Annexure A, Legal Metrology (General) Rule, 2011: ANNEXUREA
Legal Metrology (General) Rule, 2011 - Annexure A
Text of annexure A
ANNEXUREA
ANNEXUREA
TEST PROCEDURES FORAVTOMATIC RAIL·
WEIGH BRIDGES
(Mandatory)
Meaning of symbols:
I ; Indication
L = Load
D.L = Additional load to next changeover point
P = I + 0.5d • D-L; Indication prior to rounding
Pn = nth indication prior to rounding
compliance with the requirements ,in
peraqraphs 3(3) and 3(4) of this part.
(iv) Provisions of means for. testing:
For the purpose oftesting,lthe.applicantmay
be required to.furnlsh the ,Metrological
Enforcement Authority with .the test vehicles,
material,.qualified personnel and a control
instrument. The Instrument under test may
be used as a controlinstrument provided it
complies with the.requirements in paragraph
2(8)(111)(b)of this part.
(v) Place of.testing:
Instruments submitted for pattern approval
may be.tested at the following places:
• a:slte at which all necessary tests can
be conducted and agreed upon
between the Metrological Authority
and the applicant,
• a 'laboratory considered appropriate
by the Metrological Authority,
• any other suitable place mutually
agreed upon by the Metrological
Authority and the applicant.
(2) Verification
(i) Tests :
Instruments shall comply with the
requirements In paragraph 2 [except 2(9)]
and paragraph 3 of this part for any
product(s) for which they are Intended and
when operated under normal conditions of
use
Tests shall be carried out in-situ, In a normal
installation. Toe instruments shall be Installed
so that an automatic weighing operation will
be virtuallythe same for testing as It Is for a
transaction.
(ii) In-motion tests :
In-motion tests shall be conducted in
accordance with paragraph 5(1)(11i)(a)of this
part with the exception that the types of
vehicles and, for test with coupled wagons
the number of vehicles in the test train shall
be in accordance with the normal operation
of the Instruments and used in accordance
with paragraph6 of this part.
(Iii) Provisions of means of testing:
For the purposes of testing, the applicant
may be required to furnish the enforcement
authority with the test vehicles, material,
Total number of
wagons In test train(n)
n < 10
10 < n < 30
30 < J\
TABLEN
Minimum number of
reference wagons
n
10
15
441
d = Scale interval
E = P - L = error
Eo = Error calculated at zero
Ee = Corrected error
rnpe = Maximum permissible error
EUT = Equipment under test
Max= Maximum capacity
Min = Minimum capacity
1.. Documentation [5(1.)(i)J
Review the documentation that is submitted,
includingnecessary photographs, drawings, diagrams,
general software information, relevant technical and
functional description of main components, devices
etc. to determine if It Is adequate and correct. Consider
the operational manual.
2. Comparing construction with documentation
5(1.)(i)
Examine the various devices of the Instrument
to ensure compliance with the documentation.
Examine the simulator to ensure that it is
adequate for train movement simulation. It shall
provide the signals from track switches, or other
vehicle type identification devices, normally
transmitted when a vehicle passes over the weighing
system. It is not expected to simulate effects such as
dynamic loading.
3. Initial examination
(1) Metrologicalcharacterlstics:
Note metrological characteristics in the test report.
(2) Descriptive markings [3(6)]:
Check the descriptive markings.
(3) Verification marks [3(1)]:
Check the arrangements for verification marks.
4. General
( 1) General requirements for electronic
instruments under test (EUT) :
Energize the EUT for a time period equal to or
greater than the warm-up time specified by the
manufacturer and· maintain it energized for the
duration of the test.
Adjust the EUT as closely as practicable to zero
prior to each test, and do not readjust it at anytime
during the test, except to reset it if a significant fault
has been indicated.
The deviation of the no-load indication due to
any test condition shall be recorded, and any load
indication shall be corrected accordingly to obtain the
weighing results.
The handling of the instrument shall be such that
no condensation of water occurs on the instrument.
(i) Indication .with a scale Interval smaller
thand
If an instrument with digItal indication has a
device for displaying the Indication with a
smaller scale Interval, this device may be
used to calculate the error. If a device Is
used It should be noted In the test report.
(2) calculatton of error:
For instruments with digitalIndication with a test
scale Interval [see note in S(l)(IIIXb)Jless or equal to
0.1 d, the device may be used directly to read the
error.
For Instruments with digital Indication and with
a test scale Interval greater than 0.1.d, changeover
points are used to determine the Indication of the
instrument, prior to rounding, as follows:-
At a certain load, L, the Indicated value, I, is
noted. Additional weights of for example O. ld are
successivelyadded until the indication of the instrument
is increased unambiguouslyby one scale interval (I+d).
The additional load -Ladded to the. load receptor gives
theindlcatlon, P,prKlrto rounding by using the following
formula:-
P=I+O.Sd-t,L
The error prior to rounding is:
E=P-L
Thus E = (I + 0.Sd - L',L)- L
Example : An instrument with a scale interval,
d, of 10 kg Is loaded with 1000 kg and thereby indicates
1000 kg. After adding successive weights of l kg. the
indication changes from 1000 kg to 1010 kg at an
additional load of 3 kg. Inserted in the above formula
these observations give:
P = (1000 + 5 - 3) kg = 1002 kg
Thus the true Indication prior to rounding is 1002 kg,
and the error Is:
E = (1002 - 1000) kg = 2 kg
5, Testprogram
(1) Pilttern evaluation [5(1)]:
All tests of paragraphs 6 to 10 of Annexure A
shall normally be applied for pattern evaluation.
The tests for paragraphs 6 to 10 of Annexure A
shall be performed with static load, a wheel movement
442 THE GAZETTE OF ,INDIA: EXTRAORDINARY [PART11-SEC. 3(i)]
simulator (switches) may be used if necessarvfor the
calculation of the weighing results.
(2) Initial verifieatlon_[5(2)}:
Initial verification tests normally.only:fndudesthe
tests in paragraph 11 of Annexure A.
The test shall· Include all dynamic 'ln-rnotlon
effects corresponding to normal operation of the
instrument
6. Perfonnancetuts.during pattem evaluation
The equipment under test should have the
following items associated with it, for the purpose of
testing:-
• Weightslmulator.
• Train wheel simulator.
• For practlcal reasons, the weight simulator
may take·varlous forms. For example, it may
be a weigh pan or platform scale of
approximately 1/ 1000th of the weight range
of a site Installation.
Whichever method is adopted, It must be
independently calibrated and readable to at least O. ld.
{1}General conditions
(i) Temperature :
The tests shall be performed at a steady
ambient temperature, usually normal room
temperature unless otherwise specified.
The temperature is deemed to be steady
when the difference between the extreme
temperatures noted during the test does not
exceed one-fifth of the temperature range
of the given Instrument without being
greater than 5°C and the rate of change
does not exceed S°C per hour.
(Ii) Power supply;
Instruments using electric power shall
normally be connected to the power supply
and "on" throughout the tests.
(iii) Automatic zero-setting:
During the tests, the automatic zero-setting
device may be switched off. When this is
done it should be mentioned In the test
report.
'
For certain tests the test descriptionspecifies
whether the automatic zero-setting shall be
operative or inoperative.
(iv) Recovery:
After each test the instrument should be
allowed to recover sufficiently before the
following test.
(2) Checkingof zero
(I) Range of zero-setting {J(J)(iv)]
(a) Semi-automatic zero-setting :
This test shall not be carried out
during the span stability test.
With the load receptor empty, set the
instrument to zero. Place a test load
on the load receptor and use the zero
setting device. Continue incrementing
the test load until the zero-setting
device fails to re-zero when activated.
The maximum load that can be
rezeroed is the positive portion of the
zero-setting range.
To test the negative portion of the
zero-setting range, recalibrate the
instrument with an additional weight
on the load receptor at the zero and
span points. This additional weight
should be greater than the negative
zero-setting range. Then remove
weights and after each weight Is
, removed, use the zero-setting device.
The maximum load that can be
removed while the instrument can still
be re-zeroed by the zero-setting
device, is the negative portion of the
zero-setting range.
Recalibrate the instrument without
this additional weight.
(b) Automatic zero-setting :
This test shall not be carried out
during the span stability test.
With the load receptor empty, add
weights in small quantities and after
each weight is added allow time for
the automatic zero-setting device to
function to see if the instrument Is re
zeroed automatically. Repeat this.
procedure until the instrument will not
be re-zeroed automatically. The
maximum load that can be re-zeroed
is the positive portion of the automatic
zero-setting range.
["WI II ? 3(i)] 443
To test the negative portion of the
zero-setting range recalibrate the
Instrument with an additlona Iweight
on the load receptor at the zero and
span points. This additional weight
should be greater than the negative
zero-setting range. Then remove
weights and after each weight is
removed, allow time for the automatic
zero-setting device to function to see
If the Instrument is re-zeroed
automatically.The maximum load that
can be removed while the Instrument
can still be re-zeroed by the zero
setting device, is the negative portion
of the zero-setting range.
Re-zero the instrument without this
additional weight.
(3) Setting to zero before loading:
For instruments with digital Indication, the
adjustment to zero, or the determination of the zero
point is carried out as described in paragraph 6(5)(1)
of this Annexure.
(4) Association of indicating and printing device [2(7)J
If the instrument has more than one indicating
device, the indications of the various devices (both
indicating and printing) are compared during the test.
(5) Non-automatic tests of the control Instrument
Note : This sub-clause is only applicable to
instruments which are to be used as control
instruments.
(i) Accuracy of zero-setting [2(8)(/J(b)J
(a) Semi-automatic zero-setting :
The accuracy of the zero-setting device is
tested by setting the instrument to zero and
then detennlning the additional load at which
the indication changes from zero to one
scale interval above zero. The error at zero
is calculated according to the description in
paragraph 4(2) of this part.
(b) Automatic zero-setting :
The Indication is either brought outside the
zero range, or the automatic zero-setting Is
disabled. Then the additional load at which
the indication changes from one scale
interval to the next above Is determined and
the error is calculated according to the
description in paragraph 4(2) of this
Annexure. It is assumed that the error at
zero load would be equal to the error at the
load in question.
(II) Determination of weighing performance
(a) Preloading :
·
Before the first weighing test the
Instrument shall be preloaded once to
Max.
(b) Weighing test[2(8)(1)(e-1)]:
Applysubstitution material from zero up to
and Including Max, and then remove the
material back to zero. When determining the
initial intlinsic error, at least ten different load
values are selected, and for other weighing
tests at least five are selected. The values of
the loads selected shall include Max and Min,
and values at or near those at which the
maximum permsslble error(mpe) changes.
It should be noted that when loading or
unloading weights the load must be
respectively increased or decreased In a
uniform progression.
(111)Discrimination test [2(8)(i)(vi)J:
The following tests are performed with three
different loads, e.g. Min,0.5 Max and Max.
A load plus suffidentsubstitution material (e.g.
10 times 0.1 d) is placed on the load receptor.
The additional material is then successively
removed until the indication, I, is decreased
unambiguouslyby one actual scale interval, I·
d. Replacesubstitution material equivalent to
0.1 d and then a load equal to 1.4 d shall be
gently placed on the load receptor and the
result will be increased by one actual scale
Interval above the initial indication, I+ d,
Example : instrument with d = 10g
0
210g
I
.,.
I - d = l 90 g I = 200 g
I+ d
Add 1/lOd = Lg
add?i:
d = 14 g,.,... lJ,
0
190(
11
200 g
I
jj
--
J
--- -------------I
444 THE GAZETTE OF INDIA: EXTRAORDINARY [PARTII-SEc.3(i)]
TABLEV
Environmental Phenomena Test Specification
Reference of 20°C
(1) Static temperature tests [2(9)(i)}:
Static temperaturetests are carried out according
to Table V.
Supplementary information to the test procedure
Object of the test : To verify compliance with
the provisions In paragraph 2(9)(1) of part II under
?onditlonsof dry heat (non-condensing) and cold.
Spedfledhighfor 2 hours
Specified low for 2
hours
5°C
Reference of 20°c
Temperature
Test procedures in brief :
Precondition : 16 hours
·
Condition of the EUT : Normal power supplied and
"on" for a time period equal to or greater than the
warm-up time specified by the manufacturer. Power Is
to be "on" for the duration of the test.
Stablllzatlon : 2 hours at each temperature under
"free air" conditions.
Temperature : As specified in paragraph 2(9){1)of
part II.
Temperature sequence : Reference tempera·
ture of 20°C;
Specified high
temperature;
Specified low
temperature;
A temperature of
5°C;
Reference
temperature of 20°C.
Number of test cycles : At least one cycle.
Weighing test : Adjust the EUT as close to zero
indication as practicableprior to the test (ifan automatic
zero-tracking device is connected, adjust it to a value
near zero). The EUT shall not be readjusted at any
time during the test.
After stabilization at the reference temperature
and again at each specified temperature, apply at least
five different test loads or simulated loads and record:
(a) date and time;
(b) temperature;
(c) relative humidity;
(1) Static temperature Influence factor mpe(*)
(2) Damp heat, steady Influence factor mpe
state
(3) Mains power supply Influence factor mpe
voltage variation (AC)
(4) Battery power supply Influence factor mpe
Voltage variation (DC)
(*) mpe : maximum permissible error
The indicatlon.-at start is I-= 20Qg. Remove
additional weights until the indication.changes to I - d
= 190g. Add O.ld -;lg and thereafter 1.4 d .= 14g.
The Indications must then be
I+ d = 210g.
7. Additional functionality
(1) Warm-uptimetest[4(3)(v)J
(i) Disconnect.the instrument from the power
supply.for a;period•ofat least 8 hours prior
to the:test.
(ii) Reconnect the instrument and switch on while
observing the indicating device. Verify that
it is not possible to initiate automatic
weighing or printout until the indication has
stabilized or until completion of the warm
up time If this is specified by the
manufacturer.
(iii) As soon as the indication of the Indicating
device has stabilized, set the instrument to
zero and determine the error of zero-setting.
(iv) Apply a load close to Max. Determine the
error by the method in paragraph 4(2) of
this Annexure.
(v) Repeat stages (3) and (4) after S, 15 and
30 minutes.
(2) Agreement between indicating and printing devices
[2(7)}.
During the course of the tests verify that for the
same load, thedfference between any two Indicating
devices having the same scale interval is as follows:
Zero for digital indicating or printing devices;
Not greater than the maximum permissible error
for weighing-in-motion for analogue devices.
8. Influence factor tuts
SUMMARY OF TESTS
Test Characteristics Conditions
under test applied
t::t,t .i?.--- ...
__
[ "l1'T II -1glr,l" 3 ( i ) ] 445
(d) test load;
(e) indications (as applicable);
(f) errors;
(g) functional performance.
Maximum allowable variations : All functions shall
operate as designed. All errors shall be within the
maximum permissible errors specified In Tobie II,
(2) Damp heat, steadystate[4(3)(ili)}
Damp heat, steady state tests are carried out
according to Table VI.
SUPPLEMENTARY INFORMATION TO THE TEST
PROCEDURE
Object of the test : To verify compliance with the
provisions in paragraph 4(1)(1) of part II under
conditions of high humidity and constant temperature.
Precondition : None required.
Condition of the EUT: Normal power supplied and
"on" for a time period Equal to or greater than the
warm-up time specified by the manufacturer. Power is
to be "on" for the duration of the test.
Adjust the EUT as close to zero Indication as
Practicable prior to the test (if an automatic zero
tracking device Is connected, adjust it to a value near
zero). The EUT shall not be readjusted at any time
during the test.
The handling of the EUT shall be such that no
condensation of water occurs on the EUT.
Stabilization : 3 hours at reference temperature and
50% humidity.
2 days (48 hours) at the upper limit temperature
as specified In paragraph 2(9)(i) of part II.
Temperature: Reference temperature of 20°C
and at the upper limit as specified in paragraph 2(9)(1)
of part II.
Relative humidity : 50% at reference temperature;
BS% at upper limit temperature.
Temperature-humidity sequence : Reference
temperature of 20°<:: at 50% humidity;
TABLE VI
TABLE VII
Environmental phenomena
Voltagevariation
Test specification .
Reference voltage
Reference
voltage+l0%
Reference voltage-
15%
Reference voltage
SUPPLEMNTARY INFORMATION TO THE TEST
PROCEDURES
Object of the test : To verify compliance with the
provisions In paragraph 2(9)(11) of part II under
cond itlons of voltage variations.
Test procedures in brief:
Precondition : None required.
The upper limit temperature at 85%
humidity;
Reference temperature of 20°C at 50%
humidity.
Number of tes\cydes : At least one cycle.
Weighing test and test sequence : After
stabilization of the EUT at reference temperature and
50% humidity,apply at least five different test loads or
simulated loads and record:
(a) date and time;
(b) temperature;
(c} relative humidity;
(d) test load;
(e) Indications (as applicable);
(f) errors;
(g) functional performance
Increase the temperature in the chamber to the
upper limit and increase the relative humidity to 85%.
Maintain the EUT at no load for a period of 2 days (48
hours), Following the 2 days, apply at least five test
loads and record the data as Indicated above. Allow
full recovery of the EUT before any other tests are
performed.
''Maximum allowable variations : All errors shall be
within the maximum permissible errors specified In
Table II.
(3) Mains power supply voltage variation {AC)[2(9)(/i)}
Power voltage variation tests are carried out
according to Table VII.
Test specif'lcatton
Upperlimlt
temperature and
relative humidity of
85% for 2 days (48
hours)
Environmental Phenomena
Damp heat, Steady state
-----------
____
l.
_
446 THE GAZETTE OF fNDIA:EXTRAORDfNARY [PARTII-----SEC.3(i)]
Test equipment :
Condition of the EUT : Normal power supplied and
"on" for a time· period ·equal to or, greater than the
warm-up time speclfled·by the manufacturer
..
Power
Is to be "on" for the duration of the test.
Adjust the EUT as·closeto.zero Indication as
practicable prior to.the test. :If. it has an .autornatlc
zero-setting function then the instrument should be
set to zero after applying eachlevel of voltage.
Number of test.cydes : At:least one cycle.
Weighing test.: The EUT shall be.tested at no load
and with one test load.or simulated load between 50%
and maximum:capaclty cifthe'ELIT.
Test sequence·: Stablllze the power supply at the
reference voltage within the defined llmlts and record:
(a) date and time;
(b) temperature;
(c) relative humidity;
(d) power supply voltage;
.
(e) test loads;
(f) indications (as applicable);
(g) errors;
(h) functional performance.
Repeat the test weighing for each of the voltages
and record the indications.
Maximum allowable variations : All functions shall
operate as designed. All errors shall be within the
maximum permissible errors specified In Tobie II.
8(4) Battery power supply voltage variation (DC)
[2(9){iii) and 4(3)(viii)J
Test method : Variation in DC power supply. Where
the ElJT continues to operate below the stated battery
voltage, the followingtest shall be conducted using an
equivalent variable DC power source.
Object of the test : To verify compliance with the
provisions in paragraph 2(9)(iii) and 4(3)(vlll) of part
II under conditions of varying DC power supply. The
requirements shall be met either by use of an
equivalent variable DC power source or by allowing
the battery voltage to fall by use.
Reference to standard : No reference to
international standards can be given at the present
time.
Test procedures In brief : The test consists of
subjecting the EUT to DC power.variations when the
former Is operating under normal atmospheric
conditions with one test load or simulated load between
50% and maximum capacity of the ElJT.
Test severity : Supplyvoltage: lower limit,the voltage
at which the EUT clearly ceases to function (or Is
automatically put out of service) + 2% of this voltage.
Number of test cydes : At least one cycle.
Conduct of the test :
Precondition : None required.
Variable DC power source;
calibrated voltmeter;
Load cell simulator, if
applicable.
Condition of the EUT : Normal power supplied and
"on" for a time period equal to or grater than the warm
up time specified by the manufacturer.
Adjust the EUT as close to zero indication as
practicable prior to the test. If it has an automatic
zero-setting function as part of the automatic weighing
process, then the instrument should be set to zero
after applying each level of voltage.
Test sequence : Stabilize the power supply at
nominal battery voltage ±2% and record the
following data at no load and with one load or
simulated load between 50% and maximum capacity
of the EUT:
(a) date and time;
(b) temperature;
(c) relative humidity;
(d) power supply voltage;
(e) test load;
(f) Indications (as applicable);
(g) errors;
(h) functional performance.
Reduce the power supply to the EUT until the
equipment clearly ceases to function and note the
voltage. Switch the EUT "off" and increase the power
supplyvoltage to nominal battery voltage ±2%. Switch
the EUT"on" and reduce the power supply voltage to
the above noted voltage (out of service voltage) + 2%
of the noted voltage. Record the data indicated
above.
Maximum allowable variations : All functions shall
operate as designed. All errors shall be within the
maximum permissible errors specified in Table II.
447
TABLE VIII
(*) sf : value of the significant fault [see 4{2)(v) of
part I.]
9. Disturbance tests 4(1)(/i) and 4(3)(/v)
SUMMARY OF TESTS
(1) Voltage dips and short Interruptions
Short time power reduction (voltage dips and
short interruptions) tests are carried out according to
Table VJII.
Adjust the EUT as close to zero indication as
practicable prior to the test. The EUT shall not be.re
adjusted at anytime during the test except to reset If
a significant fault has been indicated.
Number oftest cycles: At least one cycle.
Weighing test and test sequence : Stabilize all
factors at nominal reference conditions. Applyone load
or simulated load between 50% and maximum capacity
of the ELJTand record:
(a) date and time;
(b) temperature;
(c) relative humidity;
(d) power supply voltage;
(e) test load;
(f) indications (As applicable);
(g) errors;
.
(h) functional performance.
Interrupt the power supply to zero voltage for a
period equal to one half cycle and conduct the test.
During Interruption observe the effect on the EUT and
record as appropriate.
Reduce the power supply to 50% of nominal
voltage for a periodequal to two half cycles and conduct
the test. During reductions observe the effect on the
EUT and record, as appropriate.
·
Maximum allowable variations : The difference
between the weight indication due to the disturbance
and the Indication without the d;sturbance either sha II
not exceed the values given in paragraph 4(2)(v) of
part I, or the EUT shall detect and act upon a significant
fault.
.
(2) ElectriClJI fast transients/burst immunity
Electrical fast transients/burst immunity tests are
carried out for 2 minutes with a positive polarity and
for 2 minutes with a negative polarity and according
to Tables IX, X and XI.
sf(*)
CtJnditions
applied
sf
sf
sf
Characteristic
under test
Disturbance
Disturbance
Disturbance
Disturbance
Environmental phenomena Test Specification
Interruption from
reference voltage
·
to zero voltage for
one half cycle
Interruption from
reference voltage
to 50% of
Voltage dips and short reference voltage
for two ha If cycles
These mains
voltage interrup
tions shall be
repeated ten times
with a time interval
of at least 10
seconds
Test
1. Voltage dips and
short Interruptions
2. Electrical fast
transients/burst
immunity
3. Electrostatic
discharge
4. Electromagnetic
susceptibility
Note : Applicable only to ports or interfacing
with cables whose total length may exceed 3m
accordingto the manufacture's functional specification.
TABLE IX
PORTS FOR SIGNAL LINES AND CONTROL
UNES
SUPPLEMENTARY INFORMATION TO lrlE TEST
PROCEDURES
Object of the test : To verify compliance with the
provisions in paragraph 4(1)(ii) of part II under
conditions of short time mains voltage Interruptions
and reductions.
Test proceduru In brief:
Precondition : None required.
Condition of the EUT : Normal power supplied and
"on" for a time period equal to or greater than the
warm-up time specified by the manufacturer.
Environmental phenomena
Fast transient common mode
Test specification
0.5 kV (peak)
5/50 ns T1,'Th
5 H rep. frequency
448 THE GAZETTE OF INDIA: EXTRAORDINARY [PART11-SEc. 3(i)]
Maximum allowable variations: The
difference between the weight Indication due to the
disturbance and the indication without the
disturbance either shall not exceed the value given
in paragraph 4(2)(v) of part I or the Instrument shall
detect and act upon a significant fault.
(J) Electrostatic discharge
Electrostatic discharge tests are carried out with
test signals and conditions as given in Table XII.
TABLE XII
Note. The 6 kV contact discharge shall be
applied to accessible conductive parts. Metallic
contacts e.g. In battery compartments or in socket
outlets are excluded from this requirement.
Contact discharge is the preferred test method.
20 discharges (10 with positive and 10 with negative
polarity) shalr be applied on each accessible metal
part of the enclosure. The time interval between
successive discharges shall be at least 10 seconds.
In the case of a non-conductive enclosure, discharges
shall be applied on the horizontal or vertical coupling
planes. Air discharges shall be used where contact
discharges cannot be applied. Tests with other (lower)
voltages than those given in Table X are not required.
SUPPLEMENTARY INFORMATION TO THE TEST
PROCEDURES
Object of the test: To verify compliance with
the provlslons in paragraph 4(1)(ii) of part II under
conditions where electrostatic discharges are applied.
Test procedures in brief:
Precondition: None required.
Condition of the EUT: Normal power supplied
and "on" for a time period equal to or greater than
the warm-up time specified by the manufacturer.
Adjust the EUT as close to zero indication as
practicable prior to the test. The EUT shall not be re
adjusted at any time during the test except to reset
if a significant fault has been indicated.
Stabilization: Before any test stabilize the EUT
under constant environmental conditions.
Weighing test; Stabilize all factors at nominal
reference conditions. Apply one load or simulated
load between 50% and maximum capacity of the
EUT and record the following with and without
electrostatic discharge:
(a) date and time;
(b) temperature;
(c) relative humidity;
:TABLE ·X
INPUT AND OUPUT DC POWER PORTS
environmental phenomena Test specification
Fast transient common mode 0.5 k V (peak)
5/SO·ns T1/Th
5 kHz rep. frequency
Note: Not applicable to battery operated
appliances that cannot be connected to the mains
while in use.
TABLE XI
INPUT AND OUTPUT AC POWER PORTS
Environmentalphenomena Test specification
Fast transient common mode 0.5 kV (peak)
5/50 ns T 1/Th
5 kHzrep. frequency
f(ote'. Not applicable to battery operated appliances
that cannot be connected to the mains while In use.
A coupling/decoupling network shall be applied
for testing AC power ports.
SUPPLEMENTARY INFORMATION TO THE TEST
PROCEDURES
Object of the test: To verify compliance with
the provisions In paragraph 4(1)(ii) of part II under
conditions where fast transients are super-Imposed
on the mains voltage, Interruptions and reductions.
Ten procedures In brief:
Precondition: None required.
Condition of the EUT: Normal power supplied
and "on" for a time period equal to or greater than
the warm-up time specified by the manufacturer.
Adjust the EUT as close to zero Indication as
practicableprior to the test. The EUT shall not be re
adjusted at any time during the test except to reset
If ii significant fault has been indicated.
Stabilization: Before any test stabilize the EUT
under constant environmental conditions.
Weighing test: Stabilize ail factors at nominal
reference conditions. Apply one load or simulated
load between 50% and maximum capacity of the
EUT and record the following with and without the
transients:
(a) date and time;
(b) temperature;
(c) relative humidity;
{d) test load;
(e) indications (as applicable);
(f) errors;
(g) functional performance.
Environmental phenomena
Electrostatic discharge
Test specification
8 kV air discharge
6 kV contact
discharge
['WT II = 3(i)] 449
SUMMARY OF TEST
specification.
In the conduct of this test, the operating
Instructions for the Instrument as supplied by the
manufacturer shall be considered.
··
The EUT shall be stabilized at sufficiently
constant ambient conditions after switch-on for at
4 7 l/ c; rl/ fl - I! 3
(d) test load;
(e) indications (as applicable);
(f) errors;
(g) functional performance.
Maximum allowable variations: The
difference between the weight indication due to the
disturbance and the indication without the
disturbance either shall not exceed the value given
in paragraph 4(2)(v) of part I or the instrument shall
detect and act upon a significant fault.
tD. Span stablllty test (4)(iii)
least five hours, and at least 16 hours after the
temperature and damp heat tests have been
performed. ·
Test severities:
Test duration:
28 days or the time period necessary to conduct
the perfonmance tests, whichever Is less.
Time (t) between tests (days): 0.5 s t ? 10.
Test load: near maximum capacity (Max): the
same test weights shall be used throughout the test.
Maximum allowable variations: The
variation In the errors of indication shall not exceed
half the absolute value of the maximum permissible
error In Table II for the test load applied on any of
the n measurements.
Number of test(n): At least 8 except where
the differences of the results indicate a trend more
than half the allowable variation specified, the
measurements shall be continued until the trend
comes to rest or reverses Itself, or until the error
exceeds the maximum allowable variation.
Precondition: None required.
Condition of the EUT: Normal power supplied
and "on" for a time period equal to or greater than
the warm-up time specified by the manufacturer.
Adjust the EUT as close to zero indication as
practicable before each test. The automatic zero
tracking should be made inoperative during the test
(if the EUT Is so equipped).
Test sequence: Stabilize all factors at nominal
reference conditions.
Adjust the Elff as close to zero as possible.
Automatic zero-tracking shall be made
inoperative and the automatic built-in span
;adjustment device shall be made operative.
Initial measurement
Determine the span error using the following
method:
1. Determine the initial zero error (E0)
If necessary disable any automatic zero-setting
or zero-tracking devices, by placing a "zero weight"
of for example 10 times the scale interval on the
load receptor. Note the indication atzero (Io).
Either by use of an indicator with a suitable
. higher resolution scale interval or using the change
, point weight method in paragraph 4(2)(ii) of
. Annexure A (noting the total addition change point
weight • Lo) determine and record the initial zero
error (E,).
Condition
applied
1/2 absolute
mpe(*)
Test Characteristic
under test
(*) mpe: maximum penmissibleerror on initial
verification in Table JI. Note: The maximum
penmissible error for the zero point shall also be taken
into consideration.
Test method: Span stability.
Object of the test; To verify compliance with
the provisions In paragraph 4(4)(iii) after the EUT
has been subjected to the performance tests.
Reference to standard: No reference to
International standards can be given at the present
time.
Test procedures In brief: The test consists
of observing the variations of the error of the EUT or
simulator under sufficiently constant ambient
conditions (reasonable constant conditions in a
nonmal laboratory environment) at various Intervals:
before, during, and after the EUT has been subjected
to performance tests.
The performance tests shall include the
temperature test and, if applicable, the damp heat
test; an endurance test shall not be included. Other
performance tests listed in this Annex may be
performed.
The Elff shall be disconnected twice from the
mains power supply (or battery supply where fitted)
for at least 8 hours during the period of the test. The
number of disconnections may be increased if so
specified by the manufacturer or at the discretion of
the approval authority in the absence of any
Span stability Stablllty
450 THE GAZET,J:E OF INDIA: EXTRAORDINARY [PARTII-SEC. 3(i)]
2. Determlne'the·error,at•near'Max capacity
(EL)
Carefully remove the change point weights (if
used) and apply the test load (or simulated load)
and note the Indication (IL)·
Either by use of an indicator with a suitable
higher resolution scale interval.or using the change
point weight method In paragraph 4(2)(11) of
Annexure A (noting the .total addition change point
weight • Lo) determine and record the error at near
Max capacity (EL),
Record;
(a) date and time;
(b) temperature;
(c) barometric pressure;
(d) relative humidity;
(e) value of 0.1 d;
(f) testload;
(g) total of added change point weights at
zero load <!Loi
(h) total of added change point weights at
test load oL;
(i) the followingIndications:
o Indication at zero (Io);
o the indication of test load (ILL)i
U) calculate;
o Initial zero error E0;
o error at test load (EL)·
(k) change in location
and apply all necessary corrections
resulting from variations of temperature,
pressure, etc. between the various
measurements.
Immediately repeat steps 1 and 2 four
more times and determine and record the
average value of the error for the five
tests.
Subsequent measurements
After observing the time between
measurements requirement repeat the
test sequence 1 to 2 once recording the
data above unless:
o either the result is outside the
maximum allowable variation, or
e the range of the five readings of the
initial measurement is more than 0.1.
d, In which case continue four more
times repeating steps 1 and 2
recording the data above and
determine and record the average
value of the error of the five tests.
The measurements shall continue until there
are at least 8 measurements except where the
differences of the results indicate a trend more than
half the allowable verletion specified, the
measurements shall be continued until the trend
comes to rest or reverses itself, or until the error
exceeds the maximum allowable variation.
1.1.. In-situ tests
(1) General:
Note the accuracy class required for wagon
weighing and train weighing.
Ensure that the desired scale interval and the
maximum wagon weight comply with Table III Check
that the minimum capacity complies with paragraph
2(4) of part II.
Check that the minimum wagon weight
complies with paragraph 2(5) of part II.
For pattern evaluation, testing may be carried
out for coupled weighing, uncoupled weighing or train
weighing depending on the approval required by the
applicant.
For Initial verification, tests shall be carried out
corresponding to the normal site operation of the
instrument.
(2) Rail alignment:
· If exemption from the alignment calibration at
Annexure S Is required, alignment of ±1 mm must
be demonstrated to the satisfaction of the
metrologlcal authority by means of:
• survey by the authority;
• submission of independent survey results;
o other mutually acceptable procedure.
(3),Control Instrument:
Establish whether or not the instrument is to
be used as the control instrument. Jf it is to be used
as the control instrument it shall comply with
paragraph 2(8)(iii)(b) of part IJ.
If not the static weighing [2(8)(i)] shall be
omitted.
If wagons have to be moved over some distance
from the control instrument to the EUT,the conditions
must be closely controlled. Differences in weather
conditions will cause errors which will not be
determinable and so this should be avoided where
poss Ible.
['WT II-- 3(i)] 451
(4) Static weighing[2(8)(i)}:
Apply test loads from zero up to and including
Max, then remove the test loads back to zero. Ensure
that the error Is recorded at change in load (normally
In steps of one tonne). Ensure that errors are
recorded in line with paragraph 4(2) of Annexure A.
Record the errors and compare them to the limits in
Table II.
··
(5) Eccentridty tests [2(B)(i)(e-III)]:
Apply the eccentricity test only where it is
practical and safe to do so. On installations where
the weighed length Is very short it may not be possible
to apply this test fully.If so note the reduced load.
(6) Coupled weighing [5(1){iii)(a-Il)}
(i)Static weighing (empty}:
Select the required number of reference wagons
as in Table IV and weigh them individually,statically
and uncoupled on. the control instrument.
Anyerrors shall not exceed the appropriate mpe
as given in Table I.
(ii)Coupled wagon weighing (empty):
Record the wagon weights of the coupled
reference wagons, In-motion (6(2)] as they are
displayed or printed by the instrument under test at
a variety of speeds up to the maximum. Record the
errors.
The instrument under test must behave
correctly for the condition of operating speed
[3(4)(iv)]. The printer shall not print the weight of
any wagon that has travelled over the load receptor
at a speed outside the range of operating speeds.
An appropriate indication shall be included on the
printout for any wagon weight not printed. A sub·
total may be printed exclusive of wagons which have
not been weighed provided that it is Indicated it is
not the total train weight.
The instrument shall also operate In the
following manner In the case of rollback [3(4)(v)].
The roll-back condition shall be detected
automatically and the weighing process shall be
aborted, i.e. the instrument shall either:
(a) not Indicate or print any weighing result,
or
(b} give a clear Indication that any Indicated
or printed weighing result may. not be
correct.
Repeat the In-motion tests to obtain not less
than 60 wagon weights.
(iii)Static weighing (full):
Select the required number of reference wagons
·
(see Table IV) and weighs them individually,statically
and uncoupled on the control instrument.
- -· ----------------
Take note of the need to only particularly fill
some of the wagons.
Any errors shall not exceed the appropriate mpe
as given in Table I,
(lv)Coupled wagon weighing (full):
Record the wagon weights of the coupled
reference wagons, in-motion as they are displayed
or printed by the instrument under test at a variety
of speeds up to the maximum. Record the errors.
The Instrument under test must behave
correctly for the condition of.operating speed 3(4)(iv).
The printer shall not print the weight of any wagon
· that has travelled over the load receptor at a speed
outside the range of operating speeds. An appropriate
indication shall be included on the printout for any
wagon weight not printed. A sub-total may be printed
exclusive of wagons which have not been weighed
provided that it is indicated it is not the total train
weight.
The Instrument shall also operate In the
followingmanner In the case of rollback 3(4)(v). The
roll-back condition shall be detected automatically
and the weighing process shall be aborted I.e. the
instrument shall either:
(a} not indicate or print any weighing result,
or
(b) give a clear indication that any indicated
or printed weighing result may not be
correct.
(7) Uncoupled weighing [S(l)(iil)(a-1)].
(I) Static weighing:
Select not less than 5 wagons and fill them
with a range of loads from zero load to
that of a fully loaded wagon and weigh
them statically on the control instrument.
·
(II) Uncoupled wagon weighing;
Record the wagon weights of the
uncoupled reference wagons, in-motion,
as they are displayed or printed by the
Instrument under test at variety of
(controlled) speeds including near Max,
near Min and the typical site operating
speed. Record the errors.
The instrument under test must behave
correctly for the condition of operating
speed 3( 4)(iv). The printer shall not print
the weight of any wagon that has travelled
over the load receptor at a speed outside
the range of operating speeds. An
appropriate indication shall be included
on the printout for any wagon weight not
_
J
-·-- - - -
452 THE GAZETTE OF INDIA: EXTRAORDINARY [PARTII-SEC. 3(i)]
.
(1) While motionless, each axle shall be
weighed in the centre and at each end of
the load receptor.
(2) The six weight indications noted in para
4(1) of Annexure B shall be summed up
and the result divided by three.
(3) Standard weights, as specified in para 4(4)
of Annexure B shall be evenly distributed
over the empty wagon and the weighing
procedure in para 4(1). and (2) of
Annexure B shall be repeated.
(4} The sum of the standard weights referred
to In para 4(3) of annexure B shall be at
least equal to the larger of the following
values:-
• the difference between the
maximum capacity and 1.5 times the
weight of the wagon as determined
in para 4(2) of Annexure B with the
result rounded down to the nearest
1 tonne;
• 10 tonne.
(5) The difference between the results of the
computations in accordance with para 4(2)
and (3} of Annexure B shall be subtracted
from the total value of the standard
weights, the result being the calibration
correction.
5. Example of alignment calibration test sheet
Accuracy class: 1
Maximum capacity: a = 35t
'
'fyplcal wagon tare weight: b = 11.St
Mass of standard weights required: c = 17t
(a - l.Sb, rounded down)
Scale Interval: 0.1 t
Scale Interval for stationary load: 0.01 t
TABLE XII
printed. •A sub-total may be printed
exclusive of wagons which have not been
welghed·provided.that:it:ls;lndlcated .it Is
not the total train weight.
A mlnlmum.of.5,weight printouts of each
wagon shall be·obtained.
(8) Train weighing
Tests.are the sameasin paragraph 11(6)
of thlsAnnexure and may be carried out
In conjunction with,paragraph 11(6) of this
Annexure ·.without .repeating the test, if
both modes ot.operation are required.
The weights of the reference wagons shall
be summed and any errors shall not
exceed the appropriate mpe as given in
Table I and applied to the summation.