Annexure, Legal Metrology (General) Rule, 2011: ANNEXURE

Legal Metrology (General) Rule, 2011 - Annexure

Text of annexure

ANNEXURE

ANNEXURE Tests for single pan balances. 1. Sensitivenes!r The balance shall be tested for sensitiveness near zero, middle and extreme position of the projection scale under three conditions 426 THE GAZETTE OF INDIA:EXTRAORDINARY [PARTII-SEC. 3(i)] of loadingnamely no load, half load and full load. Toe balance shall be.such so as to record the change in mass of the order of tsub-dlvlslon of the·projection scale accurately within the limits of the value equivalent to the least count of.the vernier or micrometre scale if provided, otherwise. within half a division of the projection scale. 2. Accuracy of projection scale-The-accuracy of the projection scale shall be examined at 10 points of the scale under.three different conditions of loading namely at no load, half load and full load. The maximum error at any point shall not exceed the value of the one half-sub-division-of the projection scale, if no vernier or micrometer scale is provided or two divisions of the vernier or micrometre scale. 3. Consistency of performance-Ten consecutive readings shall be noted by releasing and arresting the balance in the unloaded condition followed by another ten readings when the balance is in fully loaded condition. The standard deviation from the mean of the rest point shall be calculated separately for each condition, which in no case shall exceed half the division of the projection scale if no vernier or micrometer scale Is·provided or two divisions of the vernier or micrometer scale. PARTII·COUNTER MACHINES 1.. Oeflnltlon A counter machine is an equal arm weighing instrument of capacity not exceeding 50 kg, the pans of which are above the beam. Figure33 K illustrates a typical counter machine. FIGURE33 K 2. Capacities The machines may be of the followingmaximum capacities:- 500g, lkg, 2kg, 3kg, Skg, 10kg, 1Skg,20kg,25kg, 30kg and 50kg. 3. General Requirements (a) When the beam of body has two sides, they shall be connected further by not less than two cross-bars. The supports for the pans shall be of a suitable rigid structure such as cross members strengthened by straps. Central pieces or forks shall be fixed so that they are not twisted or dislocated. (b) Bearing surfaces, knife-edges and points of contact of all stays, hooks and loops shall be of hard steel or agate. The knife-edges and bearings shall be so fitted as to allow the - -- --------------------------? 427 4, Tests (a) The machine shall be tested on a horizontal level plane. (b) Sensitiveness and Error (i) The machine shall be tested for sensitiveness at full load with the beam In a horizontal position. The TABLE21J beam to move freely. The knee-edges shall rest upon the bearings along the whole length of their working part. (c) A counter machine may have a balance box for minor adjustments. In such a case, the balance box shall be permanently fixed beneath the weight pan and shall be large enough to contain loose material to an amount upto one per cent of the capacity of the machine. No other adjusting contrivance shall be used. (d) The pans may be made of any suitable material such as mild steel, stainless steel, brass or bronze, aluminium or its alloys, porcelain, enamel coated steel, glass or plastic material. They may be of any convenient shape. (e) The minimum fall of the extremities of the beam, either way, of counter machines shall be as follows:- Cepedty 500 g, 1 kg and 2 kg 3 kg, S"kg,10 kg, 15 kg 20 kg, 25 kg and 30 kg 50 kg Minimum fall 6mm 10mm 12mm 13mm addition of the weight specified in col. 2 or 4 as the _case may be of Table 21 J shall cause the pointer to have a displacement corresponding to the minimum limits specified In paragraph 3(e). (ii) The error that is the weight required to bring the beam of the instrument to horizontal position when fully loaded with weights each equal to Its capacity on both pans shall not exceed the limits specified In cols. 3 and s as the case may be of Table 21 JK. (c) Shift Test (I) When the goods pan is in the form of a scoop, the counter machine shall be correct to the prescribed limits of error if half the full load is placed against the middle of the back of the scoop and the other half at any position on the scoop. (II) When the goods pan is not in the form of a scoop, the counter machine shall indicate the same weight within half the prescribed limits of error, if the centre of a load equal to half the capacity placed on the goods pan Is moved any where within a distance from the centre equal to one-third of the length of the pan. If the pan has a vertical side against the middle of that side, the weight being entirely on the weight pan, but in any position on it. SENSITlVENESS AND MAXIMUM PERMISSIBLE ERRORS FOR COUNTER MACHINES Verification Inspection Capacity Sensitiveness Maximum Sensitiveness Maximum when fully permissible error, when fully permissible error loaded in excess or loaded in excess or deficiency, when deficiency, when fully loaded fullyloaded 1 2 3 4 5 500g 1.5g ' 2.2 g 4.Sg 4.5 g 1 kg 2.0g 3.0 g 6.0g 6.0g 2kg 3.0g 4.5 g 9.0g 9.0g 3 kg 4.0g 6.0g 12.0g 12.0g 5 kg 6.0g 9.0g 18.0g 18.0g 10kg 7.0 g 10.Sg 21.0g 21.0 g. 15kg 8.0g 12.0g 24.0g 24.0g 20 kg 9.0g 13.Sg 27.0g 27.0g 25 kg 10.0g 15.0g 30.0g 30.0g 30kg 11.0 g 20.0g 33.0g 40.0g 50kg 15.0 g 30.0g 45.0 g 60.0g 4 - 428 THE GAZETTE:Of INDIA: EXTRAORDINARY [PAJITfl-SEC. 3(i)] 5. Sealing Each madnne.shall'be provided with a,plug or stud of soft metal on a conspicuous part of the.beam or body to receive.the stamp orseal of.the verification authority. Such a· plug. or .stud -shall 'be made irremovable by undercutting itor by .sorne suitable method. SEVENTll'SCHEDULE HEADING-< AIJTOMAllC .RAIL-,WEIGHBRIDGES .'l!ARTI TERMINOLOGY (Terms and definitions) 1. Genera/deHnltlons (1) Weighing instrument Measuring instrument that serves to determine the mass of a load by using the action of gravity. (2) Automatic weighing instrument An instrument that weighs without the intervention of an operator and follows a predetermined program of automatic processes characteristic of the Instrument. (J) Rail·welghbn'dge: A weighing Instrument having a load receptor, inclusive of rails for conveying railway vehicles. (4) Electronic instrument An Instrument equipped with electronic devices. (5) Control Instrument A non-automatic weighing instrument used to determine the mass of a reference wagon. (6) Weigh zone Zone in which a wagon must be located when it is weighed. 2. ConstnJctlon Note: In this Recommendation the term "device" is applied to any part which uses any means to perform one or more specific functions. (1) Load receptor The part of the weigh zone that Is intended to . receive the load and which realizes a change In the balance of the instrument when a load Is placed upon it. {I) Multiple load receptors Two or more load receptors placed In series that are used as a single load receptor for . full draught weighing. (Ii) Aprons The parts of the weigh zone that are not the load receptor nor part of the load receptor. (2) Electronic device A device comprised of electronic subassemblles and performing a specificfunction. An electronic device is usually manufactured as a separate unit and Is capable of being independently tested. {I) Electronic sub-assembly A part of an electronic device comprised of electronic components and having a recognizable function of its own. (ii) Electronic component The smallest physical entity that uses electron or hole conduction in semiconductors, gases, or In a vacuum. {J) Indicating device The part of the instrument that displays the value of a weighing result in units of mass. (4) Anclllary devices (I) Zero-setting device The means used to set to zero the weight Indicating device when the load receptor Is ery,pty: (a) Non-automatic zero-setting device : A zero­ setting device that must be operated manually. ' (b) Semi-automatic zero-setting device: A zero· setting device that operates automatically following a manual command. (c) Automatic zero-setting device : A zero· setting device that operates automatically and without the intervention of an operator. (ii) Printing device The means to print the weight values of wagons weighed on the instrument and/or a summation of those wagon weights. 3. Hetrolog/cal charactertstlcs (1) Weighing {I) Full draught weighing Weighing a wagon that is entirely supported on the load receptor(s). (ii) Partialweighing Weighing a wagon in two or more parts on the same load receptor. The results are ['IT'TII =3(i)J 429 automatically added to indicate or print the wagon weight. (iii) Weighing-In-motion {wlm} Weighing objects that are in motion. {a) Uncoupled wagon weighing : Weighing-In-motion of wagons that travel Independently across a load receptor. (This is usually achieved by means of an incline of the approach to the load receptor) {b) Coupled wagon weighing : Weighing-in-motion of a train of coupled wagons to obtain a weight indication or printout of the individual wagons. (c) Train weighing : Weighing-in-motion of a number of coupled wagons to obtain a totalized weight of all the wagon weights. (iv) Static weighing Weighing a wagon while stationary and · uncoupled to obtain a weight for the purposes of testing. (2) Copacfty (i) Maximum capacity (Max) The largest load that an Instrument is designed to weigh-in-motion without totalizing. (ii) Minimumcapadty(Min) The load below which a weighing-In-motion result before totalizing may be subject to an excessive relative error. (3) Wagon weight (i) Maximum wagon weight The largest in-motion load that the installation is approved to weigh for a particular site. (ii) Minimum wagon weight The wagon weight below which a weighing­ In-motion result may be subject to an excessive relative error. (4) Scale Interval (d) A value expressed in units of mass for weighing­ in-motion that is the difference between: the values corresponding to two consecutive scale marks for analogue Indication, or two consecutive indicated or printed values for digital indication. (i) Scale interval for stationary load : The scale interval used for static tests. (SJ Speed (i) Maximum operating speed : The greatest velocity of a wagon that the Instrument Is designed to weigh-In-motion and above which the weighing results may be subject to an excessive relative error. (ii) Minimum operating speed : The lowest . velocity of a wagon that the instrument is designed to weigh-In-motion and below which the weighing results may be subject to an excessive relative error. (iii) Range of operating speeds : The difference between the minimum and maximum operating speeds at which a wagon may be weighed-in-motion. (Iv) Maximum transit speed : The maximum speed that a railway vehlcle can travel on the weigh zone without producing a permanent shift In the performance characteristics of a weighing instrument beyond those specified. · . (6) Wann-up time The time between the moment that power Is applied to an instrument and the moment at which the Instrument is capable of complying with the requirements. (7) Durability Ablllty of an instrument to maintain Its performance characteristics over a period of use. 4. Ind/a,t/ons and errors (1) Methods of indication (i) Analogue indication: An Indication allowing the determination of an equilibrium position to a fraction of the scale interval. · (II) OlgitalIndication: An Indication in which the scale marks are a sequence of aligned figures that do not permit interpolation to a fraction of a scale Interval. (2) Errors (I) Error{ofindication): The indication of an instrument minus the (conventional) true value of the mass. --.J.?--?? - ---- 430 THE GAZETTE OF INDIA: EXTRAORDINARY [PARTll SEc.3(i)] (ii) Intrinsicerror: The error of an Instrument.under.reference conditions (Iii) Initial intrinsic error: The Intrinsic .error -of .an :instrument as determlned.prlorto performance tests and durabilityevaluations. (iv) FauH: The difference between the error of indication and.the:intrinsic error of a weighing instrument. Note .t.: •Principally,a fault is the result of an undesired change of data contained in or flowlngthrough an electronic instrument. Note 2.: ,From the definition it follows that in this specification a "fault" is a numerical value. (v) Significant fauH: A fault greater than d. The following are not considered to be slgnlficant faults: * faults that result from simultaneous and mutually independent causes in the Instrument or in Its checking facility, · * faults that imply the Impossibility of performing any measurement, *• transitory faults that are momentary variations in the indications which cannot be interpreted, memorized or transmitted as a measurement result, * faults that are so serious that they will Inevitably be noticed by those interested in the measurement. (vi) Span stability : The capabil!ty of an instrument to maintain the difference between the Indication of weight at maximum capacity and the indication at zero within specified limits over a period of use, (vii) Maximum span stability error : A span stablllty error greater than one half of the absolute value of (he maximum permissible error applicable to the load. (viii) Rounding error: The difference between a digital measurement result (indicated or printed) and the value of the measurement result with an analogue Indication. 5. Influence quantity (J) A quantity that is not the subject of the measurement but which influences the value of the measurand and or the indication of the instrument. (I) Influence factor: An Influence quantity having a value within the specified rated operating conditions of the Instrument. · (11)Olsturbance: An Influence quantity havinga value that falls within the limits specified in this specification but that falls outside the rated operating conditions of the instrument. (2) Rated operating conditions: Conditions of use which give the ranges of the influence quantities for which the metrological characteristics are intended to lie within the specified maximum permissible errors. (J) Reference conditions: A set of specified values of influence factors fixed to ensure valid inter-comparison of the results of measurements. 6. Tests (1) Static test: A test with standard weights or a load that remains stationary on the load receptor to determine an error. (2) In-motion(dynamic) test: A test with reference wagons that are In motion on the load receptor to determine an error. (J) Simulation test: A test carried out on a complete instrument or part of an instrument in which any part of the weighing operation is simulated. (4) Performance test: A test to verify that the equipment under test (EUT) is capable of accomplishing its Intended functions. (S) Span stability test: A test to verify that the EUT is capable of maintaining its performance characteristics over a period of use. 7. Rall vehicles (1) Wagon: A loaded or unloaded railway goods vehicle that is recognized by the instrument as a vehicle to be weighed. 431 :i: 1.0d 500 ,; m ,; 2000 ± 1.Sd 2000 ,; m ,; 10000 (3) Scale Interval (d) For a particular method of weighing-in-motion and combination of bad receptors, all weight indicating and printing devices on an instrument shall have the · same scale Interval. The relationship between the accuracy class, the scale Interval and the maximum wagon weight divided by the scale interval shall be as specified in Table III. (a) A minimum of 500 metre rail track length shall be available with a gradient of not more than 1 : 400 on either side of automatic rail weighbridge. (II) Static weighing : The maximum permissible errors on static weighing for increasing or decreasing loads shall be the appropriate values In Table II. TABLE II Maximum Load (m)expressed in permissible errors number of scale intervals ± 0.Sd O ,; m ,; 500 2500 5000 1250 maximum 250 500 1000 minimum TABLE m (Maximum wagon weight)ld s 50 s 200 s 100 1 Accuracy d(kg} class 0.5 0.2 (2) Reference wagon : A wagon of known weight that is typicalof those to be used for weighing on the instrument and which has been selected for the purposes of in-motion testing. (3) Total train : A number of coupled wagons whose totalized weight Is to be obtained. PART II METROI.OGICALANDTECHNICAL REQUIREMENTS 1..SCope (1) This part specifies the requirements and . test methods for automatic rail-weigh-bridges, hereinafter referred to as "Instruments" that are used to determine the mass of railway wagons, when they are weighed in motion. 2. Metro/oglcal Requirements (1) Accuracy dasses: Instruments are divided into four accuracy classes as follows:- 0.2, 0.5, 1, 2 An instrument may be in a different accuracy class for wagon weighing than that for train weighing. (2) Maximum permissible errors (i) Weighing In motion: The maximum permissible errors for weighing-in-motion shall be as specified in Table I TABLE I The scale intervals of the indicating or printing devices shall be in the form of 1 x 10k,2 x 10k,or 5 x 10k,"k" being a positive or a negative whole number or zero. (4) Minimum capacity The minimum capatity shall not be less than It, and not greater than the value of the result of the minimum wagon weight divided by the number of partial weighings. (5) Minimum wagon weight The minimum wagon weight shall not be less than 50d. (6) Single axle or boqie weights Singleaxle or bogie weights shall not be indicated or printed without an associated warning that these weighing results cannot be verified. Accuracy Percentage of mass of single wagon or class total train as appropriate Verification Inspection 0.2 ±0.10% :1:0.2% 0.5 ±0.25% ±0.5% 1 ±0.50% :I: 1.0% 2 ±1.00% :i:2.0% Note : For the application of maximum permissible errors refer to paragraphs 2(8)(11)(a)and 2(8)(ii)(b) of this part. On i;;itial verification of an instrument weighing coupled wagons, the errors of not more than 10% of the weighing results taken from one or more passes of the test train may exceed the appropriate maximum _ permissible error given in Table I but shall not exceed two time that value. 2 s 500 100 600 - - - ? -------- 432 THE GAZETTE OF INDIA: EXTRAORDINARY [PARTll-SEC.3(i)] (7) Agreement between indicating& printing devices. I For the same load, the difference between the weighing results provided by any two devices having the same scale intervals shall be as follows:- • zero for digital.devices • i'bt greater than the.absolutevalue of the maximuri1;permissibleerrorfor weighing in motion for analog•devices. (8) Weighingtestmethods The appropriate ;procedures are specified in paragraph 5(1) .(pattern evaluation) 5(2) (Initial verification) and 5(3)(1nspection) of this part. (i) Static weighing : An instrument to be used as a control instrument shall meet the requirements of paragraph 2(8)(iii)(b) and paragraph 2(8){1){a) to 2{8)(i)(f) inclusive. The maximum permissible errors shall comply with Table II. (a) Multipleload receptors : Each load receptor shall be tested by the static-weighing method both independently and In combination. (b) Zero-setting : The Instrument shall be capable of setting to zero within± 0.25% of the scale Interval for a stationary load. (c) Eccentric loading : The indications for different positions of the load shall comply with the maximum permissible errors for the given load. (d) Verlflcatlon standards : The error of the standard weights or masses used shall not be greater than one-third of the maximum permissible error for the load, as specified in Table II. (e) Test loads (e-i) Load at which errors are to be determined Errors shall be determined for test loads of: • zero • minimum capacity, • maximum capacity, and at or near a load where the maximum permissible error changes · (e-ll) Distribution of test load Except for eccentricity test, standard weights or masses shall be evenly distributed on the load receptor. (e-lll) Eccentricity tests : Tests shall be carried out without excessive stacking or overlapping of the load on the load receptor provided that the conditions are practical and safe. The test load shall be equal to one half of the maximum capacity rounded up to the next 1 tonne increment. The test load shall be on the rails covering the least area as is practicable and stacked across each pair of supports of the load receptor. (f) Discrimination tests : An additional load that is equal to 1.4 times the scale interval for a stationary load, when gently placed on or withdrawn from each load receptor in turn when at equilibrium at any load shall change the initial indication. (Ii) Weighing-in-motion (a) Wagon weighing : The maximum permissible error for coupled or uncoupled wagon weighing shall be one of the followingvalues, whichever is greater: • the value calrulated according to Table ' !,' rounded to the nearest scale interval. o the value calrulated according to Table I, rounded to the nearest scale interval for the weight of a single wagon equal to 35% of the maximum wagon weight (as inscribed on the descriptive markings), or • ld. (b) Train weighing : The maximum permissible error for train weighing shall be one of the followingvalues, whichever is greater: • the value calculated according to Table I, rounded to the nearest scale interval, • the value calculated according to Table I, for the weight of a single wagon equal to 35% of the maximum wagon weight (as Inscribed on the descriptive markings) multiplied by the number of reference 433 wagons in the train (not exceeding 10 wagons) and rounded to the nearest scale interval, or • ld for each wagon in the train but not exceeding 10d See Figure 1 for an Illustration of this requirement. (iii) Verificetlon standard (a) Separate control instrument: A control instrument capable of being used to determine the mass of each reference wagon when stationary and uncoupled shall be available for In­ motion tests, The error of that instrument shall not be greater than either of thefollowing values:- • one-third of the appropriate maximum permissible error for in-motion­ . weighing in dause paragraph 2(8)(11) of this part and Table I if the control instrument is verified immediately prior to the in-motion tests. • one-fifth of the.maximum permissible error if. the control instrument Is verified at any other time. Figure I AlJTOMATIC RAIL-WELGHBRIDGES Illustration of maximum permissible errors and in-motion test requirements for a sample train --- ·--·· -·?- ?- .. Ern,...(1) {wnnu) 1.0 . --··-·----'!'-· .. --1--··-----r 0 B se ll)(I ·· Lnad (1t1n11??) - 20 · . - . - Numbc;r.i( wns,.lln?" SO; nurnl,c-r of refu?ocr w•ai:111• .. IS Wn1on w11l1jht• I OCH lS · U111,I {tonnes) An instrument constructed only for partial weighing of two axle wa·gonsmay be used as the control instrument provided that the alignment calibration in Annex B has been successfullyapplied. (b) Integral control Instrument An instrument under test may be used as the control Instrument provided that It meets the following requirements :- • it must have an appropriate scale interval or scale interval for stationary land, and • it must comply with the requirements in paragraph 2(8)(1) and para 2(8)(iii)(a) of this part. An Instrument constructed only for partial weighing of two-axle wagons may be used as the _ _j ·---. 434 THE GAZETTE OF INDIA: EXTRAORDINARY [PARTll--SEC. 3(i)] • control instrument, provided· that· the alignment calibration in Annex 8 has been successfullyapplied. (9) Influence qUiJntlties (i) Temperature : · Instruments ,shall. comply ·with .the appropriate -metrological and technical requirements at temperatures,from - 10°C to+ 45°C. For speciaLapplicatlons;however, the .limits of the. temperatureranee may differ,providedthat this range shall not be.less than ·30?Cand.shall'be specified in the descriptive markings. Instruments shall be tested in accordance with the static temperature tests in .paragraph 8(1) of Annexure A. (ii) Main Power Supply (AC): Instruments that are powered by an AC supply .shall comply with the appropriate metrologic.aland technical requirements when operated under variations ofvoltage from - 15% to + .10% of the value marked on the instruments. Instruments shall be tested in accordance with main power supply (AC} test in paragraph 8(3) of AnnexureA. (iii) Battery.Power supply (DC) : Instruments that are powered by a DC supply shall, comply with appropriate metrological paragraph and technical requirements in accordance with paragraph 4(3)(viii) of this pa rt. Instruments shall be tested in accordance with the DC power supply test in paragraph 8(4) of AnnexureA. (1O)Conditions of use (i) Use as a non-automatic weighing instru- ment t An Instrument that can be used as a non-automatic weighing Instrument shall meet the requirements for accuracy class III (medium accuracy) and IIII (ordinary accuracy) under specification for "non-automatic weighing instruments" given In Heading A of the Seventh Schedule. (ii) Scale Interval for stationary load: If the scale interval for stationary load Is not equal to the verification scale interval (d), it shall be automatically out of-service when the instrument is in use for weighing-in-motion. In addition, if the instrument is not verified for use as a non-automatic weighing instrument, the verification scale interval for stationary load shall not be readilyaccessible and shall only be used for static testing. 3. Technical requirement (1) Composition : Instruments shall include the following: • one or more load receptors, • aprons, • vehicle type identification devices (e.g. track switch, load cells, transponder, etc.), • Indicating devices, • printer, • control unit. (2) Sultab//lty for use : Instrument shall be designed to suit the vehicle, site and method of operation for which they are intended. Instruments constructed only for partial weighing shall not be used to weigh liquid loads or any other load that may be subjected to fluctuations in its gravity centre, unless there is a possibility to anticipate and compensate for such fluctuations. ' (J) Security of operation (I) Ace/dental ma/adjustment: Instruments shall be constructed so that maladjustments likely to disturb their metrological performance cannot normally take place without the effect being easily detected. (ii) Interlocks: Interlocks shall prevent the use of any control device that may alter a weighing operation. (Ill) Uncoupled wagon weighing: Instruments used for uncoupled wagon weighing shall recognize and indicate the following situations: • the passage of two or more coupled wago_ns, the passage of two or more uncoupled wagons that is sufficiently close to ['lfll II-? 3(i)] 435 cause either a malfunction of the Instruments or errors exceeding the appropriate maximum permissible errors. (iv) Use as a non-automatic weighingInstrument: An instrument to be used as a non-automatic weighing Instrument shall • comply with the requirements of accuracy class III or class IV of "specification for non-automatic weighing instruments prescribed in Seventh Schedule Heading A, and • be equipped with an enabling device for non-automatic operation that prevents both automatic operation and In-motion weighing. (v) zero setting device: An Instrument shall be equipped with a semi-automatic or automatic zero setting device for each load receptor. Its operation shall be possible only when the Instrument Is In stable equillbrlum and when the rate of correction is no more than O.Sd/s. The range of the zero-setting device shall not exceed 4% of the maximum capacity. (4) Indicating and pn'nting devices (i) Quality of/ndication: The weight indication shall be the self­ Indicatingtype. Indicatingand printing device shall allow reliable, simpleand unambiguous reading of the results by simple juxtaposition and shall bear the name or symbolof the appropriate unit of mass. (ii) Printing: The minimum printout resulting from each normal weighing operation shall be each wagon weight in the case of wagon weighing and total train weight in the case of train weighing. (iii) Weighing range: Instruments shall not indicate or print: • the weight of any wagon, or • a totalized weight Inclusive of any wagon, that will cause a weighing results less than Min or greater than Max+ 9d. - ---- ?-··----?- (Iv) Operating speed: The printer shall not print the weight or any wagon that has travelled over the load receptor at a speed outside the range of operating speeds. An appropriate indication shall be included on the printout for any wagon weight not printed and subtotal may be printed exclusive of unweighed wagons provlded that an indication clearly specified that it is not the total train weight. (v). Roll back: The weight indication and printout shall not be altered due to any part of any wagon travelling over the load receptor more than once. If Rollback takes place at the time of testing the entire weighing operation has to be repeated a new. The speed difference shall not be more tlhan 3 kmph. (5) Installation (i) Ease of static testing : The instrument shall be accessible to venicle for moving test weights if It Is to be used as the control instruments. (II) Drainage: If the weighing mechanism is contained in a pit, there shall be a provision for drainage to ensure that no portion of the instrument becomes submerged or partiallysubmerged In water or any other liquid. (6J Descriptive markings Instruments shall bear the following basic markings at each location having a weight indicating or printing device. (I) Markings shown in full • identification mark of the manufacturer • identification mark of the importer (If applicable) • designations of the instrument • serial number of the instrument on each load receptor (if applicable) • weighing method (see item 3(1) under part I) • maximum wagon weight-kg or t • minimum wagon weight-kg or t _ _L __ ? ? --------- 436 THE GAZETTE.OF INDIA:EXTRAORDINARY [PARTII-SEC. 3(i)] • • • • It shall be possible to seal the plate bearing the markings, unless it cannot be removed without being destroyed. (7) Verification marks (I) Position : Instruments shall have a place for the application of verification marks. Toe followingapplies for this place: The part on which the marks are located cannot be removed from the instruments without damaging the marks. The place shall permit the easy application of the marks without changing the metrologicalqualities of the instruments. The marks shall be visible when the instrument Is In service. (II) Mounting: Instruments required to bear verification marks shall have a . verification mark support located as specified above, which shall ensure the conservation of the marks as follows:- • When the mark Is made with a stamp, the support may consist of a strip of lead or any other material with similar qualities inserted into a plate fixed to , the instrument or a cavity bored into the instrument. • When the mark consists of an adhesive transfer, a space shall be provided for this purpose. 4. Requirements for electronic Instruments Electronic instruments shall comply with the following requirements, in addition to the applicable requirements of all other clauses. (1) General requirements (I) Rated operating conditions: Electronic weighing instruments shall be designed and manufactured so that they do not exceed the maximum permissible errors· under rated operating conditions. (II) Disturbances: Electronic instruments shall be designed and manufactured so that when they are exposed to disturbances, either • • • • not to be used to weighliquidproducts (Ifapplicable) • full draught or number of partial weighing per wagon • not to be used to weigh liquidproduct (Ifapplicable) • maximum transit speed-km/h • direction of weighing (Ifapplicable) • wagons.pushed/pulled (whichever is applicable) • scale :Interval for stationary load (if appllcable),kg • ort • electric power supply voltage and frequency-v ...... Hz (ii) Marklng·shown In code (a) For all instruments pattern approval mark accuracy class (for each weighing method, if applicable) 0.2, 0.5, 1 or 2. . maximum capacity Max-kg or t minimum capacity Min·kg or t scale inter d-kq or t maximum operating speed V max· km/h minimum operating speed V mln·km/h (b) For coupled wagon and train weighing : Markingrequired for each weighingmethod applicable: • maximum number of wagons per train -n max. • minimum number of wagons per train -n min. (iii) Other markings: The designation of the liquid(s)which the Instruments is designed to weight (Ifapplicable) (iv) Presentation of descriptive markings: Descriptive markings shall be indelible and of a size, shape and clarity that permit legibility under normal conditions of use of the instrument Markings shall be grouped together In a clearly visible place on the Instrument, either on a descriptive plate fixed r.ear the Indicatingdevice or on the Indicating device itself T f-····,· 437 {a) significant faults do not occur, or (b) significant faults are detected and acted upon. Note: A fault equal to or less than the significant fault {Id) is allowed Irrespective of the value of the error of indication. {iii) Durability: The requirements in paragraphs 4(1){1)and 4(1)(ii) of this part shall be met durably in accordance with the intended use of the instrument (iv) Evaluation for compliance: A pattern of an electronic instrument is presumed to comply with the requirements in paragraphs 4(1)(i), 4(1)(ii) and 4(1)(iii) of this part if it passes the examination and tests specified in Annexure A. (2) Application (i) The requirements in paragraph 4( l)(il) of this part may be applied separately for the following:- each Individual cause of significant fault, and/or each part of the electronic instrument (ii) The choice as to whether to apply paragraph 4{l)(ii)(a) or paragraph 4(1)(11)(b)of this part is left to the manufacturer. (J) Functional requirements (i) Acting upon a slgniflcant fault: When a significantfault has been detected, a visual or audible Indication shall be provided and shall continue until the user takes action or the fault disappears. Means shall be provided to retain any totalized load Information contained in the instrument when a significant fault occurs. (ii) Switch-on procedure: Upon switch-on (in the case of electronic instruments permanently connected to the mains at switch-on of indication), a special procedure shall be.performed that Indicates all the relevant signs of the Indicator In their active and non-active state for a sufficient time to be easily observed by the operator. (iii) Influence factors: An electronic Instrument shall comply with the requirements of paragraph 2(9) of this l17lt ?6)1!?/11:>c part and In addition It shall maintain its metrological and technical characteristics at a relative humidity of 85% at the upper limit of the temperature range of the instrument. (iv) Disturbances : When an electronic Instrument is subjected to the disturbances specified in Annex A, either of the followl ng shall apply: The difference between the weight indication due to the disturbance and the indication without the disturbances (intrinsic error) shall not exceed the significant fault ( ld). The instrument shall detect and act upon a significant fault. (v) Warm up time: During the warm-up time of an electronic Instrument, there shall be no indication or transmission of the weighing result and automatic operation shall be inhibited. (vi) Interface: An Instrument may be. equipped with an Interface permitting the coupling of the instrument to external equipment. When an Interface is used, the Jnstrument shall continue to function correctly and its metrological functions shall not be Influenced. (vii) Mains power supply {AC) : An Instrument that operates from the mains shall In the event of a power failure retain the metrological information contained in the Instrument at the time of failure for at least 24 hours. A switch-over to an emergency power supply shall not cause a significant fault. (viii) Battery power supply (DC) : An Instrument that operates from a battery power supply shall, whenever the voltage drops below the manufacturer's specified value, either continue to function correctly or automatically be put out of service. (4) Examination and tests : The examination and testing of an electronic weighing Instrument is intended to verify compliance with the applicable requirements of this specification and especially with the requirements in paragraph 4 of this part. 438 THE GAZETTE OF INDIA: EXTRAORDINARY [PARTIl-.5Ec. 3(i)] (i) Examinations: An electronic weighing instrument shall be examined to obtain a general appralsal of the design and construction. (ii) Performance tests: An electronic weighing instrument or electronic device, as appropriate, .shall be tested as specified in.the Annex to determine their correct functioning. Tests are to be conducted .on the whole instrument except when the size and/or configurationof the instrument does not lend itself to testing as a unit. It such cases, the separate electronic devices shall be subjected to testing. It is not Intended that electronic devices be further dismantled for separate testing of components. In addition, an examination shall be carried out on the fully operational weighing instrument or, if necessary, on the electronic devices in a simulated set-up that sufficientlyrepresents the weighing instrument. The equipment shall continue to function correctly as specified in Annexure A. (iii) Span stability tests: The instrument shall be subjected to span stability tests at various intervals, before during and after being subjected to performance tests. When the Instrument is subjected to the span stability tests specified in paragraph 10 of AnnexureA. The maximum allowable variation in the errors of Indication shall not exceed half the absolute value of the maximum permissible error in Table II for the test load applied on any of the measurements. Where the difference of the results indicate a trend more than half the allowable variation specified above, the test shall be continued until the trend comes to rest or reverses Itself,or until the error exceeds the maximum allowable variation. 5. Hetrologlcal rx,ntrols The metrologlcal controls of instruments shall consist of the following:- • pattern evaluation • initial verification • in-service Inspection (1) Pattem evaluation (i) Documentation : The application for pattern evaluation shall Include documentation which provides the following information:- • metrological characteristics of the instrument; • a standard set of specifications for the instruments; • a functional description of the components and devices; • drawings, diagrams and general soft ware information (if applicable}, explaining the construction and operation; • any document or other evidence demonstrating that the design and construction of the instrument complies with the requirements of this specification. (II) General Requirements: Pattern evaluation shall be carried out on at least one and normally, not more than three Instruments that represent the definitive pattern. At least one of the instruments shall be completely installed at a typical site and at least one of the instruments or the major component of an instrument shall be submitted in a form suitable for simulation testlrig in a laboratory. Pattern evaluation shall be carried out as prescribed under the Model Approval Rules. The evaluation shall consist of the tests specified in paragraph S{l){iii}of this part. {Ill) Pattem evaluation tests: Instruments shall comply with the following:- • the metrological requirements In paragraph 2 of this part, particularly with reference to maximum permissible errors and, if appropriate, when the instrument is operated In accordance with the manufacturer's specification for products; Note: Evaluation for static weighing [paragraph 2(8)(1)] shall be excluded unless the instrument Is constructed for use as a control instruments [paragraph 2(B)(iii)(b)]. • the technical requirement in paragraph 3 of this part. 439 Additionally, electronic Instruments shall comply with the requirements in paragraph 4 of this part. (a) In-motion tests : The Instrument shall be tested in accordance with the provisions in paragraph 2(8)(11) of this part by reference to "initial verification" as appropriate and shall complywith the requirements in paragraph2(7) of this part. Errors shall be determined by comparing the tests results with the reference wagon mass derived as in paragraph 2(8)(ili) of this part. The range of speeds used during these tests shall be in accordance with the pattern specification. (a-I) Uncoupled wagons : Instruments for weighingIndividual un coupled wagons shall be tested using not less than five reference wagons having a range of loads from zero load (wagon tare weight) to that ofa fully loaded wagon. A minimum of five weight Indications or print outs of each wagon shall be used for assessing compliance with. the requirements in paragraph 2(8)(ii)(a) of this part. (a-Il) Coupled wagons : Instruments designed to weigheither Individual coupled wagons or a total train of coupled wagons shall be tested In either of the following manners:- • By using a test train of empty reference wagons and a test train of both full and partially filled reference wagons. Each test train shall be comprised of not less than five (and normally not more than 15) reference wagons and shall be weighed repeatedly and in each direction (if applicable) to yield not less than 60 wagon weights or the equivalent in total train weight. • According to the requirements in paragraph 5(2)(11)of this part. • Each weight indication and printout obtained in the test shall be used for assessing compliance with requirements in either paragraphs 2(8)(ii)(a) or 2(8)(ii)(b) of this part, as appropriate: PROVIDED that where test wagons are not available loaded wagon weighed on a static machine fulfillingconditions laid down in paragraph 2(8)(i)(d) of this part shall be used as a reference wagon. {b) Simulation tests: Influence factors shall be applied during simulation tests in a manner that will reveal an alteration of the weighing results for any weighing process to which the instrument could be applied, in accordance with the following:- Paragraph 2(9) of this part for all Instruments Paragraph 4 of this part for electronic Instruments. Wheh conducting such tests on a load cell or on an electronic device equipped with an analogue component, the maximum permissibleerror for the device under test shall be 0.7 times the appropriate value specified in Table II. If the metrological characteristics of the load cell, or other major component has been evaluated in accordance with the requirements of the OIML International Recommendation R60 or any other applicable Recommendation that evaluation shall be used to aid in the pattern evaluation, If so requested by the applicant. Note : Since the requirements of this clause apply only to the instrument submitted for pattern evaluation and not to those subsequently submitted for verification, the means used to determine if the appropriate maximum permissible error or maximum allowable variation has been exceeded will be decided and mutually agreed upon between the Metrological Authority and the applicant. Following are examples of these means:- • an adaptation of the Indicating device to give greater resolution tha11 that of the scale Interval; • the use of the scale lnterva I for stationery load; • the use of change points; • any other means mutually agreed upon. (3) Tests for compliance to determine 440 THE GAZETTE·0F INDIA:EXTRAORDINARY [PARTII---SEC. 3(i)] qualifiedpersonnel and a control instrument. . Toe instrument under test may be used as a control instrument provided it complies with the requlrements In paragraph 2(8)(1ii)(b)of this part. (Iv) Place of testing: Verification test shall be conducted entirely at the place of installation and during testing, the Instrument shall Include all parts which form the assembly as Intended for normal use. (3) Inspection : Inspection shall be carried out In accordance with the same provisions as in paragraph 5(2) of this part for verification with the exemption that the maximum permissible errors applicable for inspection shall be applied. 6. On site test methods (1) Proporiion of reference wagons in a test train Toe proportion of reference wagons to the remaining wagons in a test train shall be in accordance with Table-IV. (2) In-motion tests with coupled wagons If the number of reference wagons is less than the total number of wagons in a test train, the reference wagons shall be distributed throughout the train. (3) In motion tests with liquid load When using liquid load, tests shall be consistent with the Intended use of the instrument.

Previous and next

All provisions of Legal Metrology (General) Rule, 2011

Questions people ask

Is annexure of Legal Metrology (General) Rule, 2011 still in force?

No repeal of Legal Metrology (General) Rule, 2011 is recorded.