Annexure B, Legal Metrology (General) Rule, 2011: ANNEX B
Legal Metrology (General) Rule, 2011 - Annexure B
Text of annexure B
ANNEX B
ANNEX B
ALIGNMENT CALIBRATION OF SINGLE:AXLE
WEIGHING INSTRUMENT
The alignment calibration is not recommended
as an adequate substitute for verification of reference
wagons by full draught weighing and shall only be
applied under the conditions in paragraph 2(8)(iii)(b)
of part II.
1. The altgnment calibration applies to
Instruments that operate by partial weighing of two
axle wagons and the same instrument Is required
for use as the control Instrument for the purposes of
determining the mass of reference wagons in
paragraph 2(8)(111){b}of part II.
2. Instruments that operate by partial weighing
are exempt from the alignment calibration procedure
provided the follow Ing:-
• the top surface of both rails along the
length of the weight zone are vertically
aligned to ±1 mm, and
• the alignment has been checked along
both rails at not less than two positions
on the load receptor and .not less than
two positions within a wagon length from
the load receptor on each associated
apron.
3.For instruments not covered by para 2 of
Annexure B, a calibration correction shall be
determined by applicationof the alignment calibration
procedure in para 4 of Annexure B. This calibration
correction shall be added to each totallzed wagon
weight to determine the mass of each reference
wagon.
·
4. The alignment calibration is conducted with
the use of a single empty uncoupled wagon of a
wheelbase similar to those wagons used for in-motion
testing.· An example of an alignment calibration is
given in para 5 of Annexure B.
Position on load receptor
First axle Leading end
Middle
Trailingend
Second axle
Total of six weighings
Divide total by three
Derived mass of
standard weights
Calibration
Indicated weight (t}
Empty Loaded wagon
5.76 14.27
5.75 14.26
5.75 14.26
5.75 14.25
5.75 14.25
5.74 14.24
34.50 85.53
d = 11.50e = 28.51
f - e - d - 17.01
C·f= ·0.01
I.?--- .:.::.::.
[ <rl'T II <'J1!,!; 3 ( i) ] 453
The calibratlon correction shall be added to the
totalized indicated weight of each reference wagon
weighed while stationary and uncoupled. In the case
given above, the minus sign should be noted.
Therefore, if the totallzed indicated weight Is 41.38,
the corrected weight will be:
41.38 + (·0.01) = 41.37
Note: The calibration correction computed In
this example is not Intended to be typical.
SEVENTH SCHEDULE
HEADING-D
AUTOMATIC GRAVIMETRIC FILLING
INSTRUMENTS
PART I
:z. General definitions
(1) Weighing instrument:
Measuring Instrument that serves to determine
the mass of a load by using the action of gravity on
that load. According to its method of operation, a
weighing instrument Is classified as automatic or non·
automatic.
(2) Automatic weighing Instrument:
An instrument which weighs without the
Intervention of an operator and follows a
predetermined programme of automatic process
characteristic of the Instrument.·
(3) Automatic gravimetric filling instrument:
.
An instrument which fills containers with
predetermined and virtually constant mass of product
from bulk by automatic weighing, and which
comprises essentially an automatic feeding device
or devices associated with one or more weighing units
and the appropriate control and discharge devices.
(i) Associative (selective combination)
weigher:
Automatic gravimetric filling Instrument
comprising one or more weighing units
and which computes an appropriate
combination of the loads and combines
them for subsequent discharge as a fill.
(ii) Cumulative weigher:
Automatic gravimetric filflng Instrument
with one welahlna unit with the facility to
affect the fl{bymore than one weighing
cycle.
(iii) Subtractive weigher:
Automatic gravimetric filling instrument
for which the fill is determined by
4 7 4 t; v1 II - It 'f
controlling the output feed from the weigh
hopper.
(4) Fill:
One or more loads discharged into a single
container to make up the predetermined
mass.
(5) E/ectronlc instrument:
An Instrument equipped with electronic devices
(6) Control instrument:
A weighing instrument used to determine the
mass of the test fills delivered by the filling
instrument.
2. Construction
Note. In this Part the term device is applied to
any part which uses any means to perform one or
more specific functions.
(1) Principal parts
(I) Weighing units:
A device which provides information on
the mass of the load to be measured. This
devise may consist of all or parts of non·
automatic weighing instruments.
(II) Local receptor:
.
Part of measurement Instrument intended
;1.:;;
.
to receive the load.
(111)Feeding device:
Device which providesthe supply of the
product from bulk to the weighing unit. It
may operate in one or more stages.
(iv) Control devices
(a) Feed control device-Device which
regulates the rate of the feed of the
feeding device
(b) Fifi setting device-Device which
allows the setting of the pre-set
value
(c) Final feed cut off device-Device
which controls the cut off the final
feed so that the average mass of
the fills corresponding to the preset
value. This device may include an
adjustable compensation for the
material in flight.
(d) Correction device-Device, which
automatically corrects the setting of
the filling instrument.
(2) Electron le parts
(I) Electronic device:
A device comprising electronic assemblies
4S4 •THE GAZETTE '.OF :INDIA : EXTRAORDINARY [PARTII-SEC. 3(i)]
and: performiogaaaspeciflc ;function.
Electronic devices are usually
manufactured .as separate•unlts and
capable of belngdndependently .tested.
(ii} Electronic sub7ii/ssembly:
A part of an electronic device employing
electronjc. components .and having a
recognizable function of its .own.
(iii) Electronic components:
The smallest physical.entitythat uses the
electron.orvhole .conductlon In semi
conductors,,,gasesor.m :a vacuum.
(3) India,ting:device:
The part ofa measuring instrument that
displays an.lndication.
/Votes-. 1. Fora weighing instrument, the indicating
device is a -set of components, which
displays the value, in units of mass, of
the.result of a weighing operation.
2. For;fflflngInstrument, the indicating device
may indicate either the mass of the load
or difference between the mass and the
preset value provided the Intention is
made ,clear.
(4) Ancilliilrydevices
(i) Zero setting device:
A device for setting the indicating device
to zero when the load receptor Is empty.
(a) Non automatic zero setting device-
A device for setting the Indicating
device to zero by an operator
(b) Semi automatic zero setting device-
A device for setting the indicting
device to zero automatically
following a manual command
(c) Automatic zero setting device-A
device for setting the Indicating
device to zero automatically without
the Intervention of an operator.
( ii}Tare device.
A device for setting the indication to zero
when a load is on the load receptor
(a) without altering the weighing range for
net loads (additive tare device),
or
(b) reducing the weighing range for net loads
(subtractive tare device).
3. Metrologlca/ characteristics
(1) Scale Interval (d):
Value expressed In units of mass, of the
difference between-
(a) the values corresponding to two
consecutive scales mark for analogue
indication.
(b) two consecutive indicated values for digital
indication.
(2) Reference psrtide mess of iii product:
Mass equal to the mean of ten of the
largest elementary particles or pieces of
the product taken from one or more loads.
(3) Pre set value:
Value, expressed In units of mass, preset
by the operator by means of the fill setting
device, in order to define the nominal
value of the fills.
(4) Static set point:
Value of the test weights or masses,
which, in static tests, balance the value,
selected on the indication of the fill setting
device.
(5) Weighing cycle:
The sequence of the operation which
Includes:
(a) delivery of the material to the load
·
receptor;
(b) a weighing operation; and
(c) the discharge of a single discrete
load.
(6) Fina/ feed time:
'
Time taken to complete the last stage of
delivery of the product to a load receptor.
(7) Minimum capiilcity (min):
Smallest discrete load that can be weighed
a'utomatically on a load receptor.
(8) Maximum a,padty (msx):
The largest discrete load that can be
weighed automatically on a load receptor.
(9) Warm up time:
The time between the moment power is
applied to an instrument and the moment
at which the instrument is capable of
complying with the requirements.
(10) Average number of loads per fill:
Half the sum of the maximum and
minimum number of loads per fill that can
be set by the operator or, in cases where
the number of loads per fill is not directly
determined by the operator, either the
["ll'l II-? J(i)] 455
means of the actual number of loads per
fill (If known) In a period of normal
operation, or the optimum number of
loads per fill as may be specified by the
manufacturer for the type of product
which Is to be weighed.·
(11) Rated minimum fill:
The rated value of the fill below which
the weighing results may be subject to
errors outside the limits specified in this
part.
(12) Minimum discharge:
The smallest load that can be discharged
by a subtractive weigher.
4, Indications and errors
(1) Methods of indication
.
(i) Analogue indication:
An Indication allowing the evaluation
of equilibrium position to a fraction
of the scale interval.
(ii) Digital Indication:
An Indication in which the scale mark
comprises a sequence of an aligned
figure that do not permit
Interpolation to fraction of a scale
Interval.
(2) Errors
(i) Error of indication:
The Indication of an instrument
minus the (conventional) true value
of the mass.
(II) Intrinsic error:
The error of an instrument under
reference conditions
(iii) Initial Intrinsic error:
The Intrinsic error of an Instrument
as determined prior to performance
and span stability tests.
(iv) Fault:
The difference between the error of
Indication of an instrument and the
intrinsic error.
Note:Prlncipally, a fault is the result of an
undesired change of data contained in or
flowing an electronic instrument.
(v) Significant fault:
A fault greater than 0.25 of the maximum
permissible deviation of each fill [as
specified in sub-paragraph (2) of para 2]
for an in-service verification, for a fill equal
to the rated minimum fill.
Note: The following are not considered to be
significant faults even when they exceed
the value defined above:
(a) Faults arising from simultaneous and
mutually independent causes in the
instrument.
(b) Faults implying the impossibility to
perform any stage of operation.
(c) Faults being so serious that they are
bound to be noticed by an operator.
(d) Transitory faults bring momentary
variations In the indication or
operation which cannot affect the
final results of the automatic cycles .
· Note. For Instruments where the fill may be greater
than one load, the value of significant fault
applicable for a test on one static load
shall be cakutated in accordance with the
test procedures.
(vi) Span stability.
The capability of an instrument to
maintain the difference between the
Indication of weight at maximum capacity
and the indication at zero within specified
limits over a period of use. ·
(3) Reference value for accuracy class
The value of accuracy class determined
by static testing of the weighing unit
during Influence quantity testing at pattern
approval stage. The reference value for
accuracy class is equal to the best
accuracy class for which the Instrument
may be verified for operational use.
5. Influences and reference conditions
(1) Influence quantity
A quantity, which is not the subject of the
measurement but which influences the
.
value of measurand or the indication of
the instrument.
(i) Influence factor:
An Influence quantity having a value within
the specified rated operating conditions
of the instrument.
(I!) Disturbance:
An Influence quantity having a value within
the limits specified in this Part but outside
the rated operating conditions of the
Instrument
.
J.
_
456 THE·GAZET'FE :oF ,JNDIA: EXTRAORDINARY [PARTII-SEC 3(i)]
(Z) Rated operating conditions
Condltions.of use, ._giving.the ranges of
the· measurand and '.the 'Influence
quantities for which. the ,metrologlcal
charactertsncs.ere: intended .to .lie ·within
the-maxlrnurn permissible :deviation
specified.
(3J Reference condltions
A set of specified values •of 'influence
factors ;f11<ed Jo ,ensure valid Inter
comparison .of lhe results of
measurements.
6. Tests
(1) Mater/a/test:
A test carried out on a complete
instrument using the type of material,
which ,itJs Intended to weigh.
(Z) SimukJ.tion tests:
A test carried out on a complete
instr.ument or a part of an Instrument in
which ariy part of the weighing operation
is simulated.
(3) Pelformance test:
A test to verify whether the equipment
under test (EUT) is able to accomplish its
intended functions
(4) Span stability test:
A test to verify that the EUT Is capable of
maintaining Its perfonnance characteristics
over a period of use.
PART II
1. General
accordance with sub-paragraph (2) of
paragraph 2 and marked Q.T1 the
Instrument in accordance with sub·
paragraph (10) of paragraph 3.
Accuracy class shall be specified for intended
usage, I.e. nature of the product to be weighed, type
of Installation, value of the fill, and operating rate.
Note. The limitation of accuracy classes to
certain application may be determined by these
Rules.
(Z) Limits of error
(I) Maximum permissible error for static tests:
This instrument shall have a reference
value for accuracy class Ref (x), applicable
for static testing only, for which the
maximum permissible error for influence
factor tests shall be as specified in sub
paragraph (4) of paragraph 2, multiplied
by the class designation factor (x).
(II) Maximum permissible deviation of each
fills:
The instrument shall have a specified
accuracy class X(x) for which the
maximum permissible deviation of each
fill from the average shall be equal to the
limits specified in Table I given below,
multipliedby the dass designation factor (x),
(x) shall be 1 x 10k,2 x 10k,5 x 10k,k being
a positive or negative whole number or
zero,
TABLE 1
(See Table 2 under sub-paragraph (3) of para
Initial In service
verification Inspection
M < 50 6.3% 9%
50 < M < 100 3.15g 4.5g
100 < M < 200 3.15% 4.5%
200<M<300 6.3g 9g
300< M < 500 2.1 o/o 3%
500 < M < 1000 10.Sg 15g
1000 < M < 10000 1.05% 1.5%
10000 < M < 15000 105g 150%
15000 < M 0. 7% 1o/o
(1) Scope:
This Part specifies the metrologlcal and
technical requirements for automatic gravimetric
filling instruments (hereafter called "filling
instruments"), which sub-divide a bulk product into
fills of predetermined and virtually constant mass by
automatic weighing, the fills being kept separate.
Note. (1) This specification places no constrain
on the maximum or minimum capacities of the
instruments for which this specification Is applicable.
(2) Fillinginstruments may also be required to comply
with certain requirements e.g. an instrument which
could operate as a non-automatic instrument will
need to comply with the specification for non·
automatic weighing Instruments.
2. Metro/ogla/ requirements
(1) Accuracy dasses
The accuracy class and reference value
for accuracy class shall be specified in
Value of the
mass of the
fill M(g)
Maximum permissible deviation of
each fill from the average for
classX{l}
['IIT'Tn-= J(ilJ 457
6 for the number of fills required to find the average
value)
For in-service testing, when the reference
particle mass exceeds 0.1 of the maximum
permissible in-service deviation, the values derived
from Table 1 shall be Increased by 1.5 times the value
of the reference particle mass. However, the
maximum value of the maximum permissible
deviation shall not exceed (x) x 9%.
Note. Partlde mass correction is not appllcable
to limits, which are derived from Table 1, e.g.
influence quantity test, zero setting, etc.
(3) Maximum permissible preset value error
(setting error):
For instruments where it is possible to preset a
fill weight the maximum difference between the
preset value and the average mass of the fill shall
not exceed 0.25 of the maximum permissible
deviation of each fill from the average, as specified
for in service inspection in item (ii) of sub-paragraph
(2) of this paragraph. This limit will apply for initial
verification and for ln•,ervice inspection testing.
(4) Maximum permissible error far influence
factor tests:
The maximum permissible error for any static
test load during influence factor tests shall be 0.25
of the maximum permissible deviation (as specified
in item (ii) of sub-paragraph (2) of this paragraph
for in-service inspection, corresponding to the
value of a fill equal to that load.
Note. For instruments where the fill may not
be equal to one load, the maximum permissible error
applicable for a test on one static load shall be
calculated in accordance with the test procedures.
(5) Influence factors: Refer ta Annex A for test
conditions.
(i) Static temperatures:
Instruments shall comply with the
appropriate metrological and technical
requirements at temperatures from • 10°C
to +400C. However, for special applications
the limits of .the temperature range may
differ from those given above but such a
range shall not be less than 30'C and shall
be specifiedIn the descriptive markings.
(ii) Power supply {AC}:
Instruments which are powered by an AC
electricity supply shall comply with the
appropriate metrological and technical .
requirements when operated at voltages
from • 15% to + 10% of the reference
voltage.
(iii) Tilting:
Instruments which are not intended for
installation in a fixed position and which
do not have a level Indicter shall comply
with the appropriate metrological and
technical requirements when tilted by 5%.
Where a level indictor Is present it shall
enable the instrument to be set to a tilt of
1% or less.
(6) Units of measurement:
The units of mass to be used on an instrument
· are the milligram (mg), the gram (g}, the kilogram
(kg) and the tonne (t).
3. Technical requirements
(1) Suitability for use
A filling instrument shall be designed to
suit the method of operation and the
products for which It is intended. It shall
be of adequately robust construction so
that it maintains its .netrological
characteristics.
(2) Security of operation
(I) Accidental mat-adjustment:
An instrument shall be so constructed that
an accidental. breakdown or a mal
adjustment of control elements likely to
disturb its correct functioning cannot take
place without its effect being evident.
(ii) Use of a printer:
Any print-out is for information purposes
only and not for use in a commercial
transaction, except for preset values and
number of weighings.
(iii) Andflary devices:
Any ancillary device provided for use with
a filling instrument shall not affect the
correct functioning of the instrument.
(Iv) Salle interval (d):
Scale intervals of ail indicating devices
associated with a weighing unit shall be
the same.
(3) Fill setting device
If fill setting is by means of a scale, it
shall be graduated in units of mass.
If fill setting is by means of weights, they
shall be either weights in accordance with
these rules or purpose-designed of any
nominal value, distinguishable by shape
and identified with the filling instrument.
458 THE GAZETTE OF INDIA: EXTRAORDINARY [PARTJI-SEC. 3(i)]
(4) Final feed cut-off device
The final feed .cut-off device shall be
clearly differentiated .from any other
device.
The direction of movement corresponding
to the sense-of the desired result shall be
shown, where applicable.
(5) Feeding device:
. The feeding device shall be designed to
provide sufficient and regular flow rate(s).
An adjustable feeding device shall be fitted
with an Indication of the direction of
movement corresponding to the sense of
the adjustment of the feed where
appllcable.
(6) Load receptor
(i) The load receptor, and feed and discharge
devices as appropriate, shall be designed
to ensure that residual material retained
after each discharge is negligible.
(ii} Instruments using the subtractive
· weighing principle shall be designed to
ensure that residual material retained at
feed from the discharge gate Is negligible.
(iii) The load receptor shall provide access and
facilities so that where necessary test
weights or masses up to the maximum
capacity can be placed in position, in a
safe and secure manner. If these facilities
are not a permanent fixture of the
instrument, they should be kept In the
vicinity for the instrument.
(iv) Manual discharge of the load receptor
shall not be possible during automatic
operation.
(7) Zero setting and tare devices
(i) The filling Instrument shall be provided
with a zero-setting device, which may also
be used for the setting of tare. The device
may be-
(a) Manual;
(b) Semi-automatic, or
(c) Automatic
(ii} Zero-setting and tare devices shall be
capable of setting to less than or equal to
0.25 of the maximum permissible
deviation for in-service inspection for a
fill equal to the minimum capacity for
·instruments with one weighing unit, and
for a fill equal to the rated minimum fill
for selective combination weigher.
(iii) Non-automatic or semi-automatic zero
setting and tare device should be locked
during automatic operations.
(iv) The weighing unit shall b'e in stable
equilibrium when the zero setting and tare
device is being set.
(8) Equilibrium mechanism
The equilibrium mechanism may be provided
with detachable masses which shall be either weights
in accordance with OIML requirements or purpose
designed masses of any nominal value,
distinguishableby shape and Identified with the filling
instruments.
(9) Security
A security means shall be provided for
components and pre-set controls to which access is
prohibited.
(10) Descriptive markings
FIiiing Instruments shall bear the following
markings
.
(I} Marking shown in full
(a) Name or identification mark of the
manufacturer
(b) Name or identification mark of the
Importer (if applicable)
(c) Serial number and type designation of the
Instrument
(d) Product(s) designation {i.e. materials that
may be weighed)
'
(e) Temperature range (if applicable, see
2.5.1) in the form
....... C/ ....
C
· (fl Electrical supply voltage in the form
......
V
(g} Electrical supply frequency in the form
..
,Hz
(h) Working fluid pressure (if applicable) in
the form
......
kPa
(I} Average number of loads/fill (if applicable)
(J} Maximum fill (if applicable)
(k) Minimum fill (If applicable)
(I} Rated minimum fill
(m) Maximum rate of operation (if applicable}
In the form
.......
Load per minute
(ii} Marking shown in code
(a) Pattern approval sign
{b) Indication of the accuracy class X(x)
(c) Reference value for accuracy class Ref (x}
(d) Scale interval (if applicable) in the form
d =
"""""
['WT II-? J(i)] 459
(e) Maximum capacity (or minimum discharge
where applicable) in the form Max=
.....
(f) Maximum additive tare in the form
T= +
......
(g) Maximum subtractive tare In the form
T = -
.
An instrument may be verified for different
materials for which different classes shall apply or
which require different operating parameters to
maintain limits of error. Marking shall be such that
the alternative class or operating parameters are
clearly associated with the appropriate material
designation.
In the case of subtractive weigher the minimum
load to be discharged shall be specified.
(iii) Presentation of descriptive markings:
The descriptive marking shall be indelible and
of a size, shape and clarity to enable legibility under
normal conditions of use of the filling instrument.
They shall be grouped together in a clearly visible
place on the fillinginstrument, either on a date plate
fixed to the instrument or on the filling instrument
itself.
Where the markings are placed on a date plate,
it shall be possible to seal the plate bearing the
markings. Where they are marked on the filling
instrument Itself it shall not be possible to remove
them without destroying them.
The descriptive markings may be shown on a
programmable display, which is controlled by
software. In this case, means shall be provided for
any access to reprogramming of the markings to be
automatically and non-erasable, recorded, e.g. by
traceable access software when a programmable
display is used, the plate on the instrument shall
bear at least the following markings:
(a) Type and designation of the instrument;
(b) Name or identification mark of the
manufacturer;
(c) Pattern approval number;
(d) Electrical supply voltage;
(e) Electrical supply frequency; and
(f) Pneumatic pressure
(11) Verification marks
(i) Position:
The filling instruments shall have a place for
the application of verification marks. This place shall:
(a) be such that the part of which it is located
cannot be removed from the filling
instrument without damaging the marks
(b) allow easy application of the mark without
changing the metrological qualities of the
filling instruments
(c) be visible without the filling instrument
having to be moved when it is in service.
(ii) Mounting:
Filling instruments required to bear verification
marks shall have a verification mark support, at the
place provided for above, which shall ensure the
· conservation of the marks.
When the mark is made with a stamp, this
support may consists of a strip of lead or any together
material with similar qualities, Inserted into a plate
fixed to the filling instrument or a cavity bored In the
filling instrument itself.
When the marks consist of an adhesive transfer,
a space shall be prepared for this purpose.
(12} Control instrument:
The control instrument may be separate from
an integral part of the filling Instrument.
4. Requirements for electronic Instruments
Electron le filling instruments shall comply with
the following requirements, In addition to the
-appllcable requirements of all other clauses of this
specification.
(1) General requirements
(i) Rated operating conditions:
Electronic operating conditions
Electronic instrument shall be so designed
and manufactured that they do not exceed
the maximum permissible errors under
rated operating conditions.
(ii) Influence fadors:
An electronic Instruments shall comply
with the requirements of sub-paragraph
(5) of paragraph 2 and shall also comply
with appropriate metrological and
technical requirements at a relative
humidity of 85% at the upper limit of the
temperature range of the Instrument.
(iii) Disturbances:
Electronic instruments shall be so
designed and manufactured that when
exposed to disturbances, elther:-
(a) significant faults do not occur, i.e.
the difference between the weight
indication due to the disturbance and
- - -
__l.
__ ----
460 THE iG)\:ZEirIT'E or INDIA: EXTRAORDINARY [PARTII-SEC. 3(i)]
the · indication .wtthout the
dlsturbance'(intrinsic error) shall not
exceed 'the value specified in
paragraph %2)(\i),Of Part I; or
(b) significant .faults .are detected and
acted .upon.
Note: A fault:equalto erlessthan the value in
paragraph 4(2)(v),of;Part.I,is:allowed irrespective of
the value of the error-of.lndkatlon.
(iv) Evaluation for ,compliance:
The pattern ,of :an electronic instrument is
presumed to comply with the requirements of
paragraphs 4(1)(i),4(1)(ii) and 4(1)(i1i), If it passes
the examination and tests specified in Annex A.
(v) Application:
Therequlrernents in paragraph 4(1) (iii) may
be applied separately-
(a) either to each individual cause of
significant fault; or
(b) to eadi part of the electronic Instrument;
or
(c) to both.
The choice of whether sub-item (a) or sub
item (b) above should be applied is left to
the manufacturer.
(2) Functional requirements
(i) Indicator display tests:
If the failure of an indicator display
element can cause a false weight
indication, then the instrument shall have
a display test facility which is automatically
initiated at switch-on of indication, e.g.
indication of all the relevant signs of the
indicator in the active and non-active
states for a sufficient time to be easily
observed by the operator.
(ii) Acting upon a significant fault:
When a significant fault has been
detected, the Instrument shall either be
automatically made inoperative or a visual
or audible Indication shall be provided
automatically and shall continue until such
time as the user takes action or the fault
disappears.
(iii) Warm-up time:
During the warm-up time of an electronic
instrument there shall be no indication or
transmission of the result of weighing, and
automatic operation shall be Inhibited.
(iv) Interface:
An instrument may be equipped with an
interface within allows it to be coupled to
external equipment. When an interface is
used, the instrument shall continue to
function correctly and its metrologlcal
functions shall not be influenced.
(v) Battery power supply:
An instrument that operates from a
battery power supply shall, whenever the
voltage drops below the manu_facturer's
specified minimum value, either continue
to function correctly or automatically be
put out of service,
(3) Examination and tests:
The examination and testing of an electronic
instrument is intended to verify compliance with the
applicable requirements of this specification and
especially with the requirements of this paragraph.
(i) Examinations
An electronic instrument shall be ·
examined to obtain a general appraisal
of the design and construction.
(ii) Performance tests:
An electronic instrument or electronic
device, as appropriate, shall be tested as
specified in Annex A to determine the
correct functioning of the instrument.
Tests are to be carried out on the whole
instrument except when either the size
or the configuration or both of the
instrument does not lend itself to testing
as a unit. In such cases the electronic
devices shall be tested, where possible
as a simulated instrument including all
electronic elements of a system which can
affect the weighing result. In addition, an
examination shall be carried out on the
fully operational instrument.
Susceptibility that would result from the
use of electronic Interfaces to other
equipment shall be simulated in the tests.
(iii) Span stability:
When an electronic Instrument is subjected
to the span stability test specified in
paragraph 7 of Annexure A, the absolute
value of the difference between the errors
obtained for any two measurements shall
not exceed half the maximum permissible
error for influence factor tests for a near
maximum capacity load.
5. Metrologica/ controls
(1) General
(i) Provision of means of testing:
["!T'TII-- 3( i)]
··-- ---·------- --
--------
461
For the purposes of testing, the
metrological authority may require from
the applicant the product (i.e. the material
to be weighed), the handling equipment,
the appropriately qualified personnel and
a control Instrument.
(ii) Material tests:
(a) For pattern evaluation-The material used
as the test load for pattern evaluation shall
be representative of a product for which
the Instrument is designed. The test shall
be conducted in accordance with the test
procedure in paragraph 8(1) of Annexure
A.
(b) For initial verification and in-service
verification-The in-situ material tests
shall be done in accordance with the
descriptive marking under the normal
conditions of which the instrument Is
intended. The test shall be conducted in
accordance with the test procedure laid
down In paragraph 8(2) of Annexure A.
(2) Pattern approval
(IJ Documentation:
The application for pattern approval shall
include documentation comprising:
(a) metrologlcal characteristics of the
Instrument,
.
(b) a set of specifications for the
Instrument,
(c) a functional description of the
components and services,
(d) drawings, diagrams and general soft
ware information (if applicable),
explaining the construction and
operation, including interlocks,
(e) any document or other evidence that
the design and construction of the
Instrument complies with the
requirements of this specification.
(ii) General requirements:
The pattern evaluation shall be carried out
on one or more (and normally not more
than three) units that represent the
definitive pattern. One or more of the units
shall be complete and fully operational of
the purposed of paragraph 5(2)(iii)(a).
One or more of the units shall be
submitted in a form suitable for simulation
testing In a laboratory and shall include
the whole of the electronics which affect
li74 t. {)11-1/6
the weighing result except In the case of
an associative weigher where only one
representative weighing unit may be
included.
The Instrument or simulated Instrument
shall have a load Indicator, or an interface
allowing access to a quantity that can be
calibrated to provide an indication of load,
so that the requirements of paragraph
2(2)(i) (maximum permissible errors for
static tests) may be tested and so that
the instrument may be tested for influence
quantities. The scale Interval of the load
indicator shall not exceed 0.125 of the
maximum permissible deviation for In
service verification for a fill equal to the
minimum capacity.
The evaluation shall consist of the tests
specified in paragraphs S(l)(ii)(a) and
5(2)(iii).
(iii) Pattern evaluation:
The submitted documents shall be
examined and tests carried out to verify
that the instrument complies with:
(a) the requirements specified for static
tests in paragraph 2,
(b) the technical requirements in
paragraph 3,
(c) the requirements In paragraph 4 for
electronic Instruments, where
applicable.
The appropriate metrologlcal authority
shall conduct the test In a manner, which
prevents an unnecessary commitment of
resources.
Note: The appropriate metrological authority
may accept with the consent of the applicant,
equivalent test data obtained from other metrological
authorities.
(a) Tests for compliance with technical
requirements
Operational test with material shall be
done in accordance with the procedure In
paragraph 8(1) of Annex A on a complete
instrument to assess compliance with the
technical requirements of paragraph 3.
(b) Influence factor tests
Influence factors shall be applied during
simulation tests in a manner that will reveal
a corruption of the weighing results of any
weighing process to which the instrument
should applied, in accordance with-
462 THE GAZETTE OF INDIA: EXTRAORDINARY [PARTII-SEC. 3(i)]
(a) Paragraph 2(5) for all instruments;
and
(b) Paragraph 4 for electronic
Instruments
(c) Apportioning of errors
Where parts of an instrument are
examined separately in the process of
pattern approval, the following
requirements .shall apply-
The error limits applicable in a part which
is examined separately are equal to a
fraction P1of the maximum permissible
errors or the allowed variations of the
indication of the complete instruments.
The fractions for any part have to be taken
for the same accuracy class as for the
complete Instrument incorporating the
part.
The fractions Pi shall satisfy the following
equation:
(P12+ Pi2 + Pi+ .... ) s 1
'
The fraction Pi, shall be chosen by the
manufacturer of the part and shall be
verified by an appropriate test. However,
the fraction shall not exceed 0.8 and shall
not be less than 0.3 when more than one
part contributes to the effect In question.
Note: As the requirements of this sub
paragraph only apply to the instrument submitted
for pattern evaluation and not to those subsequently
submitted for verification, the means by which it will
be possible to determine whether the appropriated
maximum permlss!ble error or maximum allowable
variation has been exceeded while decided mutually
between the metrologlcal authority and the applicant.
The.means may be for example:
(a) the provision or adoption of the
indicating device to give the required
resolution,
(b) the use of change point weights, or
(c) any other means mutually agreed.
(iv) Place to testing:
Instruments submitted for pattern
approval may be tested either-
{a) on the premises of the metrological
authority to which the application
has been submitted; or
(b) In any other suitable place agreed
to between the metrologlcal
authority concerned and the
applicant.
(v) Certificate of approval and determination
of paragraph [2(2)(1) and paragraph 5 of
Annex A}:
The pattern approval certificate shall state
the reference value for the accuracy class
as determined by the static tests In
paragraph 5 of Annex A and shall state
that the actual class (equal to the greater
than the reference value) shall be
determined by compliance with the
metrological requirements at initial
verification.
(3) Initial verification
(i) General requirements:
Instruments shall be examined for
conformity with the approved pattern
where applicable and shall be tested for
compliance with paragraph 2 [excluding
paragraph 2(2)(i) and paragraph 2(5)] for
the intended products and corresponding
accuracy classes under normal conditions
of use.
Tests shall be carried out by the
appropriate metrological authority, in-situ,
with the instrument fully assembled and
fixed in the position in which it is intended
to be used. The installation of an
instrument shall be so designed that an
automatic weighing operation will be the
same whether for the purposes of testing
or for use for a transaction.
(ii) Material tests:
Material test shall be carried out In
compliance with paragraph S(l)(ii) using
the test methods specified in paragraph
6.
(iii) Conduct of the test:
The appropriate metrological authority:
(a) shall conduct the test in a manner
which prevents an unnecessary
commitment of resources,
(b) may, where appropriate and to avoid
duplicating tests previously done on
the instrument for pattern evaluation
under paragraph 5(2)(iii) (a) use the
resu Its of observed tests to assess
for initial verification.
(Iv) Determination of accuracy class:
[ '1T'T II-? 3 (i)]
... '--.------- - - --
----,------,---
463
The appropriate metrological authority
shall:-
(a) determine the accuracy class for the
materials used in the tests in
accordance with paragraph 5(2)(v}
by reference to the material test
results and the limits of error
specified in paragraph 2(2)(ii) and
paragraph 2(3} for initial verification,
(b) Verify that accuracy classes
determined as above.
(4) In-service verification
In-service verification shall be as specified in
para 5(3)(1) and para 5(3)(ii). The maximum
permissible errors shall be as specified in para 2(2)(ii)
for in-service verification.
The appropriate metrological authority shall
conduct the test in a manner, which prevents an
unnecessary commitment of resources.
Quite often a question is asked as to how to
calculate R(x). The method is as follows:
Ref. (x) or simply x, is a numeral indicating the
class of the gravimetric filling machine, which is to
be assigned at the time of model approval. Once the
value of x or Ref. (x) Is assigned, one can determine
its MPD by multlplylngthe figures given in Table I of
this part of this book, At the time of verification, one
has to see only for the compliance to the MPD and
other requirements.
For calculating x or Ref. (x), one should calculate
(1) value of setting error se and (2) maximum
deviation md as explained in this part on. Remember
mpse(1) is 1/4 of mpd1given in table of 5(3)(i) and
last column of the observation sheet gives deviation
from the mean fill, so maximum deviation md is
calculated from here. Number of fills to be taken is
chosen with the help of Table 2.
Then examine the values of se/mpsel and md/
mpd(l), these may be a whole number (including
zero) plus a fraction like 2 + 0.325 and 2 + 0.545
then class of the Instrument is one plus the whole
number irrespective of the value of fraction, i.e. 3 in
the given example. In case whole number is zero
then Ref. (x} or x is 1.
If the values of se/mpse 1 and md/mpd1 are
3.575 and 3.265 then Ref. (x} or x will be 5 and not
4, as x has to be either of 1, 2, 5 or multiples of 10 of
1, 2 and 5.
6. Test methods
(1) Determination of the mass of individual
fills
The mass of the individual fills is
determined using one of the methods
specified in paragraph 6(5)(1) or in
paragraph 6(5)(1iJ.
(2) Conduct of material tests
(a) The tests shall be carried out on fills using
loads at or near to, the maximum capacity
and also at, or near to, the minimum
capacity. Material tests should only be
carried out with the products the
instrument is intended to be used for.
(b) Cumulative welghers shall be tested as
above with the maximum practicalnumber
of load per fill and also with the minimum
number of loads per fill, and associative
weighers as above with the average (or
optimum) number of loads per fill.
(c) If the minimum capacity is less than one
third of the maximum capacity then tests
shall also be carried out near the centre
of the load weighing range preferably at
a value close to, but not above, lOOg,
300g, 1000g or 15000g as appropriate.
(d) All tests shall be conducted with any
adjustable parameter critical to
inetrologicalintegrity e.g. final feed time
or rate, set to the most onerous condition
allowed by the manufacturer's printed
instructions and incorporated in the
descriptive markings.
(i) Testing the effect of a correction device
(a) Any correction device e.g. in-flight
correction or automatic zero setting fitted
to an instrument shall be operated during
the tests according to the manufacturer's
printed instructions.
(b} If the correction device is not activated
during each filling operation, then tests
at minimum capacity shall be arranged to
Include the effect of one or more regular
operations of the correction device, e.g.
buy including in the test at least three
fills immediately before and after the
activation of the device.
(c) The Initial fills after the charge between
maximum capacity and minimum capacity
shall be included In the test unless the
instrument bears a clear warning to
discard the stated number of fills after a
change to the instrument settings.
---··
----4-?--
464 "i'HE rGAZEif'f,E OF !NOIA: EXTRAORDINARY [PARTJI SEC. 3(i)]
(3) Number of fills
The number df individual tests fill depends upon
the preset value (m):as specified inTable2 below:
TABlE;2
(4) Accuracy of stendsrds
The control 'Instrument :and standard weights
used in testing shalhensurethat·checklng of the test
fills to an error m>tigreater than either-
(a) one-third of the maximum permissible
deviation and maximum permissible
preset value error (as appropriate) for
automatic weighing (paragraph 2(2) and
paragraph 2(3) respectively) if the control
instrument or the device used for control ·
purposes Is verified immediately prior to
the material test; or
(b) one-fifth of the maximum permissible
deviation and maximum permissible
preset value error (as appropriate) for
automatic weighing (paragraph 2(2) and
paragraph 2(3) respectively] in all other
cases;
(5) Material test methods
(i) Separate verification method:
The separate verification method requires
the use of a (separate) control Instrument
to find the conventional true value of the
mass of the test fill.
(ii) Integral verification method:
With this method the instrument being
tested Is used to determine the
· conventional true value of the mass of the
test fill. The Integral verification method
/
shall be conducted using either:-
(a) an appropriately designed Indicating
device, or ·
(b) an indicating device with standard
weights to assess the rounding error.
The total uncertainty of the test method
(separate or integral verification) shall be
not greater than one-third of the
maximum permissible error for the
instrument.
Notes-; (1) The integral verification method
depends on determining the masses of the loads.
Limits of error as specified in paragraph 2(2) are for
the mass of the fill. If it Is not possible to ensure that
In normal operation, all load is discharged at. each
m < 10kg
10kg s m ?,25kg
25kg < m < ,lOOkg
100kg < m
60 fills
32 fills
.20 fills
10 fills
cycle of operation, l.e, that the sum of the loads is
equal to the fill, then the separate verification method
[paragraph 6(5)(1)] should be used.
(2) When using the integral verification method
for a cumulative weighing instrument a sub-division
of the test fill is unavoidable. When calculating the
conventional true value of the mass of the test fill, it
is necessary-to consider the Increased uncertainty
due to the division of the test fill.
(a) Interruption of automatic operation
An automatic fillingoperation of a test fill
be initiated as for normal operation.
However, the automatic operation shall be
interrupted twice during each fillingcycle
l.e, after the load is assembled and after
the load is discharged.
An _automatic operation shall not be
Interrupted during consecutive weighing
cycles if the speed of operation is so high
that the interruption would significantly
affect the mass of the fill.
(i) Pre-discharge (full) Interrupt
The automatic operation shall be
interrupted after the load receptor(s) has
(have) been loaded and the feed of
material has ceased. Wheh the load
receptor(s) has {have) stabilized, the net
weight of the fill indicated or determined
by balancing with standard weights shall
be recorded and the instrument switched
back to automatic operation.
(ii) Post-discharge (empty) interrupt
The automatic operation shall be
' .
Interrupted after the load(s) has (have)
been discharged and the load receptor(s)
Is (are) ready to receive a further load.
When the load receptor(s) has (have)
stabilized, the empty load receptor weight .
Indicated or determined by balancing with
standard weights shall be recorded and
the instrument switched back to automatic
operation.
(6) Preset value
The indicated preset value of the fill shall be
noted where applicable.
(7) Mass and average value of the test fill
The test fill shall be weighed on a control
instrument and the result shall be considered as being
the conventional true value of the test fill. The
average value of all the fills in the test shall be
calculated and noted.
(8) Deviation for automatic weighing
The deviation for automatic weighing used to
determine compliance of each fill with the maximum
['lT'T Il-1s1""6 3(i)] 465
permissible deviation for automatic weighing
paragraph 2(2) shall be the difference between the
conventional true value of the mass of the test fill
[as defined in paragraph 6(7)] and the average value
of all the fills In the test.
(9) Preset value error for automatic weighing
The preset value error for automatic weighing
used to determine compliance with paragraph 2(3)
shall be the difference between the average value of
the conditional true value of the mass of the test fill
(as defined in paragraph 6(7) and the preset value
for the fills.
Previous and next