. From this curve, power corresponding to the maximum energy dissipation that will cause a change in indicated temperature by 0.01°C for reusable, interchangeable, or single-use probes shall be determined. This value is the maximum power that may be provided by an indicating unit for the type of probe tested and the manufacturer's specified value shall be equal to or less than the value determined. 2. Electrical Insulation resistance of the probe (1) The resistance of the temperature probe shall be determined at one or more temperatures using the procedure specified in paragraphs 2(1)(i) or 2(2)(ii) of Annexure A. The probe shall then be immersed to a length equal to that intended to be in contact with the body, or 50 mm, whichever Is greater, in a physiological saline solution (9.Sg of NaCl per . litre of distilled water}. (2) After at least one minute, the resistance between the electrical connections of the probe ta ken together and an electrode immersed In the physiologicalsaline solution shall be measured using an instrument that applies a voltage of 10 V ± 1 V between the probe connections and the electrode. The resistance measured shall be greater than the shunt resistance that would correspond to a change in indicated temperature of 0.02°C. (3) The probe shall be left in the physiological saline solution for 24 hours, after which Its resistance shall be re-measured as specified in paragraph 2(1) of Annexure B. The difference in indicated temperature between measurements shall not be greater than 0.02°c. 3. Cleaning and disinfecting the probe (1) The applied part of the temperature probe or of the complete compact thermometer shall be cleaned and disinfected twenty times according to the manufacturer's Instructions. (2) After cleaning and disinfecting as specified in paragraph 3(1) of Annexure B, the requirements of paragraph 4(2} shall be met. 4, Low battery indication Note: Paragraphs 4 to 9 of Annexure Bit is to be understood that the temperature indication of a complete thermometer shall be generated within the measuring range by inserting the probe in a reference water bath or in another bath with similar qualities. The temperature indication of an indicating unit designed for use with interchangeable probes shall be generated by replacing the probe by an auxiliary device, such as an appropriate precision resistor simulating the temperature of a resistance probe. The reference temperature Indication is that obtained under the reference conditions described in paragraph · 4(3}. ( 1) The battery shall be replaced by a variable DC voltage source. (2} The voltage of the source shall be reduced unit a low battery indication or warning signal Is activated at the level specified by the manufacturer. The test shall be carried out at three different temperatures; 37°C ± 1 °C, and the lower and upper limits of the measuring range. 5. Ambient temperature (1)The complete thermometer or indicating unit shall be placed in a test chamber, and the temperature of the chamber varied from 10°C to 40°C with each temperature setting constant within± 2°C. Sufficient time shall be allowed at each temperature setting to permit the complete thermometer or indicating unit to reach thermal equilibrium with the chamber. (2) At each temperature tested, the requirements specified in paragraph 4(2) shall be met. 6. Thermal shock (1) The Indicating unit shall be placed in a test chamber at - S°C ± 2°c. · · (2) After thermal equilibrium has been. established, the complete thermometer or Indicating ---------? - --- 590 THE .GkZETTE OF INDIA: EXTRAORDINARY [PARTII-SEC. 3(i)] unit shall be placed in a test chamber at50°C.± 2°c until thermal equilibrium has been-established and all traces of condensed molsture.haveevaporated. · (3) The process.described,in;paragraphs 6(1) and 6(2) of Annexure B shall be performed five times. (4) The indicating unit shall ,be allowed to achieve thermal· equilibrium.at room temperature after which the· Indicated temperature shall not change by more than.± 0.1 °Casa result of exposure to the thermal shocks described ,in paragraphs 6(1) and 6(2) by Annexure.B. Note : Thermal equilibrium may be achieved more quickly and completely iby opening the casing of the thermometer, If possible. 7. Humidity (1) The complete thermometer or Indicating unit shall be stabilized atatemperature Jwithin the range 20°C to 32°C for-4 hours or more. During this time, t shall remain constant within ± 2°c. (2) After achieving a stable temperature as specified in paragraph 7(1) of Annexure B, the complete thermometer or indicating unit shall be placed in a humidity test chamber containing air at a temperature between .t and t + 4°C and a relative humidity between 91 % and 95% for a period of 48 hours. (3) After exposure as specified in paragraph 7(2) of Annexure B, the complete thermometer or indicating unit shall be removed from the test chamber and allowed to stabilize at room temperature for 48 hours. The indicated temperature shall not vary by more than ± 0.1 °Casa result of this test. 8. Electro-magnetic: radiation interference (1) The complete thermometer or indicating unit shall be exposed to an electromagnetic field with a field strength of 10 V/m at frequencies between 150 kHz and 500 MHz modulated by a l kHz sine wave and 80% amplitude modulation. (2) The specific field strength shall be established prior to testing and without the instrument being placed In the electro-magnetic field. The field strength may be generated as follows :- a strip line for low frequencies (below 3 MHz)or In some cases 150 MHz) for small instruments, dipole antennas, or antennas with circular polarization,placed lm from the Instrument at higher frequencies. (3) The field shall be generated with two orthogonal polarizations and then slowly scanned through the frequency range. Antennas with circular polarization may be used to generate the electro­ magnetic field without a change in their positions. The test shall be carried out in a shielded enclosure to comply with international laws prohibiting interference with radio communications, but care shall be taken to minimize reflections. (4) During the test, the requirements specified in paragraph 5(3)(vi) shall be met. 9. Mec:hanlc:al shock (1) The complete thermometer or indicatingunit shall be allowed to fall from a height of lm on to a hard surface (for example, a block of hard wood of density greater than 700 kg/m3 and of suitable size lying flat on a rigid base). The drop shall be performed once for three different orientations of the complete thermometer or indicating unit. (2) After the test, the requirements specified in paragraph 5(3)(vii) shall be met.

Previous and next

All provisions of Legal Metrology (General) Rule, 2011

Questions people ask

Is annexure B of Legal Metrology (General) Rule, 2011 still in force?

No repeal of Legal Metrology (General) Rule, 2011 is recorded.