Annexure A, Legal Metrology (General) Rule, 2011: ANNEXUREA

Legal Metrology (General) Rule, 2011 - Annexure A

Text of annexure A

ANNEXUREA

ANNEXUREA TEST PROCEDURES FORAVTOMATIC RAIL· WEIGH BRIDGES (Mandatory) Meaning of symbols: I ; Indication L = Load D.L = Additional load to next changeover point P = I + 0.5d • D-L; Indication prior to rounding Pn = nth indication prior to rounding compliance with the requirements ,in peraqraphs 3(3) and 3(4) of this part. (iv) Provisions of means for. testing: For the purpose oftesting,lthe.applicantmay be required to.furnlsh the ,Metrological Enforcement Authority with .the test vehicles, material,.qualified personnel and a control instrument. The Instrument under test may be used as a controlinstrument provided it complies with the.requirements in paragraph 2(8)(111)(b)of this part. (v) Place of.testing: Instruments submitted for pattern approval may be.tested at the following places: • a:slte at which all necessary tests can be conducted and agreed upon between the Metrological Authority and the applicant, • a 'laboratory considered appropriate by the Metrological Authority, • any other suitable place mutually agreed upon by the Metrological Authority and the applicant. (2) Verification (i) Tests : Instruments shall comply with the requirements In paragraph 2 [except 2(9)] and paragraph 3 of this part for any product(s) for which they are Intended and when operated under normal conditions of use Tests shall be carried out in-situ, In a normal installation. Toe instruments shall be Installed so that an automatic weighing operation will be virtuallythe same for testing as It Is for a transaction. (ii) In-motion tests : In-motion tests shall be conducted in accordance with paragraph 5(1)(11i)(a)of this part with the exception that the types of vehicles and, for test with coupled wagons the number of vehicles in the test train shall be in accordance with the normal operation of the Instruments and used in accordance with paragraph6 of this part. (Iii) Provisions of means of testing: For the purposes of testing, the applicant may be required to furnish the enforcement authority with the test vehicles, material, Total number of wagons In test train(n) n < 10 10 < n < 30 30 < J\ TABLEN Minimum number of reference wagons n 10 15 441 d = Scale interval E = P - L = error Eo = Error calculated at zero Ee = Corrected error rnpe = Maximum permissible error EUT = Equipment under test Max= Maximum capacity Min = Minimum capacity 1.. Documentation [5(1.)(i)J Review the documentation that is submitted, includingnecessary photographs, drawings, diagrams, general software information, relevant technical and functional description of main components, devices etc. to determine if It Is adequate and correct. Consider the operational manual. 2. Comparing construction with documentation 5(1.)(i) Examine the various devices of the Instrument to ensure compliance with the documentation. Examine the simulator to ensure that it is adequate for train movement simulation. It shall provide the signals from track switches, or other vehicle type identification devices, normally transmitted when a vehicle passes over the weighing system. It is not expected to simulate effects such as dynamic loading. 3. Initial examination (1) Metrologicalcharacterlstics: Note metrological characteristics in the test report. (2) Descriptive markings [3(6)]: Check the descriptive markings. (3) Verification marks [3(1)]: Check the arrangements for verification marks. 4. General ( 1) General requirements for electronic instruments under test (EUT) : Energize the EUT for a time period equal to or greater than the warm-up time specified by the manufacturer and· maintain it energized for the duration of the test. Adjust the EUT as closely as practicable to zero prior to each test, and do not readjust it at anytime during the test, except to reset it if a significant fault has been indicated. The deviation of the no-load indication due to any test condition shall be recorded, and any load indication shall be corrected accordingly to obtain the weighing results. The handling of the instrument shall be such that no condensation of water occurs on the instrument. (i) Indication .with a scale Interval smaller thand If an instrument with digItal indication has a device for displaying the Indication with a smaller scale Interval, this device may be used to calculate the error. If a device Is used It should be noted In the test report. (2) calculatton of error: For instruments with digitalIndication with a test scale Interval [see note in S(l)(IIIXb)Jless or equal to 0.1 d, the device may be used directly to read the error. For Instruments with digital Indication and with a test scale Interval greater than 0.1.d, changeover points are used to determine the Indication of the instrument, prior to rounding, as follows:- At a certain load, L, the Indicated value, I, is noted. Additional weights of for example O. ld are successivelyadded until the indication of the instrument is increased unambiguouslyby one scale interval (I+d). The additional load -Ladded to the. load receptor gives theindlcatlon, P,prKlrto rounding by using the following formula:- P=I+O.Sd-t,L The error prior to rounding is: E=P-L Thus E = (I + 0.Sd - L',L)- L Example : An instrument with a scale interval, d, of 10 kg Is loaded with 1000 kg and thereby indicates 1000 kg. After adding successive weights of l kg. the indication changes from 1000 kg to 1010 kg at an additional load of 3 kg. Inserted in the above formula these observations give: P = (1000 + 5 - 3) kg = 1002 kg Thus the true Indication prior to rounding is 1002 kg, and the error Is: E = (1002 - 1000) kg = 2 kg 5, Testprogram (1) Pilttern evaluation [5(1)]: All tests of paragraphs 6 to 10 of Annexure A shall normally be applied for pattern evaluation. The tests for paragraphs 6 to 10 of Annexure A shall be performed with static load, a wheel movement 442 THE GAZETTE OF ,INDIA: EXTRAORDINARY [PART11-SEC. 3(i)] simulator (switches) may be used if necessarvfor the calculation of the weighing results. (2) Initial verifieatlon_[5(2)}: Initial verification tests normally.only:fndudesthe tests in paragraph 11 of Annexure A. The test shall· Include all dynamic 'ln-rnotlon effects corresponding to normal operation of the instrument 6. Perfonnancetuts.during pattem evaluation The equipment under test should have the following items associated with it, for the purpose of testing:- • Weightslmulator. • Train wheel simulator. • For practlcal reasons, the weight simulator may take·varlous forms. For example, it may be a weigh pan or platform scale of approximately 1/ 1000th of the weight range of a site Installation. Whichever method is adopted, It must be independently calibrated and readable to at least O. ld. {1}General conditions (i) Temperature : The tests shall be performed at a steady ambient temperature, usually normal room temperature unless otherwise specified. The temperature is deemed to be steady when the difference between the extreme temperatures noted during the test does not exceed one-fifth of the temperature range of the given Instrument without being greater than 5°C and the rate of change does not exceed S°C per hour. (Ii) Power supply; Instruments using electric power shall normally be connected to the power supply and "on" throughout the tests. (iii) Automatic zero-setting: During the tests, the automatic zero-setting device may be switched off. When this is done it should be mentioned In the test report. ' For certain tests the test descriptionspecifies whether the automatic zero-setting shall be operative or inoperative. (iv) Recovery: After each test the instrument should be allowed to recover sufficiently before the following test. (2) Checkingof zero (I) Range of zero-setting {J(J)(iv)] (a) Semi-automatic zero-setting : This test shall not be carried out during the span stability test. With the load receptor empty, set the instrument to zero. Place a test load on the load receptor and use the zero­ setting device. Continue incrementing the test load until the zero-setting device fails to re-zero when activated. The maximum load that can be rezeroed is the positive portion of the zero-setting range. To test the negative portion of the zero-setting range, recalibrate the instrument with an additional weight on the load receptor at the zero and span points. This additional weight should be greater than the negative zero-setting range. Then remove weights and after each weight Is , removed, use the zero-setting device. The maximum load that can be removed while the instrument can still be re-zeroed by the zero-setting device, is the negative portion of the zero-setting range. Recalibrate the instrument without this additional weight. (b) Automatic zero-setting : This test shall not be carried out during the span stability test. With the load receptor empty, add weights in small quantities and after each weight is added allow time for the automatic zero-setting device to function to see if the instrument Is re­ zeroed automatically. Repeat this. procedure until the instrument will not be re-zeroed automatically. The maximum load that can be re-zeroed is the positive portion of the automatic zero-setting range. ["WI II ? 3(i)] 443 To test the negative portion of the zero-setting range recalibrate the Instrument with an additlona Iweight on the load receptor at the zero and span points. This additional weight should be greater than the negative zero-setting range. Then remove weights and after each weight is removed, allow time for the automatic zero-setting device to function to see If the Instrument is re-zeroed automatically.The maximum load that can be removed while the Instrument can still be re-zeroed by the zero­ setting device, is the negative portion of the zero-setting range. Re-zero the instrument without this additional weight. (3) Setting to zero before loading: For instruments with digital Indication, the adjustment to zero, or the determination of the zero point is carried out as described in paragraph 6(5)(1) of this Annexure. (4) Association of indicating and printing device [2(7)J If the instrument has more than one indicating device, the indications of the various devices (both indicating and printing) are compared during the test. (5) Non-automatic tests of the control Instrument Note : This sub-clause is only applicable to instruments which are to be used as control instruments. (i) Accuracy of zero-setting [2(8)(/J(b)J (a) Semi-automatic zero-setting : The accuracy of the zero-setting device is tested by setting the instrument to zero and then detennlning the additional load at which the indication changes from zero to one scale interval above zero. The error at zero is calculated according to the description in paragraph 4(2) of this part. (b) Automatic zero-setting : The Indication is either brought outside the zero range, or the automatic zero-setting Is disabled. Then the additional load at which the indication changes from one scale interval to the next above Is determined and the error is calculated according to the description in paragraph 4(2) of this Annexure. It is assumed that the error at zero load would be equal to the error at the load in question. (II) Determination of weighing performance (a) Preloading : · Before the first weighing test the Instrument shall be preloaded once to Max. (b) Weighing test[2(8)(1)(e-1)]: Applysubstitution material from zero up to and Including Max, and then remove the material back to zero. When determining the initial intlinsic error, at least ten different load values are selected, and for other weighing tests at least five are selected. The values of the loads selected shall include Max and Min, and values at or near those at which the maximum permsslble error(mpe) changes. It should be noted that when loading or unloading weights the load must be respectively increased or decreased In a uniform progression. (111)Discrimination test [2(8)(i)(vi)J: The following tests are performed with three different loads, e.g. Min,0.5 Max and Max. A load plus suffidentsubstitution material (e.g. 10 times 0.1 d) is placed on the load receptor. The additional material is then successively removed until the indication, I, is decreased unambiguouslyby one actual scale interval, I· d. Replacesubstitution material equivalent to 0.1 d and then a load equal to 1.4 d shall be gently placed on the load receptor and the result will be increased by one actual scale Interval above the initial indication, I+ d, Example : instrument with d = 10g 0 210g I .,. I - d = l 90 g I = 200 g I+ d Add 1/lOd = Lg add?i: d = 14 g,.,... lJ, 0 190( 11 200 g I jj -- J --- -------------I 444 THE GAZETTE OF INDIA: EXTRAORDINARY [PARTII-SEc.3(i)] TABLEV Environmental Phenomena Test Specification Reference of 20°C (1) Static temperature tests [2(9)(i)}: Static temperaturetests are carried out according to Table V. Supplementary information to the test procedure Object of the test : To verify compliance with the provisions In paragraph 2(9)(1) of part II under ?onditlonsof dry heat (non-condensing) and cold. Spedfledhighfor 2 hours Specified low for 2 hours 5°C Reference of 20°c Temperature Test procedures in brief : Precondition : 16 hours · Condition of the EUT : Normal power supplied and "on" for a time period equal to or greater than the warm-up time specified by the manufacturer. Power Is to be "on" for the duration of the test. Stablllzatlon : 2 hours at each temperature under "free air" conditions. Temperature : As specified in paragraph 2(9){1)of part II. Temperature sequence : Reference tempera· ture of 20°C; Specified high temperature; Specified low temperature; A temperature of 5°C; Reference temperature of 20°C. Number of test cycles : At least one cycle. Weighing test : Adjust the EUT as close to zero indication as practicableprior to the test (ifan automatic zero-tracking device is connected, adjust it to a value near zero). The EUT shall not be readjusted at any time during the test. After stabilization at the reference temperature and again at each specified temperature, apply at least five different test loads or simulated loads and record: (a) date and time; (b) temperature; (c) relative humidity; (1) Static temperature Influence factor mpe(*) (2) Damp heat, steady Influence factor mpe state (3) Mains power supply Influence factor mpe voltage variation (AC) (4) Battery power supply Influence factor mpe Voltage variation (DC) (*) mpe : maximum permissible error The indicatlon.-at start is I-= 20Qg. Remove additional weights until the indication.changes to I - d = 190g. Add O.ld -;lg and thereafter 1.4 d .= 14g. The Indications must then be I+ d = 210g. 7. Additional functionality (1) Warm-uptimetest[4(3)(v)J (i) Disconnect.the instrument from the power supply.for a;period•ofat least 8 hours prior to the:test. (ii) Reconnect the instrument and switch on while observing the indicating device. Verify that it is not possible to initiate automatic weighing or printout until the indication has stabilized or until completion of the warm­ up time If this is specified by the manufacturer. (iii) As soon as the indication of the Indicating device has stabilized, set the instrument to zero and determine the error of zero-setting. (iv) Apply a load close to Max. Determine the error by the method in paragraph 4(2) of this Annexure. (v) Repeat stages (3) and (4) after S, 15 and 30 minutes. (2) Agreement between indicating and printing devices [2(7)}. During the course of the tests verify that for the same load, thedfference between any two Indicating devices having the same scale interval is as follows: Zero for digital indicating or printing devices; Not greater than the maximum permissible error for weighing-in-motion for analogue devices. 8. Influence factor tuts SUMMARY OF TESTS Test Characteristics Conditions under test applied t::t,t .i?.--- ... __ [ "l1'T II -1glr,l" 3 ( i ) ] 445 (d) test load; (e) indications (as applicable); (f) errors; (g) functional performance. Maximum allowable variations : All functions shall operate as designed. All errors shall be within the maximum permissible errors specified In Tobie II, (2) Damp heat, steadystate[4(3)(ili)} Damp heat, steady state tests are carried out according to Table VI. SUPPLEMENTARY INFORMATION TO THE TEST PROCEDURE Object of the test : To verify compliance with the provisions in paragraph 4(1)(1) of part II under conditions of high humidity and constant temperature. Precondition : None required. Condition of the EUT: Normal power supplied and "on" for a time period Equal to or greater than the warm-up time specified by the manufacturer. Power is to be "on" for the duration of the test. Adjust the EUT as close to zero Indication as Practicable prior to the test (if an automatic zero­ tracking device Is connected, adjust it to a value near zero). The EUT shall not be readjusted at any time during the test. The handling of the EUT shall be such that no condensation of water occurs on the EUT. Stabilization : 3 hours at reference temperature and 50% humidity. 2 days (48 hours) at the upper limit temperature as specified In paragraph 2(9)(i) of part II. Temperature: Reference temperature of 20°C and at the upper limit as specified in paragraph 2(9)(1) of part II. Relative humidity : 50% at reference temperature; BS% at upper limit temperature. Temperature-humidity sequence : Reference temperature of 20°<:: at 50% humidity; TABLE VI TABLE VII Environmental phenomena Voltagevariation Test specification . Reference voltage Reference voltage+l0% Reference voltage- 15% Reference voltage SUPPLEMNTARY INFORMATION TO THE TEST PROCEDURES Object of the test : To verify compliance with the provisions In paragraph 2(9)(11) of part II under cond itlons of voltage variations. Test procedures in brief: Precondition : None required. The upper limit temperature at 85% humidity; Reference temperature of 20°C at 50% humidity. Number of tes\cydes : At least one cycle. Weighing test and test sequence : After stabilization of the EUT at reference temperature and 50% humidity,apply at least five different test loads or simulated loads and record: (a) date and time; (b) temperature; (c} relative humidity; (d) test load; (e) Indications (as applicable); (f) errors; (g) functional performance Increase the temperature in the chamber to the upper limit and increase the relative humidity to 85%. Maintain the EUT at no load for a period of 2 days (48 hours), Following the 2 days, apply at least five test loads and record the data as Indicated above. Allow full recovery of the EUT before any other tests are performed. ''Maximum allowable variations : All errors shall be within the maximum permissible errors specified In Table II. (3) Mains power supply voltage variation {AC)[2(9)(/i)} Power voltage variation tests are carried out according to Table VII. Test specif'lcatton Upperlimlt temperature and relative humidity of 85% for 2 days (48 hours) Environmental Phenomena Damp heat, Steady state ----------- ____ l. _ 446 THE GAZETTE OF fNDIA:EXTRAORDfNARY [PARTII-----SEC.3(i)] Test equipment : Condition of the EUT : Normal power supplied and "on" for a time· period ·equal to or, greater than the warm-up time speclfled·by the manufacturer .. Power Is to be "on" for the duration of the test. Adjust the EUT as·closeto.zero Indication as practicable prior to.the test. :If. it has an .autornatlc zero-setting function then the instrument should be set to zero after applying eachlevel of voltage. Number of test.cydes : At:least one cycle. Weighing test.: The EUT shall be.tested at no load and with one test load.or simulated load between 50% and maximum:capaclty cifthe'ELIT. Test sequence·: Stablllze the power supply at the reference voltage within the defined llmlts and record: (a) date and time; (b) temperature; (c) relative humidity; (d) power supply voltage; . (e) test loads; (f) indications (as applicable); (g) errors; (h) functional performance. Repeat the test weighing for each of the voltages and record the indications. Maximum allowable variations : All functions shall operate as designed. All errors shall be within the maximum permissible errors specified In Tobie II. 8(4) Battery power supply voltage variation (DC) [2(9){iii) and 4(3)(viii)J Test method : Variation in DC power supply. Where the ElJT continues to operate below the stated battery voltage, the followingtest shall be conducted using an equivalent variable DC power source. Object of the test : To verify compliance with the provisions in paragraph 2(9)(iii) and 4(3)(vlll) of part II under conditions of varying DC power supply. The requirements shall be met either by use of an equivalent variable DC power source or by allowing the battery voltage to fall by use. Reference to standard : No reference to international standards can be given at the present time. Test procedures In brief : The test consists of subjecting the EUT to DC power.variations when the former Is operating under normal atmospheric conditions with one test load or simulated load between 50% and maximum capacity of the ElJT. Test severity : Supplyvoltage: lower limit,the voltage at which the EUT clearly ceases to function (or Is automatically put out of service) + 2% of this voltage. Number of test cydes : At least one cycle. Conduct of the test : Precondition : None required. Variable DC power source; calibrated voltmeter; Load cell simulator, if applicable. Condition of the EUT : Normal power supplied and "on" for a time period equal to or grater than the warm­ up time specified by the manufacturer. Adjust the EUT as close to zero indication as practicable prior to the test. If it has an automatic zero-setting function as part of the automatic weighing process, then the instrument should be set to zero after applying each level of voltage. Test sequence : Stabilize the power supply at nominal battery voltage ±2% and record the following data at no load and with one load or simulated load between 50% and maximum capacity of the EUT: (a) date and time; (b) temperature; (c) relative humidity; (d) power supply voltage; (e) test load; (f) Indications (as applicable); (g) errors; (h) functional performance. Reduce the power supply to the EUT until the equipment clearly ceases to function and note the voltage. Switch the EUT "off" and increase the power supplyvoltage to nominal battery voltage ±2%. Switch the EUT"on" and reduce the power supply voltage to the above noted voltage (out of service voltage) + 2% of the noted voltage. Record the data indicated above. Maximum allowable variations : All functions shall operate as designed. All errors shall be within the maximum permissible errors specified in Table II. 447 TABLE VIII (*) sf : value of the significant fault [see 4{2)(v) of part I.] 9. Disturbance tests 4(1)(/i) and 4(3)(/v) SUMMARY OF TESTS (1) Voltage dips and short Interruptions Short time power reduction (voltage dips and short interruptions) tests are carried out according to Table VJII. Adjust the EUT as close to zero indication as practicable prior to the test. The EUT shall not be.re­ adjusted at anytime during the test except to reset If a significant fault has been indicated. Number oftest cycles: At least one cycle. Weighing test and test sequence : Stabilize all factors at nominal reference conditions. Applyone load or simulated load between 50% and maximum capacity of the ELJTand record: (a) date and time; (b) temperature; (c) relative humidity; (d) power supply voltage; (e) test load; (f) indications (As applicable); (g) errors; . (h) functional performance. Interrupt the power supply to zero voltage for a period equal to one half cycle and conduct the test. During Interruption observe the effect on the EUT and record as appropriate. Reduce the power supply to 50% of nominal voltage for a periodequal to two half cycles and conduct the test. During reductions observe the effect on the EUT and record, as appropriate. · Maximum allowable variations : The difference between the weight indication due to the disturbance and the Indication without the d;sturbance either sha II not exceed the values given in paragraph 4(2)(v) of part I, or the EUT shall detect and act upon a significant fault. . (2) ElectriClJI fast transients/burst immunity Electrical fast transients/burst immunity tests are carried out for 2 minutes with a positive polarity and for 2 minutes with a negative polarity and according to Tables IX, X and XI. sf(*) CtJnditions applied sf sf sf Characteristic under test Disturbance Disturbance Disturbance Disturbance Environmental phenomena Test Specification Interruption from reference voltage · to zero voltage for one half cycle Interruption from reference voltage to 50% of Voltage dips and short reference voltage for two ha If cycles These mains voltage interrup­ tions shall be repeated ten times with a time interval of at least 10 seconds Test 1. Voltage dips and short Interruptions 2. Electrical fast transients/burst immunity 3. Electrostatic discharge 4. Electromagnetic susceptibility Note : Applicable only to ports or interfacing with cables whose total length may exceed 3m accordingto the manufacture's functional specification. TABLE IX PORTS FOR SIGNAL LINES AND CONTROL UNES SUPPLEMENTARY INFORMATION TO lrlE TEST PROCEDURES Object of the test : To verify compliance with the provisions in paragraph 4(1)(ii) of part II under conditions of short time mains voltage Interruptions and reductions. Test proceduru In brief: Precondition : None required. Condition of the EUT : Normal power supplied and "on" for a time period equal to or greater than the warm-up time specified by the manufacturer. Environmental phenomena Fast transient common mode Test specification 0.5 kV (peak) 5/50 ns T1,'Th 5 H rep. frequency 448 THE GAZETTE OF INDIA: EXTRAORDINARY [PART11-SEc. 3(i)] Maximum allowable variations: The difference between the weight Indication due to the disturbance and the indication without the disturbance either shall not exceed the value given in paragraph 4(2)(v) of part I or the Instrument shall detect and act upon a significant fault. (J) Electrostatic discharge Electrostatic discharge tests are carried out with test signals and conditions as given in Table XII. TABLE XII Note. The 6 kV contact discharge shall be applied to accessible conductive parts. Metallic contacts e.g. In battery compartments or in socket outlets are excluded from this requirement. Contact discharge is the preferred test method. 20 discharges (10 with positive and 10 with negative polarity) shalr be applied on each accessible metal part of the enclosure. The time interval between successive discharges shall be at least 10 seconds. In the case of a non-conductive enclosure, discharges shall be applied on the horizontal or vertical coupling planes. Air discharges shall be used where contact discharges cannot be applied. Tests with other (lower) voltages than those given in Table X are not required. SUPPLEMENTARY INFORMATION TO THE TEST PROCEDURES Object of the test: To verify compliance with the provlslons in paragraph 4(1)(ii) of part II under conditions where electrostatic discharges are applied. Test procedures in brief: Precondition: None required. Condition of the EUT: Normal power supplied and "on" for a time period equal to or greater than the warm-up time specified by the manufacturer. Adjust the EUT as close to zero indication as practicable prior to the test. The EUT shall not be re­ adjusted at any time during the test except to reset if a significant fault has been indicated. Stabilization: Before any test stabilize the EUT under constant environmental conditions. Weighing test; Stabilize all factors at nominal reference conditions. Apply one load or simulated load between 50% and maximum capacity of the EUT and record the following with and without electrostatic discharge: (a) date and time; (b) temperature; (c) relative humidity; :TABLE ·X INPUT AND OUPUT DC POWER PORTS environmental phenomena Test specification Fast transient common mode 0.5 k V (peak) 5/SO·ns T1/Th 5 kHz rep. frequency Note: Not applicable to battery operated appliances that cannot be connected to the mains while in use. TABLE XI INPUT AND OUTPUT AC POWER PORTS Environmentalphenomena Test specification Fast transient common mode 0.5 kV (peak) 5/50 ns T 1/Th 5 kHzrep. frequency f(ote'. Not applicable to battery operated appliances that cannot be connected to the mains while In use. A coupling/decoupling network shall be applied for testing AC power ports. SUPPLEMENTARY INFORMATION TO THE TEST PROCEDURES Object of the test: To verify compliance with the provisions In paragraph 4(1)(ii) of part II under conditions where fast transients are super-Imposed on the mains voltage, Interruptions and reductions. Ten procedures In brief: Precondition: None required. Condition of the EUT: Normal power supplied and "on" for a time period equal to or greater than the warm-up time specified by the manufacturer. Adjust the EUT as close to zero Indication as practicableprior to the test. The EUT shall not be re­ adjusted at any time during the test except to reset If ii significant fault has been indicated. Stabilization: Before any test stabilize the EUT under constant environmental conditions. Weighing test: Stabilize ail factors at nominal reference conditions. Apply one load or simulated load between 50% and maximum capacity of the EUT and record the following with and without the transients: (a) date and time; (b) temperature; (c) relative humidity; {d) test load; (e) indications (as applicable); (f) errors; (g) functional performance. Environmental phenomena Electrostatic discharge Test specification 8 kV air discharge 6 kV contact discharge ['WT II = 3(i)] 449 SUMMARY OF TEST specification. In the conduct of this test, the operating Instructions for the Instrument as supplied by the manufacturer shall be considered. ·· The EUT shall be stabilized at sufficiently constant ambient conditions after switch-on for at 4 7 l/ c; rl/ fl - I! 3 (d) test load; (e) indications (as applicable); (f) errors; (g) functional performance. Maximum allowable variations: The difference between the weight indication due to the disturbance and the indication without the disturbance either shall not exceed the value given in paragraph 4(2)(v) of part I or the instrument shall detect and act upon a significant fault. tD. Span stablllty test (4)(iii) least five hours, and at least 16 hours after the temperature and damp heat tests have been performed. · Test severities: Test duration: 28 days or the time period necessary to conduct the perfonmance tests, whichever Is less. Time (t) between tests (days): 0.5 s t ? 10. Test load: near maximum capacity (Max): the same test weights shall be used throughout the test. Maximum allowable variations: The variation In the errors of indication shall not exceed half the absolute value of the maximum permissible error In Table II for the test load applied on any of the n measurements. Number of test(n): At least 8 except where the differences of the results indicate a trend more than half the allowable variation specified, the measurements shall be continued until the trend comes to rest or reverses Itself, or until the error exceeds the maximum allowable variation. Precondition: None required. Condition of the EUT: Normal power supplied and "on" for a time period equal to or greater than the warm-up time specified by the manufacturer. Adjust the EUT as close to zero indication as practicable before each test. The automatic zero­ tracking should be made inoperative during the test (if the EUT Is so equipped). Test sequence: Stabilize all factors at nominal reference conditions. Adjust the Elff as close to zero as possible. Automatic zero-tracking shall be made inoperative and the automatic built-in span ;adjustment device shall be made operative. Initial measurement Determine the span error using the following method: 1. Determine the initial zero error (E0) If necessary disable any automatic zero-setting or zero-tracking devices, by placing a "zero weight" of for example 10 times the scale interval on the load receptor. Note the indication atzero (Io). Either by use of an indicator with a suitable . higher resolution scale interval or using the change , point weight method in paragraph 4(2)(ii) of . Annexure A (noting the total addition change point weight • Lo) determine and record the initial zero error (E,). Condition applied 1/2 absolute mpe(*) Test Characteristic under test (*) mpe: maximum penmissibleerror on initial verification in Table JI. Note: The maximum penmissible error for the zero point shall also be taken into consideration. Test method: Span stability. Object of the test; To verify compliance with the provisions In paragraph 4(4)(iii) after the EUT has been subjected to the performance tests. Reference to standard: No reference to International standards can be given at the present time. Test procedures In brief: The test consists of observing the variations of the error of the EUT or simulator under sufficiently constant ambient conditions (reasonable constant conditions in a nonmal laboratory environment) at various Intervals: before, during, and after the EUT has been subjected to performance tests. The performance tests shall include the temperature test and, if applicable, the damp heat test; an endurance test shall not be included. Other performance tests listed in this Annex may be performed. The Elff shall be disconnected twice from the mains power supply (or battery supply where fitted) for at least 8 hours during the period of the test. The number of disconnections may be increased if so specified by the manufacturer or at the discretion of the approval authority in the absence of any Span stability Stablllty 450 THE GAZET,J:E OF INDIA: EXTRAORDINARY [PARTII-SEC. 3(i)] 2. Determlne'the·error,at•near'Max capacity (EL) Carefully remove the change point weights (if used) and apply the test load (or simulated load) and note the Indication (IL)· Either by use of an indicator with a suitable higher resolution scale interval.or using the change point weight method In paragraph 4(2)(11) of Annexure A (noting the .total addition change point weight • Lo) determine and record the error at near Max capacity (EL), Record; (a) date and time; (b) temperature; (c) barometric pressure; (d) relative humidity; (e) value of 0.1 d; (f) testload; (g) total of added change point weights at zero load <!Loi (h) total of added change point weights at test load oL; (i) the followingIndications: o Indication at zero (Io); o the indication of test load (ILL)i U) calculate; o Initial zero error E0; o error at test load (EL)· (k) change in location and apply all necessary corrections resulting from variations of temperature, pressure, etc. between the various measurements. Immediately repeat steps 1 and 2 four more times and determine and record the average value of the error for the five tests. Subsequent measurements After observing the time between measurements requirement repeat the test sequence 1 to 2 once recording the data above unless: o either the result is outside the maximum allowable variation, or e the range of the five readings of the initial measurement is more than 0.1. d, In which case continue four more times repeating steps 1 and 2 recording the data above and determine and record the average value of the error of the five tests. The measurements shall continue until there are at least 8 measurements except where the differences of the results indicate a trend more than half the allowable verletion specified, the measurements shall be continued until the trend comes to rest or reverses itself, or until the error exceeds the maximum allowable variation. 1.1.. In-situ tests (1) General: Note the accuracy class required for wagon weighing and train weighing. Ensure that the desired scale interval and the maximum wagon weight comply with Table III Check that the minimum capacity complies with paragraph 2(4) of part II. Check that the minimum wagon weight complies with paragraph 2(5) of part II. For pattern evaluation, testing may be carried out for coupled weighing, uncoupled weighing or train weighing depending on the approval required by the applicant. For Initial verification, tests shall be carried out corresponding to the normal site operation of the instrument. (2) Rail alignment: · If exemption from the alignment calibration at Annexure S Is required, alignment of ±1 mm must be demonstrated to the satisfaction of the metrologlcal authority by means of: • survey by the authority; • submission of independent survey results; o other mutually acceptable procedure. (3),Control Instrument: Establish whether or not the instrument is to be used as the control instrument. Jf it is to be used as the control instrument it shall comply with paragraph 2(8)(iii)(b) of part IJ. If not the static weighing [2(8)(i)] shall be omitted. If wagons have to be moved over some distance from the control instrument to the EUT,the conditions must be closely controlled. Differences in weather conditions will cause errors which will not be determinable and so this should be avoided where poss Ible. ['WT II-- 3(i)] 451 (4) Static weighing[2(8)(i)}: Apply test loads from zero up to and including Max, then remove the test loads back to zero. Ensure that the error Is recorded at change in load (normally In steps of one tonne). Ensure that errors are recorded in line with paragraph 4(2) of Annexure A. Record the errors and compare them to the limits in Table II. ·· (5) Eccentridty tests [2(B)(i)(e-III)]: Apply the eccentricity test only where it is practical and safe to do so. On installations where the weighed length Is very short it may not be possible to apply this test fully.If so note the reduced load. (6) Coupled weighing [5(1){iii)(a-Il)} (i)Static weighing (empty}: Select the required number of reference wagons as in Table IV and weigh them individually,statically and uncoupled on. the control instrument. Anyerrors shall not exceed the appropriate mpe as given in Table I. (ii)Coupled wagon weighing (empty): Record the wagon weights of the coupled reference wagons, In-motion (6(2)] as they are displayed or printed by the instrument under test at a variety of speeds up to the maximum. Record the errors. The instrument under test must behave correctly for the condition of operating speed [3(4)(iv)]. The printer shall not print the weight of any wagon that has travelled over the load receptor at a speed outside the range of operating speeds. An appropriate indication shall be included on the printout for any wagon weight not printed. A sub· total may be printed exclusive of wagons which have not been weighed provided that it is Indicated it is not the total train weight. The instrument shall also operate In the following manner In the case of rollback [3(4)(v)]. The roll-back condition shall be detected automatically and the weighing process shall be aborted, i.e. the instrument shall either: (a) not Indicate or print any weighing result, or (b} give a clear Indication that any Indicated or printed weighing result may. not be correct. Repeat the In-motion tests to obtain not less than 60 wagon weights. (iii)Static weighing (full): Select the required number of reference wagons · (see Table IV) and weighs them individually,statically and uncoupled on the control instrument. - -· ---------------- Take note of the need to only particularly fill some of the wagons. Any errors shall not exceed the appropriate mpe as given in Table I, (lv)Coupled wagon weighing (full): Record the wagon weights of the coupled reference wagons, in-motion as they are displayed or printed by the instrument under test at a variety of speeds up to the maximum. Record the errors. The Instrument under test must behave correctly for the condition of.operating speed 3(4)(iv). The printer shall not print the weight of any wagon · that has travelled over the load receptor at a speed outside the range of operating speeds. An appropriate indication shall be included on the printout for any wagon weight not printed. A sub-total may be printed exclusive of wagons which have not been weighed provided that it is indicated it is not the total train weight. The Instrument shall also operate In the followingmanner In the case of rollback 3(4)(v). The roll-back condition shall be detected automatically and the weighing process shall be aborted I.e. the instrument shall either: (a} not indicate or print any weighing result, or (b) give a clear indication that any indicated or printed weighing result may not be correct. (7) Uncoupled weighing [S(l)(iil)(a-1)]. (I) Static weighing: Select not less than 5 wagons and fill them with a range of loads from zero load to that of a fully loaded wagon and weigh them statically on the control instrument. · (II) Uncoupled wagon weighing; Record the wagon weights of the uncoupled reference wagons, in-motion, as they are displayed or printed by the Instrument under test at variety of (controlled) speeds including near Max, near Min and the typical site operating speed. Record the errors. The instrument under test must behave correctly for the condition of operating speed 3( 4)(iv). The printer shall not print the weight of any wagon that has travelled over the load receptor at a speed outside the range of operating speeds. An appropriate indication shall be included on the printout for any wagon weight not _ J -·-- - - - 452 THE GAZETTE OF INDIA: EXTRAORDINARY [PARTII-SEC. 3(i)] . (1) While motionless, each axle shall be weighed in the centre and at each end of the load receptor. (2) The six weight indications noted in para 4(1) of Annexure B shall be summed up and the result divided by three. (3) Standard weights, as specified in para 4(4) of Annexure B shall be evenly distributed over the empty wagon and the weighing procedure in para 4(1). and (2) of Annexure B shall be repeated. (4} The sum of the standard weights referred to In para 4(3) of annexure B shall be at least equal to the larger of the following values:- • the difference between the maximum capacity and 1.5 times the weight of the wagon as determined in para 4(2) of Annexure B with the result rounded down to the nearest 1 tonne; • 10 tonne. (5) The difference between the results of the computations in accordance with para 4(2) and (3} of Annexure B shall be subtracted from the total value of the standard weights, the result being the calibration correction. 5. Example of alignment calibration test sheet Accuracy class: 1 Maximum capacity: a = 35t ' 'fyplcal wagon tare weight: b = 11.St Mass of standard weights required: c = 17t (a - l.Sb, rounded down) Scale Interval: 0.1 t Scale Interval for stationary load: 0.01 t TABLE XII printed. •A sub-total may be printed exclusive of wagons which have not been welghed·provided.that:it:ls;lndlcated .it Is not the total train weight. A mlnlmum.of.5,weight printouts of each wagon shall be·obtained. (8) Train weighing Tests.are the sameasin paragraph 11(6) of thlsAnnexure and may be carried out In conjunction with,paragraph 11(6) of this Annexure ·.without .repeating the test, if both modes ot.operation are required. The weights of the reference wagons shall be summed and any errors shall not exceed the appropriate mpe as given in Table I and applied to the summation.

Previous and next

All provisions of Legal Metrology (General) Rule, 2011

Questions people ask

Is annexure A of Legal Metrology (General) Rule, 2011 still in force?

No repeal of Legal Metrology (General) Rule, 2011 is recorded.