Annexure A, Legal Metrology (General) Rule, 2011: ANNEXUREA

Legal Metrology (General) Rule, 2011 - Annexure A

Text of annexure A

ANNEXUREA

ANNEXUREA ESTABLISHING REFERENCE TEMPERA1URES AND DITTRMINING MAXIMUM PERMISSIBLE ERRORS (Mandatory) · 1. Reference temperatures (1) A well-regulated and stirred water bath containing at least one litre in volume shall be used 588 ' THE GAZETirE OF .:INDIA.: ·EXTRAORDINARY [PART11-SEc. 3(i)] ? 10°c s (2) Inter-changeable and single-use probe (i) An inter-changeable or single-use probe shall be Immersed In a reference water bath as specified In paragraph 2(1)(1) of Annexure A measured physical property of the probe shall be converted to a to establish reference temperatures .over the measuring range for conducting various. performance tests on an instrument. The bath shall becontrolled to a temperature stability oLbetter:than ± '0.02°C over the specified.temperature·range:and shall not have a temperature_gradient_greater-.than .±'0.01°C within Its workl!)gspace·at a specified.temperature. This temperature gradient shalLbe assured under all conditions and. methods of loading temperature probes. Note : The ,water 'bath .descnbed above is referred to as•.a··'.'reference ·.water bath" In this specification. (2) A referencethermometer with an expanded uncertainty not greater.than 0.03°C (calculated for a coverage factor k·= 3) .shall be used to determine the temperature of!the .water bath. The calibration shall be traceable to national measurement standards, 2. Detl!rminlng maximum permissible errors (1) Complete tnermometer (i) The temperature probe of a complete thermometer shall be immersed in a reference water bath at a constant temperature until temperature equilibrium is established. The temperature Indicated by the thermometer shall be compared to that indicated by the reference thermometer. The bath temperature shall then be increased or decreased, the temperature equilibrium re-established, and the measurement process repeated. The difference between the measured and reference temperatures shall meet the requirements for maximum permissible errors as specified in paragraph 4(2), (ii) The number of measurements at different temperatures depends on the measuring range of the Instrument; however, measurements shall be carried out for at least the followingnumber of temperatures within the mtasuring range:- Measuring range s 10°C Number of temperatures 3 temperature value by using an appropriate instrument to measure a change in that property as a function of temperature. For a resistance-type probe, an appropriate instrument for measuring Its output signal may be an ohm meter that can apply power to the probe at a level below the limit specified in paragraph 5(2)(i) and the temperature value Is obtained from the manufacturer's data of resistance versus temperature. The expanded measurement uncertainty of the appropriate instrument shall not be greater than a value equivalent to 0.01°C (calculated for a coverage factor k., 3), referring to the manufacturer's data at a temperature of 37°C. The calibration shall be traceable to national measurement standards. Each temperature value obtained for the probe In this way shall be compared to that Indicated by the reference thermometer in the bath. The difference between these temperature values shall meet the requirements for maximum permissible errors as specified in paragraph 4(2). (II) The number of measurements required shall be the same as specified in paragraph 2{l)(ii) of Annexure A. (3) Indicating unit (I) The performance of an indicating unit shall be tested using a device that simulates the relevant physical properties of the appropriate probe type. The expanded measurement uncertainty of the simulating device shall not be greater than a value equivalent to 0.01 °C (calculated for a coverage factor k = 3), referring to the manufacturer's data at a temperature of 37°C. The calibration shall be traceable to national measurement standards. ·, Note : For example, a calibrated decade resistance box may be used to provide a variable resistance to simulate a resistance-type probe. Values of resistance for input to the Indicating unit over its specified measuring range shall be selected from the manufacturer's data of resistance· versus temperature. Similarly, variable voltage sources may be used to simulate a thermocouple. (ii) The difference between the temperatures displayed by the indicating unit and the corresponding simulated values of [ "lT'1 11-<Y"s 3( i)] 589 temperature shall meet the requirements for maximum permissible errors specified in paragraph 4(2). (iii) The number of measurements shall be the same as specified in paragraph 2(1)(ii) of Annexure A.

Previous and next

All provisions of Legal Metrology (General) Rule, 2011

Questions people ask

Is annexure A of Legal Metrology (General) Rule, 2011 still in force?

No repeal of Legal Metrology (General) Rule, 2011 is recorded.