Annexure B, Legal Metrology (General) Rule, 2011: ANNEXURE B
Legal Metrology (General) Rule, 2011 - Annexure B
Text of annexure B
ANNEXURE B
ANNEXURE B
TESTING OF GAS ELIMINATION DEVICES
(Informative)
1. Testing of a gas elimination device as a unit
separate from the measuring system for which
It is intended
(1) General provisions
In order to examine whether the pattern of a
gas elimination device complies with the
requirements in, paragraph 2(10), a specimen of the
pattern must be installed on a suitable test bench
equipped with a meter and a conventional proving tank.
Note: In tests on gas elimination devices, the
proving tank may be replaced by any appropriate
standard.
The efficiency of the gas elimination device is
determined with reference to the meter error at the
same flowrate.
The test bench must, as far as possible, comply
with the following provisions:-
(i) The capacity of the proving tank should be
at least equal to the greater of the following
two values: volume delivered in one minute
at maximum flowrate, or 1000 times the
scale interval of the meter of the test bench.
(ii) It is recommended that an adjustable non
return valve be installed downstream of the
meter in order to prevent back flow of the
liquid which has been measured and to
obtain the minimum back pressure required
for the proper operation of the gas
elimination device.
(iii) There should be no reverse gradient in the
pipework downstream of the meter so that
gas bubbles are allowed to escape in the
normal way in order to keep this pipework
filled to the same level at the beginning
and end of the test.
(iv) The liquid used for the tests should either
be the same as that for which the device is
intended or should be of a viscosity which
is at least equal to that of the liquid for
which it is intended.
Tests on gas elimination devices should be
carried out for flowrates up to a maximum of 100
m'/h. For higher flowrates, characteristics may be
determined by analogy with equipment of the same
design and smaller dimensions. "By analogy" means
that parameters like Reynolds number, Froude
number, etc., are to be taken into account for the
gas elimination device.
(2) Tests on gas separators
The volume of air or gas continuously entering
may be measured by a gas meter and isothermally
converted to atmospheric pressure on the basis of
the indication of a pressure gauge fitted upstream
of the gas meter.
A pressure gauge positioned upstream of the
meter for liquid makes it possible to determine the
lowest pressure at which the gas separator still meets
the efficiency requirements.
['11'1 n-= J(i)J 569
Before starting a test, the whole apparatus is
made to operate at the desired liquid and gas
flow rates so that all parts of the apparatus (except
the proving tank) fill up under set conditions as regards
the entry of air or gas.
The air may be introduced either by injection
downstream of the pump or by suction upstream of
it (see Figures 1, 2 and 3, which are given as
examples).
In the former case, which makes it possible to
operate without changing the performance of the
pump due to the entry of air, the liquid and gas flows
are adjusted by means of control valves. The air or
gas is introduced through a tube positioned in the
centre of the pipework for the liquid, for example at
an elbow.
In the latter case, which reproduces the
conditions encountered in reality (pressure reduction
by suction), the pump must be set to the maximum
flowrate of the separator. If the pump has too great
a flowrate,.it must be possible to regulate it with a
speed reducer. The pump should preferably be of
the volumetric type but it may also be of the
centrifugal type if the supply tank feeds the pump
by gravity. The pressure reduction must then be
regulated by a valve positioned upstream of the
pump, and the air Inlet must be equipped with a
non-return valve which prevents any leakage at the
movement of switching off.
(3) Test on gas extractors
An example of test bench is shown in Figure 4.
It includes a container for creating a pocket of
air to be removed with a volume equal to the
minimum measured quantity of the gas extractor (the
minimum measured quantity of the system being not
yet specified). When the test is carried out with a
proving tank having such a great capacity that the
maximum permissible error cannot be evaluated
correctly on the basis of a single operation of the
gas extractor, the number of operations during the
same test shall be multiplied by 2, 3 or 4, to obtain
the required accuracy.
(4) Tests on special gas extractors
Special gas extractors, mainly used for
measuring systems on road tankers, are principally
intended to prevent measurement errors which may
arise from the complete emptying of one
compartment. They must also separate and
continuously remove introduced air, although to a
lesser degree than a gas separator.
In the case of separate approval, they should
be tested on a test bench which corresponds in
principle to Figure 5.
This test bench is similar to that in Figure 1 but
it differs In order to reproduce the actual conditions of
delivery from road tankers to underground tanks, as
is the case when service stations are supplied with
petrol, and domestic fuel oil deliVerles are made. Thus,
1 the supply tank is located above the special gas
extractor and the meter, l.e, at a level corresponding
to that of road tanker, and the proving tank is
approximately 4 m below the meter.
In order to determine the additional error arising
from the complete emptying of a road tanker, which
may be caused by the air drawn In at the end of the
delivery by the formation of a vortex, the supply tank
is filled with a volume of liquid equivalent to that of
the proving tank. The liquid is then emptied through
the meter into the proving tank without operating
the shut-off valve.
For delivery by gravity, pipework is used which
bypasses the pump.
Due to the existence of the automatic shut-off
valve in the hydraulic circuit connected to the gas
extractor, the pocket of air to be evacuated (described
in Annexure-8, paragraph 1(3)) may be created by
emptying the pipework between the tank and the
gas extractor.
Furthermore, given that a special elimination
device should also perform the function of a separator
,._for a small percentage of air introduced continuously
it is necessary to carry out this test in a manner
comparable to the tests described for separators In
Annexure-B, 1(2), with air being either injected into
the supply pipework or drawn In upstream of the
.
pump by creating an entry of air and partly closing
the valve of the supply tank.
2. Tests on gas elimination devices forming
part of a measuring system during pattern
approval
The tests are carried out with a proving tank of
the capacity specified in Annexure B, paragraph 1(1)
or any appropriate standard.
(1) Tests on a gas separator
This examination particularly applies to
patterns of separators included in
measuring systems which can be mass
produced and transported without
dismantling, such as petrol pump fed by
their own supply pumps.
The essential part of the test bench (Figure
6) is the measuring system itself (in this
case, the fuel dispenser).
In accordance with conditions encountered
in actual use, the liquid is drawn up from a
- - --? --? __
J??---
570 THE GAZETTE OF INDIA: EXTRAORDINARY [PARTl!?SEC. 3(i)]
tank on a lower level than the meter. The
air is drawn in by suction through a special
Inlet equipped with a control valve. The air
can be measured by.a gas•meter, :However,
It is not necessary.to.use a
..gas meter if the
separator is capable .of ,separating and
eliminating ,the .air Introduced in any
proportion, as provided in, paragraph
2(10)(vlil).
The.tequlrements :In, paragraph 2(10)(1)
and ?(lO)(vlll) should be complied with
under test :conditions such that the
maxlmumflowrate of the measuring system
Is reached when no air enters.
(2) Tests on a gas extractor and a special gas
extractor
The measuring system comprising the gas
elimination device must be constructed so
that the tests can be carried out as
described in Annexure B, Paragraph 1(3)
or 1(4).
3. Tests on gas elimination devices forming
part of a measuring system during verification
The gas elimination devices are tested without
It being necessary to verify that the maximum
permissible errors for the individual units are compiled
with.
Gas separator to be tested
., ,!l"'
:, .,
"' E"'
i'!
@
i(l
? "'
li'!
0..
?
{g
"
?
?
FIGURE-1
TEST BENCH FOR GAS SEPARATORS
[ 'IPl Il-<!i"S 3( i) ] 571
proving tank
Gas meter
(as required)
Vacuum gau e
Calibrated air inlet
Gas separator
to be tested)
Flow meter
Non-return valve
(adjustable)
,.,--------+--t---+-t--? Supply tank
Non-return valve
FIGURE-2
TEST BENCH FOR GAS SEPARATORS
572 THE GAZETTE OF INDIA: EXTRAORDINARY [PART!I-SEC. 3(i)]
Gas Cylinder CompressorSupply tank
Gas meter
: Pressure gau e
as separator
(with float to
be tested
[:--+ntainer for Iquid
??--·-
Pressure gauge
Flow meter Tarottle Pump for··
valve gas Suction
Flow meter
i
I Supply tank
-------,-
L..a:x1-.J--->i--.J
Pump
FIGURE-3
TEST BENCH FOR GAS SEPARATORS
..............................
with gas injection
.......
with gas suction
.f
I'."111:,1Y.:;;\;0;l l. "'" ·.J"'l>rn·•N
(<) ht.
.
1,,,1,,,1
T?11l o»
i
I
i
--J ·
I ·.Ii
.
i..
(/.-•... ·
la,i
·J11·11scpa11?dn11
of ?i,
l1M?<'.IO
/'lllHJ! fol ri\/ht/',
,upply 1"nk
l
..
(J·-1c,[
:;,,pp)y ! .
l"""I'
FIGURE-4
TEST BENCH FOR SPECIAL GAS EXTRACTORS
['WT n-= 3(i)J 573
Supply tank
Connection for automatic
closing
??
Gas indicator
Empty hose
compressor
Proving tank
I
Pressure
gauge
···f9-
Special gas
extractor
to
be tested
Gas meter
(for injection)
Flow meter
- Pressure gauge
Pump
\
Gas meter
(Forsuction)
Non return valve
FIGURE-5 : Test bench for special gas extractors
Pressure gauge
Gas indicator (If
necessary}
Flow meter
Gas meter if necessary
(see point 2.1 of
Annex A)
Valve
Air inlet
Gas separator to be
tested
Pressure gauge
Valve
11
r:ov,ny:ank
?
------·········--·····
Supply tank
FIGURE-6
TESTING FACIUTY FOR GAS SEPARATORS IN FUEL DISPENSERS
574 THWGAZETTE'OF J!NDIA: EXTRAORDINARY [PARTII-SEC. 3(i)]
PARTV
VOLUMETRIC CONTAINER TYPE UQUIO
,MEASURING DEVICE
1, General
The volumetric container type.liquid measuring
device consists of.a·butket, afloat ·and a:dlp stick
suitably graduated to Indicate the volume.cf liquid
at different heights In the bucket. The •device is
generally usedforrneesurtnq.the quantltv.otmllkat
milk purchasing centres.
2. Nominal ?pacltles
(a) The unit .of .volurne -shall 'be the cubic
dedmetre.or.litre.ror the cubic centimetre
(cm) or mlllllltre.
(b) The volumetric container type liquid
measuring device may be one of the
following,capacities .:
5 dm3, 10 dmi, 20 dm3 and 50 dm3
(c) The value of ,the smallest graduation on
the dipstick shall be of the form of 1 x !On,
2 x 10n or 5 x 10n where 'n' represents a
whole number, positive or negative or is
equal to zero.
3, General requirements
(a) The bucket shall be made of suitable metal
or alloy. The metal or alloy used shall be
thermally stable, shall resist deformation,
shall not have an unduly high coefficient of
cubical expansion, and shall not affect the
liquid being measured in any way or be
injurious to health. Some of the materials
considered suitable are :
(i) mild steel,
(ii) stainless steel,
(iii) brass sheet,
(iv) copper sheet,
(v) aluminium alloy.
(b) The wall thickness of the bucket shall be
so selected that the bucket will not get
dented In normal use or become unusable
after a few years in service.
(c) The bucket shall be free from surface
defects and indentation. External and
internal surfaces of the bucket made of mild
steel, copper sheet and brass sheet shall
be well tinned or tin plated.
(d) All seams shall be filled and smoothened
to prevent the entrapment of air or liquid ..
(e) The bucket shall be provided with a well
·
formed and proportionate spout to facilitate
pouring of liquid.
(f) The bucket shall be cylindricalin form. The
bottom of the bucket shall be slightly
concave to prevent change of shape due
to the weight of the liquid. The maximum
depth of the concavity shall not be more
than 20 mm.
(g) The bottom of the bucket shall be reinforced
with angle strip of thickness not less than
5 mm.
(h) The upper edge of the bucket shall be
reinforced round the circumference with a
reinforcing band having a thickness not less
than 5 mm.
(i) The bucket shall be provided with a suitable
handle on the side opposite the spout.
(j) The top of the bucket shall be provided with
a cross band across the diameter. The band
shall have groove of appropriate size and
centrally located for Inserting the dip stick.
(k) The float shall be suitably fabricated so as
to be free from holes, pockets, dents or
crevices. A dip stick shall be firmly welded
on the upper centre of the float. The dip
stick together with the float shall be so
constructed that the device so formed
maintains verticality In all positions.
(I) The dip stick shall have a rectangular cross·
section of minimum dimensions 20mm x
10mm. The graduations shall be made by
engraving or other means on both sides of
the vertical surface.
(m) The dip stick shall be graduated at suitable
Intervals throughout the nominal capacity
of the measure.
(n) The graduation lines on the dip stick shall
be clear straight, perpendicular to the axis
of the stick and of uniform thickness not
exceeding 1 mm.
(o) The dip stick shall be Identified with the
bucket by a number of Identifications, which
shall be clearly legible and indelible.
4. Marking
The following inscriptions shall be clearly and
indeliblymarked at a conspicuous place on the bucket
or on a special plate securely attached to the
bucket:-
(a) nominal capacity,
. (b) manufacturer's name or trade mark,
(c) the words "for edible liquids".
['WT II-? 3(i)] 575
5.Sealing
A suitable plate or other device shall be provided
to receive the stamp or seal of the verification
authority.
PART VI
CLINICAL THERMOMETER
PART A
CLINICAL THERMOMETER-SOLID STEM
TYPE
Scope : This part specifies the requirements
and methods of tests for solid stem type mercury in
glass clinical thermometers having a maximum
Indicating device.
1. Terminology
For the purpose of this standard the definitions
given in IS 2527-1979 Glossary of terms relating to
liquid-in-glass thermometers (first Revision) as
revised from time to time In addition to the following,
shall apply.
2. Types
The thermometers shall be of the solid stem
mercury-in-glass type.
3. Temperature Scale
·
The thermometers shall be graduated in
degrees Celsius (°C) and shall have a range from
35° to 42°C or 35° to 43°C.
4. Testing
Thermometers shall be tested for total vertical
immersion.
S, Requirements
(1) Pattern.?-There shall be two patterns of
bulb, namely, oral and rectal as follows:
(a) Oral-for thermometers for use in
mouth and
(b) Rectal-for thermometers for use In
rectum
Note 1: Oral thermometers may also be used
in armpit or groin.
Note 2: Rectal thermometers may also be used
in the mouth, armpit or groin after proper disinfection.
Rectal thermometers if used orally, may however give ·
slightly lower reading.
(2) Materials
(i) Glass tubing-The thermometers
shall be made from glass tubing
conforming to IS 4529-1958
Specification for glass tubes for
medical thermometers.
(ii) Bulb-The thermometer bulb shall be
made from a type of glass which
assures that the depression of zero,
determined in accordance with the
procedure given in Appendix D, does
not exceed 0.07°C. The glass shall
be identified· visillle and indelible,
either by the glass manufacturer or
by the manufacturer of the
thermometer.
(iii) The types of glass used for the
maximum device, capillary tube, and
bulbs shall meet the following
requirements:-
When the glass is analyzed
according to the requirements of IS
2303 (Part I/Section 1), the quantity
of alkali passed Into solution for lg
of glass must correspond to not more
than 263 .5g of Na,O.
(iv) Thermometric liquid ? The
thermometric liquid shall be pure, dry
mercury.
(3) Construction
(I) The stem shall be in alignment with
the bulb.
(a) The free end of the stem shall be
finished smooth, preferably
hemispherical in shape, as shown
in Figure 1.
FIGURE 1 STEM OF CLINICAL THERMOMETER
576 ··THE;GAZET,'f.E <OF INDIA: EXTRAORDINARY [PARTJJ-SEC. 3(i)]
(ii)• Maximum ,iRdiGating devlce
· (constridion)-:The thermometers
.
.shall-have -e.maxlmum .indicating
·:,device (constriction) constructed ,in
!'the capillary,· between the ·mercury
bulb and the scale in order to prevent
the mercury :in the .stern from
returning to the bulb on -cooling. It
shall be locatedin the bore between
: the top of the bulb and the bottom
::of the. scale.
'(a) The:maximum indicating device
shall •pass the hardness test
prescnbedfn Appendix A when an
·acceleration of 600 m/sec2 is
applied at the dosed end of the
bulb for a period of at least 2
minutes.
(iii) Locator-A.locator may be drawn on
the stem before the start of the scale
to facilitate positioning the
thermometer so that the mercury
column Is in full view. The locator
may comprise a thick ( 1 mm of
more) dot or line or a rectangle of
1 x 3 mm
.
(Iv) Bu/1>-The bulb shall be cylindrical
with a smooth rounded end and shall
be in alignment with the stem,
(a) The diameter of bulb and the Joint
of bulb with stem shall nowhere
exceed that of the stem. The joint
shall be smooth and regular.
(b) The bulb shall be hemispherical
at the tip.
(v) The bulb, capillarytube and mercury
shall be sufficiently free from
entrapped gas, debris and foreign
bodies in order to ensure the correct
functioning of the thermometer.
(4) Dimensions
The dimensional and scale requirements for
solid stem type clinical thermometers, shall
be as given in Table 1.
TABLE 1
DIMENSIONAL AND SCALE REQUIREMENTSFOR SOLID STEM CLINICAL THERMOMffiRS
SI. No. Char1:1cteristic
(1) (2)
(i) Scale range, °C
(ii) Smallest scale division, °C
(iii) Over all length, mm
(iv) Scale length, mm
(v) External diameter of stem, mm
(vi) Bulb length, mm
(a) Oral thermometer
(b) Rectal thermometer
(vii) External diameter of bulb, mm
(a) Oral thermometer
(b) Rectal thermometer
(viii) Distance from top of the
constriction to the 35.5°C mark,
mm, Min,
(ix) Distance from the
highest graduation line
to top of stem, mm, Min
(x) Scale spacing, mm Min
Note: See Figure 3 in IS 8757 : 1977
Requirements
-----· ------ -----
(3)
35.0 to 42.0 or 35.0 to 43.0
0.1
100 to 115
40 to 65
4 to 6
12 to 18
Not more than 9
2.0 to 3.5
3.0 to 6.0 (But not exceeding that of stem)
10
8
0.5
[ '1l'T u-iil,rg 3(;) l 577
(5) Graduation and Numbering
(i) The thermometer scale shall be sub
divided in 0.1 °C as shown In Figure
SSQ.
(ii) The graduation lines shall be equally
spaced and at right angles to the axis
of the thermometer.
(iii) All graduation lines shall be of equal
and uniform thickness of not more
than 1/4 of the length of the smallest
scale division or 0.15 mm whichever
Is less.
(iv) The pattern of graduation and
numbering shall be as follows:
(a) Every tenth graduation line
shall be a long llne (about 2
mm) which shall be numbered.
(b) There shall be a medium line
(about 1.5 mm) midway
between two consecutive Jong
lines, and
(c) There shall be four short Jines
(about 1 mm) equally spaced
between consecutive medium
and Jong lines.
(v) When the thermometer Is held in a
vertical position and viewed from the
front, the left-hand ends of the
graduation lines shall Ile on an
imaginary vertical ltne (see Figure
55Q).
(vi) The numbers shall be pla?ed to the
right of the axis in such a way that
extension of the line to which they
refer, would bisect them. The figures
shall be placed parallel to the axis of
the thermometer.
(vii) A distinguishing mark (say, an arrow)
shall be marked at 37.0°C mark to
indicate the normal temperature.
(viii) All graduation lines and numbering
shall be clearly and permanently
marked or etched on the stem and
filled with a pigment. The marking
shall pass the test for permanenc.y
given in Annex B.
Note : The pigment may be black, red, green
or blue as agreed to between the purchaser and the
supplier.
6, Petformance requirements
(1) Appearance
When heated to 42°C or so the mercury
column shall look like a strip of uniform
width with clarity and brightness
throughout.
· Note : This test takes care of all visual
defects, for example, twist in the glass,
aberrations, cloudiness, devitrification, etc.
(2) Ageing and Accuracy
(i) The clinical thermometers shall meet all the
requirements after ageing by natural or
artificial means.
Natural Ageing involves holding the
thermometers at room temperature for four
months after completion of the
construction. Artificial ageing method
involves heat treatment of the bulb and
.
constriction of the unfilled thermometer at
a temperature and duration to be specified
by the manufacture of the glass tubing,
(II) Clinical thermometers shall after ageing
ensure that their accuracy as measured by
the method prescribed In Appendix C shall
be ± 0.1 °C for at least one year after the
first official verification.
(3) lnffuence of Immersion Time
If a thermometer at temperature t, ( 15°C t, <
30°C) is suddenly immersed in a well-stirred water
· bath having a constant temperature t, (35.S°C < t,
< 42°C) and is withdrawn after 20 seconds the
thermometer reading, after cooling to ambient
temperature ( 15°C to 30°C)
(I} shall comply with maximum permissible
error requirement, and
(II} shall not deviate from its stabilized reading
for temperature t, •c by more than 0.005
(t, - t,)°C.
This stabilized reading is the thermometer
reading obtained when the thermometer has been
cooled to ambient temperature, after reaching
complete thermal equilibrium with the water bath at
a temperature t2°C.This reading shaU also meet the
maximum permissible error requirements stipulated
In paragraph 6(2)(ii).
Note : A free chclce of test method Is
permitted, provided the law of the variation of the
Indication of the thermometer as a function of
immersion time is known.
7. Marking and Packing
(1) Marking
Each clinical thermometer shall be legiblyand
indeliblymarked on its stem with the following:-
L--?
--··- ·-----?----------,
578 THE,GAZE'JTE OF .INDIA :.EXTRAORDINARY [PARTII-SEC. 3(i)]
(a) TheJetter ?c near the top of the.scales;
(b) The word.Oral or Rectal,..as the case may
be;
,.,
(c) The manutacturers name or his recognized
trade-mark,"lf any;
(d) A code number.totrace the batch of
manufacture;
(e) A recognizedmark of the verifying.authority
showing the year:of initial .verification;
(f) An Indication ldentifying'.theglass,used for
the bulb; lftheglass is not already Identified
by its maker;:and
(g) The StandartfMark ofthe:Bureau oflndian
Standards.
(2) Packing
Clinical thermometers shall be.securely packed,
individuallyandcollectlvely, in anymanner acceptable
to the purchaser soas to minimize the risk of damage
in handling, transport and .storage.