Annexure B, Legal Metrology (General) Rule, 2011: ANNEXURE B

Legal Metrology (General) Rule, 2011 - Annexure B

Text of annexure B

ANNEXURE B

ANNEXURE B TESTING OF GAS ELIMINATION DEVICES (Informative) 1. Testing of a gas elimination device as a unit separate from the measuring system for which It is intended (1) General provisions In order to examine whether the pattern of a gas elimination device complies with the requirements in, paragraph 2(10), a specimen of the pattern must be installed on a suitable test bench equipped with a meter and a conventional proving tank. Note: In tests on gas elimination devices, the proving tank may be replaced by any appropriate standard. The efficiency of the gas elimination device is determined with reference to the meter error at the same flowrate. The test bench must, as far as possible, comply with the following provisions:- (i) The capacity of the proving tank should be at least equal to the greater of the following two values: volume delivered in one minute at maximum flowrate, or 1000 times the scale interval of the meter of the test bench. (ii) It is recommended that an adjustable non­ return valve be installed downstream of the meter in order to prevent back flow of the liquid which has been measured and to obtain the minimum back pressure required for the proper operation of the gas elimination device. (iii) There should be no reverse gradient in the pipework downstream of the meter so that gas bubbles are allowed to escape in the normal way in order to keep this pipework filled to the same level at the beginning and end of the test. (iv) The liquid used for the tests should either be the same as that for which the device is intended or should be of a viscosity which is at least equal to that of the liquid for which it is intended. Tests on gas elimination devices should be carried out for flowrates up to a maximum of 100 m'/h. For higher flowrates, characteristics may be determined by analogy with equipment of the same design and smaller dimensions. "By analogy" means that parameters like Reynolds number, Froude number, etc., are to be taken into account for the gas elimination device. (2) Tests on gas separators The volume of air or gas continuously entering may be measured by a gas meter and isothermally converted to atmospheric pressure on the basis of the indication of a pressure gauge fitted upstream of the gas meter. A pressure gauge positioned upstream of the meter for liquid makes it possible to determine the lowest pressure at which the gas separator still meets the efficiency requirements. ['11'1 n-= J(i)J 569 Before starting a test, the whole apparatus is made to operate at the desired liquid and gas flow rates so that all parts of the apparatus (except the proving tank) fill up under set conditions as regards the entry of air or gas. The air may be introduced either by injection downstream of the pump or by suction upstream of it (see Figures 1, 2 and 3, which are given as examples). In the former case, which makes it possible to operate without changing the performance of the pump due to the entry of air, the liquid and gas flows are adjusted by means of control valves. The air or gas is introduced through a tube positioned in the centre of the pipework for the liquid, for example at an elbow. In the latter case, which reproduces the conditions encountered in reality (pressure reduction by suction), the pump must be set to the maximum flowrate of the separator. If the pump has too great a flowrate,.it must be possible to regulate it with a speed reducer. The pump should preferably be of the volumetric type but it may also be of the centrifugal type if the supply tank feeds the pump by gravity. The pressure reduction must then be regulated by a valve positioned upstream of the pump, and the air Inlet must be equipped with a non-return valve which prevents any leakage at the movement of switching off. (3) Test on gas extractors An example of test bench is shown in Figure 4. It includes a container for creating a pocket of air to be removed with a volume equal to the minimum measured quantity of the gas extractor (the minimum measured quantity of the system being not yet specified). When the test is carried out with a proving tank having such a great capacity that the maximum permissible error cannot be evaluated correctly on the basis of a single operation of the gas extractor, the number of operations during the same test shall be multiplied by 2, 3 or 4, to obtain the required accuracy. (4) Tests on special gas extractors Special gas extractors, mainly used for measuring systems on road tankers, are principally intended to prevent measurement errors which may arise from the complete emptying of one compartment. They must also separate and continuously remove introduced air, although to a lesser degree than a gas separator. In the case of separate approval, they should be tested on a test bench which corresponds in principle to Figure 5. This test bench is similar to that in Figure 1 but it differs In order to reproduce the actual conditions of delivery from road tankers to underground tanks, as is the case when service stations are supplied with petrol, and domestic fuel oil deliVerles are made. Thus, 1 the supply tank is located above the special gas extractor and the meter, l.e, at a level corresponding to that of road tanker, and the proving tank is approximately 4 m below the meter. In order to determine the additional error arising from the complete emptying of a road tanker, which may be caused by the air drawn In at the end of the delivery by the formation of a vortex, the supply tank is filled with a volume of liquid equivalent to that of the proving tank. The liquid is then emptied through the meter into the proving tank without operating the shut-off valve. For delivery by gravity, pipework is used which bypasses the pump. Due to the existence of the automatic shut-off valve in the hydraulic circuit connected to the gas extractor, the pocket of air to be evacuated (described in Annexure-8, paragraph 1(3)) may be created by emptying the pipework between the tank and the gas extractor. Furthermore, given that a special elimination device should also perform the function of a separator ,._for a small percentage of air introduced continuously it is necessary to carry out this test in a manner comparable to the tests described for separators In Annexure-B, 1(2), with air being either injected into the supply pipework or drawn In upstream of the . pump by creating an entry of air and partly closing the valve of the supply tank. 2. Tests on gas elimination devices forming part of a measuring system during pattern approval The tests are carried out with a proving tank of the capacity specified in Annexure B, paragraph 1(1) or any appropriate standard. (1) Tests on a gas separator This examination particularly applies to patterns of separators included in measuring systems which can be mass produced and transported without dismantling, such as petrol pump fed by their own supply pumps. The essential part of the test bench (Figure 6) is the measuring system itself (in this case, the fuel dispenser). In accordance with conditions encountered in actual use, the liquid is drawn up from a - - --? --? __ J??--- 570 THE GAZETTE OF INDIA: EXTRAORDINARY [PARTl!?SEC. 3(i)] tank on a lower level than the meter. The air is drawn in by suction through a special Inlet equipped with a control valve. The air can be measured by.a gas•meter, :However, It is not necessary.to.use a ..gas meter if the separator is capable .of ,separating and eliminating ,the .air Introduced in any proportion, as provided in, paragraph 2(10)(vlil). The.tequlrements :In, paragraph 2(10)(1) and ?(lO)(vlll) should be complied with under test :conditions such that the maxlmumflowrate of the measuring system Is reached when no air enters. (2) Tests on a gas extractor and a special gas extractor The measuring system comprising the gas elimination device must be constructed so that the tests can be carried out as described in Annexure B, Paragraph 1(3) or 1(4). 3. Tests on gas elimination devices forming part of a measuring system during verification The gas elimination devices are tested without It being necessary to verify that the maximum permissible errors for the individual units are compiled with. Gas separator to be tested ., ,!l"' :, ., "' E"' i'! @ i(l ? "' li'! 0.. ? {g " ? ? FIGURE-1 TEST BENCH FOR GAS SEPARATORS [ 'IPl Il-<!i"S 3( i) ] 571 proving tank Gas meter (as required) Vacuum gau e Calibrated air inlet Gas separator to be tested) Flow meter Non-return valve (adjustable) ,.,--------+--t---+-t--? Supply tank Non-return valve FIGURE-2 TEST BENCH FOR GAS SEPARATORS 572 THE GAZETTE OF INDIA: EXTRAORDINARY [PART!I-SEC. 3(i)] Gas Cylinder CompressorSupply tank Gas meter : Pressure gau e as separator (with float to be tested [:--+ntainer for Iquid ??--·- Pressure gauge Flow meter Tarottle Pump for·· valve gas Suction Flow meter i I Supply tank -------,- L..a:x1-.J--->i--.J Pump FIGURE-3 TEST BENCH FOR GAS SEPARATORS .............................. with gas injection ....... with gas suction .f I'."111:,1Y.:;;\;0;l l. "'" ·.J"'l>rn·•N (<) ht. . 1,,,1,,,1 T?11l o» i I i --J · I ·.Ii . i.. (/.-•... · la,i ·J11·11scpa11?dn11 of ?i, l1M?<'.IO /'lllHJ! fol ri\/ht/', ,upply 1"nk l .. (J·-1c,[ :;,,pp)y ! . l"""I' FIGURE-4 TEST BENCH FOR SPECIAL GAS EXTRACTORS ['WT n-= 3(i)J 573 Supply tank Connection for automatic closing ?? Gas indicator Empty hose compressor Proving tank I Pressure gauge ···f9- Special gas extractor to be tested Gas meter (for injection) Flow meter - Pressure gauge Pump \ Gas meter (Forsuction) Non return valve FIGURE-5 : Test bench for special gas extractors Pressure gauge Gas indicator (If necessary} Flow meter Gas meter if necessary (see point 2.1 of Annex A) Valve Air inlet Gas separator to be tested Pressure gauge Valve 11 r:ov,ny:ank ? ------·········--····· Supply tank FIGURE-6 TESTING FACIUTY FOR GAS SEPARATORS IN FUEL DISPENSERS 574 THWGAZETTE'OF J!NDIA: EXTRAORDINARY [PARTII-SEC. 3(i)] PARTV VOLUMETRIC CONTAINER TYPE UQUIO ,MEASURING DEVICE 1, General The volumetric container type.liquid measuring device consists of.a·butket, afloat ·and a:dlp stick suitably graduated to Indicate the volume.cf liquid at different heights In the bucket. The •device is generally usedforrneesurtnq.the quantltv.otmllkat milk purchasing centres. 2. Nominal ?pacltles (a) The unit .of .volurne -shall 'be the cubic dedmetre.or.litre.ror the cubic centimetre (cm) or mlllllltre. (b) The volumetric container type liquid measuring device may be one of the following,capacities .: 5 dm3, 10 dmi, 20 dm3 and 50 dm3 (c) The value of ,the smallest graduation on the dipstick shall be of the form of 1 x !On, 2 x 10n or 5 x 10n where 'n' represents a whole number, positive or negative or is equal to zero. 3, General requirements (a) The bucket shall be made of suitable metal or alloy. The metal or alloy used shall be thermally stable, shall resist deformation, shall not have an unduly high coefficient of cubical expansion, and shall not affect the liquid being measured in any way or be injurious to health. Some of the materials considered suitable are : (i) mild steel, (ii) stainless steel, (iii) brass sheet, (iv) copper sheet, (v) aluminium alloy. (b) The wall thickness of the bucket shall be so selected that the bucket will not get dented In normal use or become unusable after a few years in service. (c) The bucket shall be free from surface defects and indentation. External and internal surfaces of the bucket made of mild steel, copper sheet and brass sheet shall be well tinned or tin plated. (d) All seams shall be filled and smoothened to prevent the entrapment of air or liquid .. (e) The bucket shall be provided with a well · formed and proportionate spout to facilitate pouring of liquid. (f) The bucket shall be cylindricalin form. The bottom of the bucket shall be slightly concave to prevent change of shape due to the weight of the liquid. The maximum depth of the concavity shall not be more than 20 mm. (g) The bottom of the bucket shall be reinforced with angle strip of thickness not less than 5 mm. (h) The upper edge of the bucket shall be reinforced round the circumference with a reinforcing band having a thickness not less than 5 mm. (i) The bucket shall be provided with a suitable handle on the side opposite the spout. (j) The top of the bucket shall be provided with a cross band across the diameter. The band shall have groove of appropriate size and centrally located for Inserting the dip stick. (k) The float shall be suitably fabricated so as to be free from holes, pockets, dents or crevices. A dip stick shall be firmly welded on the upper centre of the float. The dip stick together with the float shall be so constructed that the device so formed maintains verticality In all positions. (I) The dip stick shall have a rectangular cross· section of minimum dimensions 20mm x 10mm. The graduations shall be made by engraving or other means on both sides of the vertical surface. (m) The dip stick shall be graduated at suitable Intervals throughout the nominal capacity of the measure. (n) The graduation lines on the dip stick shall be clear straight, perpendicular to the axis of the stick and of uniform thickness not exceeding 1 mm. (o) The dip stick shall be Identified with the bucket by a number of Identifications, which shall be clearly legible and indelible. 4. Marking The following inscriptions shall be clearly and indeliblymarked at a conspicuous place on the bucket or on a special plate securely attached to the bucket:- (a) nominal capacity, . (b) manufacturer's name or trade mark, (c) the words "for edible liquids". ['WT II-? 3(i)] 575 5.Sealing A suitable plate or other device shall be provided to receive the stamp or seal of the verification authority. PART VI CLINICAL THERMOMETER PART A CLINICAL THERMOMETER-SOLID STEM TYPE Scope : This part specifies the requirements and methods of tests for solid stem type mercury in glass clinical thermometers having a maximum Indicating device. 1. Terminology For the purpose of this standard the definitions given in IS 2527-1979 Glossary of terms relating to liquid-in-glass thermometers (first Revision) as revised from time to time In addition to the following, shall apply. 2. Types The thermometers shall be of the solid stem mercury-in-glass type. 3. Temperature Scale · The thermometers shall be graduated in degrees Celsius (°C) and shall have a range from 35° to 42°C or 35° to 43°C. 4. Testing Thermometers shall be tested for total vertical immersion. S, Requirements (1) Pattern.?-There shall be two patterns of bulb, namely, oral and rectal as follows:­ (a) Oral-for thermometers for use in mouth and (b) Rectal-for thermometers for use In rectum Note 1: Oral thermometers may also be used in armpit or groin. Note 2: Rectal thermometers may also be used in the mouth, armpit or groin after proper disinfection. Rectal thermometers if used orally, may however give · slightly lower reading. (2) Materials (i) Glass tubing-The thermometers shall be made from glass tubing conforming to IS 4529-1958 Specification for glass tubes for medical thermometers. (ii) Bulb-The thermometer bulb shall be made from a type of glass which assures that the depression of zero, determined in accordance with the procedure given in Appendix D, does not exceed 0.07°C. The glass shall be identified· visillle and indelible, either by the glass manufacturer or by the manufacturer of the thermometer. (iii) The types of glass used for the maximum device, capillary tube, and bulbs shall meet the following requirements:- When the glass is analyzed according to the requirements of IS 2303 (Part I/Section 1), the quantity of alkali passed Into solution for lg of glass must correspond to not more than 263 .5g of Na,O. (iv) Thermometric liquid ? The thermometric liquid shall be pure, dry mercury. (3) Construction (I) The stem shall be in alignment with the bulb. (a) The free end of the stem shall be finished smooth, preferably hemispherical in shape, as shown in Figure 1. FIGURE 1 STEM OF CLINICAL THERMOMETER 576 ··THE;GAZET,'f.E <OF INDIA: EXTRAORDINARY [PARTJJ-SEC. 3(i)] (ii)• Maximum ,iRdiGating devlce · (constridion)-:The thermometers . .shall-have -e.maxlmum .indicating ·:,device (constriction) constructed ,in !'the capillary,· between the ·mercury bulb and the scale in order to prevent the mercury :in the .stern from returning to the bulb on -cooling. It shall be locatedin the bore between : the top of the bulb and the bottom ::of the. scale. '(a) The:maximum indicating device shall •pass the hardness test prescnbedfn Appendix A when an ·acceleration of 600 m/sec2 is applied at the dosed end of the bulb for a period of at least 2 minutes. (iii) Locator-A.locator may be drawn on the stem before the start of the scale to facilitate positioning the thermometer so that the mercury column Is in full view. The locator may comprise a thick ( 1 mm of more) dot or line or a rectangle of 1 x 3 mm . (Iv) Bu/1>-The bulb shall be cylindrical with a smooth rounded end and shall be in alignment with the stem, (a) The diameter of bulb and the Joint of bulb with stem shall nowhere exceed that of the stem. The joint shall be smooth and regular. (b) The bulb shall be hemispherical at the tip. (v) The bulb, capillarytube and mercury shall be sufficiently free from entrapped gas, debris and foreign bodies in order to ensure the correct functioning of the thermometer. (4) Dimensions The dimensional and scale requirements for solid stem type clinical thermometers, shall be as given in Table 1. TABLE 1 DIMENSIONAL AND SCALE REQUIREMENTSFOR SOLID STEM CLINICAL THERMOMffiRS SI. No. Char1:1cteristic (1) (2) (i) Scale range, °C (ii) Smallest scale division, °C (iii) Over all length, mm (iv) Scale length, mm (v) External diameter of stem, mm (vi) Bulb length, mm (a) Oral thermometer (b) Rectal thermometer (vii) External diameter of bulb, mm (a) Oral thermometer (b) Rectal thermometer (viii) Distance from top of the constriction to the 35.5°C mark, mm, Min, (ix) Distance from the highest graduation line to top of stem, mm, Min (x) Scale spacing, mm Min Note: See Figure 3 in IS 8757 : 1977 Requirements -----· ------ ----- (3) 35.0 to 42.0 or 35.0 to 43.0 0.1 100 to 115 40 to 65 4 to 6 12 to 18 Not more than 9 2.0 to 3.5 3.0 to 6.0 (But not exceeding that of stem) 10 8 0.5 [ '1l'T u-iil,rg 3(;) l 577 (5) Graduation and Numbering (i) The thermometer scale shall be sub­ divided in 0.1 °C as shown In Figure SSQ. (ii) The graduation lines shall be equally spaced and at right angles to the axis of the thermometer. (iii) All graduation lines shall be of equal and uniform thickness of not more than 1/4 of the length of the smallest scale division or 0.15 mm whichever Is less. (iv) The pattern of graduation and numbering shall be as follows:­ (a) Every tenth graduation line shall be a long llne (about 2 mm) which shall be numbered. (b) There shall be a medium line (about 1.5 mm) midway between two consecutive Jong lines, and (c) There shall be four short Jines (about 1 mm) equally spaced between consecutive medium and Jong lines. (v) When the thermometer Is held in a vertical position and viewed from the front, the left-hand ends of the graduation lines shall Ile on an imaginary vertical ltne (see Figure 55Q). (vi) The numbers shall be pla?ed to the right of the axis in such a way that extension of the line to which they refer, would bisect them. The figures shall be placed parallel to the axis of the thermometer. (vii) A distinguishing mark (say, an arrow) shall be marked at 37.0°C mark to indicate the normal temperature. (viii) All graduation lines and numbering shall be clearly and permanently marked or etched on the stem and filled with a pigment. The marking shall pass the test for permanenc.y given in Annex B. Note : The pigment may be black, red, green or blue as agreed to between the purchaser and the supplier. 6, Petformance requirements (1) Appearance When heated to 42°C or so the mercury column shall look like a strip of uniform width with clarity and brightness throughout. · Note : This test takes care of all visual defects, for example, twist in the glass, aberrations, cloudiness, devitrification, etc. (2) Ageing and Accuracy (i) The clinical thermometers shall meet all the requirements after ageing by natural or artificial means. Natural Ageing involves holding the thermometers at room temperature for four months after completion of the construction. Artificial ageing method involves heat treatment of the bulb and . constriction of the unfilled thermometer at a temperature and duration to be specified by the manufacture of the glass tubing, (II) Clinical thermometers shall after ageing ensure that their accuracy as measured by the method prescribed In Appendix C shall be ± 0.1 °C for at least one year after the first official verification. (3) lnffuence of Immersion Time If a thermometer at temperature t, ( 15°C t, < 30°C) is suddenly immersed in a well-stirred water · bath having a constant temperature t, (35.S°C < t, < 42°C) and is withdrawn after 20 seconds the thermometer reading, after cooling to ambient temperature ( 15°C to 30°C) (I} shall comply with maximum permissible error requirement, and (II} shall not deviate from its stabilized reading for temperature t, •c by more than 0.005 (t, - t,)°C. This stabilized reading is the thermometer reading obtained when the thermometer has been cooled to ambient temperature, after reaching complete thermal equilibrium with the water bath at a temperature t2°C.This reading shaU also meet the maximum permissible error requirements stipulated In paragraph 6(2)(ii). Note : A free chclce of test method Is permitted, provided the law of the variation of the Indication of the thermometer as a function of immersion time is known. 7. Marking and Packing (1) Marking Each clinical thermometer shall be legiblyand indeliblymarked on its stem with the following:- L--? --··- ·-----?----------, 578 THE,GAZE'JTE OF .INDIA :.EXTRAORDINARY [PARTII-SEC. 3(i)] (a) TheJetter ?c near the top of the.scales; (b) The word.Oral or Rectal,..as the case may be; ,., (c) The manutacturers name or his recognized trade-mark,"lf any; (d) A code number.totrace the batch of manufacture; (e) A recognizedmark of the verifying.authority showing the year:of initial .verification; (f) An Indication ldentifying'.theglass,used for the bulb; lftheglass is not already Identified by its maker;:and (g) The StandartfMark ofthe:Bureau oflndian Standards. (2) Packing Clinical thermometers shall be.securely packed, individuallyandcollectlvely, in anymanner acceptable to the purchaser soas to minimize the risk of damage in handling, transport and .storage.

Previous and next

All provisions of Legal Metrology (General) Rule, 2011

Questions people ask

Is annexure B of Legal Metrology (General) Rule, 2011 still in force?

No repeal of Legal Metrology (General) Rule, 2011 is recorded.