Annexure A., Legal Metrology (General) Rule, 2011: ANNEXUREA

Legal Metrology (General) Rule, 2011 - Annexure A.

Text of annexure A.

ANNEXUREA.

ANNEXUREA. PERFORMANCE TEST FOR ELECTRONIC MEASURING SYSTEMS. (Mandatory). 1. General This Annexure defines the program of performance tests intended to verify that electronic measuring systems may perform and function as intended in a specified environment and under specified conditions. Each test indicates, where appropriate, the reference conditions for determining the intrinsic error. These tests supplement any other prescribed test. When the effect of one influence quantity is being evaluated, all other influence quantities are to be held relatively constant, at values close to reference conditions. 564 THE GAZETTE .OF INDIA: EXTRAORDINARY [PARTII-SEC. 3(il) The above tests' involve the electronic part of the measuring system or its devices. The following .rules .sha!l be .taken tnto consideration for these tests:- (i) Tests volumes Some: influence .quantities should have a constant effect on•measurementresults and not a -prcportlonal -effect related to the measured volume. 'the value of the slgnificant:faliltis related to the measured volume; ttherefore, ;in order to be able to compare.results obtained In different . laboratories, .it ,is necessary to perform a test on ,a volume corresponding to that delivered 'In one minute at the maximum flowrate, liut not less than the minimum measured ·quantity. Some tests, however, may require more than one minute, in which case they shall be carried out in the shortest possible time. (ii) Influence of the liquid temperature Temperature tests concern the ambient temperature and not the temperature of the liquidused. It is therefore advisable to use a simulation test method so that the temperature of the liquid does not influence the test results. (1) Dry heat (2) Cold Cold Test severities: To verify compliance with the provisions In paragraph 4(1)(1) under conditions of low temperature. Test procedure in brief: The test consists of exposure of EUT to a temperature of-25°C (classes C or I) or -10°C (class B) under "free air" conditions for a 2- hour period after the ELJT has reached temperature stability. The ELJT shall be tested at least one nowrate (or simulated flow rate): at the reference temperature of 20°C following conditioning, at a temperature of -25°C or -10°C, 2 hours after temperature stabilization, after recovery of the EUT at the reference temperature of 20°C. ( 1) Temperature : severity level 2: -10°C severity level 3: -25°C (2) Duration : 2 hours Number of test cycles: One cycle Maximum allowable All functions shall operate as designed. variations: All errors shall be within the maximum permissible errors. (3) Damp heat, cyclic Test method: Damp heat, cyclic(condensing) Object of the test: To verify compliance with the provisions in 4.1.1 under conditions of high humidity when combined with cyclic temperature changes. Test procedure in brief: The test consists of exposure Test method: Object of the test: (1) Temperature : severity level 2: 40°C severity level 3: 55°C (2) Duration : 2 hours Number of test Cycles: One cycle Maximum allowable All functions shall operate as designed. variations: All errors shall be within the maximum permissible errors. Test severities: Dry heat (non-condensing) To verify compliance with the provisions in paragraph 4(1)(i) under conditions of high temperature. The test consists of exposure of the ELJT to a temperature of ss-c (classes C or I) or 40°C (class BJ under "free air" conditions for a 2-hour period after the EUT has reached temperature stability. The ELJT shall be tested at least one flowrate (or simulated flowrate): at the reference temperature of 20°C following conditioning, at the temperature of 55°C or 40°C, 2 hours after temperature stabilization, after recovery of the ELJT at the reference temperature of 20°C. Test procedure in brief: Test method: Object of the test: ['WT II-? 3(i)] 565 Test severities: Test method: Test method: of the ELIT to cydic temperature variations between, 25°C and , the upper temperature of 55°C (class C or I) or 40°C (class B), maintaining the relative humidity : above 95% during the temperature changes and during the phases at low temperature, and at 93% at ; . the upper temperature phases. Condensation should occur on the EUT during the temperature rise. Standard . stabilizing period before and · recovery after the cyclic exposure are indicated in !EC . Publication 68-2-30. The power supply is not on when the ' influence factor is applied. (1) Upper temperature: severity level 1: 40°C severity level 2: SS°C (2) Humidity : > 93% (3) Duration: 24 hours Number of test cydes: Two cycles Maximum allowable After the application of the variations : influence factor and recovery: all functions shall operate as designed, and all errors shall be within the 1 maximum permissible errors. (4) Vibration Test method : Sinusoidal vibration Object of the test: To verify compliance with the . provisions in paragraph 4(1)(i) under conditions of sinusoidal vibration. This test should normally apply to mobile measuring systems · only. Test procedure in brief: The EUT shall be tested by sweeping the frequency in the specified frequency range, at 1 octave/minute, at the specified acceleration level with . a specified number of sweep cycles per axis. The EUT shall · be tested in its three, mutually ' perpendicular main axes, mounted on a rigid fixture by its normal mounting means. It i shall normally be mounted so that the gravitational force acts in the same direction as it would in normal use. The instrument Is non­ operationa I when the influence factor is applied. Test severities: (1) Frequency range: 10-150 Hz (2) Max. acceleration level 20 m.s-2 Number of test cycles: 20 sweep cycles per axis Maximum allowable After the application of the · variations: . influence factor and recovery: all functions shall operate as designed and all errors shall be within the maximum permissible errors. (5) Power voltage variation Variation in a.c, mains power supply (single phase) Object of the test: To verify compliance with the provisions in paragraph 4(1)(1) under conditions of varying a.c, mains power supply. Test procedure in brief: The test consists of exposure of the EUT to power voltage variations, while the EUT is operating under normal atmospheric conditions. Test severities: Main voltage : upper limit Vnom +10% lower limit: Vnom -15% Number of test cydes: One cycle Maximum allowable All functions shall operate as variations: designed. All errors shall be within the maximum permissible errors. (6) Short time power reduction Short time interruptions and reductions in mains voltage. Object of the test: To verify compliance with the provisions in paragraph 4(1}(i) under conditions of short time mains voltage interruptionsand reductions. Test procedure in brief: The test consists of subjecting the EUT to voltage interruptions from nominal voltage to zero voltage for a duration equal to half a cycle 566 THWGAZETTE OF IND.IA: EXTRAORDINARY '==="'[P==A==RT==l=J==SE=c=.3=(i?)] Maximum allowab' variations: bf line frequency, and reductions from nominalvoltage , to 50% of nominal for a duration equal to one .cvcte .of .llne frequency, The.rnalns voltage · intt)r;uptions and reductions ··shat! t.e repeatedten times with a time interval of.atleast ten Test severities: 11ifYo voltage interruptionfor a , ptcnxi, equal to half a cycle: · 5 J% ..voltage reductlon for a · ;:2/iod -equal to one cycle. · Number of test cycle:;: least ten lnterruptons and te;, reductions, each with a "rtnirnurn of ten seconds b;;tween tests. The interruptions and reducttons are repeated ,hroughout the time necessary to perform the whole test; for this reason, more than ten interruptions and reductions •nay be necessary. Maximum allow, (a) for interruptlr.e measuring variations: sy?:ems, either \ne difference be ·1?n th•e vo!v·•;c, indication d•, •ng the tbt. and the ir .iicat.on uncer reference c,ndit!ons shall ,,ot exceed the ,dues given in rnragraph 3(12) ·;:•rt-! or the mra,uring system ,:,al! detect and act upon a . i!g\·,ncant fauif; in compliance wit·: p?11Jgrar\h4(3)(1). (L'• For nun-interruptible :T c:;:3surin9 systems, the r:".:?rence bEtween the volume ::<Hr:otlon during the test and indication under reference on.:H:ions shall not exceed the '.'R!ues given in paragraph 3(12) :.:11t-l. (7) Bursts Test method: dectrical bursts Object of the test: Io verify compliance with the provisions in paragraph 4( 1 )(i) under conditions where clectrtcal bursts are supenmposed on the mains vo!tage. Test procedureIn brt:f The test consists of subjecting the EIJT to bursts of double exponential waveform transient voltages. Each spike shall have a rise time of 5 ns and a half amplitude duration of SO ns. The burst length shall be 15 ms, the burst period ( repetition time interval) shall be 300 ''11S. All bursts shall be applied during the same measurement or simulated measurement in symmetrical mode ?nd asymmetricalmode. Test severities : .l\mphlu\'.le(peak value) 1000 V Number of test cycks: /\t !c%t ten positive and ten nc92tive randomly phased t,ursts shall be applied at .,,,,., burc;ts are applied during ?! tl,e time necessary to ,>;.f.:,r m the test; to that p·.,, uosc more bursts than :nGicated above may be t·ecc·ssary. For interruptible measuring stems, either the difference -tween the vo'ume indication .:dng the lest and the dication under reference · »nditions sh&ii not exceed the . Jiuesgiven:,, paragraph 3(12) .,crt·l or the measuring system :,all detect and act upon a ':,;nificant fault, in compliance 'h para;,r;;ph 4(3)(1). • 'o For :ion-interruptibie ·,asurt ·g systems, the c, . ,renc '·;.2tween the volume 11·· :atic during the test and tt ·, :nd' .tion under refetence :, .·w:. , shall not exn°··,d the ,,., ,,,en in paragra,.h 3(12) pc . (8) Electrosta!,c· discha1, Test method: Ele: ,static discharge (ESD) Object of the test· Tc · ify compliance with the ::Cons in paragraph 4(1)(1) conditions of direct and :nc: . · >:electrostatic disc! ,arges. Test procedure in b!ici A ·:3citor of 15C pF Is ch: ,;ed by a sultaiile DC vc 'J9e source. The capacitor i, '•,en discharged through the !T by connecting one ·• :ninal to ground (chassis) ······-?--- .. -- ···--··?--· . [<Ji'! II-? J(i)] 567 'Iest sever.Jes: F,eqi.iency range 26-500 MHz 500-lOOOMHz field strength 3 V/m l V/m Modulation 80°/o AM lkHz Sine wave (a) For interruptible measuring systems, either the difference between the volume indication during the test and the indication under reference conditions shall not exceed the values given in paragraph 3(12) of part-I or the measuring system shall detect and act upon a significant fault, in electromagnetic field strength as spe?ifled by the severity leve!. The field strength can be generated In various ways: the strip line is used at low frequencies below 30 MHz (or in some cases 150 MHz)for small ELJTs; the long wire is used at low frequencies (below 30 MHz) for larger EUTs; dipole antennas or antennas with circular polarization placed 1 m from the EVT are used at high frequencies. The specified filed strength shall be established prior to the actual testing (without EVT in the field). The field shall be generated in two orthogonal polarizations and the frequency range shall be scanned slowly.If antennas with circular polarization i.e. tog-spiral or helical antennas are used to generate the electromagnetic field,a change in the position of the antennas is not required. When the test is carried out in a shielded enclosure to comply with international laws prohibiting interference to radio communications, care should be taken to handle reflections from the walls. Anechoic shielding may be necessary. "-13ximun: allowable variatlons: and the other via 330 ohms to surfaces which are normally accessible to the operator. The test includes the paint penetration method, if appropriate. For direct discharges the air discharge sha!I be used where the contact · discharge method cannot be applied. 8 kV for air discharges 6 kV for contact discharges N,:c,1r:::rc/ te<:t cycles: At each test point, at least ten direct discharges shall be' applied at intervals of at least ten seconds between discharges, during the same, measurement or simulated'. measurement. As for indirect discharges, a total of ten discharges shall be applied on · the horizontal coupling plane, · and a total of ten discharges for the various positions of the vertical coupling plane. :,,xim,Jm ;/owable (a) For interruptible measuring •;riation:,-. systems, either the difference ' between the volume indication during .tne test and the indication under reference conditions shall not exceed the' values given in paragraph 3(12) of part-I or the measuring system shall detect and act upon a significant fault, In compliance with paragraph 4(3)(i). (b) For non-interruptibl · measuring systems, th.. difference between the volume indication during the test and the indication under reference conditions shall not exceed the values given in paragraph 3(12), of part-I. (.'!) Ele·.:Uomagnetic susceptibility 1i'·,t n ,,thod: Electromagneticfields (radiated) 0!1',,ct of the test: To verify compliance with the provisions in paragraph 4(1)(i) under conditions of electromagnetic fields. Tcs!. prore:lure in brief: The EUT shall be exposed to 568 THE GAZETTE OF INDIA: EXTRAORDINARY [PARTIl-e-Ssc. 3(i)] compliance With paragraph 4(3)(i). · (b) F.or non-interruptlble measuring .svstems, the .difference betweenthe volume indication -duringthe test and .the indication under reference conditions shall not exceed the values given in paragraph 3(12) .otpart-I. (lO)· Perturbaticns ,on de voltage powered instruments Electronic measuring systems supplied with de voltage sha!Lfulfll the tests in Annexure A 4(1) to Annexure A, paragraph 4(9), with the exception of Annexure A, ·paragraph 4(5), 4(6) and 4(7) which are to be replaced by the following provisions:- General ·provision For under-voltages or over-voltages all errors.shall be within maximum permissible errors when the instrument is still operating. The under-voltage or over-voltage is applied for a complete measurement or part of a measurement. Provision applicable to instruments fed by the battery of a vehicle Tests pulses 1, 2 and 3 of the relevant part of ISO 7637 Electrical disturbance by conduction and coupling, Part 1: Vehicles with nominal 12 V supply voltage, and Part 2: Commercial vehicles with nominal 24 V supply voltage, are applied at the various severity levels specified in this Standard. Pulses are repeated for as long as necessary to complete the test. The pattern approva I certificate shall indicate, for each type of pulse, the maximum severity level met by the instrument.

Previous and next

All provisions of Legal Metrology (General) Rule, 2011

Questions people ask

Is annexure A. of Legal Metrology (General) Rule, 2011 still in force?

No repeal of Legal Metrology (General) Rule, 2011 is recorded.