Annexure A, Legal Metrology (General) Rule, 2011: ANNEXURE A
Legal Metrology (General) Rule, 2011 - Annexure A
Text of annexure A
ANNEXURE A
ANNEXURE A
TESTING PROCEDURES FOR AUTOMATIC
GRAVIMETRIC FILLING
(Mandatory)
[See paragraph 2(5))
Meaning of symbols
I = Indication
In = nth Indication
L = Load
AL = Additional load to next changeover point
P = I+ 1/2 d • 6 L + Indication prior to rounding
(digital indication)
E = P • L = Error
mpd = Maximum permissible deviation of each
fill from the average
EUT = Equipment under test
se = Setting error
mpse = Maximum permissible setting error
md = Maximum deviation of each fill from the
average
1.. Examination for pattem approval
(1) Administrative examination [paragraph 5(2)]
Review the documentation that is submitted to
determine if it Is adequate and correct. For pattern
approval the documentation shall include:-
1(l){a) metrologlcal characteristics of the
instrument,
l(l)(b) a set of specifications for the
instrument,
l{l)(c) a functional description of the
components and devices, and
l(l)(d) drawings, diagrams and general
software Information (If applicable),
explaining the construction and
operation, Including interlocks.
Consider any document or other evidence that
, the design and construction of the instrument
complies with the requirementsof this specification.
(2) Compare construction with documentation
[paragraphs 4(3) and 5(2)]
Examine the various devices of the instrument
to ensure conformity with the documentation.
(3) Technical requirements (Paragraph 3)
Examine the instrument for conformity with
technical requirements according to the checklist
given In the test report format.
·
(4) Functional requirements [Paragraphs 4(2} and
4(3))
Examine the instrument for conformity with
functional requirements according to the checklist
given in the test report format.
2. Examination for initial verlncation
(1) Compare construction with documentation
[paragraph S(3}(i}}
Examine the instrument for conformity with the
approved pattern.
(2) Descriptive markings [paragraph 3(10}]:
Check the descriptive markings according to the
checklist given in the test report format.
3. General test requirements
(1) Power supply
Power up the equipment under test (EUT) for
a time period equal to or greater than the warm-up
time specified by the manufacturer and maintain the
EUT energized for the duration of each test.
(2) Zero-setting
Using the manual or semi-automatic zero·
setting facility,adjust the EUT as closely as practicable
to zero prior to each test, and do not readjust it at
any time during the test, except to reset if a significant
fault has been indicated.
Status of automatic zero facilities shall be as
specified for each test.
(3) Temperature
The tests shall be performed at a steady
ambient temperature, usually normal ambient
temperature unless otherwise specified. The
temperature is deemed to be steady when the
difference between the extreme temperatures noted
during the test does not exceed one-fifth of the
temperature range of the instrument without being
greater than S°C, and the rate of change does not
exceed S°C per hour.
__J.
-
466 .THE .GAZETTE <OF :JNDIA: EXTRAORDINARY [PARTIl=-Ssc. 3(i)]
The handllng:of the instrument shall .be such
that no condensation .of -water .occurs on the
instrument.
(4) Control instruments
(l)Accuracrof tesf.system[paragraph 6(4)}
The control Instrument and standard weights·
used in testing shail.ensuretne-determmation of the
weight of test loads,andJills toan error not greater
than either:
(a) one-third ,of :the .maximum permissible
error of:theJnstrument t.a, In the case of
matertaltests, one-third of the maximum
permissible deviation and maximum
permissible preset value error (as
appropriate) for automatic weighing, if the
control Instrument or the device used for
control purposes is verified immediately
prior to the material test; or
(b) one-fifth of the maximum permissible
error of the Instrument i.e., In the case of
.
material tests, one-fifth of the maximum
permissible deviation. and maximum
permlsslble preset value error (as
appropriate) for automatic weighing, in
all other cases.
Note: Accuracy requirements for the test
system depend on the limits of error which depend
on the accuracy class. However, the class is
determined from the results of the tests. It is,
therefore, necessary that the metrological authority
responsible for testing should be informed of the best
accuracy class that may be achieved, prior to
commencement of testing.
(ii)Use of standard weights to assess rounding
error
(a) General method to assess error prior to
rounding;
For Instruments with digital indication
having a scale interval d, changeover
points may be used to interpolate between
scale Intervals, i.e. to determine the
indication of the instrument, prior to
rounding, as follows:-
·
At a certain load, L, the Indicated value,
I, is noted. Additional weights of say 0.1
d are successively added until the
indication of the instrument is Increased
unambiguously by one scale interval (I
+d). The additional load el, added to the
load receptor gives the indication, P, prior
to rounding by using the following
formula:-
P = I + 0.5 d - t. L
The error prior to rounding is:
E = P • L = I + 0.5 d - t. L - L
Example. An instrument with a scale interval,
d, of 5g Is loaded 1kg and thereby Indicates 1000g.
After adding successive weights of O.Sg, the indication
changes from 1000g to 1005g at an additional load
of 1.Sg. Inserted in the above formula these
observations give:
P = (1000 + 2.5 • 1.5) g = 10019
Thus the true indication prior to rounding is
1001 g, and the error prior to rounding is:
.
E = (1001 • 1000) g : + 1 g
(b) Correction for error at zero
Evaluate the error at zero load, (E 0) by the
method specified in paragraph 3(4)(ii)(a) of this
Annex.
Evaluate the error at load L, (E) by the method
specified in paragraph 3(4)(1i)(a) of this Annex
.
The corrected error prior to rounding, (Ee) Is:
Ec:E-Eo
Example: if, for the example in paragraph 3(4)(ii)(a),
the error calculated at zero load was:
· Eo = + 0.5 g
The corrected error is:
Ee = + 1 - ( + o.5) = + o.sg
4. Test program
(1) Patte,n evaluation [paragraph 5(2}(iil)}
(I) Paragraphs 1, S, 6, 7 and 8(1) of this
Annex shall normally be applied for
pattern evaluation.
(Ii) For instruments In which the weighing
function is provided by a non-automatic
weighing instrument that has been
approved of its model, the tests specified
In paragraph 4(l)(i) may be omitted.
(2) !nit/al verification [Paragraph 5(3)}
Paragraph 2 and paragraph 8(2) of this Annex
shall be applied for initial verification tests.
Static weighing test method (paragraph 5(4) of this
Annex) may also be used if necessary to verify the
indicator for the integral verification method of
material tests.
5. Static tests (pattern approval stage)
(1) General Paragraph 5{2}(il)
Electronic instruments or Instrument simulators
are required to have a load indicator, or an interface
['Wf Jl-- J(i)] 467
allowing access to quantity that can be calibrated to
provide an indication of load so that the effect of
influence quantities may be tested and the reference
accuracy class determined. This facility also enables
testing of warm-up time and zero-and tare-setting
where applicable. This static weighing tests are
normally done as part of influence quantity testing.
Limits for warm-up time tests and for accuracy
of zero-and tare-setting are derived from Paragraph
2(2), and are therefore dependent on the reference
accuracy class. Therefore the results of these tests
must be evaluated after the reference accuracy class
has been determined.
(2) Warm-up time [Paragraph 4(2)(iii)J
This rest Is to verify that in the period
immediately after switch-on, operation is inhibited
until the warm-up time is finished, I.e. until the
metrological performance can be maintained. The
method is to check that automatic operation is
inhibited until a stable Indication is obtained and to
verify that zero and span errors comply with the
specified requirements during the first 30 minutes of
operation. If the zero Is set as part of the normal
automatic weighing cycle then this function shall be
enabled or simulated as part of the test.
Other test methods which verify that
metrological performance is maintained during the
first 30 minutes of operation may be used:
(i) Disconnect the instrument from the power
supply for a period of at least 8 hours prior
to the test.
(ii) Reconnect the instrument and switch on
while observing the load indicator.
(iii) Check that It is not possible to initiate
automatic weighing until the Indicator has
stabilized.
(iv) As soon as the indication has stabilized,
set the Instrument to zero if this is not
done automatically.
(v) Determine the error at zero by the method
of para 3(4)(ii)(a) of Annexure A.
(vi) Apply a static load close to Max.
Determine the error by the method of
paragraphs 3(4)(ii)(a) and 3(4}(11)(b)of
this Annex.
·
(vii) Repeat steps (v) and (vi) after 5, 15 and
30 minutes.
(viii) From (v) verify that the zero-setting error
is not greater than the limit specified in
paragraph 3(7);
(ix) From steps (vi) and (vii) verify that:
(a) the error (corrected for zero error)
for a static load close to Max is not
greater than the limit specified In
·
paragraph 2( 4).
(b) after each tlme Interval the error at
zero is not greater than twice the
.
limit specified In paragraph 3(7).
Note. Zero-setting accuracy Is specified as 0.25
mpd so the additional allowance of 0.25 mpd is added
for variation of zero after the initial zero-setting. This
is consistent with paragraph 2(4) (rnpe for a static
test load) and paragraph 6(2)(11) of this Annex
(temperature effect on no load indication).
(3) Zero- and tare-setting paragraph 3(7)
(l)Generaf.
Unless It is clear that zero and tare functions
are performed by the same process then both
function shall be tested separately.
Zero-and tare-setting maybe by more than one
mode, for example:
(a) non-automatic or semi-automatic,
(bl automatic at switch-on,
(c) automatic at start of automatic operation,
(d) automatic as part of weighing cycle.
It is normally only necessary to test the accuracy.
ofzero-and tare-setting in one mode if it is clear
that the same process is used for each mode. If zero
or tare is set as part of the automatic weighing cycle
then this mode shall be tested. To test automatic
zero-or tare-setting it is necessary to allow the
instrument to operate through the appropriate part
of the automatic cycle and then to halt the instrument
before testing.
(ii) Accuracy of zero-settrng [paragraph 3(7)]:
( 1) Set the instrument to zero in a mode as
determined by paragraph 5(3)(i) of this
Annex.
(2) Add weights to the load receptor to
determine the additional load at which the
Indication changes from zero to one scale
interval above zero.
(3) Calculate the error at zero according to
the description in paragraph 3(4)(ii)(a) of
this Annex.
(iii) Accuracyoftare-settfng {paragraph 3(7)}:
Accuracy of tare shall be tested at the maximum
tare as specified by the manufacturer.
(1) Place the tare load on the load receptor
and allow the tare function to operate In
a mode as determined by paragraph
----?----------------?
---
468 THE ',GAZETTE OF INDIA: EXTRAORDINARY (PART!I-SEC 3(i)]
5(3)(1) of this Annex and In accordance
with the manufacturer's instructions.
(2) Add weights .to ·the load receptor to
determine the additional'load at which the
indication chanqes.frorn zero to one scale
interval .above zero.
(3) Calculate the error according to the
description In paragraph 3(4)(il)(a)of this
Annex.
(4) Static weighmg'test,method[paragraph S(2)(iii)}
Apply test.leadsfrom zero up to and including
Max, and simila?yremove the test loads back to zero.
The test loads selected shall include values close to
Max and Min and other critical loads as specified in
paragraph 6(2)(a), subject to requirements of this
Annex.
Determine :tbe error at each test load, using
the procedure of paragraph 3(4)(ii) of this Annex, if
necessary, to obtain the accuracy requirements of
paragraph 3(4)61)of this Annex.
It should be noted that when loading or
unloading weights, the load shall be progressively
Increased or progressively decreased.
(S) Determination of reference accuracy class
[paragraph S(2)(v)J
The static weighing tests during application of
Influence factors (as appropriate) shall be used at
pattern approval stage to establish the reference
value for accuracy class, l.e. Ref(x), as follows :-
(i) Perform static weighing tests for Influence
factors and loads as specified in this
Annex.
(ii) For each load determine the maximum
permissible error for influence factor tests
for class X(l), mpe(l)
(Refer to para 2(4) and to this Annex
where appropriate).
(iii) Calculate [Error/mpe(l)lfor each load.
(iv) From (Ill)determine the maximum value
of [Error/mpe(t)lfor all the Influence
factor tests, [Error/mpe(l)m"]
(v) Determine Ref (x) from [Error/mpe(l)m,,l
such that :
Ref (x)> [Error/mpe(lJl max and
Ref (x) = 1 x 10k
,
2 x 10k
,
or 5 x 10k,
the Index k being a positive or negative
whole number or zero. Values for
significant fault shall then be calculated
from the mpd for the reference class.
6. Influence factor and disturbance tests
(1) Test conditions
(!)General requirements:
Influence factor and disturbance tests are
Intended to verify that electronic
Instruments can perform and function as
Intended in the environment and under
the conditions specified. Each test
Indicates, where appropriate, the
reference condition under which the
Intrinsic error Is determined.
It Is generally not possible to apply the
influence factors or disturbances to an
instrument which Is processing material
automatically. The Instrument shall
therefore be subjected to the influence
factors or disturbances under static
conditions or simulated operation as
defined herein. The permissible effects of
the influence factors or disturbances,
under these conditions, are specified for
each case. When the effect of one
influence factor Is being evaluated, all
other factors are to be held relatively
constant, at a value close to normal. After
each test the instrument shall be allowed
to recover sufficiently before the following
test-
Where parts of the instrument are
examined separately, errors shall be
apportioned in accordance with paragraph
5(2)(iii)(c).
The operational status of the instrument
or simulator shall be recorded for each
test. When the instrument is connected
in other than a normal configuration, the
procedure shall be mutually agreed on by
the approving authority and the applicant.
(ii) Simulator requirements
(a) General
The simulator for influence factor
and disturbance tests should include
all electronic elements of the
weighing system.
(b) Load cell
The simulator should also include the
load cell and a means to apply
standard test loads. Where this Is
not possible, e.g. for high capacity
instruments, then a load cell
simulator may be used or
alternatively the load cell interface
-=,----,--,--',,--cc,-----,=cc-cc=====r-=
•.
:::.,.=--= ...
=--
..
t??c-. -
...
-?--? ... ?.,.r.1
..
===CCTc.r,1._cr.,,,::_:::-,rc.??-•=·•cc,c: •• r,cc= ..
_,cccc,.r==::::.,.=
..
tcc
....
-====?=--
['WT!!-- 3(i)] 469
may be modified to incorporate a
.
scaling factor to give the design
output for a small test load.
Repeatability and stability of a load
cell simulator should make It possible
to determine the performance of the
instrument with at least the same
accuracy as when the instrument Is
tested with weights.
(c) Interfaces
Susceptibility that would result from
the use of electronic interfaces to
other equipment shall be simulated
In the tests. For this purpose it is
sufficient to connect 3 m of interface
cable terminated to simulate the
interface impedance of the other
equipment.
(d) Documentation
Simulators shall be defined in terms
of hardware and functionality by
reference to the instrument under
test, and by any other docu
mentation necessary to ensure
reproducible test conditions. This
Information shall be attached to, or
be traceable from the test report.
(iii) Test limits for multi-load instruments :
For instruments where the fill may consist
of more than one load, the value of a
significant fault and the limit of error for
influence factor tests must be determined
by the metrological authority after
considering the design of the Instrument
and the method of test, such that the
effect on the fill is equivalent to the values
specified In paragraph 4(2)(v) specified
in Part I and paragraph 2(4) of Part II.
(a) Significant fault for multi-load
Instruments
The following examples show how
to determine the value of a
significant fault on selective
combination weighers and
cumulative weighers when testing.
(a-1) Significant fault for selective
combination weighers
A fault greater than 0.25 of the
maximum permissible deviation of
each fill (as specified in Table 1) for
in-service verification divided by the
square root of the average ( or
optimum) number of loads in a fill,
for a fill equal to the minimum
capacity multiplied by the average
(or optimum) number of loads In a
fill.
Example : For .a class X(l)
instrument with Min = 200g
designed for an average of 8 loads
per fill, fill = 1600g, the maximum
permissible deviation of each fill
from the average fill (as specified In
Table 1) for in-service verification is
1.5% = 24g. Hence the value of
significant fault Is :
0.25 X (24/ \ 8) = 2.12g
(a·ii) Significant fault for cumulative
weighers:
A fault greater than 0.25 of the
maximum permissible deviation of
each fill (as specified in Table 1) for
in-service verification, for a fill equal
to the rated minimum fill,divided by
the square root of the minimum
number of loads per fill.
Example : For a class X ( 1)
instrument with Max = 1200g and
rated minimum fill of 8 kg: 8 kg/1.2
kg = 6.67; therefore the minimum
number of loads per fill is 7.
The maximum permissible deviation (as
specified in Table 1) for the minimum fill
of 8 kg is 1.5% or 120g. Hence the value
of significant fault Is :
0.25 X ( 120/ Y7) = 11.34g
Note : This definition of significant fault for
cumulative weighers does not include Min.
A cumulative weigher would normally be
used at or near to Max.
(b) Limits of error for influence factor tests
The following examples show how to
determine the limit of error for influence
factor testing for selective combination
weighers and cumulative weighers when
testing.
(b-i) For selective combination weighers
the maximum permissible error for
470 THE GAZETTE OF INDIA: EXTRAORDINARY [PARTII-SEC. 3(i)]
Environmental Phenomena Test specification
Reference of 200c
Temperature Specified high for 2 hours
Specified low for 2 hours
s0c
Reference of 200c
(*) mpe : maximum permissible error
(!)Static temperatures (paragraph 2(5){i)
Static temperature tests are carried out
according to Table 3 given below :
TABLE 3
Object of the test: to verify complaince with
the provisions in paragraph 2(5)(1) under conditions
of dry heat (non condensing) and cold. The test
specified in 6(2)(ii) of this Annex may be conducted
during this test.
Test procedures in brief
Precondition : 16 hours
Mpe(*)
Mpe(*)
Mpe(*)
· Mpe(*)
Mpe(*)
Conditions
applied
Normal power supplied
and "on" for a time period
equal to or greater than
the warm-up time
specified by the
manufacturer. Power Is to
be "on" for the duration
of the test. The automatic
zero-setting should be
disabled.
2 hours at each
temperature under "free
air" conditions.
As specified in paragraph
2(5)(1)
Characteristic under test
(2) Influence factor tests
SUMMARY OF TESTS
Test
Stabilization :
Temperature sequence : Reference temperature of
20°C
Specified high tem
perature
Specified low temperature
Temperature of soc
Reference temperature of
20°c
Temperature :
Condition of the EUT :
Static temperatures Influence factor
Temperature effect on Influence ractor
no-load indication
Damp heat, steady state Influence factor
Power voltage variation Influence factor
Tilting In"uence factor
any:statlctest load during influence
factor Jests shall be 0.25 of the
maximum permissible deviation for
ln-.service verification for the
appropriate rnass of the fill divided
by the square mot of the average
(or optimum) number of loads per
fill.
Example : Class X(1) selective
combination weigher, where the
average number of loads per fill =
4. For a .static test load = 100g the
appropriate mass of the fill will be
400g for which the maximum
permissible deviation for in-service
verification is 3% i.e. 12g. Hence the
maximum permissible error for
Influence factor tests Is :
0.25' (12g/4) = 1.Sg
(b-il) For cumulative weighers the
maximum permissible error for any
static test load during influence
factor tests shall be 0.25 of the
maximum permissible deviation for
In-service verification for the rated
minimum fill divided by the square
root of the minimum number of
loads per fil I.
Example : For a class X(1)
Instrument with Max = 1200g and
rated minimum fill of s·kg : 8 kg/1.2
kg = 6.67; therefore the minimum
number of loads per fill = 7.
The maximum permissible deviation
(as specified in Table 1) for the
minimum fill of 8 kg is 1.5%, i.e.
120g. Hence the maximum
permissible error for influence factor
tests Is :
0.25' (120/ ,J 7) = 11.35g
Note: For cumulative welghers the
average number of loads per fill is
not known. Therefore It is not
possible to define the limit of error
for Influence factors in terms of
average loads per fill and
appropriate mass of the fill. The
above definition is based on
maximum load and rated minimum
fill.
["!T'Tn-= 3(i)J 471
Maximum allowable
variations :
(iii) Damp heat, steady state [paragraph 4(J)(ii)j
Damp heat, steady state test are carried out
according to Table 4.
The zero-setting and zero- tracking facilities shall be
enabled as for normal operation. Adjust the EUT as
i
close to zero Indication as is practicable, prior to the
test. The handling of the EUT shall be such that no
condensation of water occurs on the EUT.
· Maximum permissible variations : The change
in zero Indication shall not yary by more than the
maximum permissible error for Infiuence factor tests
tor· a load equal to the rated minimum fill, for a
temperature difference of soc.
Condition of EUT : Normal power supplied and
"on" for a time period equal to
or greater than the warm-up
time specified by the
manufacturer. Power is to be
"on" for the duration of the test.
Test specification
Upper limit tem
perature and
relative humidityof
85% for 2 days
To verify compliance with
the provrsrons of
paragraph 4(3)(i) under
conditions of high
humidity and constant
temperature.
None required.
One static test load close
to minimum capacity.
Normal power supplied
and "on" for a time
period equal to or
greater than the warm·
up time specified by the
manufacturer. Power Is
to be "on" for the
duration of the test.
TABLE 4
Environmental phenomena
Damp heat, steady state
Supplementary information :
Object of the test :
Condition of the EUT :
Precondition :
Test load:
Number of test cycles : At least one cycle.
Weighing test : After stabilization at the
reference· temperature
and again at each
specified temperature
conduct the following :
Adjust the EUT as close
to zero indication as
practicable. It is important
to ensure that the test
result is unaffected by the
automatic zero-setting
function which should
therefore be disabled. The
EUT shall be tested with
at least five different static
test loads (or simulated
loads) including Maximum
and Minimum capacities.
When loading or
unloading weights the
load must be respectively
increased or decreased
monotonically.
Record : (a) date and time
(b) temperature
(c) relative humidity
(d) test load
(e) indications
(f) errors
(g) functional perform-
ance
All functions shall operate
as designed. All errors
shall be within the
maximum permissible
errors specified in
paragraph 2(4 ).
(II) Temperature effect on no-load indication
Note : This test should not be performed for
instruments that have automatic zero-setting as part
of the automatic weighing process.
The instrument Is set to zero, the temperature
is then changed to the prescribed highest and lowest
temperatures as well as at S°C. After stablllzatiori,
the error of the zero Indication is determined. The
change in zero Indication per S°C is calculated. The
· changes of these errors per S°C are calculated for
any two consecutive temperatures of this test.
This test may be performed during the
temperature test [paragraph 6(2)(i) of this Annex].
___
_j
_
472 ;THE :G:A:ZETTE :of INDIA: EXTRAORDINARY · [PART!I-SEC. 3(i)]
Maximum allowable variations : All errors shall
be within the maximum permissible errors specified
In paragraph 2(4).
(Iv) Power voltage variation [paragraph 2(5)(iiJJ
Power voltage variation tests are carried out
•
according to Table 5.
Test specification
To verify compliance
with the provisions of
paragraph 2(5)(11)
under conditions of
voltage variations.
TABLE 5
None required.
Normal power supplied
and "on" for a time
period equal to or
greater than the warm
up time specified by
the manufacturer.
Adjust the EUT as close
to zero indication as
practicable, prior to the
test. If it has an
automatic zero-setting
function then the
Instrument should be
set to zero after
applying each level of
voltage.
At least one cycle.
The EUT shall be tested
with a test load
approximately equal to
the minimum capacity.
Zero-setting function
shall be in operation.
Stabilize the power
supply at the reference
voltage within the
defined limits and
apply the test load.
Record the followingdata :(a) date and time
(b) temperature
(c) power supply voltage
(d) test load
(e) indications (as
applicable)
(f) errors
(g) functional performance
Environmental phenomena
Reference voltage
Voltage variation Reference voltage + 10%
Reference voltage - 15%
Reference voltage
SUPPLEMENTARY INFORMATION :
Test sequence :
Test procedures in brief
Precondition :
Condition of the EUT :
Number of test.cycles :
Weighing test :
Object of the test :
-50% at reference
temperature. 85% at
upper limit temperature.
The reference tem
perature of 20°C at 50%
humidity. The upper limit
temperature at 85%
humidity. The reference
temperature of 20°c at
50% humidity.
'Reference temperature
0
-of.20 C and atthe upper
limit as specified In
:paragraph 2(S)(i).
Three hours.at reference
-temperature and '50%
humidity:n:wodays atthe
, upper .lirnlt temperature
-as specitledin ,paragraph
2{5){i).
Stabilization :
Relative humidity::
Temperature :
Record:
Temperature
humidity sequence :
Number of test cvdes :
Weighing-test and test
sequence:
At least one cycle.
Alter stabilization of the
EUT at reference
temperature and 50 %
humidity apply the test
load.
(a) date and time
(b) temperature
(c) relative humidity
(d) test load
(e) indications
(f) errors
Increase the temperature in. the chamber to
the upper limit and increase the relative humidity to
85%. Maintain the EUT at no load for a period of
2 days. Following the 2 days, apply the static test
load and record the data as indicated above. Allow
full recover of the EUT before any other tests are
performed.
/
['WT u-= J(i)J 473
(v) Tilting [paragraph 2(5}(iii)]
Note: This test only applies to instruments that will
not be penmanently Installed.
This test fs not required for mobile instruments
with a level indicator If It can be established that the
tilt can be adjusted to 1% or less.
Test method : Static tests whilst the
EUT is tilted.
Test sequence :
Maximum allowable
variations :
Object of the test :
Test procedure In brief :
Test severities :
Maximum allowable
variations :
Condition of EUT :
All functions shall
operate as designed.
All errors shall be
within the. maximum
permissible errors
specified in paragraph
2(4).
To verify compliance
with the provisions in
paragraph 2(5)(iii)
under conditions of tilt.
The test consists of
tilting the EUT both
forwards and back
wards, longitudinally
and from side to side
(transversely), while
observing the weight
indications for a static
test load.
Instruments without
level indicatorsshall be
tested at a tilt of 5%.
All indications shall be
within maximum
permissible errors
specified In paragraph
2(4).
Normal power supplied
and "on" for a time
period equal to or
greater than the warm
up time specified by
the manufacturer.
Power Is to be "on" for
the duration of the test.
Adjust the EUT In its
reference position (not
tilted) as close to zero
indication as practi
cable. If the instrument
is provided with
automatic zero-setting
it shall not be in
. operation.
The test sha 11 be
performed with a test
load approximately
equal to the maximum
capacity.
Record the zero
Indication. Apply the
.
test load and record
the Indication. Remove
the test load.
TIit the EUT longitu
dinallyto the appropriate
extent and record the
zero Indication. Apply
the test load and
record the indication.
· Remove the test load.
Without further adjust·
ment to any control
affecting metrologlcal
performs nee tilt the
EUT to the appropriate
extent in the opposite
direction and repeat
the static weighing
tests as above.
TIit the EUT in the
transverse direction to
the appropriate extent
and repeat the tests.
TIit the EUT in the
opposite direction and
repeat the tests.
In order to determine
the infiuence of tilting
on the loaded instru
ment, the indication
obtained at each tilt
shall be corrected for
the deviation from zero
which the instrument
had prior to loading.
(3) Disturbance tests [paragraph 4(1){iii)J
·(f)Short time power reduction
Short time power reduction (voltage dips and
474 THE:GAZETTE :Of 'INDIA: EXTRAORDINARY IPART II-SEC 3(i)]
short lnterruptionsj.tests are.carrled out.accordlnq
to Table 6.
Object of the test : To verifycompliance
with the provisions
in paragraph 4(1)
(iii)under conditions
of short mains
voltage Interrup
tions and reductions
while observing the
weight Indication for
a static load
approximately equal
to the minimum
capacity. -
Test procedures In brief:
Precondition : None required.
Condition of the EUT: Normal power
supplied and "on"
for a time period
equal to or greater
than the warm-up
time specified by
the manufacturer.
Adjust the EUT as
close to zero
indication as practi
cable, prior to the
test. Zero-setting
functions shall not
be in operation. Not
to be adjusted or
readjusted at any
time during the test
except the reset if a
significant fault has
been indicated.
At least one cycle.
The EUT shall be
tested with a test
load approximately
equal to the mini·
mum capacity.
Stabilize all factors
at nominal reference
con-ditions. Apply
the test load and
record the following
data:
•
(a) date and time
(b) temperature
(c) power supplyvoltage
(d) test load
(e) indications
(f) errors
(h) functional perfor
mance Interrupt the
power supply to
zero voltage for a
period equal to one
half cycle and
conduct the test.
During interruption
observe the effect
on the EUT and re
cord, as appropriate.
Reduce the power
supply to 50% of
nominal voltage far
a period equal to
two half cycles and
conduct the test.
During reductions
observe the effect
·
on the EUT and re·
cord, as appropriate.
Number of test cycles :
Weighing test and test
variations :
,, Testspeciflcation
TAB1:E6
-Interruptlcn from
.reference -voltage
.to. zero voltage ,for
.one-half.cycle.
Interruption from
-reference voltage
to 500/oofreference
vciltagefortwo half
.cvcles.
These mains vcltage
interruptions shall
be repeated ten
times with a time
interval of at least 10
seconds
Supplementary information to the test
procedures :
Environmental phenomena
Voltage dips and short
interruptions.
[ ,wr II -19'16 3 (i ) ] 475
Maximum allowable variations: The difference
between the weight
indication due to the
disturbance and the
indication without
the disturbance
either shall not
exceed the values
given in paragraph
4(2)(v), or the EUT
shall detect and act
upon a significant
fault.
Supplementary information to the test procedures :
Object of the test : To verify compliance with the
provisions in paragraph
4( 1)(111)under conditions
where electrical bursts (fast
transients) are super·
imposed on the mains
voltage while observing the
weight Indication for a static
test load approximately
equal to the minimum
capacity.
Test procedures in brief:
Test specification
8 kV air discharge
6 kV contact
discharge
Precondition :
Condition of the EUT :
Weighing test :
Stabilization :
Environmental phenomena
None required.
Normal power supplied and
"on" for a time period equal
to or greater than the warm·
up time specified by the
manufacturer. Reset the EUT
if a significant fault has been
indicated.
Before any test stabilize the
EUT under constant
environmental conditions.
With the single static load In
place record the following
with and without the
transients :-
(a) date and time
(b) temperature
(c) test load
(d) indications (as
applicable).
Maximum allowable The difference between the
variations : weight Indication due to the
disturbance and the
indication without the
disturbance either shall not
exceed the value given In
paragraph 4{2)(v) of Part !,
or the Instrument shall
detect and act upon a
significant fault.
(iii) Electrostatic discharge
Electrostatic discharge tests are carried out with
test signals and conditions as given in Table 8 below:
TABLE 8
Electrostatic discharge
A coupling/decoupling network shall be applied
for testing AC power ports.
TABLE 7.2
Input and Output DC Power Ports
Environmental phenomena Test specification
Fast transient common mode 0.5 kV (peak)
5/50 ns T 1/Th
5 kHz rep.
frequency
Environmental phenomena Test specification
Fast transient common mode 0.5 kV (peak)
5/50 ns T 1/Th
5 kHz rep. Frequency
(ii) Eledrica/ btJrsts{fast transient tests)
Electrical bursts tests (fast transient tests) are
carried out for two minutes with a positive polarity
and for 2 minutes with a negative polarity according
to Tables 7.1, 7.2 and 7.3 given below :
TABLE 7.1
Ports for signal lines and control lines
Environmental phenomena Test specification
Fast transient common mode 0.5 kV (peak)
5/50 ns T 11'Th
5 kHz rep.
Frequency
TABLE 7.3
Input and Output DC Power Ports
Note : Not applicable to battery operated
appliances that cannot be connected to the mains
while in use.
Note : Applicable only to ports or interfacing
with cables whose total length may exceed 3m
according to the manufactures functional
specification.
476 THE GAZETTE OF INDIA: EXTRAORDINARY [PARTII-SEC. 3(i)]
Stabilization :
Weighing test :
Maximum allowable : The difference between the
variations : weight indication due to the
disturbance and the
indication without the
disturbance either shall not
exceed the value given in
paragraph 4(2)(v) of Part I
or the instrument shall
Test method :
Object of the test :
Span stability.
To verifycompliance with the
provisions of paragraph
4(3)(111)after the EUT has
been subjected to the
performance tests.
Reference to standard : No reference to international
standards can be given.
The EUT shall be stabilized
at sufficiently constant
ambient conditions after
detect and act upon a
significant fault.
7. Span stability test [paragraph 4(3)(iii)]
Test procedure in brief :Thetestconslstsofobserving
the variations of error of the
EUT under sufficiently
constant ambient conditions
(reasonably constant
conditions in a normal
laboratory environment) at
various intervals, before,
during and after the EUT has
been subjected to
performa nee tests.
The performance tests shall
include the temperature test
and, if applicable, the damp
heat test. Other performance
tests listed in this Annex may
be performed.
The EUT shall be
disconnected from the
mains power supply, or
battery supply where fitted,
two times for at least eight
hours during the period of
the test. The number of
disconnections may be
increased if the manu
facturer specifies so or at the
discretion of the approval
authority in the absence of
any such specification.
In the conduct of this test,
the operating instructions for
the Instrument as supplied by
the manu-facturer shall be
considered.
Note : The 6 kV contact discharge shall be
applied to conductive accessible perts. Metallic
contacts e.g. In battery compartments or In socket
outlets are excluded from this requirement. Contact
discharge is the preferred.test method. 20 discharges
(10 with positive and 10 with negative polarity) shall
be applied on each accessible metal part of the
enclosure. The time :Interval between successive
discharges shall be at •feast 10 s. In the case of a
non conductive encosure, discharges shall be applied
on the horizontal .or vertical coupling planes. Air
discharges shall be used where contact discharges
cannot be applied. Tests with other (lower) voltages
than those given In :fable 8 are not required.
Supplementary Information to the test
procedures :
Object of the test : To verify compliance with
the provisions of paragraph
4(1)(iii) under conditions
where electrostatic
discharges are applied while
observing the weight
Indication for a static test
load approximately equal to
the minimum capacity.
Test procedures In brief:
Precondition : None required.
Condition of the EUT : Normal power supplied and
"on" for a time period equal
to or greater than the
warm-up time specified by
the manufacturer. Reset the
EUT if a significant fault has
been indicated.
Before any test stabilize the
EUT under constant
environmental conditions.
With the single static load
in place, record the
following with and without
electrostatic discharge :
(a) date and time
(b) temperature
(c) testload
(d) indications (as appli
cable).
['1T'TII-@="s==3(=i)?J================="lml===<!>T==(l=?q="='=mn==m(Ul=========================4==77
variations :
Test severities :
Test sequence :
Maximum allowable
(e) test load
(f} Indication
(g) errors
(h) changes in test location
and apply all necessary
corrections resulting from
verlatlcnsof temperature,
pressure, etc. between the
various measurements.
At the first measurement
immediately repeat zeroing
and loading four times to
determine the average
value of error. For the next
measurements perform only
one, unless either the result
is outside the specified
tolerance or the range of the
five readings of the initial
measurement was more
than 1/10 of the maximum
permissible variation.
Repeat this test at periodic
intervals during and after
the conduct of the various
performance tests.
Allow full recovery of the
EUT before any other tests
are performed.
8. Procedure for material tests
(1) Material tests at pattern evaluation [paragraph
5(1}{/i)(a) and paragraph 5(2)]
· Operational tests with material shall be done
on a complete instrument to assess compliance with
the technical requirements of sub-paragraph (3) with
material for the test load as specified in paragraph
S(l)(il)(a).
(i) Feeding device [paragraph 3(5)}
Check that the feeding device provides
sufficient and regular flow rate.
Check that any adjustable feed device has
an Indication of the direction of movement
corresponding to the sense of the
adjustment of the feed (where applicable).
For instruments using the subtractive
weighing principle check that residual
material retained at the feeding device
after each load is delivered, is negligible
relative to limits of error.
switch-on for at least five
hours, and at least sixteen
hours atter the temperature
and damp heat tests have
been performed.
Test duration : Twenty eight
days or over the period
necessary for the conduct of
the performance tests,
whichever is less. Time t
(days) between tests: 0.5 S:
t S: 10.
Test load : a static test load
near maximum capacity
(Max); the same test
weights shall be used
throughout the test.
The variation In the
indication of the test load
shall not exceed 1/ 2 the
absolute value of the mpe
for influence factor tests
paragraph 2(4) for the test
load a pp lied on any of the
(n) tests conducted.
n .2'..8. If the test results
indicate a trend more than
half the permissible
variation specified above,
conduct additional tests
until the trend comes to rest
or reverses itself, or until the
error exceeds the maximum
permissible variation.
None required.
Verified mass standards.
Adjust the EUT as close to
zero indication as
practicable before each test.
Stabilize all factors at
nominal reference
conditions. If the Instrument
Is provided with automatic
zero-setting It shall not be
In operation.Applythe test
load (or simulated load) and
record the following data :?
(a) date and time
(b) temperature
(c) barometric pressure
(d) relative humidity
lt74 4/711..- t zi o
Precondition :
Test equipment :
Number of tests {n):
·
Condition of the EUT :
I
478 THE 'GAZETTE ;OF INDIA: EXTRAORDINARY [PART!I-SEC 3(i)]
Number of fills [para 6(3)] :
Correction devices
[paragraph 6(2)(1)] :
Mass of test loads and fills
[paragraph 6(2)] :
Types of loads [paragraph
5(3Xi)and paragraph 6(2)(a)] :
(a) Separateverification method : the separate
verification method Is as defined In
paragraph 6(5)(1).
(b) Integral verification method: the integral
verification method Is as defined in
paragraph 6(5)(11).
(iii) Procedure for metrologicel material tests
STEPS:
(1) Set up the instrument In accordance with
paragraph 6(2) (d) and 6(2)(i).
(2) Select a preset value for the fill and set
the load value if different from the fill, In
accordance with paragraph 6(2). Record
the indicated preset value.
(3) Run the instrument to produce a number
of fills as specified in paragraph 6(3).
(4) Weigh all the fills by one of the methods
In paragraphs 6(5)(i) or 6(5)(ii).
(5) Calculate the average value of all the fills
In the test and the preset value error
[paragraph 2(3)].
(6) Calculate the deviation of each fill from
the average paragraph [2(2)].
(7) Repeat stages (2) to (6) for other loads
as specified in paragraph 6(2).
Concerning paragraph 6(7) : The result of
weighing the test fill
on the control
instrument shall be
considered as the
conventional true
value of the test fill.
Concerning Paragraph 6(8) : The deviation for
automatic weighing
used to determine
compliance of each
fill with the
maximum per·
mtsslble deviation for
automatic weighing
[paragraph 2(2)]
shall be the
difference between
the conventional true
value of the mass of
the test fill as defined
in paragraph 6(7)
and the average
value of all the fills in
the test.
Concerning paragraph 6(9) : The preset value
error for automatic
The materials
used as the test
load shall be as
specified in
paragraph 6(2)
(a).
The mass of the
test loads and fills
shall be as
specified in
paragraphs 6(2)
(a, band c),
Adjustments [paragraph 6(2) (d)] :The adjustments
shall be set as
specified in
paragraph 6.(2)
(d).
Any correction
device shall be
operated as
specified in
paragraph 6(2)(i).
The number of
fills shall be as
specified in
Paragraph 6(3).
(ii) Methods for metrological material tests
(paragraph 6(5)]
One of the following verification methods shall
be used:
(ii) Load receptoi: [paragraph 3(6)]
For instruments·thatweigh material in a
separate lcedreceotor.prlor to discharge
to a :·container, check ·that 'the :residual
materlal·retalned at the load receptor·after
each dlschai:ge, is :negligible ·,relative to
limits·ofcerror.
Check-.that,manual discharqe.of the load
receptor.ls.notposslbte during automatic
operation.
(2) Material tests. at Initial verification {pilragraph
S(J){ii)(b) and par.agrtJph.5(3)]
Metrologlcaltests with -material shall be done
on a complete Instrument, fully.assembled and fixed
In the position In •which It is intended to be used and
as specified in paragraph S(l)(ii)(b).
The accuracy class (or classes) shall be
determined from .the results,
(i)Requlrements formetrologica/ material tests
[ 'lT'T II - i;;ri:s 3 ( i ) ] 479
weighing used to
determine com
pliance with
. paragraph 2(3) shall
be the difference
between the average
value of the
conventional true
value of the mass of
the test fills, as
defined In para 6(7)
and the preset value
for the fills.
(iv)DetermlNJtlon of accuracy class [paragraph
S(2}{v)J
(1) For each preset value of the test fill,
determine the preset value error (i.e. the
setting error, se) and the maximum
permissible preset value error for class
X(l), mpse :
l
Then calculate [se/mpse ] for each preset
value of the test fill.
1
(2) For each preset value of the test fill
determine the maximum actual deviation
from the average (md) and the maximum
permissible deviation from the average for
class X(l), mpd .
l
Then calculate md/mpd for each preset
value of the test fill.
1
·
(3) From (1) determine the maximum value
of [se./mpse ], [se/mpse J max
l l
(4) From (2) determine the maximum value
of md/mpd 1 [md/mpd ] max.
1 l
(5) Determine the accuracy class oo such that
(x) > [se/mpse ] max
l
and (x) > [md/mpd J max
I
and (x) = 1 x 10k,2 X 10k,or 5 X 10k,
the Index k being a positive or negative
whole number or zero.
SEVENTH SCHEDULE
HEADING-E
DISCONTINUOUS TOTALIZING AUTOMATIC
WEIGHING INSTRUMENTS
(TOTALIZING HOPPER WEIGHERS)
PART I
1. General definitions
(1) Weighing instrument
A measuring Instrument that serves to
determine the mass of a load by using the action of
gravity. According to its method of operation, a
weighing instrument is classlfJed as automatic or non
automatic
.
(2) Automatic weighing instrument
An Instrument that weighs without the
intervention of an operator and follows a
predetermined programme of automatic processes
characteristic of the instrument.
(3) Discontinuous totalizing automatic weighing
instrument (totaliZing hopper weigher)
An automatic weighing Instrument that weighs
·
a bulk product by dividing It into discrete loads,
determining the mass of each discrete load in
sequence, summing the weighing results and
delivering the discrete loads to bulk.
(4) Electronic instrument
An instrument equipped with electronic devices.
(5) Control instrument
A non-automatic weighing Instrument used to
determine the mass of the product used as the test
load during material tests.
2. Construction
Note : Jn this Part and Part II, the term "device"
is applied to any part which uses any means to
perform one or more specific functions.
(1) Load receptor
The part of the instrument intended to receive
the load.
(2) Electronic parts
(I) Electronic device
A device comprised of electronic sub
assemblies and performing a specific
function. An electronic device is usually
manufactured as a separate unit and Is
capable of being Independently tested.
(Ii) Electronic sub-assembly
A part of an electronic device comprised
of electronic components and having a
recognizable function of its own.
(iii) Electronic component
The smallest physical entity that uses
electron or hole conduction in semi
conductors, gases or In a vacuum.
(3) Indicating device
The part of the instrument that displays the
value of a weighing result in units of mass.
(i) Totalization indicating device
The part of the instrument that indicates
480 ,THECGAZETif.E ,OF INDIA: EXTRAORDINARY [PARTII-SEC. 3(i))
the sum of consecutive loads weighed and
discha_rgedto:bulk.
(a) Princlpaltatalization· indicating device
The part.of the instrumenttnatlnolcates
the surnrof all.the loadswelqhed and
di sch a rged,to .bulk.
(b) Partial tatallzation.indicating device
The part of theinstrument that indicates
the sum ,of a ,limited number of
consecutlvelloads.dellvered to bulk.
(c) Supplementary .totalization indicating
device
An indicating.device-with a scale interval
greater than ,that of the principal
totalizatlon ·indicating device and
indicating the sum of consecutive loads
weighed over a fairly long period of time.
(ii)Control indicating device
An indicati11gdevice that enables the use of
the instrument as a control instrument to weigh
discrete loads for control purposes.
(4) Ancillarydevices
(i) Zero-setting device
The means used to set the weight
indicating device to zero when the load
receptor Is empty.
(a) Non-automatic zero-setting device
A zero-setting device that must be
operated manually.
(b) Semi-automatic zero-setting device
A zero-setting device that operates
automatically following a manual
command.
(c) Automatk zero-setting device
A zero-setting device that operates
automatically and without the Intervention
of an operator.
(ii) Printing device
The means to print the value of each discrete
load weighed in the load receptor, or the sum of
consecutive loads weighed and discharged to bulk
or both.
3. Metrologlcal characteristics
(1) Scale-interval
A value expressed in units of mass that is the
difference between
(a) the values corresponding to two
consecutive scale marks for analogue
indication, or
(b) two consecutive indicated values for digital
indication.
(I) Totalization scale Interval (dtl
The scale interval of a principaltotalizatlon
Indicating device.
(II) Control scale interval (d)
The scale interval on a control indicating
device.
·
(2) Weighing cycle
The sequence of weighing operations that
includes the following :-
(I) one delivery of a load to the load receptor,
(II) a single weighing operation,
(Ill) the discharge to bulk of a single discrete
load.
(3) Automatic weighing range
The range from minimum capacity to maximum
capacity.
(i) Maximum capacity (Mex)
The largest discrete load that can be
weighed automatically.
(ii) Minimum capacity (Min)
The smallest discrete load that can be
weighed automatically.
(iii) 11Jrgetload
The' preset value of the load in the load
receptor that causes the flow to stop In
each weighing cycle.
(4) Minimum totsttzed /o?d(?minl
The value of the smallest bulk load that can be
totallzed without exceeding the maximum permissible
error when the automatic operation is comprised of
discrete loads, each within the automatic weighing
range.
3(5) Warm-up time
The time between the moment that power Is
applied to an instrument and the moment at which
the instrument is capable of complying with the
requirements.
4. Indications and errors
(1) Methods of indication
(I) Analogue indication
An indication allowing the determination
of an equilibrium position to a fraction of
the scale interval.
[qpr [I-? 3(i)] 481
(ii) Digital Indication
An Indication In which the scale marks are
a sequence of aligned figures that do not
permit Interpolation to a fraction of a scale
interval.
(2) Errors
(i) Error (of Indication)
The Indication of an instrument minus the
(conventional) true value of the mass.
(ii) Intrinsic error
·
The error of an instrument under
reference conditions.
(iii) Initial Intrinsic error
The Intrinsic error of an instrument as
determined prior to perfonnance tests and
durablllty evaluations.
(iv) Fault
The difference between the error of
indication and the intrinsic error of a
weighing instrument.
Note 1 : Principally, a fault is the result of an
undesired change of data contained In or flowing
through an electronic Instrument.
Note 2 : From the definition it follows that in
this heading a "fault" Is a numerical value.
(v) Significant fault
A fault greater than dt,
The following are not considered to be
significant faults, namely :-
(a) faults that result from simultaneous
and mutually independent cause in
the Instrument or in Its checking
faclllty,
(b) faults that imply the Impossibility of
performing any measurement,
(c) transitory faults that are momentary
variations in the indications which
cannot be interpreted, memorized
or transmitted as a measurement
result, and
(d) faults that are so serious that they
will Inevitably be noticed by those
Interested in the measurement.
(vi) Span stability
The capability of an Instrument to
maintain the difference between the
indication of weight at maximum capacity
and the Indication at zero within specified
limits over a period of use.
'-/74 c,071,_,2,
(vii) Maximum span stability error
A span stability error greater than one half'
of the absolute value of the maximum
permissible error applicable to the load.
5. Influences and reference conditions
(1) Influence quantity
A quantity that is not the subject of the
measurement but which influences the value of the
measurand or the indication of the Instrument.
(i) Influence factor
.
An Influence quantity having a value within
the specified rated operating conditions
of the instrument.
(ii) Disturbance
An influence quantity having a value that
falls within the limits specified in this
International Recommendation but that
falls outside the rated operating conditions
of the instrument.
(2) Rated operating conditions
Conditions of use which give the ranges of the
influence quantities for which the metrologlcal
characteristics are intended to lie within the specified
maximum permissible errors.
CJ) Reference conditions
A set of specified values of Influence factors
fixed to ensure valid inter-comparison of the results
of measurements.
6. Tests
(1) Material test
A test carried out on a complete instrument
using the type of material that it is Intended to weigh.
(2) Simulation test
A test carried out on a complete instrument or
part of an Instrument in which any part of the
weighing operation is simulated.
(3) Performance test
A test to verify that the equipment under test
(EUT) Is capable of accomplishing its intended
functions.
(4) Span stability test
A test to verify that the EUT is capable of
maintaining Its performance characteristics over a
period of use.
PART II
1. General
{l)Scope
This specification specifies the requirements and
482
. :THE· GAZETi[E •OF INDIA: EXTRAORDINARY [PARTII-SEC. 3(i)}
TABLE 2
+ l.Odt 500 < m < 2000
+ 1.5 dt 2000 < m < 10000
+ 0.5 dt O < m < 500
1000 x dt for class· 0.2 Instruments,
400 x dt for class 0.5 instruments,
200 x dt for class 1 Instruments; and
100 x dt for class 2 Instruments.
Example : maximum capacity = 1000 kg
minimum capacity = 200 kg
totalization scale Interval = 0.2 kg
(see paragraph 2.4)
.
accuracy class of Instrument = 0.5
Emin 400 x 0.2 = 80 kg
But to satisfy the second indent above :
Emin ? min = 200 kg
Therefore, in this example the minimum value
of the minimum totalized load Is 200 kg.
(The values used in this example are not
intended to be typical.)
(6) Agreement between indicating and printing
devices
For the same load, the difference between the
Digital indications and printed results shall be
corrected for the rounding error, and the error shall
be determined with an accuracy of at least 0.2 dt,
(3) Form of the scale interval
The scale intervals of the indicating and printing
devices shall be in the form of 1 x 10k,2 x 10k, or
5 x 10k
,
"k" being a positive or negative whole
number 'or zero.
(4) Totalization scale interval (dt)
The totalization scale Interval shall be :
(a) not less than 0.01 % of maximum
capacity;and
(b) not greater than 0.2 % of maximum
capacity.
(S)Minimum value of minimum totalized load(Lm1n)
The minimum totalized load shall not be less
than :
(a) the value of the load. for which the
maximum permissible error for automatic
weighing on initial verification is equal to
the totalization scale interval (dt); and
(b) the minimum capacity (Min).
Note : It results from the first indent above
that the maximum permissible error on initial
verification (mpe in the follow Ing example) for a load
equal to I:min shall not be less than dt,
Therefore, using table 1, l:mfn shall not be less
than:
+ 1.0 %
+ 2.0 %
+ 0.25 % + 0.5 %
± 0.10 % · ± 0.2 %
Verification Inspection
Percentage of the mass of
the totalized load
2 + 1.00 %
1 + 0.50 %
0.2
0.5
test methods for. discontinuous totalizing automatic
weighing instruments {totalizil}g,hopper welghers),
[hereafter referred. to.as "mstrurnentstl.
It is intended to .provlde -standerdlzed
requirements and'test procedures to -evaluate the
metrological endttechntcal ·characteristics of an
Instrument in a unlform·andttraceable way.
(Z) Application
This specification appllesto instruments having
a load receptor ln·the'.form·cif,a:hopper.
This Part doesnotapply.tothe.following types
of Instruments :-
(a) "welghll}g,ln,motion",instwments,
(b) instruments thattotalize the bulk load by
multiplying the weight of a preset constant
load by;the number of weighing cycles.
2, Hetrologicahrequlrements
(1) Accuracy classes
Instruments shall be divided into four accuracy
classes as follows :-
0.2, 0:5 1, 2
(2) Maximum permtssibleerrors
(i) Automatic weighing
The maximum permissible errors for each
accuracy class shall be the appropriate values
in Table 1 given below rounded to the nearest
totalizatlon scale interval. Maximum
permissible errors apply to loads not less than
the minimum totaiized load (Iminl· ·
TABLE 1
(ii) Influence quantities
The maximum permissible errors applied
in tests to assess the effect of influence
quantities shall be as specified in Table 2
below:-
Maximum permlsslble errors Loed (m) expressed
in totallzatlon scale
intervals ·
Accuracy class
?._;::: -===-
['1T'TII-? 3(i)] 483
weighing results provided by any two devices having
the same scale Interval shall be as follows :-
(a) zero for dlgltal indicating or printing
devices;
(b) not greater than the absolute value of the
maximum permissible error for automatic
weighing for analogue devices.
(7) Influence qu11ntltles
(i) Static temperature
Instruments shall comply with the
appropriate metrological and technical
requirements at temperatures from - 100c
to+ 450c.
For special applications, however, the
limits of the temperature range may differ
from the above provided that this range
is not less than 30°C and Is specified in
the descriptive markings.
Instruments shall be tested In accordance
with the static temperatures test in as
specified In paragraph 8(3)(1) of Annex A.
(ii) Mains power supply (AC}
Instruments that are powered by an AC
supply shall comply with the appropriate
metrologlcal and technical requirements
when operated under variations of voltage
from - 15 % to + 10 % of the value
marked on the instrument.
Instruments shall be tested In accordance
with the mains power supply (AC) test as
specified In paragraph 8(3)(111)of Annex
A.
(iii) Battery power supply (DC)
Instruments that are powered by a DC
supply shall comply with the appropriate
metrologlcal and technical requirements
in accordance with paragraph 4(3)(viii).
Instruments shall be tested in accordance
with the battery power supply (DC) test
as specified In paragraph 8(3)(1v)of Annex
A.
3. Technical requirements
(1) Suitability for use
Instruments shall be designed to suit the
method of operation and the materials for which they
are intended.
(2) Security of operation
(i) Accidental ma/adjustment
Instruments shall be constructed so that
a maladjustment likely to disturb their
metrological performance cannot normally
take place without the effect being easily
detected.
(ii) Purging of load receptor
The design of the load receptor and the
operation of the Instrument shall be such
that the weighing results are not adversely
affected by any variation in the quantity
of the load remaining In the load receptor
after discharge during a weighing cycle.
(Iii) Automatic weighing conditions
An automatic operation shall be
Interrupted, printing shall be prevented
or marked and a warning signal shall be
given in the following cases :-
(a) if the maximum capacity (Max) has
been exceeded by more than 9 d;
and
(b) if the value of the load to be weighed
and discharged to bulk is less than
minimum capacity (Min), unless
processed as the last discrete load
of the transaction.
(Iv) Operational adjustments
It shall not be possible to make operating
adjustments nor to reset the indicating
devices during an automatic weighing
operation, with the exception of the
possibility to interrupt the weighing cycle
during testing as described in paragraph
6.3.1.
(v) Dust extraction
The operation of a dust extractor shall not
affect the result of the measurement.
(vi) Zero-setting device
Instruments that do not tare-weigh after
each discharge shall be provided with a
zero-setting device.
An interlock shall be provided to stop an
automatic operation If the zero indication
varies by:
(a) 1dt on instruments with an
automatic zero-setting device; or
(b) O.Sdt on instruments with a semi
automatic or non-automatic zero
setting device.
A zero-setting device shall be
capable of setting zero to± 0.25 of
the smallest scale interval of all the
l
- - . -
------ - -- ---
------"-?--- ------------------
484 THE •,GAZETTE OF INDIA: EXTRAORDINARY [PARTfl-SEC. 3(i)]
·Indicating devices of the instrument
.:and have.a ra?ge of,ac;IJustmentnot
··exceeding 4% of maximum capacity.
(vii) Fraudulent use
Instruments shall not have characteristics
likely-tofacilitate theirJrauduient use.
(3) Instruments with control indicating devices
For an instrument ·with a control indicating
device, the load ·receptor shall have the facility to
support a quantlty.ohtandard weights ln accordance
with Table 3 givend,elow::-
JABl:E 2
(4) Totalization Indicating and printing devices
Instrumentsshall lndude a principal totalization
Indicating device end may include a supplementary
totalization indicating device, partial totallzation
indicating devices, and printing devices.
On an instrument equipped with a printing
device, the following shall apply :-
(a) It shall not be possible to reset the
principal totalization indicating device to
zero unless the printing device
automatically prints the last total indicated
before resetting to zero.
(b) An automatic printout of the last total shall
be generated if the automatic operation
is interrupted and operating adjustments
can be made.
(i) Qua/tty of indication
A totallzation indicating and printing
device shall allow reliable, simple
and unambiguous reading of the
results by simple juxtaposition and
shall bear the name or symbol of
the appropriate unit of mass.
(ii) Scale interval
Except for a supplementary
totallzation indicating device, the
scale Intervals of all totalization
Indicating devices shall be the same.
{iii)Supplementary totettzetton
Indicating devices
The scale interval of a
o maximum capacity
Max =
... g or kg or t
o minimum capacity
Min =
... g or kg or t
supplementary totalization indicating
device shall be at least equal to ten
times the totallzatlon scale interval
indicated In the descriptive
markings.
(iv) Combined indicating devices
Two or more types of indicating
devices may be combined so that
the indication required can be
displayed on demand provided that
it is clearly Identified.
(S) Ancillary devices
Ancillarydevices shall not affect the indicated
totallzation(s) representing a bulk load for a
transaction.
(6) Sealing
Components that are not intended to be
adjusted or removed by the user shall either be fitted
with a sealing device or be enclosed. When enclosed,
it shall be possible to seal the enclosure.
(7) Descriptive markings
Instruments shall bear the following markings:
(i) Marking shown in full
o identification mark of the
manufacturer
o identification mark of the importer
(if applicable)
, serial number and type designation
of the instrument
o nature of product
o control scale interval
(if applicable) .... g or kg or t
o Electrical supply voltage .. ,
V
o Electrical supply frequency .. ,
Hz
o Working fluid pressure (if applicable)
kPa or bar
o Operational speed Max and min
(ii) Markings
o pattern approval sign in accordance
with national requirements
0.2, 0.5, 1 or 2• accuracy class
Minimum quantity of
standarti weights
Max
5 t
20 °/o Max
10 t
Max? 5 t
5 t < Max ? 2S·t
25 t < Ma x ·?'50 t
50t<Max
Maximum caf)dcl/y (Max)
['WT II-? J(i)J 485
• minimum totalized load
•min =
... g or kg or t
• totallzation scale interval
dt= .. gorkgort
• Output per hour at maximum
operating speed
• Model approval mark
(iii) Presentstton of descriptive markings
Descriptive markings shall be Indelible and
of a size, shape and clarity that permit
legibllltyunder rated operating conditions.
Markings shall be grouped together in a
clearly visible place on the Instrument,
either on a descriptive plate fixed near
the Indicating device or on the Indicating
device Itself.
It shall be possible to seal the plate
bearing the markings, unless the plate
cannot be removed without being
destroyed.
(8) Verification marks
(i) Position
Instruments shall have a place for the
application of verification marks. The
followingapplies for this place :
(a) the part on which the marks are
located cannot be removed from the
Instrument without damaging the
marks;
(b) the place shall permit the easy
appllcatlon of the marks without
changing the metrologlcal qualities
of the instrument; and
(c) the marks shall be visible without
requiring that the instrument or its
protective covers be moved when it
Is In service.
(ii) Mounting
Instruments required to bear verification
marks shall have a verification mark
support located as specified above, which
shall ensure theconse'rvation of the marks
as follows:-
(a) When the mark is made with a
stamp, the support may consist of a
strip of lead or any other material
with similar qualities inserted into a
plate fixed to the instrument or in a
cavity bored into the instrument;
(bl When the mark consists of an
adhesive transfer, a space shall be
provided for this purpose.
4. Requirements for electronic Instruments
Electronic instruments shall comply with the
following requirements, in addition to the applicable
requirements of all other clauses.
(1) General requirements
(i) Rated operating conditions
Electronic instruments shall be designed
and manufactured so that they do not
exceed the maximum permissible errors
under rated operating conditions.
(ii) Disturbances
Electronic instruments shall be designed
and manufactured so that when they are
exposed to disturbances, either-
(a) significant faults do not occur; or
(b) significant faults are detected and
acted upon.
Note : A fault equal to or less than the value
specified in paragraph 4(2)(v) of Part I
(1 dtl is allowed Irrespective of the value
of the error of indication.
(iii) Durability
The requirements In paragraphs 4(1)(1)
and 4( l)(ii) shall be met durably In
accordance with the intended use of the
Instrument.
(iv) Evaluation for compliance
A pattern of an electronic instrument Is
presumed to comply with the
requirements in 4(1)(1), 4(1)(11) and
4(1)(iii) if it passes the examination and
tests specified in Annex A.
(2) Application of requirements for disturbances
(i) The requirements In paragraph 4(1)(1i)
may be applied separately to :
(a) each individual cause of significant
fault, and/or
(b) each part of the electronic
instrument.
(ii) The choice as to whether to apply
paragraphs 4(1)(1i)(a) or (b) is left to the
manufacturer.
(3) Functional requirements
(I) Acting upon a significant fault
When a significant fault has been
detected, a visual or audible indication
-- -?-? -
486 .THE GAZETTE OF ,INDIA: EXTRAORDINARY [PARTII-SEC. 3(i)]
shall be provided and shall .continue until
the user takes action .or .the fault
disappears.
Means .shall be .provtded to .retain any
totallzedloed inforrnatiorncontalned in the
instrumentwhen a significant.fault occurs.
(ii) Switch-onprocedure
Upon switch-on .(in thecase of electronic
instruments permanently.connected to the
mains etswltch-onot indication), a special
procedure .shall :be ·performed that
indicates ·all the relevant signs of the
indicator :in :their active and non-active
states for a sufficient time to be easily
observed by the operator.
(iii) Influence quantities
An electronic instrument shall comply with
the requirementsparagraph 2(7) and in
addition •It shall maintain its metrological
and technical characteristics at a relative
humidityof 85 % at the upper limit of the
temperature range of the instrument. ·
(iv) Disturbances
When an electronic instrument is
subjected to the disturbances specified in
Annex A, either of the following shall apply :
(a) the difference between the weight
Indication due to the disturbance and
the indication without the disturbance
(intrinsic error) shall not exceed the
value specified in paragraph
4(2)(v) of part I (1 dtl
(b) the instrument shall detect and act
upon a significant fault.
(v) Warm-up time
During the warm-up time of an electronic
instrument, there shall be no indication
or transmission of the weighing result and
automatic operation shall be Inhibited.
(vi) Interface
An instrument may be equipped with an
interface permitting the coupling of the
instrument to external equipment. When
an Interface Is used, the instrument shall
continue to function correctly and its
metrologlcal functions shall not be
influenced.
(vii) Mains power supply (AC}
An Instrument that operates from the
mains shall, in the event of a power failure,
?t
--
retain the metrologlcal information
contained in the Instrument at the time
of failure for at least 24 hours. A switch
over to an emergency power supply shall
not cause a significantfault.
(viii) Battery power supply (DC)
An instrument that operates from a
battery power supply shall, whenever the
voltage drops below the manufacturer's
specified value, either continue to function
correctly or automatically be put out of
service.
(4) Examination and tests
The examination and testing of an electronic
weighing Instrument is intended to verify compliance
with the applicable requirements of these
specifications and especially with the requirements
in paragraph 4.
(i) Examination
An electronic weighing instrument shall
be examined to obtain a general appraisal
of the design and construction.
(II) Performance tests
An electronic weighing instrument or
electronic device, as appropriate, shall be
tested as specified in Annex A to
determine its correct functioning.
Tests are to be conducted on the whole
instrument except when either the size
or confiquratlonor both of the instrument
does not lend itself to testing as a unit. In
such cases, the separate electronic
devices shall be subjected to testing. It Is
not intended that electronic devices be
further dismantled for separate testing of
components. In addition, an examination
shall be carried out on the fullyoperational
weighing instrument or, if necessary, on
the electronic devices in a simulated set·
up that sufficiently represents the
weighing instrument. The equipment shall
continue to function correctly as specified
in Annex A.
(iii) Span stability tests
The instrument shall be subjected to span
stability tests at various intervals, I.e.
before, during and after being subjected
to performance tests.
When the instrument is subjected to span
stability test specified in paragraph 9 of
Annex A-
['WT II-1W'6 3(i)] 487
(a) the maximum a!!owable variation in
the errors of indication shall not
exceed half the absolute value of the
maximum permissible error specified
In paragraph 2(2)(ii), Table 2 for the
test load applied on any of the n
measurements;
(b) where the differences of the results
Indicate a trend more than half the
allowable variation specified above,
the test shall be continued until the
trend comes to rest or reverses itself,
or until the error exceeds the
maximum allowable variation.
5. Metrologlcal controls
(1) Pattem evaluation
(i) Documentation
The application for pattern evaluation shall
include documentation comprising :
(a) metrological characteristics of the
Instrument;
(b) a standard set of specifications for
the Instrument;
(c) a functional description of the
components and devices;
(d) drawings, diagrams and general
software information (Ifapplicable),
explaining the construction and
operation;
(e) any document or other evidence
demonstrating that the design and
construction of the Instrument
complies with the requirements of
these specifications.
(ii) General requirements
Pattern evaluation shall be carried out on
at least one and normally, not more than
three instruments that represent the
definitive pattern. At least one of the
instruments shall be completely installed
at a typical site and at least one of the
instruments or the major component of
an instrument shall be submitted in a form
suitable for simulation testing in a
laboratory. The evaluation shall consist of
the tests specified in paragraph S(l)(iii).
(iii) Pattern evaluation tests
(a) Instruments shall comply with :
• The metrological requirements in
paragraph 2,· particularly with
reference to maximum permissible
errors, when the instrument Is
operated in accordance with the
manufacturer's specifications for
range and product(s);
• The technical requirements in
paragraph 3 including the
requirement for security of operation
in paragraph 3(2). Additionally
electronic Instruments shall comply
with the requirements in paragraph
4.
(b) The appropriate metrological
authority :
• shall conduct the tests in a manner
that prevents unnecessary
commitment of resources;
• shall permit, when the same
instrument Is involved; the result of
these tests to be assessed for initial
verification; is advised to accept,
with the consent of the applicant,
test data obtained from other
metrological authorities without
repeating the tests;
• shall ensure that an instrument that
can be operated as a non-automatic
weighing Instrument meets the
relevant requirements for class III
or class IV of non-automatic
weighing instruments.
(c) Material tests
Instruments shall be subjected to in·
situ material tests In accordance with
either the separate verification
method as specified in paragraph
6(2)(ii) of Annex A or the integral
venrlcation method as specified In
paragraph 6(2)(1il) of Annex A
Where the material test is conducted
using the integral control instrument
the integral verification method
weighing test in paragraph
6(2)(iii)(a) of Annex A shall be
performed.
In-situ material tests shall be carried
out as follows :
• in accordance with the descriptive
markings;
• under the rated operating conditions
for the instrument;
• not less than three material tests
shall be conducted, one at minimum
488 THE GAZETTE OF INDIA: EXTRAORDINARY [PART11--SEC. 3(i))
·
..
capacity, one at maximum capacity
.and one at.dose .to the ·minimum
totalized load :(Smin);
• each test shall be conducted at the
maximum .rate of ,weighing cycles
per·hour;
.
• with a test load representative of the
: range and type otproducts for which
the Instrument is 'likely to be used
.or a product for which the
'Instrument is specified to be used;
• ,with a quantity of material not less
than the minimum totalized load .
(Smln) marked on the instrument;
• when the quantity of material equal
to the minimum totalized load
(Smin) can be totalized in less than
five weighing cycles, the following
additional material tests shall be
conducted, five cycles each at
maximum capacity (Max) and five
cycles at minimum capacity (Min);
• equipment near the automatic
weighing instrument, including
conveyors, dust collection systems
etc. that are in use when the
instrument is in normal operation,
shall be in use;
• if the Instrument can divert weighed
material through alternative
discharge facilities, the test program
shall be performed for each
alternative unless It can be ·
established that the weigh hopper
is not affected for example by
different air flow.
When a load receptor cannot be loaded
with sufficient standard weights to verify
and determine the rounding error of the
control Instrument indicating device or
partial indicating device, then the
instrument shall be subjected to material
tests by the separate verification method.
For this method an appropriately designed
control Instrument shall be available so
that the material tests can be effectively
and efficiently conducted.
The error for automatic weighing shall be
·
the difference between the conventional
true value of the mass of the test load as
defined In paragraph 6(2)(1i) or 6(3)(iii)
as appropriate and the indicated weight
observed and recorded as defined in
paragraph 6(2)(1) or 6(3)(ii) as
appropriate.
The maximum permissible error for
automatic weighing shall be as specified
in paragraph 2(2)(1) Table 1 for initial
verification and as appropriate for the
class of instrument.
(ill) Simulation tests
Influence quantities shall be applied
during simulation tests in a manner that
will reveal an alteration of the weighing
result for any weighing process to which
the instrument could .be applied, In
accordance with :
• paragraph 2(7) for all instruments;
• paragraph 4 for electronic
instruments.
When conducting such tests on a load cell
or on an electronic device equipped with
an analogue component, the maximum
permissible error for the device under test
shall be 0.7 times the appropriate value
specified in Table 2.
Note: Since the requirements of this paragraph
apply only to the instrument submitted for pattern
evaluation and not to those subsequently submitted
for verification, the means used to determine the
appropriate maximum permissible error or maximum
allowable variation has been exceeded will be decided
and mutually agreed upon between the Metrological
Authority and the applicant. Following are examples
of these means :
(a) an adaptation of the totalizatlon
indicating device to give greater
resolution than that of the
totalization scale Interval;
(b) the use of change point weights;
(c) any other means mutually agreed
upon.
(iv) Provision of means for testing
For the purposes of testing, the applicant
may be required to furnish the
Metrological Authority with the material,
handling equipment, qualified personnel,
and a control instrument.
(v) Place of testing
Instruments submitted for pattern
approval may be submitted at the
following places :
(a) the premises of the Metrologlcal
Authority to which the application
['WT II-? 3(i)J 489
has been submitted;
(b) any other suitable place mutually
agreed upon by the Metrologlcal
Authority and the applicant.
{2}Tests to be carried out during verification and re
verification
(i) Tests
Instruments shall comply with the
requirements in paragraph 2 [except
paragraph 2(7)] and paragraph 3 for any
product(s) for which they are Intended and
when operated under normal conditions
of use.
Tests shall be carried out insitu, in a
normal Installation. The instrument shall
be installed so that an automatic weighing
operation will be virtually the same for
testing as It Is for a transaction.
(a) Non-automatic weighing instru
ments
When an instrument can be
operated as a non-automatic
weighing instrument, it shall meet
the relevant requirements for class
III or class IV of non-automatic
weighing instruments.
(b) Material tests
Instruments shall be subjected to in
situ material tests in accordance with
either the separate verification
method as specified In paragraph
6(2)(11) of Annex A or the integral
verification method as specified in
paragraph 6(2)(iii) of Annex A.
Where the material test is conducted using
the Integral control instrument the integral
verification method weighing test in
paragraph 6(2)(iii)(a) of Annex A shall be
performed.
Insltu material tests shall be carried out
as follows:-
• In accordance with the descriptive
markings;
• under the rated operating conditions
for the instrument;
• not less than three material tests
shall be conducted, one at minimum
capacity, one at maximum capacity
and one at close to the minimum
totallzed load (•min);
• each test shall be conducted at the
f..t7 4 c,6) 11- I??,
maximum. rate of weighing cycles
per hour;
• with a test load of products or
products for which the instrument
is intended;
• with a quantity of material not less
than the minimum totalized load
(•min) marked on the instrument;
• when the quantity of material equal
to the minimum totalized load
(•min) can be totallzed in less than
five weighing cycles, the following
additional material tests shall be
conducted, five cycles each at
maximum capacity (Max) and five
cycles at minimum capacity (Min);
• equipment near the automatic
weighing Instrument, including
conveyors, dust collection systems
etc. that are In use when the
instrument ts in normal operation,
shall be in use;
• if the instrument can divert weighed
material through alternative
discharge facilities, the test program
shall be performed for each
alternative unless it can be
established that the weigh hopper
is not affected for example by
different air flow. Testing for the full
range of products only needs to be
done for one discharge facility.
When a load receptor cannot be loaded
with sufficient standard weights to verify
and determine the rounding error of the
control instrument indicating device or
partial indicating device, then the
Instrument shall be subjected to material
tests by the separate verification method.
For this method an appropriately designed
control instrument shall be available so
that the material tests can be effectively
and efficiently conducted.
The error for automatic weighing shall be
the difference between the conventional
true value of the mass of the test load as
defined in paragraph 6(2)(ii) or paragraph
6(3)(iii), as appropriate and the indicated
weight observed and recorded as defined
In paragraph 6(2)(1) or paragraph 6(3)(11),
as appropriate.
The maximum permissible error for
automatic weighing shall be as specified
490 THE' GAZETTE <(i)F INDIA: EXTRAORDINARY [PARTll-SEC. 3(i)]
in paragraph 2(2)(i) Table 1 .for Initial
verification as appropriate for.the class of
instrument.
(ii) Provlslon.otmeansfor:testlng
For the.purposes of.testing, .the.applrcant
shall be, required to furntsh with the
material, handling equipment, qualified
personnel.ard a-control instrument.
(3) Tests to.becemea out during inspection
Inspection ·shall .be carried out :in accordance
with the same provisions as .in paraqraph 5(2) for
verification, with the exception that the inspection
maximum permissible errors shall 'be applied.
6. Test methods
(1) Contro!lnst;ument and test standards
The control instrument and standard
weights ·used for tests shall ensure the
checkln_gof the test load to an error not
greater .than :
(a) one-third of the maximum
permissible error for automatic
weighing when the control
Instrument or the device used for
control purposes Is verified
immediately prior to the material
test, or
(b) one-fifth of the maximum
permissible error for automatic
weighing in all other cases.
Note : When using the integral
verification method, a sub-division
of the test load is unavoidable and
this may also be true when using
the separate verification method.
When calculating the conventional
true value of the mass of the test
load, It Is necessary to consider the
Increased uncertainty due to sub
dividing the test load.
(2) Separate verification method
.
With this method, an instrument other than the
Instrument being verified is used to determine the
conventional true value of the mass of the test load.
'
.
.
(i) Indicated weight
A test load shall be weighed as an
automatic bulk to bulk weighing operation
and the Indicated weight value on the
principal totalization indicatingdevice shall
be observed and recorded.
(ii) Mass of the test load
The test load shall be weighed on a control
instrument and the result shall be
considered as the conventional true value
of the mass of the test load.
(3) Integral verificationmethod
With this method, the instrument being verified
is used to determine the conventional true value of
the mass of the test load.
The Integral verification method shall be
conducted by using either of the following :
o a partial totallzation indicating
device with standard weights to
assess the rounding error; or
o an appropriately designed control
Indicating device.
(I) Interruption of automatic operation
An automatic weighing operation of a test
load shall be initiated following the same
procedure. as for weighing bulk to bulk.
However, the automatic operation shall be
Interrupted twice during each weighing
cycle necessary to weigh and discharge a
sub-division of the test load.
An automatic operation shall not be
Interrupted during consecutive weighing
cydes if the instrument is installed as an
air-enclosed system.
(i)(a)Pre-discharge (gross) interrupt
After the load receptor has been loaded
and the instrument has automatically
processed a gross weight, the automatic
operation shall be Interrupted. When the
load receptor has stabilized, the gross
weight indicated or determined by
balancing with standard weights shall be
recorded and the Instrument switched
back to automatic operation.
(b) Post-discharge (tare) Interrupt
After the load has been discharged and
the instrument has automatically
processed a tare weight, the automatic
operation shall be interrupted. When the
discharged load receptor has stabilized,
the tare weight indicated or determined
by balancing with standard weights shall
be recorded and the Instrument switched
back to automatic operation.
(Ii) Indicated weight
The principal totallzatlon indicating device
shall be used in obtaining the indicated
weight of the test load.
(Iii) Mass of the test load
For each discharge, the tare weight value
subtracted from the gross weight value is
['ll'T II-1s!Us 3(i)] 491
the net weight of the material discharged.
A summation of the net welght values of
all the discharges in the test load shall be
the conventional true value of the mass
of the test load.