Annexure B, Legal Metrology (General) Rule, 2011: ANNEX B

Legal Metrology (General) Rule, 2011 - Annexure B

Text of annexure B

ANNEX B

ANNEX B ALIGNMENT CALIBRATION OF SINGLE:AXLE WEIGHING INSTRUMENT The alignment calibration is not recommended as an adequate substitute for verification of reference wagons by full draught weighing and shall only be applied under the conditions in paragraph 2(8)(iii)(b) of part II. 1. The altgnment calibration applies to Instruments that operate by partial weighing of two­ axle wagons and the same instrument Is required for use as the control Instrument for the purposes of determining the mass of reference wagons in paragraph 2(8)(111){b}of part II. 2. Instruments that operate by partial weighing are exempt from the alignment calibration procedure provided the follow Ing:- • the top surface of both rails along the length of the weight zone are vertically aligned to ±1 mm, and • the alignment has been checked along both rails at not less than two positions on the load receptor and .not less than two positions within a wagon length from the load receptor on each associated apron. 3.For instruments not covered by para 2 of Annexure B, a calibration correction shall be determined by applicationof the alignment calibration procedure in para 4 of Annexure B. This calibration correction shall be added to each totallzed wagon weight to determine the mass of each reference wagon. · 4. The alignment calibration is conducted with the use of a single empty uncoupled wagon of a wheelbase similar to those wagons used for in-motion testing.· An example of an alignment calibration is given in para 5 of Annexure B. Position on load receptor First axle Leading end Middle Trailingend Second axle Total of six weighings Divide total by three Derived mass of standard weights Calibration Indicated weight (t} Empty Loaded wagon 5.76 14.27 5.75 14.26 5.75 14.26 5.75 14.25 5.75 14.25 5.74 14.24 34.50 85.53 d = 11.50e = 28.51 f - e - d - 17.01 C·f= ·0.01 I.?--- .:.::.::. [ <rl'T II <'J1!,!; 3 ( i) ] 453 The calibratlon correction shall be added to the totalized indicated weight of each reference wagon weighed while stationary and uncoupled. In the case given above, the minus sign should be noted. Therefore, if the totallzed indicated weight Is 41.38, the corrected weight will be: 41.38 + (·0.01) = 41.37 Note: The calibration correction computed In this example is not Intended to be typical. SEVENTH SCHEDULE HEADING-D AUTOMATIC GRAVIMETRIC FILLING INSTRUMENTS PART I :z. General definitions (1) Weighing instrument: Measuring Instrument that serves to determine the mass of a load by using the action of gravity on that load. According to its method of operation, a weighing instrument Is classified as automatic or non· automatic. (2) Automatic weighing Instrument: An instrument which weighs without the Intervention of an operator and follows a predetermined programme of automatic process characteristic of the Instrument.· (3) Automatic gravimetric filling instrument: . An instrument which fills containers with predetermined and virtually constant mass of product from bulk by automatic weighing, and which comprises essentially an automatic feeding device or devices associated with one or more weighing units and the appropriate control and discharge devices. (i) Associative (selective combination) weigher: Automatic gravimetric filling Instrument comprising one or more weighing units and which computes an appropriate combination of the loads and combines them for subsequent discharge as a fill. (ii) Cumulative weigher: Automatic gravimetric filflng Instrument with one welahlna unit with the facility to affect the fl{bymore than one weighing cycle. (iii) Subtractive weigher: Automatic gravimetric filling instrument for which the fill is determined by 4 7 4 t; v1 II - It 'f controlling the output feed from the weigh hopper. (4) Fill: One or more loads discharged into a single container to make up the predetermined mass. (5) E/ectronlc instrument: An Instrument equipped with electronic devices (6) Control instrument: A weighing instrument used to determine the mass of the test fills delivered by the filling instrument. 2. Construction Note. In this Part the term device is applied to any part which uses any means to perform one or more specific functions. (1) Principal parts (I) Weighing units: A device which provides information on the mass of the load to be measured. This devise may consist of all or parts of non· automatic weighing instruments. (II) Local receptor: . Part of measurement Instrument intended ;1.:;; . to receive the load. (111)Feeding device: Device which providesthe supply of the product from bulk to the weighing unit. It may operate in one or more stages. (iv) Control devices (a) Feed control device-Device which regulates the rate of the feed of the feeding device (b) Fifi setting device-Device which allows the setting of the pre-set value (c) Final feed cut off device-Device which controls the cut off the final feed so that the average mass of the fills corresponding to the preset value. This device may include an adjustable compensation for the material in flight. (d) Correction device-Device, which automatically corrects the setting of the filling instrument. (2) Electron le parts (I) Electronic device: A device comprising electronic assemblies 4S4 •THE GAZETTE '.OF :INDIA : EXTRAORDINARY [PARTII-SEC. 3(i)] and: performiogaaaspeciflc ;function. Electronic devices are usually manufactured .as separate•unlts and capable of belngdndependently .tested. (ii} Electronic sub7ii/ssembly: A part of an electronic device employing electronjc. components .and having a recognizable function of its .own. (iii) Electronic components: The smallest physical.entitythat uses the electron.orvhole .conductlon In semi­ conductors,,,gasesor.m :a vacuum. (3) India,ting:device: The part ofa measuring instrument that displays an.lndication. /Votes-. 1. Fora weighing instrument, the indicating device is a -set of components, which displays the value, in units of mass, of the.result of a weighing operation. 2. For;fflflngInstrument, the indicating device may indicate either the mass of the load or difference between the mass and the preset value provided the Intention is made ,clear. (4) Ancilliilrydevices (i) Zero setting device: A device for setting the indicating device to zero when the load receptor Is empty. (a) Non automatic zero setting device-­ A device for setting the Indicating device to zero by an operator (b) Semi automatic zero setting device-­ A device for setting the indicting device to zero automatically following a manual command (c) Automatic zero setting device-A device for setting the Indicating device to zero automatically without the Intervention of an operator. ( ii}Tare device. A device for setting the indication to zero when a load is on the load receptor (a) without altering the weighing range for net loads (additive tare device), or (b) reducing the weighing range for net loads (subtractive tare device). 3. Metrologlca/ characteristics (1) Scale Interval (d): Value expressed In units of mass, of the difference between- (a) the values corresponding to two consecutive scales mark for analogue indication. (b) two consecutive indicated values for digital indication. (2) Reference psrtide mess of iii product: Mass equal to the mean of ten of the largest elementary particles or pieces of the product taken from one or more loads. (3) Pre set value: Value, expressed In units of mass, preset by the operator by means of the fill setting device, in order to define the nominal value of the fills. (4) Static set point: Value of the test weights or masses, which, in static tests, balance the value, selected on the indication of the fill setting device. (5) Weighing cycle: The sequence of the operation which Includes: (a) delivery of the material to the load · receptor; (b) a weighing operation; and (c) the discharge of a single discrete load. (6) Fina/ feed time: ' Time taken to complete the last stage of delivery of the product to a load receptor. (7) Minimum capiilcity (min): Smallest discrete load that can be weighed a'utomatically on a load receptor. (8) Maximum a,padty (msx): The largest discrete load that can be weighed automatically on a load receptor. (9) Warm up time: The time between the moment power is applied to an instrument and the moment at which the instrument is capable of complying with the requirements. (10) Average number of loads per fill: Half the sum of the maximum and minimum number of loads per fill that can be set by the operator or, in cases where the number of loads per fill is not directly determined by the operator, either the ["ll'l II-? J(i)] 455 means of the actual number of loads per fill (If known) In a period of normal operation, or the optimum number of loads per fill as may be specified by the manufacturer for the type of product which Is to be weighed.· (11) Rated minimum fill: The rated value of the fill below which the weighing results may be subject to errors outside the limits specified in this part. (12) Minimum discharge: The smallest load that can be discharged by a subtractive weigher. 4, Indications and errors (1) Methods of indication . (i) Analogue indication: An Indication allowing the evaluation of equilibrium position to a fraction of the scale interval. (ii) Digital Indication: An Indication in which the scale mark comprises a sequence of an aligned figure that do not permit Interpolation to fraction of a scale Interval. (2) Errors (i) Error of indication: The Indication of an instrument minus the (conventional) true value of the mass. (II) Intrinsic error: The error of an instrument under reference conditions (iii) Initial Intrinsic error: The Intrinsic error of an Instrument as determined prior to performance and span stability tests. (iv) Fault: The difference between the error of Indication of an instrument and the intrinsic error. Note:Prlncipally, a fault is the result of an undesired change of data contained in or flowing an electronic instrument. (v) Significant fault: A fault greater than 0.25 of the maximum permissible deviation of each fill [as specified in sub-paragraph (2) of para 2] for an in-service verification, for a fill equal to the rated minimum fill. Note: The following are not considered to be significant faults even when they exceed the value defined above: (a) Faults arising from simultaneous and mutually independent causes in the instrument. (b) Faults implying the impossibility to perform any stage of operation. (c) Faults being so serious that they are bound to be noticed by an operator. (d) Transitory faults bring momentary variations In the indication or operation which cannot affect the final results of the automatic cycles . · Note. For Instruments where the fill may be greater than one load, the value of significant fault applicable for a test on one static load shall be cakutated in accordance with the test procedures. (vi) Span stability. The capability of an instrument to maintain the difference between the Indication of weight at maximum capacity and the indication at zero within specified limits over a period of use. · (3) Reference value for accuracy class The value of accuracy class determined by static testing of the weighing unit during Influence quantity testing at pattern approval stage. The reference value for accuracy class is equal to the best accuracy class for which the Instrument may be verified for operational use. 5. Influences and reference conditions (1) Influence quantity A quantity, which is not the subject of the measurement but which influences the . value of measurand or the indication of the instrument. (i) Influence factor: An Influence quantity having a value within the specified rated operating conditions of the instrument. (I!) Disturbance: An Influence quantity having a value within the limits specified in this Part but outside the rated operating conditions of the Instrument . J. _ 456 THE·GAZET'FE :oF ,JNDIA: EXTRAORDINARY [PARTII-SEC 3(i)] (Z) Rated operating conditions Condltions.of use, ._giving.the ranges of the· measurand and '.the 'Influence quantities for which. the ,metrologlcal charactertsncs.ere: intended .to .lie ·within the-maxlrnurn permissible :deviation specified. (3J Reference condltions A set of specified values •of 'influence factors ;f11<ed Jo ,ensure valid Inter­ comparison .of lhe results of measurements. 6. Tests (1) Mater/a/test: A test carried out on a complete instrument using the type of material, which ,itJs Intended to weigh. (Z) SimukJ.tion tests: A test carried out on a complete instr.ument or a part of an Instrument in which ariy part of the weighing operation is simulated. (3) Pelformance test: A test to verify whether the equipment under test (EUT) is able to accomplish its intended functions (4) Span stability test: A test to verify that the EUT Is capable of maintaining Its perfonnance characteristics over a period of use. PART II 1. General accordance with sub-paragraph (2) of paragraph 2 and marked Q.T1 the Instrument in accordance with sub· paragraph (10) of paragraph 3. Accuracy class shall be specified for intended usage, I.e. nature of the product to be weighed, type of Installation, value of the fill, and operating rate. Note. The limitation of accuracy classes to certain application may be determined by these Rules. (Z) Limits of error (I) Maximum permissible error for static tests: This instrument shall have a reference value for accuracy class Ref (x), applicable for static testing only, for which the maximum permissible error for influence factor tests shall be as specified in sub­ paragraph (4) of paragraph 2, multiplied by the class designation factor (x). (II) Maximum permissible deviation of each fills: The instrument shall have a specified accuracy class X(x) for which the maximum permissible deviation of each fill from the average shall be equal to the limits specified in Table I given below, multipliedby the dass designation factor (x), (x) shall be 1 x 10k,2 x 10k,5 x 10k,k being a positive or negative whole number or zero, TABLE 1 (See Table 2 under sub-paragraph (3) of para Initial In service verification Inspection M < 50 6.3% 9% 50 < M < 100 3.15g 4.5g 100 < M < 200 3.15% 4.5% 200<M<300 6.3g 9g 300< M < 500 2.1 o/o 3% 500 < M < 1000 10.Sg 15g 1000 < M < 10000 1.05% 1.5% 10000 < M < 15000 105g 150% 15000 < M 0. 7% 1o/o (1) Scope: This Part specifies the metrologlcal and technical requirements for automatic gravimetric filling instruments (hereafter called "filling instruments"), which sub-divide a bulk product into fills of predetermined and virtually constant mass by automatic weighing, the fills being kept separate. Note. (1) This specification places no constrain on the maximum or minimum capacities of the instruments for which this specification Is applicable. (2) Fillinginstruments may also be required to comply with certain requirements e.g. an instrument which could operate as a non-automatic instrument will need to comply with the specification for non· automatic weighing Instruments. 2. Metro/ogla/ requirements (1) Accuracy dasses The accuracy class and reference value for accuracy class shall be specified in Value of the mass of the fill M(g) Maximum permissible deviation of each fill from the average for classX{l} ['IIT'Tn-= J(ilJ 457 6 for the number of fills required to find the average value) For in-service testing, when the reference particle mass exceeds 0.1 of the maximum permissible in-service deviation, the values derived from Table 1 shall be Increased by 1.5 times the value of the reference particle mass. However, the maximum value of the maximum permissible deviation shall not exceed (x) x 9%. Note. Partlde mass correction is not appllcable to limits, which are derived from Table 1, e.g. influence quantity test, zero setting, etc. (3) Maximum permissible preset value error (setting error): For instruments where it is possible to preset a fill weight the maximum difference between the preset value and the average mass of the fill shall not exceed 0.25 of the maximum permissible deviation of each fill from the average, as specified for in service inspection in item (ii) of sub-paragraph (2) of this paragraph. This limit will apply for initial verification and for ln•,ervice inspection testing. (4) Maximum permissible error far influence factor tests: The maximum permissible error for any static test load during influence factor tests shall be 0.25 of the maximum permissible deviation (as specified in item (ii) of sub-paragraph (2) of this paragraph for in-service inspection, corresponding to the value of a fill equal to that load. Note. For instruments where the fill may not be equal to one load, the maximum permissible error applicable for a test on one static load shall be calculated in accordance with the test procedures. (5) Influence factors: Refer ta Annex A for test conditions. (i) Static temperatures: Instruments shall comply with the appropriate metrological and technical requirements at temperatures from • 10°C to +400C. However, for special applications the limits of .the temperature range may differ from those given above but such a range shall not be less than 30'C and shall be specifiedIn the descriptive markings. (ii) Power supply {AC}: Instruments which are powered by an AC electricity supply shall comply with the appropriate metrological and technical . requirements when operated at voltages from • 15% to + 10% of the reference voltage. (iii) Tilting: Instruments which are not intended for installation in a fixed position and which do not have a level Indicter shall comply with the appropriate metrological and technical requirements when tilted by 5%. Where a level indictor Is present it shall enable the instrument to be set to a tilt of 1% or less. (6) Units of measurement: The units of mass to be used on an instrument · are the milligram (mg), the gram (g}, the kilogram (kg) and the tonne (t). 3. Technical requirements (1) Suitability for use A filling instrument shall be designed to suit the method of operation and the products for which It is intended. It shall be of adequately robust construction so that it maintains its .netrological characteristics. (2) Security of operation (I) Accidental mat-adjustment: An instrument shall be so constructed that an accidental. breakdown or a mal­ adjustment of control elements likely to disturb its correct functioning cannot take place without its effect being evident. (ii) Use of a printer: Any print-out is for information purposes only and not for use in a commercial transaction, except for preset values and number of weighings. (iii) Andflary devices: Any ancillary device provided for use with a filling instrument shall not affect the correct functioning of the instrument. (Iv) Salle interval (d): Scale intervals of ail indicating devices associated with a weighing unit shall be the same. (3) Fill setting device If fill setting is by means of a scale, it shall be graduated in units of mass. If fill setting is by means of weights, they shall be either weights in accordance with these rules or purpose-designed of any nominal value, distinguishable by shape and identified with the filling instrument. 458 THE GAZETTE OF INDIA: EXTRAORDINARY [PARTJI-SEC. 3(i)] (4) Final feed cut-off device The final feed .cut-off device shall be clearly differentiated .from any other device. The direction of movement corresponding to the sense-of the desired result shall be shown, where applicable. (5) Feeding device: . The feeding device shall be designed to provide sufficient and regular flow rate(s). An adjustable feeding device shall be fitted with an Indication of the direction of movement corresponding to the sense of the adjustment of the feed where appllcable. (6) Load receptor (i) The load receptor, and feed and discharge devices as appropriate, shall be designed to ensure that residual material retained after each discharge is negligible. (ii} Instruments using the subtractive · weighing principle shall be designed to ensure that residual material retained at feed from the discharge gate Is negligible. (iii) The load receptor shall provide access and facilities so that where necessary test weights or masses up to the maximum capacity can be placed in position, in a safe and secure manner. If these facilities are not a permanent fixture of the instrument, they should be kept In the vicinity for the instrument. (iv) Manual discharge of the load receptor shall not be possible during automatic operation. (7) Zero setting and tare devices (i) The filling Instrument shall be provided with a zero-setting device, which may also be used for the setting of tare. The device may be- (a) Manual; (b) Semi-automatic, or (c) Automatic (ii} Zero-setting and tare devices shall be capable of setting to less than or equal to 0.25 of the maximum permissible deviation for in-service inspection for a fill equal to the minimum capacity for ·instruments with one weighing unit, and for a fill equal to the rated minimum fill for selective combination weigher. (iii) Non-automatic or semi-automatic zero­ setting and tare device should be locked during automatic operations. (iv) The weighing unit shall b'e in stable equilibrium when the zero setting and tare device is being set. (8) Equilibrium mechanism The equilibrium mechanism may be provided with detachable masses which shall be either weights in accordance with OIML requirements or purpose­ designed masses of any nominal value, distinguishableby shape and Identified with the filling instruments. (9) Security A security means shall be provided for components and pre-set controls to which access is prohibited. (10) Descriptive markings FIiiing Instruments shall bear the following markings . (I} Marking shown in full (a) Name or identification mark of the manufacturer (b) Name or identification mark of the Importer (if applicable) (c) Serial number and type designation of the Instrument (d) Product(s) designation {i.e. materials that may be weighed) ' (e) Temperature range (if applicable, see 2.5.1) in the form ....... C/ .... C · (fl Electrical supply voltage in the form ...... V (g} Electrical supply frequency in the form .. ,Hz (h) Working fluid pressure (if applicable) in the form ...... kPa (I} Average number of loads/fill (if applicable) (J} Maximum fill (if applicable) (k) Minimum fill (If applicable) (I} Rated minimum fill (m) Maximum rate of operation (if applicable} In the form ....... Load per minute (ii} Marking shown in code (a) Pattern approval sign {b) Indication of the accuracy class X(x) (c) Reference value for accuracy class Ref (x} (d) Scale interval (if applicable) in the form d = """"" ['WT II-? J(i)] 459 (e) Maximum capacity (or minimum discharge where applicable) in the form Max= ..... (f) Maximum additive tare in the form T= + ...... (g) Maximum subtractive tare In the form T = - . An instrument may be verified for different materials for which different classes shall apply or which require different operating parameters to maintain limits of error. Marking shall be such that the alternative class or operating parameters are clearly associated with the appropriate material designation. In the case of subtractive weigher the minimum load to be discharged shall be specified. (iii) Presentation of descriptive markings: The descriptive marking shall be indelible and of a size, shape and clarity to enable legibility under normal conditions of use of the filling instrument. They shall be grouped together in a clearly visible place on the fillinginstrument, either on a date plate fixed to the instrument or on the filling instrument itself. Where the markings are placed on a date plate, it shall be possible to seal the plate bearing the markings. Where they are marked on the filling instrument Itself it shall not be possible to remove them without destroying them. The descriptive markings may be shown on a programmable display, which is controlled by software. In this case, means shall be provided for any access to reprogramming of the markings to be automatically and non-erasable, recorded, e.g. by traceable access software when a programmable display is used, the plate on the instrument shall bear at least the following markings: (a) Type and designation of the instrument; (b) Name or identification mark of the manufacturer; (c) Pattern approval number; (d) Electrical supply voltage; (e) Electrical supply frequency; and (f) Pneumatic pressure (11) Verification marks (i) Position: The filling instruments shall have a place for the application of verification marks. This place shall: (a) be such that the part of which it is located cannot be removed from the filling instrument without damaging the marks (b) allow easy application of the mark without changing the metrological qualities of the filling instruments (c) be visible without the filling instrument having to be moved when it is in service. (ii) Mounting: Filling instruments required to bear verification marks shall have a verification mark support, at the place provided for above, which shall ensure the · conservation of the marks. When the mark is made with a stamp, this support may consists of a strip of lead or any together material with similar qualities, Inserted into a plate fixed to the filling instrument or a cavity bored In the filling instrument itself. When the marks consist of an adhesive transfer, a space shall be prepared for this purpose. (12} Control instrument: The control instrument may be separate from an integral part of the filling Instrument. 4. Requirements for electronic Instruments Electron le filling instruments shall comply with the following requirements, In addition to the -appllcable requirements of all other clauses of this specification. (1) General requirements (i) Rated operating conditions: Electronic operating conditions Electronic instrument shall be so designed and manufactured that they do not exceed the maximum permissible errors under rated operating conditions. (ii) Influence fadors: An electronic Instruments shall comply with the requirements of sub-paragraph (5) of paragraph 2 and shall also comply with appropriate metrological and technical requirements at a relative humidity of 85% at the upper limit of the temperature range of the Instrument. (iii) Disturbances: Electronic instruments shall be so designed and manufactured that when exposed to disturbances, elther:- (a) significant faults do not occur, i.e. the difference between the weight indication due to the disturbance and - - - __l. __ ---- 460 THE iG)\:ZEirIT'E or INDIA: EXTRAORDINARY [PARTII-SEC. 3(i)] the · indication .wtthout the dlsturbance'(intrinsic error) shall not exceed 'the value specified in paragraph %2)(\i),Of Part I; or (b) significant .faults .are detected and acted .upon. Note: A fault:equalto erlessthan the value in paragraph 4(2)(v),of;Part.I,is:allowed irrespective of the value of the error-of.lndkatlon. (iv) Evaluation for ,compliance: The pattern ,of :an electronic instrument is presumed to comply with the requirements of paragraphs 4(1)(i),4(1)(ii) and 4(1)(i1i), If it passes the examination and tests specified in Annex A. (v) Application: Therequlrernents in paragraph 4(1) (iii) may be applied separately- (a) either to each individual cause of significant fault; or (b) to eadi part of the electronic Instrument; or (c) to both. The choice of whether sub-item (a) or sub item (b) above should be applied is left to the manufacturer. (2) Functional requirements (i) Indicator display tests: If the failure of an indicator display element can cause a false weight indication, then the instrument shall have a display test facility which is automatically initiated at switch-on of indication, e.g. indication of all the relevant signs of the indicator in the active and non-active states for a sufficient time to be easily observed by the operator. (ii) Acting upon a significant fault: When a significant fault has been detected, the Instrument shall either be automatically made inoperative or a visual or audible Indication shall be provided automatically and shall continue until such time as the user takes action or the fault disappears. (iii) Warm-up time: During the warm-up time of an electronic instrument there shall be no indication or transmission of the result of weighing, and automatic operation shall be Inhibited. (iv) Interface: An instrument may be equipped with an interface within allows it to be coupled to external equipment. When an interface is used, the instrument shall continue to function correctly and its metrologlcal functions shall not be influenced. (v) Battery power supply: An instrument that operates from a battery power supply shall, whenever the voltage drops below the manu_facturer's specified minimum value, either continue to function correctly or automatically be put out of service, (3) Examination and tests: The examination and testing of an electronic instrument is intended to verify compliance with the applicable requirements of this specification and especially with the requirements of this paragraph. (i) Examinations An electronic instrument shall be · examined to obtain a general appraisal of the design and construction. (ii) Performance tests: An electronic instrument or electronic device, as appropriate, shall be tested as specified in Annex A to determine the correct functioning of the instrument. Tests are to be carried out on the whole instrument except when either the size or the configuration or both of the instrument does not lend itself to testing as a unit. In such cases the electronic devices shall be tested, where possible as a simulated instrument including all electronic elements of a system which can affect the weighing result. In addition, an examination shall be carried out on the fully operational instrument. Susceptibility that would result from the use of electronic Interfaces to other equipment shall be simulated in the tests. (iii) Span stability: When an electronic Instrument is subjected to the span stability test specified in paragraph 7 of Annexure A, the absolute value of the difference between the errors obtained for any two measurements shall not exceed half the maximum permissible error for influence factor tests for a near maximum capacity load. 5. Metrologica/ controls (1) General (i) Provision of means of testing: ["!T'TII-- 3( i)] ··-- ---·------- -- -------- 461 For the purposes of testing, the metrological authority may require from the applicant the product (i.e. the material to be weighed), the handling equipment, the appropriately qualified personnel and a control Instrument. (ii) Material tests: (a) For pattern evaluation-The material used as the test load for pattern evaluation shall be representative of a product for which the Instrument is designed. The test shall be conducted in accordance with the test procedure in paragraph 8(1) of Annexure A. (b) For initial verification and in-service verification-The in-situ material tests shall be done in accordance with the descriptive marking under the normal conditions of which the instrument Is intended. The test shall be conducted in accordance with the test procedure laid down In paragraph 8(2) of Annexure A. (2) Pattern approval (IJ Documentation: The application for pattern approval shall include documentation comprising: (a) metrologlcal characteristics of the Instrument, . (b) a set of specifications for the Instrument, (c) a functional description of the components and services, (d) drawings, diagrams and general soft ware information (if applicable), explaining the construction and operation, including interlocks, (e) any document or other evidence that the design and construction of the Instrument complies with the requirements of this specification. (ii) General requirements: The pattern evaluation shall be carried out on one or more (and normally not more than three) units that represent the definitive pattern. One or more of the units shall be complete and fully operational of the purposed of paragraph 5(2)(iii)(a). One or more of the units shall be submitted in a form suitable for simulation testing In a laboratory and shall include the whole of the electronics which affect li74 t. {)11-1/6 the weighing result except In the case of an associative weigher where only one representative weighing unit may be included. The Instrument or simulated Instrument shall have a load Indicator, or an interface allowing access to a quantity that can be calibrated to provide an indication of load, so that the requirements of paragraph 2(2)(i) (maximum permissible errors for static tests) may be tested and so that the instrument may be tested for influence quantities. The scale Interval of the load indicator shall not exceed 0.125 of the maximum permissible deviation for In­ service verification for a fill equal to the minimum capacity. The evaluation shall consist of the tests specified in paragraphs S(l)(ii)(a) and 5(2)(iii). (iii) Pattern evaluation: The submitted documents shall be examined and tests carried out to verify that the instrument complies with: (a) the requirements specified for static tests in paragraph 2, (b) the technical requirements in paragraph 3, (c) the requirements In paragraph 4 for electronic Instruments, where applicable. The appropriate metrologlcal authority shall conduct the test In a manner, which prevents an unnecessary commitment of resources. Note: The appropriate metrological authority may accept with the consent of the applicant, equivalent test data obtained from other metrological authorities. (a) Tests for compliance with technical requirements Operational test with material shall be done in accordance with the procedure In paragraph 8(1) of Annex A on a complete instrument to assess compliance with the technical requirements of paragraph 3. (b) Influence factor tests Influence factors shall be applied during simulation tests in a manner that will reveal a corruption of the weighing results of any weighing process to which the instrument should applied, in accordance with- 462 THE GAZETTE OF INDIA: EXTRAORDINARY [PARTII-SEC. 3(i)] (a) Paragraph 2(5) for all instruments; and (b) Paragraph 4 for electronic Instruments (c) Apportioning of errors Where parts of an instrument are examined separately in the process of pattern approval, the following requirements .shall apply- The error limits applicable in a part which is examined separately are equal to a fraction P1of the maximum permissible errors or the allowed variations of the indication of the complete instruments. The fractions for any part have to be taken for the same accuracy class as for the complete Instrument incorporating the part. The fractions Pi shall satisfy the following equation: (P12+ Pi2 + Pi+ .... ) s 1 ' The fraction Pi, shall be chosen by the manufacturer of the part and shall be verified by an appropriate test. However, the fraction shall not exceed 0.8 and shall not be less than 0.3 when more than one part contributes to the effect In question. Note: As the requirements of this sub­ paragraph only apply to the instrument submitted for pattern evaluation and not to those subsequently submitted for verification, the means by which it will be possible to determine whether the appropriated maximum permlss!ble error or maximum allowable variation has been exceeded while decided mutually between the metrologlcal authority and the applicant. The.means may be for example: (a) the provision or adoption of the indicating device to give the required resolution, (b) the use of change point weights, or (c) any other means mutually agreed. (iv) Place to testing: Instruments submitted for pattern approval may be tested either- {a) on the premises of the metrological authority to which the application has been submitted; or (b) In any other suitable place agreed to between the metrologlcal authority concerned and the applicant. (v) Certificate of approval and determination of paragraph [2(2)(1) and paragraph 5 of Annex A}: The pattern approval certificate shall state the reference value for the accuracy class as determined by the static tests In paragraph 5 of Annex A and shall state that the actual class (equal to the greater than the reference value) shall be determined by compliance with the metrological requirements at initial verification. (3) Initial verification (i) General requirements: Instruments shall be examined for conformity with the approved pattern where applicable and shall be tested for compliance with paragraph 2 [excluding paragraph 2(2)(i) and paragraph 2(5)] for the intended products and corresponding accuracy classes under normal conditions of use. Tests shall be carried out by the appropriate metrological authority, in-situ, with the instrument fully assembled and fixed in the position in which it is intended to be used. The installation of an instrument shall be so designed that an automatic weighing operation will be the same whether for the purposes of testing or for use for a transaction. (ii) Material tests: Material test shall be carried out In compliance with paragraph S(l)(ii) using the test methods specified in paragraph 6. (iii) Conduct of the test: The appropriate metrological authority: (a) shall conduct the test in a manner which prevents an unnecessary commitment of resources, (b) may, where appropriate and to avoid duplicating tests previously done on the instrument for pattern evaluation under paragraph 5(2)(iii) (a) use the resu Its of observed tests to assess for initial verification. (Iv) Determination of accuracy class: [ '1T'T II-? 3 (i)] ... '--.------- - - -- ----,------,--- 463 The appropriate metrological authority shall:- (a) determine the accuracy class for the materials used in the tests in accordance with paragraph 5(2)(v} by reference to the material test results and the limits of error specified in paragraph 2(2)(ii) and paragraph 2(3} for initial verification, (b) Verify that accuracy classes determined as above. (4) In-service verification In-service verification shall be as specified in para 5(3)(1) and para 5(3)(ii). The maximum permissible errors shall be as specified in para 2(2)(ii) for in-service verification. The appropriate metrological authority shall conduct the test in a manner, which prevents an unnecessary commitment of resources. Quite often a question is asked as to how to calculate R(x). The method is as follows: Ref. (x) or simply x, is a numeral indicating the class of the gravimetric filling machine, which is to be assigned at the time of model approval. Once the value of x or Ref. (x) Is assigned, one can determine its MPD by multlplylngthe figures given in Table I of this part of this book, At the time of verification, one has to see only for the compliance to the MPD and other requirements. For calculating x or Ref. (x), one should calculate (1) value of setting error se and (2) maximum deviation md as explained in this part on. Remember mpse(1) is 1/4 of mpd1given in table of 5(3)(i) and last column of the observation sheet gives deviation from the mean fill, so maximum deviation md is calculated from here. Number of fills to be taken is chosen with the help of Table 2. Then examine the values of se/mpsel and md/ mpd(l), these may be a whole number (including zero) plus a fraction like 2 + 0.325 and 2 + 0.545 then class of the Instrument is one plus the whole number irrespective of the value of fraction, i.e. 3 in the given example. In case whole number is zero then Ref. (x} or x is 1. If the values of se/mpse 1 and md/mpd1 are 3.575 and 3.265 then Ref. (x} or x will be 5 and not 4, as x has to be either of 1, 2, 5 or multiples of 10 of 1, 2 and 5. 6. Test methods (1) Determination of the mass of individual fills The mass of the individual fills is determined using one of the methods specified in paragraph 6(5)(1) or in paragraph 6(5)(1iJ. (2) Conduct of material tests (a) The tests shall be carried out on fills using loads at or near to, the maximum capacity and also at, or near to, the minimum capacity. Material tests should only be carried out with the products the instrument is intended to be used for. (b) Cumulative welghers shall be tested as above with the maximum practicalnumber of load per fill and also with the minimum number of loads per fill, and associative weighers as above with the average (or optimum) number of loads per fill. (c) If the minimum capacity is less than one third of the maximum capacity then tests shall also be carried out near the centre of the load weighing range preferably at a value close to, but not above, lOOg, 300g, 1000g or 15000g as appropriate. (d) All tests shall be conducted with any adjustable parameter critical to inetrologicalintegrity e.g. final feed time or rate, set to the most onerous condition allowed by the manufacturer's printed instructions and incorporated in the descriptive markings. (i) Testing the effect of a correction device (a) Any correction device e.g. in-flight correction or automatic zero setting fitted to an instrument shall be operated during the tests according to the manufacturer's printed instructions. (b} If the correction device is not activated during each filling operation, then tests at minimum capacity shall be arranged to Include the effect of one or more regular operations of the correction device, e.g. buy including in the test at least three fills immediately before and after the activation of the device. (c) The Initial fills after the charge between maximum capacity and minimum capacity shall be included In the test unless the instrument bears a clear warning to discard the stated number of fills after a change to the instrument settings. ---·· ----4-?-- 464 "i'HE rGAZEif'f,E OF !NOIA: EXTRAORDINARY [PARTJI SEC. 3(i)] (3) Number of fills The number df individual tests fill depends upon the preset value (m):as specified inTable2 below:­ TABlE;2 (4) Accuracy of stendsrds The control 'Instrument :and standard weights used in testing shalhensurethat·checklng of the test fills to an error m>tigreater than either- (a) one-third of the maximum permissible deviation and maximum permissible preset value error (as appropriate) for automatic weighing (paragraph 2(2) and paragraph 2(3) respectively) if the control instrument or the device used for control · purposes Is verified immediately prior to the material test; or (b) one-fifth of the maximum permissible deviation and maximum permissible preset value error (as appropriate) for automatic weighing (paragraph 2(2) and paragraph 2(3) respectively] in all other cases; (5) Material test methods (i) Separate verification method: The separate verification method requires the use of a (separate) control Instrument to find the conventional true value of the mass of the test fill. (ii) Integral verification method: With this method the instrument being tested Is used to determine the · conventional true value of the mass of the test fill. The Integral verification method / shall be conducted using either:- (a) an appropriately designed Indicating device, or · (b) an indicating device with standard weights to assess the rounding error. The total uncertainty of the test method (separate or integral verification) shall be not greater than one-third of the maximum permissible error for the instrument. Notes-; (1) The integral verification method depends on determining the masses of the loads. Limits of error as specified in paragraph 2(2) are for the mass of the fill. If it Is not possible to ensure that In normal operation, all load is discharged at. each m < 10kg 10kg s m ?,25kg 25kg < m < ,lOOkg 100kg < m 60 fills 32 fills .20 fills 10 fills cycle of operation, l.e, that the sum of the loads is equal to the fill, then the separate verification method [paragraph 6(5)(1)] should be used. (2) When using the integral verification method for a cumulative weighing instrument a sub-division of the test fill is unavoidable. When calculating the conventional true value of the mass of the test fill, it is necessary-to consider the Increased uncertainty due to the division of the test fill. (a) Interruption of automatic operation An automatic fillingoperation of a test fill be initiated as for normal operation. However, the automatic operation shall be interrupted twice during each fillingcycle l.e, after the load is assembled and after the load is discharged. An _automatic operation shall not be Interrupted during consecutive weighing cycles if the speed of operation is so high that the interruption would significantly affect the mass of the fill. (i) Pre-discharge (full) Interrupt The automatic operation shall be interrupted after the load receptor(s) has (have) been loaded and the feed of material has ceased. Wheh the load receptor(s) has {have) stabilized, the net weight of the fill indicated or determined by balancing with standard weights shall be recorded and the instrument switched back to automatic operation. (ii) Post-discharge (empty) interrupt The automatic operation shall be ' . Interrupted after the load(s) has (have) been discharged and the load receptor(s) Is (are) ready to receive a further load. When the load receptor(s) has (have) stabilized, the empty load receptor weight . Indicated or determined by balancing with standard weights shall be recorded and the instrument switched back to automatic operation. (6) Preset value The indicated preset value of the fill shall be noted where applicable. (7) Mass and average value of the test fill The test fill shall be weighed on a control instrument and the result shall be considered as being the conventional true value of the test fill. The average value of all the fills in the test shall be calculated and noted. (8) Deviation for automatic weighing The deviation for automatic weighing used to determine compliance of each fill with the maximum ['lT'T Il-1s1""6 3(i)] 465 permissible deviation for automatic weighing paragraph 2(2) shall be the difference between the conventional true value of the mass of the test fill [as defined in paragraph 6(7)] and the average value of all the fills In the test. (9) Preset value error for automatic weighing The preset value error for automatic weighing used to determine compliance with paragraph 2(3) shall be the difference between the average value of the conditional true value of the mass of the test fill (as defined in paragraph 6(7) and the preset value for the fills.

Previous and next

All provisions of Legal Metrology (General) Rule, 2011

Questions people ask

Is annexure B of Legal Metrology (General) Rule, 2011 still in force?

No repeal of Legal Metrology (General) Rule, 2011 is recorded.