Annexure B, Legal Metrology (General) Rule, 2011: ANNEXUREB
Legal Metrology (General) Rule, 2011 - Annexure B
Text of annexure B
ANNEXUREB
ANNEXUREB
BRIEF DESCRIPTION OF INSTRUMENT
PERFORMANCE TESTS
(Mandatory)
1, Energy dissipation of a resistance-type
Interchangeable probe
( 1) The probe shall be placed in a reference
water bath as specified In paragraph 1(1) of Annexure
A at a temperature of 37°C ± 1 °C. Measurements
shall be carried out at three or more DC currents
with the highest power being 2mW. For each applied
current, the voltage and current shall be measured.
(2) The equivalent resistance values shall be
calculated and then converted to temperature values
using the manufacturer's characteristic (resistance
versus temperature) table for the probe type. A linear
(least-squares fit) curve of temperature as a function
of applied power shall be drawn. From this curve,
power corresponding to the maximum energy
dissipation that will cause a change in indicated
temperature by 0.01°C for reusable, interchangeable,
or single-use probes shall be determined. This value
is the maximum power that may be provided by an
indicating unit for the type of probe tested and the
manufacturer's specified value shall be equal to or
less than the value determined.
2. Electrical Insulation resistance of the probe
(1) The resistance of the temperature probe
shall be determined at one or more temperatures
using the procedure specified in paragraphs 2(1)(i)
or 2(2)(ii) of Annexure A. The probe shall then be
immersed to a length equal to that intended to be in
contact with the body, or 50 mm, whichever Is greater,
in a physiological saline solution (9.Sg of NaCl per .
litre of distilled water}.
(2) After at least one minute, the resistance
between the electrical connections of the probe ta ken
together and an electrode immersed In the
physiologicalsaline solution shall be measured using
an instrument that applies a voltage of 10 V ± 1 V
between the probe connections and the electrode.
The resistance measured shall be greater than the
shunt resistance that would correspond to a change
in indicated temperature of 0.02°C.
(3) The probe shall be left in the physiological
saline solution for 24 hours, after which Its resistance
shall be re-measured as specified in paragraph 2(1)
of Annexure B. The difference in indicated
temperature between measurements shall not be
greater than 0.02°c.
3. Cleaning and disinfecting the probe
(1) The applied part of the temperature probe
or of the complete compact thermometer shall be
cleaned and disinfected twenty times according to
the manufacturer's Instructions.
(2) After cleaning and disinfecting as specified
in paragraph 3(1) of Annexure B, the requirements
of paragraph 4(2} shall be met.
4, Low battery indication
Note: Paragraphs 4 to 9 of Annexure Bit is to
be understood that the temperature indication of a
complete thermometer shall be generated within the
measuring range by inserting the probe in a reference
water bath or in another bath with similar qualities.
The temperature indication of an indicating unit
designed for use with interchangeable probes shall
be generated by replacing the probe by an auxiliary
device, such as an appropriate precision resistor
simulating the temperature of a resistance probe.
The reference temperature Indication is that obtained
under the reference conditions described in paragraph
· 4(3}.
( 1) The battery shall be replaced by a variable
DC voltage source.
(2} The voltage of the source shall be reduced
unit a low battery indication or warning signal Is
activated at the level specified by the manufacturer.
The test shall be carried out at three different
temperatures; 37°C ± 1 °C, and the lower and upper
limits of the measuring range.
5. Ambient temperature
(1)The complete thermometer or indicating unit
shall be placed in a test chamber, and the
temperature of the chamber varied from 10°C to 40°C
with each temperature setting constant within± 2°C.
Sufficient time shall be allowed at each temperature
setting to permit the complete thermometer or
indicating unit to reach thermal equilibrium with the
chamber.
(2) At each temperature tested, the
requirements specified in paragraph 4(2) shall be
met.
6. Thermal shock
(1) The Indicating unit shall be placed in a test
chamber at - S°C ± 2°c.
·
·
(2) After thermal equilibrium has been.
established, the complete thermometer or Indicating
---------? - ---
590 THE .GkZETTE OF INDIA: EXTRAORDINARY [PARTII-SEC. 3(i)]
unit shall be placed in a test chamber at50°C.± 2°c
until thermal equilibrium has been-established and
all traces of condensed molsture.haveevaporated. ·
(3) The process.described,in;paragraphs 6(1)
and 6(2) of Annexure B shall be performed five times.
(4) The indicating unit shall ,be allowed to
achieve thermal· equilibrium.at room temperature
after which the· Indicated temperature shall not
change by more than.± 0.1 °Casa result of exposure
to the thermal shocks described ,in paragraphs 6(1)
and 6(2) by Annexure.B.
Note : Thermal equilibrium may be achieved
more quickly and completely iby opening the casing
of the thermometer, If possible.
7. Humidity
(1) The complete thermometer or Indicating unit
shall be stabilized atatemperature Jwithin the range
20°C to 32°C for-4 hours or more. During this time,
t shall remain constant within ± 2°c.
(2) After achieving a stable temperature as
specified in paragraph 7(1) of Annexure B, the
complete thermometer or indicating unit shall be
placed in a humidity test chamber containing air at a
temperature between .t and t + 4°C and a relative
humidity between 91 % and 95% for a period of 48
hours.
(3) After exposure as specified in paragraph
7(2) of Annexure B, the complete thermometer or
indicating unit shall be removed from the test
chamber and allowed to stabilize at room
temperature for 48 hours. The indicated temperature
shall not vary by more than ± 0.1 °Casa result of
this test.
8. Electro-magnetic: radiation interference
(1) The complete thermometer or indicating unit
shall be exposed to an electromagnetic field with a
field strength of 10 V/m at frequencies between 150
kHz and 500 MHz modulated by a l kHz sine wave
and 80% amplitude modulation.
(2) The specific field strength shall be
established prior to testing and without the
instrument being placed In the electro-magnetic field.
The field strength may be generated as follows :-
a strip line for low frequencies (below 3
MHz)or In some cases 150 MHz) for small
instruments,
dipole antennas, or antennas with circular
polarization,placed lm from the Instrument
at higher frequencies.
(3) The field shall be generated with two
orthogonal polarizations and then slowly scanned
through the frequency range. Antennas with circular
polarization may be used to generate the electro
magnetic field without a change in their positions.
The test shall be carried out in a shielded enclosure
to comply with international laws prohibiting
interference with radio communications, but care
shall be taken to minimize reflections.
(4) During the test, the requirements specified
in paragraph 5(3)(vi) shall be met.
9. Mec:hanlc:al shock
(1) The complete thermometer or indicatingunit
shall be allowed to fall from a height of lm on to a
hard surface (for example, a block of hard wood of
density greater than 700 kg/m3 and of suitable size
lying flat on a rigid base). The drop shall be
performed once for three different orientations of
the complete thermometer or indicating unit.
(2) After the test, the requirements specified
in paragraph 5(3)(vii) shall be met.