Annexure, Legal Metrology (General) Rule, 2011: ANNEXURE
Legal Metrology (General) Rule, 2011 - Annexure
Text of annexure
ANNEXURE
ANNEXURE
Tests for single pan balances.
1. Sensitivenes!r The balance shall be tested
for sensitiveness near zero, middle and extreme
position of the projection scale under three conditions
426 THE GAZETTE OF INDIA:EXTRAORDINARY [PARTII-SEC. 3(i)]
of loadingnamely no load, half load and full load. Toe
balance shall be.such so as to record the change in
mass of the order of tsub-dlvlslon of the·projection
scale accurately within the limits of the value equivalent
to the least count of.the vernier or micrometre scale if
provided, otherwise. within half a division of the
projection scale.
2. Accuracy of projection scale-The-accuracy
of the projection scale shall be examined at 10 points
of the scale under.three different conditions of loading
namely at no load, half load and full load. The
maximum error at any point shall not exceed the value
of the one half-sub-division-of the projection scale, if
no vernier or micrometer scale is provided or two
divisions of the vernier or micrometre scale.
3. Consistency of performance-Ten
consecutive readings shall be noted by releasing
and arresting the balance in the unloaded condition
followed by another ten readings when the balance
is in fully loaded condition. The standard deviation
from the mean of the rest point shall be calculated
separately for each condition, which in no case
shall exceed half the division of the projection
scale if no vernier or micrometer scale Is·provided
or two divisions of the vernier or micrometer
scale.
PARTII·COUNTER MACHINES
1.. Oeflnltlon
A counter machine is an equal arm weighing
instrument of capacity not exceeding 50 kg, the pans
of which are above the beam. Figure33 K illustrates a
typical counter machine.
FIGURE33 K
2. Capacities
The machines may be of the followingmaximum
capacities:-
500g, lkg, 2kg, 3kg, Skg, 10kg, 1Skg,20kg,25kg,
30kg and 50kg.
3. General Requirements
(a) When the beam of body has two sides, they
shall be connected further by not less than
two cross-bars. The supports for the pans
shall be of a suitable rigid structure such as
cross members strengthened by straps.
Central pieces or forks shall be fixed so that
they are not twisted or dislocated.
(b) Bearing surfaces, knife-edges and points of
contact of all stays, hooks and loops shall be
of hard steel or agate. The knife-edges and
bearings shall be so fitted as to allow the
- --
--------------------------?
427
4, Tests
(a) The machine shall be tested on a horizontal
level plane.
(b) Sensitiveness and Error
(i) The machine shall be tested for
sensitiveness at full load with the
beam In a horizontal position. The
TABLE21J
beam to move freely. The knee-edges shall
rest upon the bearings along the whole
length of their working part.
(c) A counter machine may have a balance box
for minor adjustments. In such a case, the
balance box shall be permanently fixed
beneath the weight pan and shall be large
enough to contain loose material to an
amount upto one per cent of the capacity of
the machine. No other adjusting contrivance
shall be used.
(d) The pans may be made of any suitable
material such as mild steel, stainless steel,
brass or bronze, aluminium or its alloys,
porcelain, enamel coated steel, glass or
plastic material. They may be of any
convenient shape.
(e) The minimum fall of the extremities of the
beam, either way, of counter machines shall
be as follows:-
Cepedty
500 g, 1 kg and 2 kg
3 kg, S"kg,10 kg, 15 kg
20 kg, 25 kg and 30 kg
50 kg
Minimum fall
6mm
10mm
12mm
13mm
addition of the weight specified in col.
2 or 4 as the _case may be of Table
21 J shall cause the pointer to have a
displacement corresponding to the
minimum limits specified In
paragraph 3(e).
(ii) The error that is the weight required
to bring the beam of the instrument
to horizontal position when fully
loaded with weights each equal to Its
capacity on both pans shall not exceed
the limits specified In cols. 3 and s as
the case may be of Table 21 JK.
(c) Shift Test
(I) When the goods pan is in the form of
a scoop, the counter machine shall
be correct to the prescribed limits of
error if half the full load is placed
against the middle of the back of the
scoop and the other half at any
position on the scoop.
(II) When the goods pan is not in the form
of a scoop, the counter machine shall
indicate the same weight within half
the prescribed limits of error, if the
centre of a load equal to half the
capacity placed on the goods pan Is
moved any where within a distance
from the centre equal to one-third of
the length of the pan. If the pan has
a vertical side against the middle of
that side, the weight being entirely
on the weight pan, but in any position
on it.
SENSITlVENESS AND MAXIMUM PERMISSIBLE ERRORS FOR COUNTER MACHINES
Verification Inspection
Capacity Sensitiveness Maximum Sensitiveness Maximum
when fully permissible error, when fully permissible error
loaded in excess or loaded in excess or
deficiency, when deficiency, when
fully loaded fullyloaded
1 2 3 4 5
500g 1.5g
'
2.2 g 4.Sg 4.5 g
1 kg 2.0g 3.0 g 6.0g 6.0g
2kg 3.0g 4.5 g 9.0g 9.0g
3 kg 4.0g 6.0g 12.0g 12.0g
5 kg 6.0g 9.0g 18.0g 18.0g
10kg 7.0 g 10.Sg 21.0g 21.0 g.
15kg 8.0g 12.0g 24.0g 24.0g
20 kg 9.0g 13.Sg 27.0g 27.0g
25 kg 10.0g 15.0g 30.0g 30.0g
30kg 11.0 g 20.0g 33.0g 40.0g
50kg 15.0 g 30.0g 45.0 g 60.0g
4 -
428 THE GAZETTE:Of INDIA: EXTRAORDINARY [PAJITfl-SEC. 3(i)]
5. Sealing
Each madnne.shall'be provided with a,plug or
stud of soft metal on a conspicuous part of the.beam
or body to receive.the stamp orseal of.the verification
authority. Such a· plug. or .stud -shall 'be made
irremovable by undercutting itor by .sorne suitable
method.
SEVENTll'SCHEDULE
HEADING-<
AIJTOMAllC .RAIL-,WEIGHBRIDGES
.'l!ARTI
TERMINOLOGY (Terms and definitions)
1. Genera/deHnltlons
(1) Weighing instrument
Measuring instrument that serves to determine
the mass of a load by using the action of gravity.
(2) Automatic weighing instrument
An instrument that weighs without the
intervention of an operator and follows a
predetermined program of automatic processes
characteristic of the Instrument.
(J) Rail·welghbn'dge:
A weighing Instrument having a load receptor,
inclusive of rails for conveying railway vehicles.
(4) Electronic instrument
An Instrument equipped with electronic devices.
(5) Control Instrument
A non-automatic weighing instrument used to
determine the mass of a reference wagon.
(6) Weigh zone
Zone in which a wagon must be located when it
is weighed.
2. ConstnJctlon
Note: In this Recommendation the term "device"
is applied to any part which uses any means to perform
one or more specific functions.
(1) Load receptor
The part of the weigh zone that Is intended to
. receive the load and which realizes a change In the
balance of the instrument when a load Is placed upon
it.
{I) Multiple load receptors
Two or more load receptors placed In series
that are used as a single load receptor for
.
full draught weighing.
(Ii) Aprons
The parts of the weigh zone that are not the
load receptor nor part of the load receptor.
(2) Electronic device
A device comprised of electronic subassemblles
and performing a specificfunction. An electronic device
is usually manufactured as a separate unit and Is
capable of being independently tested.
{I) Electronic sub-assembly
A part of an electronic device comprised of
electronic components and having a
recognizable function of its own.
(ii) Electronic component
The smallest physical entity that uses electron
or hole conduction in semiconductors, gases,
or In a vacuum.
{J) Indicating device
The part of the instrument that displays the value
of a weighing result in units of mass.
(4) Anclllary devices
(I) Zero-setting device
The means used to set to zero the weight
Indicating device when the load receptor Is
ery,pty:
(a) Non-automatic zero-setting device : A zero
setting device that must be operated
manually.
'
(b) Semi-automatic zero-setting device: A zero·
setting device that operates automatically
following a manual command.
(c) Automatic zero-setting device : A zero·
setting device that operates automatically
and without the intervention of an operator.
(ii) Printing device
The means to print the weight values of
wagons weighed on the instrument and/or
a summation of those wagon weights.
3. Hetrolog/cal charactertstlcs
(1) Weighing
{I) Full draught weighing
Weighing a wagon that is entirely supported
on the load receptor(s).
(ii) Partialweighing
Weighing a wagon in two or more parts on
the same load receptor. The results are
['IT'TII =3(i)J 429
automatically added to indicate or print the
wagon weight.
(iii) Weighing-In-motion {wlm}
Weighing objects that are in motion.
{a) Uncoupled wagon weighing :
Weighing-In-motion of wagons that
travel Independently across a load
receptor. (This is usually achieved by
means of an incline of the approach
to the load receptor)
{b) Coupled wagon weighing :
Weighing-in-motion of a train of
coupled wagons to obtain a weight
indication or printout of the individual
wagons.
(c) Train weighing :
Weighing-in-motion of a number of
coupled wagons to obtain a totalized
weight of all the wagon weights.
(iv) Static weighing
Weighing a wagon while stationary and
·
uncoupled to obtain a weight for the
purposes of testing.
(2) Copacfty
(i) Maximum capacity (Max)
The largest load that an Instrument is
designed to weigh-in-motion without
totalizing.
(ii) Minimumcapadty(Min)
The load below which a weighing-In-motion
result before totalizing may be subject to an
excessive relative error.
(3) Wagon weight
(i) Maximum wagon weight
The largest in-motion load that the
installation is approved to weigh for a
particular site.
(ii) Minimum wagon weight
The wagon weight below which a weighing
In-motion result may be subject to an
excessive relative error.
(4) Scale Interval (d)
A value expressed in units of mass for weighing
in-motion that is the difference between:
the values corresponding to two consecutive scale
marks for analogue Indication, or
two consecutive indicated or printed values for
digital indication.
(i) Scale interval for stationary load : The scale
interval used for static tests.
(SJ Speed
(i) Maximum operating speed : The greatest
velocity of a wagon that the Instrument Is
designed to weigh-In-motion and above
which the weighing results may be subject
to an excessive relative error.
(ii) Minimum operating speed : The lowest
. velocity of a wagon that the instrument is
designed to weigh-In-motion and below
which the weighing results may be subject
to an excessive relative error.
(iii) Range of operating speeds : The difference
between the minimum and maximum
operating speeds at which a wagon may be
weighed-in-motion.
(Iv) Maximum transit speed : The maximum
speed that a railway vehlcle can travel on
the weigh zone without producing a
permanent shift In the performance
characteristics of a weighing instrument
beyond those specified. ·
.
(6) Wann-up time
The time between the moment that power Is
applied to an instrument and the moment at which
the Instrument is capable of complying with the
requirements.
(7) Durability
Ablllty of an instrument to maintain Its
performance characteristics over a period of use.
4. Ind/a,t/ons and errors
(1) Methods of indication
(i) Analogue indication:
An Indication allowing the determination of
an equilibrium position to a fraction of the
scale interval.
·
(II) OlgitalIndication:
An Indication in which the scale marks are
a sequence of aligned figures that do not
permit interpolation to a fraction of a scale
Interval.
(2) Errors
(I) Error{ofindication):
The indication of an instrument minus the
(conventional) true value of the mass.
--.J.?--??
- ----
430 THE GAZETTE OF INDIA: EXTRAORDINARY [PARTll SEc.3(i)]
(ii) Intrinsicerror:
The error of an Instrument.under.reference
conditions
(Iii) Initial intrinsic error:
The Intrinsic .error -of .an :instrument as
determlned.prlorto performance tests and
durabilityevaluations.
(iv) FauH:
The difference between the error of
indication and.the:intrinsic error of a weighing
instrument.
Note .t.: •Principally,a fault is the result of
an undesired change of data contained in or
flowlngthrough an electronic instrument.
Note 2.: ,From the definition it follows that
in this specification a "fault" is a numerical
value.
(v) Significant fauH:
A fault greater than d. The following are not
considered to be slgnlficant faults:
* faults that result from simultaneous
and mutually independent causes in
the Instrument or in Its checking
facility, ·
* faults that imply the Impossibility of
performing any measurement,
*• transitory faults that are momentary
variations in the indications which
cannot be interpreted, memorized or
transmitted as a measurement result,
* faults that are so serious that they
will Inevitably be noticed by those
interested in the measurement.
(vi) Span stability :
The capabil!ty of an instrument to maintain
the difference between the Indication of
weight at maximum capacity and the
indication at zero within specified limits over
a period of use,
(vii) Maximum span stability error :
A span stablllty error greater than one half
of the absolute value of (he maximum
permissible error applicable to the load.
(viii) Rounding error:
The difference between a digital
measurement result (indicated or printed)
and the value of the measurement result with
an analogue Indication.
5. Influence quantity
(J) A quantity that is not the subject of the
measurement but which influences the value of the
measurand and or the indication of the instrument.
(I) Influence factor:
An Influence quantity having a value within
the specified rated operating conditions of
the Instrument.
·
(11)Olsturbance:
An Influence quantity havinga value that falls
within the limits specified in this specification
but that falls outside the rated operating
conditions of the instrument.
(2) Rated operating conditions:
Conditions of use which give the ranges of the
influence quantities for which the metrological
characteristics are intended to lie within the specified
maximum permissible errors.
(J) Reference conditions:
A set of specified values of influence factors fixed
to ensure valid inter-comparison of the results of
measurements.
6. Tests
(1) Static test:
A test with standard weights or a load that
remains stationary on the load receptor to
determine an error.
(2) In-motion(dynamic) test:
A test with reference wagons that are In
motion on the load receptor to determine
an error.
(J) Simulation test:
A test carried out on a complete instrument
or part of an instrument in which any part of
the weighing operation is simulated.
(4) Performance test:
A test to verify that the equipment under
test (EUT) is capable of accomplishing its
Intended functions.
(S) Span stability test:
A test to verify that the EUT is capable of
maintaining its performance characteristics
over a period of use.
7. Rall vehicles
(1) Wagon:
A loaded or unloaded railway goods vehicle
that is recognized by the instrument as a
vehicle to be weighed.
431
:i: 1.0d 500 ,; m ,; 2000
± 1.Sd 2000 ,; m ,; 10000
(3) Scale Interval (d)
For a particular method of weighing-in-motion
and combination of bad receptors, all weight indicating
and printing devices on an instrument shall have the
· same scale Interval.
The relationship between the accuracy class,
the scale Interval and the maximum wagon weight
divided by the scale interval shall be as specified in
Table III.
(a) A minimum of 500 metre rail track length
shall be available with a gradient of not more
than 1 : 400 on either side of automatic rail
weighbridge.
(II) Static weighing : The maximum
permissible errors on static weighing
for increasing or decreasing loads
shall be the appropriate values In
Table II.
TABLE II
Maximum Load (m)expressed in
permissible errors number of scale intervals
± 0.Sd O ,; m ,; 500
2500
5000
1250
maximum
250
500
1000
minimum
TABLE m
(Maximum wagon
weight)ld
s 50
s 200
s 100
1
Accuracy d(kg}
class
0.5
0.2
(2) Reference wagon :
A wagon of known weight that is typicalof
those to be used for weighing on the
instrument and which has been selected for
the purposes of in-motion testing.
(3) Total train :
A number of coupled wagons whose
totalized weight Is to be obtained.
PART II
METROI.OGICALANDTECHNICAL
REQUIREMENTS
1..SCope
(1) This part specifies the requirements and
.
test methods for automatic rail-weigh-bridges,
hereinafter referred to as "Instruments" that are used
to determine the mass of railway wagons, when they
are weighed in motion.
2. Metro/oglcal Requirements
(1) Accuracy dasses:
Instruments are divided into four accuracy classes
as follows:-
0.2, 0.5, 1, 2
An instrument may be in a different accuracy
class for wagon weighing than that for train weighing.
(2) Maximum permissible errors
(i) Weighing In motion:
The maximum permissible errors for
weighing-in-motion shall be as
specified in Table I
TABLE I
The scale intervals of the indicating or printing
devices shall be in the form of 1 x 10k,2 x 10k,or
5 x 10k,"k" being a positive or a negative whole
number or zero.
(4) Minimum capacity
The minimum capatity shall not be less than It,
and not greater than the value of the result of the
minimum wagon weight divided by the number of
partial weighings.
(5) Minimum wagon weight
The minimum wagon weight shall not be less
than 50d.
(6) Single axle or boqie weights
Singleaxle or bogie weights shall not be indicated
or printed without an associated warning that these
weighing results cannot be verified.
Accuracy Percentage of mass of
single wagon or class
total train as appropriate
Verification Inspection
0.2 ±0.10% :1:0.2%
0.5 ±0.25% ±0.5%
1 ±0.50% :I: 1.0%
2 ±1.00% :i:2.0%
Note : For the application of maximum
permissible errors refer to paragraphs 2(8)(11)(a)and
2(8)(ii)(b) of this part.
On i;;itial verification of an instrument weighing
coupled wagons, the errors of not more than 10% of
the weighing results taken from one or more passes
of the test train may exceed the appropriate maximum
_
permissible error given in Table I but shall not exceed
two time that value.
2 s 500 100 600
- - - ? --------
432 THE GAZETTE OF INDIA: EXTRAORDINARY [PARTll-SEC.3(i)]
(7) Agreement between indicating& printing
devices.
I
For the same load, the difference between the
weighing results provided by any two devices having
the same scale intervals shall be as follows:-
• zero for digital.devices
• i'bt greater than the.absolutevalue of the
maximuri1;permissibleerrorfor weighing in
motion for analog•devices.
(8) Weighingtestmethods
The appropriate ;procedures are specified in
paragraph 5(1) .(pattern evaluation) 5(2) (Initial
verification) and 5(3)(1nspection) of this part.
(i) Static weighing :
An instrument to be used as a control
instrument shall meet the requirements of
paragraph 2(8)(iii)(b) and paragraph
2(8){1){a) to 2{8)(i)(f) inclusive. The
maximum permissible errors shall comply
with Table II.
(a) Multipleload receptors :
Each load receptor shall be tested by
the static-weighing method both
independently and In combination.
(b) Zero-setting :
The Instrument shall be capable of
setting to zero within± 0.25% of the
scale Interval for a stationary load.
(c) Eccentric loading :
The indications for different positions
of the load shall comply with the
maximum permissible errors for the
given load.
(d) Verlflcatlon standards :
The error of the standard weights or
masses used shall not be greater than
one-third of the maximum permissible
error for the load, as specified in
Table II.
(e) Test loads
(e-i) Load at which errors are to be
determined
Errors shall be determined for test
loads of:
• zero
• minimum capacity,
• maximum capacity, and at or near a
load where the maximum
permissible error changes
·
(e-ll) Distribution of test load
Except for eccentricity test, standard weights
or masses shall be evenly distributed on the
load receptor.
(e-lll) Eccentricity tests :
Tests shall be carried out without excessive
stacking or overlapping of the load on the
load receptor provided that the conditions
are practical and safe. The test load shall
be equal to one half of the maximum capacity
rounded up to the next 1 tonne increment.
The test load shall be on the rails covering
the least area as is practicable and stacked
across each pair of supports of the load
receptor.
(f) Discrimination tests :
An additional load that is equal to 1.4 times
the scale interval for a stationary load,
when gently placed on or withdrawn
from each load receptor in turn when
at equilibrium at any load shall change
the initial indication.
(Ii) Weighing-in-motion
(a) Wagon weighing :
The maximum permissible error for coupled
or uncoupled wagon weighing shall be one
of the followingvalues, whichever is greater:
• the value calrulated according to Table
' !,' rounded to the nearest scale
interval.
o the value calrulated according to Table
I, rounded to the nearest scale
interval for the weight of a single
wagon equal to 35% of the maximum
wagon weight (as inscribed on the
descriptive markings), or
• ld.
(b) Train weighing :
The maximum permissible error for train
weighing shall be one of the followingvalues,
whichever is greater:
• the value calculated according to Table I,
rounded to the nearest scale interval,
• the value calculated according to Table I,
for the weight of a single wagon equal to
35% of the maximum wagon weight (as
Inscribed on the descriptive markings)
multiplied by the number of reference
433
wagons in the train (not exceeding 10
wagons) and rounded to the nearest scale
interval, or
• ld for each wagon in the train but not
exceeding 10d
See Figure 1 for an Illustration of this requirement.
(iii) Verificetlon standard
(a) Separate control instrument:
A control instrument capable of being
used to determine the mass of each
reference wagon when stationary and
uncoupled shall be available for In
motion tests, The error of that
instrument shall not be greater than
either of thefollowing values:-
• one-third of the appropriate maximum
permissible error for in-motion
.
weighing in dause paragraph 2(8)(11)
of this part and Table I if the control
instrument is verified immediately
prior to the in-motion tests.
• one-fifth of the.maximum permissible
error if. the control instrument Is
verified at any other time.
Figure I
AlJTOMATIC RAIL-WELGHBRIDGES
Illustration of maximum permissible errors and in-motion test requirements for a sample train
--- ·--·· -·?- ?- ..
Ern,...(1)
{wnnu)
1.0
. --··-·----'!'-· .. --1--··-----r
0 B se ll)(I
·· Lnad (1t1n11??) -
20 ·
. -
. -
Numbc;r.i( wns,.lln?" SO; nurnl,c-r of refu?ocr w•ai:111• .. IS
Wn1on w11l1jht• I OCH
lS ·
U111,I {tonnes)
An instrument constructed only for partial
weighing of two axle wa·gonsmay be used as the
control instrument provided that the alignment
calibration in Annex B has been successfullyapplied.
(b) Integral control Instrument
An instrument under test may be used as
the control Instrument provided that It meets
the following requirements :-
• it must have an appropriate scale
interval or scale interval for stationary
land, and
• it must comply with the requirements
in paragraph 2(8)(1) and para
2(8)(iii)(a) of this part.
An Instrument constructed only for partial
weighing of two-axle wagons may be used as the
_
_j
·---.
434 THE GAZETTE OF INDIA: EXTRAORDINARY [PARTll--SEC. 3(i)]
•
control instrument, provided· that· the alignment
calibration in Annex 8 has been successfullyapplied.
(9) Influence qUiJntlties
(i) Temperature :
·
Instruments ,shall. comply ·with .the
appropriate -metrological and technical
requirements at temperatures,from - 10°C
to+ 45°C.
For speciaLapplicatlons;however, the .limits of
the. temperatureranee may differ,providedthat this
range shall not be.less than ·30?Cand.shall'be specified
in the descriptive markings.
Instruments shall be tested in accordance with
the static temperature tests in .paragraph 8(1) of
Annexure A.
(ii) Main Power Supply (AC):
Instruments that are powered by an
AC supply .shall comply with the
appropriate metrologic.aland technical
requirements when operated under
variations ofvoltage from - 15% to
+ .10% of the value marked on the
instruments.
Instruments shall be tested in accordance with
main power supply (AC} test in paragraph 8(3) of
AnnexureA.
(iii) Battery.Power supply (DC) :
Instruments that are powered by a
DC supply shall, comply with
appropriate metrological paragraph
and technical requirements in
accordance with paragraph 4(3)(viii)
of this pa rt.
Instruments shall be tested in accordance with
the DC power supply test in paragraph 8(4) of
AnnexureA.
(1O)Conditions of use
(i) Use as a non-automatic weighing instru-
ment t
An Instrument that can be used as a
non-automatic weighing Instrument
shall meet the requirements for
accuracy class III (medium accuracy)
and IIII (ordinary accuracy) under
specification for "non-automatic
weighing instruments" given In Heading
A of the Seventh Schedule.
(ii) Scale Interval for stationary load:
If the scale interval for stationary load
Is not equal to the verification scale
interval (d), it shall be automatically
out of-service when the instrument is
in use for weighing-in-motion. In
addition, if the instrument is not
verified for use as a non-automatic
weighing instrument, the verification
scale interval for stationary load shall
not be readilyaccessible and shall only
be used for static testing.
3. Technical requirement
(1) Composition :
Instruments shall include the following:
• one or more load receptors,
• aprons,
• vehicle type identification devices (e.g. track
switch, load cells, transponder, etc.),
• Indicating devices,
• printer,
• control unit.
(2) Sultab//lty for use :
Instrument shall be designed to suit the vehicle,
site and method of operation for which they are
intended.
Instruments constructed only for partial weighing
shall not be used to weigh liquid loads or any other
load that may be subjected to fluctuations in its gravity
centre, unless there is a possibility to anticipate and
compensate for such fluctuations.
'
(J) Security of operation
(I) Ace/dental ma/adjustment:
Instruments shall be constructed so that
maladjustments likely to disturb their
metrological performance cannot normally
take place without the effect being easily
detected.
(ii) Interlocks:
Interlocks shall prevent the use of any control
device that may alter a weighing operation.
(Ill) Uncoupled wagon weighing:
Instruments used for uncoupled wagon
weighing shall recognize and indicate the
following situations:
• the passage of two or more coupled
wago_ns,
the passage of two or more uncoupled
wagons that is sufficiently close to
['lfll II-? 3(i)] 435
cause either a malfunction of the
Instruments or errors exceeding the
appropriate maximum permissible
errors.
(iv) Use as a non-automatic weighingInstrument:
An instrument to be used as a non-automatic
weighing Instrument shall
• comply with the requirements of
accuracy class III or class IV of
"specification for non-automatic
weighing instruments prescribed in
Seventh Schedule Heading A, and
• be equipped with an enabling device
for non-automatic operation that
prevents both automatic operation and
In-motion weighing.
(v) zero setting device:
An Instrument shall be equipped with
a semi-automatic or automatic zero
setting device for each load receptor.
Its operation shall be possible only
when the Instrument Is In stable
equillbrlum and when the rate of
correction is no more than O.Sd/s.
The range of the zero-setting device shall not
exceed 4% of the maximum capacity.
(4) Indicating and pn'nting devices
(i) Quality of/ndication:
The weight indication shall be the self
Indicatingtype. Indicatingand printing
device shall allow reliable, simpleand
unambiguous reading of the results
by simple juxtaposition and shall bear
the name or symbolof the appropriate
unit of mass.
(ii) Printing:
The minimum printout resulting from each
normal weighing operation shall be each
wagon weight in the case of wagon weighing
and total train weight in the case of train
weighing.
(iii) Weighing range:
Instruments shall not indicate or
print:
• the weight of any wagon, or
• a totalized weight Inclusive of any
wagon, that will cause a weighing
results less than Min or greater than
Max+ 9d.
- ----
?-··----?-
(Iv) Operating speed:
The printer shall not print the weight
or any wagon that has travelled over
the load receptor at a speed outside
the range of operating speeds. An
appropriate indication shall be
included on the printout for any wagon
weight not printed and subtotal may
be printed exclusive of unweighed
wagons provlded that an indication
clearly specified that it is not the total
train weight.
(v). Roll back:
The weight indication and printout
shall not be altered due to any part of
any wagon travelling over the load
receptor more than once. If Rollback
takes place at the time of testing the
entire weighing operation has to be
repeated a new. The speed difference
shall not be more tlhan 3 kmph.
(5) Installation
(i) Ease of static testing :
The instrument shall be accessible to
venicle for moving test weights if It Is
to be used as the control instruments.
(II) Drainage:
If the weighing mechanism is
contained in a pit, there shall be a
provision for drainage to ensure that
no portion of the instrument becomes
submerged or partiallysubmerged In
water or any other liquid.
(6J Descriptive markings
Instruments shall bear the following basic
markings at each location having a weight indicating
or printing device.
(I) Markings shown in full
• identification mark of the
manufacturer
• identification mark of the importer (If
applicable)
• designations of the instrument
• serial number of the instrument on
each load receptor (if applicable)
• weighing method (see item 3(1) under
part I)
• maximum wagon weight-kg or t
• minimum wagon weight-kg or t
_
_L
__ ? ? ---------
436 THE GAZETTE.OF INDIA:EXTRAORDINARY [PARTII-SEC. 3(i)]
•
•
•
•
It shall be possible to seal the plate bearing the
markings, unless it cannot be removed without being
destroyed.
(7) Verification marks
(I) Position :
Instruments shall have a place for the
application of verification marks. Toe
followingapplies for this place:
The part on which the marks are
located cannot be removed from the
instruments without damaging the
marks.
The place shall permit the easy
application of the marks without
changing the metrologicalqualities of
the instruments.
The marks shall be visible when the
instrument Is In service.
(II) Mounting:
Instruments required to bear
verification marks shall have a .
verification mark support located as
specified above, which shall ensure
the conservation of the marks as
follows:-
• When the mark Is made with a stamp,
the support may consist of a strip of
lead or any other material with similar
qualities inserted into a plate fixed to
,
the instrument or a cavity bored into
the instrument.
• When the mark consists of an
adhesive transfer, a space shall be
provided for this purpose.
4. Requirements for electronic Instruments
Electronic instruments shall comply with the
following requirements, in addition to the applicable
requirements of all other clauses.
(1) General requirements
(I) Rated operating conditions:
Electronic weighing instruments shall
be designed and manufactured so that
they do not exceed the maximum
permissible errors· under rated
operating conditions.
(II) Disturbances:
Electronic instruments shall be
designed and manufactured so that
when they are exposed to
disturbances, either
•
•
•
• not to be used to weighliquidproducts
(Ifapplicable)
• full draught or number of partial
weighing per wagon
• not to be used to weigh liquidproduct
(Ifapplicable)
• maximum transit speed-km/h
• direction of weighing (Ifapplicable)
• wagons.pushed/pulled (whichever is
applicable)
• scale :Interval for stationary load (if
appllcable),kg
• ort
• electric power supply voltage and
frequency-v ......
Hz
(ii) Marklng·shown In code
(a) For all instruments
pattern approval mark
accuracy class (for each weighing
method, if applicable) 0.2, 0.5, 1 or
2.
.
maximum capacity Max-kg or t
minimum capacity Min·kg or t
scale inter d-kq or t
maximum operating speed V max·
km/h
minimum operating speed V
mln·km/h
(b) For coupled wagon and train weighing :
Markingrequired for each weighingmethod
applicable:
• maximum number of wagons per
train -n max.
• minimum number of wagons per train
-n min.
(iii) Other markings:
The designation of the liquid(s)which
the Instruments is designed to weight
(Ifapplicable)
(iv) Presentation of descriptive markings:
Descriptive markings shall be indelible
and of a size, shape and clarity that
permit legibility under normal
conditions of use of the instrument
Markings shall be grouped together In a clearly
visible place on the Instrument, either on a descriptive
plate fixed r.ear the Indicatingdevice or on the Indicating
device itself
T f-····,·
437
{a) significant faults do not occur, or
(b) significant faults are detected and
acted upon.
Note: A fault equal to or less than the significant
fault {Id) is allowed Irrespective of the value of the
error of indication.
{iii) Durability:
The requirements in paragraphs
4(1){1)and 4(1)(ii) of this part shall
be met durably in accordance with the
intended use of the instrument
(iv) Evaluation for compliance:
A pattern of an electronic instrument
is presumed to comply with the
requirements in paragraphs 4(1)(i),
4(1)(ii) and 4(1)(iii) of this part if it
passes the examination and tests
specified in Annexure A.
(2) Application
(i) The requirements in paragraph 4( l)(il) of
this part may be applied separately for the
following:-
each Individual cause of significant
fault, and/or
each part of the electronic instrument
(ii) The choice as to whether to apply paragraph
4{l)(ii)(a) or paragraph 4(1)(11)(b)of this
part is left to the manufacturer.
(J) Functional requirements
(i) Acting upon a slgniflcant fault:
When a significantfault has been detected,
a visual or audible Indication shall be provided
and shall continue until the user takes action
or the fault disappears.
Means shall be provided to retain any
totalized load Information contained in the
instrument when a significant fault occurs.
(ii) Switch-on procedure:
Upon switch-on (in the case of electronic
instruments permanently connected to the
mains at switch-on of indication), a special
procedure shall be.performed that Indicates
all the relevant signs of the Indicator In their
active and non-active state for a sufficient
time to be easily observed by the operator.
(iii) Influence factors:
An electronic Instrument shall comply with
the requirements of paragraph 2(9) of this
l17lt ?6)1!?/11:>c
part and In addition It shall maintain its
metrological and technical characteristics at
a relative humidity of 85% at the upper limit
of the temperature range of the instrument.
(iv) Disturbances :
When an electronic Instrument is subjected
to the disturbances specified in Annex A,
either of the followl ng shall apply:
The difference between the weight indication
due to the disturbance and the indication
without the disturbances (intrinsic error)
shall not exceed the significant fault ( ld).
The instrument shall detect and act upon a
significant fault.
(v) Warm up time:
During the warm-up time of an electronic
Instrument, there shall be no indication or
transmission of the weighing result and
automatic operation shall be inhibited.
(vi) Interface:
An Instrument may be. equipped with an
Interface permitting the coupling of the
instrument to external equipment. When an
Interface is used, the Jnstrument shall
continue to function correctly and its
metrological functions shall not be
Influenced.
(vii) Mains power supply {AC) :
An Instrument that operates from the mains
shall In the event of a power failure retain
the metrological information contained in the
Instrument at the time of failure for at least
24 hours. A switch-over to an emergency
power supply shall not cause a significant
fault.
(viii) Battery power supply (DC) :
An Instrument that operates from a battery
power supply shall, whenever the voltage
drops below the manufacturer's specified
value, either continue to function correctly
or automatically be put out of service.
(4) Examination and tests :
The examination and testing of an electronic
weighing Instrument is intended to verify compliance
with the applicable requirements of this specification
and especially with the requirements in paragraph 4
of this part.
438 THE GAZETTE OF INDIA: EXTRAORDINARY [PARTIl-.5Ec. 3(i)]
(i) Examinations:
An electronic weighing instrument shall be
examined to obtain a general appralsal of
the design and construction.
(ii) Performance tests:
An electronic weighing instrument or
electronic device, as appropriate, .shall be
tested as specified in.the Annex to determine
their correct functioning.
Tests are to be conducted .on the whole
instrument except when the size and/or
configurationof the instrument does not lend
itself to testing as a unit. It such cases, the
separate electronic devices shall be
subjected to testing. It is not Intended that
electronic devices be further dismantled for
separate testing of components. In addition,
an examination shall be carried out on the
fully operational weighing instrument or, if
necessary, on the electronic devices in a
simulated set-up that sufficientlyrepresents
the weighing instrument. The equipment
shall continue to function correctly as
specified in Annexure A.
(iii) Span stability tests:
The instrument shall be subjected to span
stability tests at various intervals, before
during and after being subjected to
performance tests.
When the Instrument is subjected to the span
stability tests specified in paragraph 10 of
AnnexureA.
The maximum allowable variation in the
errors of Indication shall not exceed half the
absolute value of the maximum permissible
error in Table II for the test load applied on
any of the measurements.
Where the difference of the results indicate
a trend more than half the allowable
variation specified above, the test shall be
continued until the trend comes to rest or
reverses Itself,or until the error exceeds the
maximum allowable variation.
5. Hetrologlcal rx,ntrols
The metrologlcal controls of instruments shall
consist of the following:-
• pattern evaluation
• initial verification
• in-service Inspection
(1) Pattem evaluation
(i) Documentation :
The application for pattern evaluation shall
Include documentation which provides the
following information:-
• metrological characteristics of the
instrument;
• a standard set of specifications for the
instruments;
• a functional description of the
components and devices;
• drawings, diagrams and general soft
ware information (if applicable},
explaining the construction and
operation;
• any document or other evidence
demonstrating that the design and
construction of the instrument
complies with the requirements of this
specification.
(II) General Requirements:
Pattern evaluation shall be carried out on at
least one and normally, not more than three
Instruments that represent the definitive
pattern. At least one of the instruments shall
be completely installed at a typical site and
at least one of the instruments or the major
component of an instrument shall be
submitted in a form suitable for simulation
testlrig in a laboratory. Pattern evaluation
shall be carried out as prescribed under the
Model Approval Rules.
The evaluation shall consist of the tests
specified in paragraph S{l){iii}of this part.
{Ill) Pattem evaluation tests:
Instruments shall comply with the
following:-
• the metrological requirements In
paragraph 2 of this part, particularly
with reference to maximum
permissible errors and, if appropriate,
when the instrument is operated In
accordance with the manufacturer's
specification for products;
Note: Evaluation for static weighing [paragraph
2(8)(1)] shall be excluded unless the instrument Is
constructed for use as a control instruments
[paragraph 2(B)(iii)(b)].
• the technical requirement in paragraph 3 of
this part.
439
Additionally, electronic Instruments shall
comply with the requirements in paragraph
4 of this part.
(a) In-motion tests :
The Instrument shall be tested in
accordance with the provisions in
paragraph 2(8)(11) of this part by
reference to "initial verification" as
appropriate and shall complywith the
requirements in paragraph2(7) of this
part. Errors shall be determined by
comparing the tests results with the
reference wagon mass derived as in
paragraph 2(8)(ili) of this part.
The range of speeds used during
these tests shall be in accordance with
the pattern specification.
(a-I) Uncoupled wagons :
Instruments for weighingIndividual un
coupled wagons shall be tested using
not less than five reference wagons
having a range of loads from zero load
(wagon tare weight) to that ofa fully
loaded wagon. A minimum of five
weight Indications or print outs of
each wagon shall be used for
assessing compliance with. the
requirements in paragraph 2(8)(ii)(a)
of this part.
(a-Il) Coupled wagons :
Instruments designed to weigheither
Individual coupled wagons or a total
train of coupled wagons shall be
tested In either of the following
manners:-
• By using a test train of empty
reference wagons and a test train of
both full and partially filled reference
wagons. Each test train shall be
comprised of not less than five (and
normally not more than 15) reference
wagons and shall be weighed
repeatedly and in each direction (if
applicable) to yield not less than 60
wagon weights or the equivalent in
total train weight.
• According to the requirements in
paragraph 5(2)(11)of this part.
• Each weight indication and printout
obtained in the test shall be used for
assessing compliance with
requirements in either paragraphs
2(8)(ii)(a) or 2(8)(ii)(b) of this part,
as appropriate:
PROVIDED that where test wagons are
not available loaded wagon weighed
on a static machine fulfillingconditions
laid down in paragraph 2(8)(i)(d) of
this part shall be used as a reference
wagon.
{b) Simulation tests:
Influence factors shall be applied during
simulation tests in a manner that will reveal
an alteration of the weighing results for any
weighing process to which the instrument
could be applied, in accordance with the
following:-
Paragraph 2(9) of this part for all
Instruments
Paragraph 4 of this part for electronic
Instruments.
Wheh conducting such tests on a load cell
or on an electronic device equipped with an
analogue component, the maximum
permissibleerror for the device under test
shall be 0.7 times the appropriate value
specified in Table II.
If the metrological characteristics of the load
cell, or other major component has been
evaluated in accordance with the
requirements of the OIML International
Recommendation R60 or any other
applicable Recommendation that evaluation
shall be used to aid in the pattern evaluation,
If so requested by the applicant.
Note : Since the requirements of this clause
apply only to the instrument submitted for pattern
evaluation and not to those subsequently submitted
for verification, the means used to determine if the
appropriate maximum permissible error or maximum
allowable variation has been exceeded will be decided
and mutually agreed upon between the Metrological
Authority and the applicant. Following are examples
of these means:-
• an adaptation of the Indicating device to give
greater resolution tha11 that of the scale
Interval;
• the use of the scale lnterva I for stationery
load;
• the use of change points;
• any other means mutually agreed upon.
(3) Tests for compliance to determine
440 THE GAZETTE·0F INDIA:EXTRAORDINARY [PARTII---SEC. 3(i)]
qualifiedpersonnel and a control instrument.
.
Toe instrument under test may be used as
a control instrument provided it complies
with the requlrements In paragraph
2(8)(1ii)(b)of this part.
(Iv) Place of testing:
Verification test shall be conducted entirely
at the place of installation and during testing,
the Instrument shall Include all parts which
form the assembly as Intended for normal
use.
(3) Inspection :
Inspection shall be carried out In accordance with
the same provisions as in paragraph 5(2) of this part
for verification with the exemption that the maximum
permissible errors applicable for inspection shall be
applied.
6. On site test methods
(1) Proporiion of reference wagons in a test train
Toe proportion of reference wagons to the
remaining wagons in a test train shall be in accordance
with Table-IV.
(2) In-motion tests with coupled wagons
If the number of reference wagons is less than
the total number of wagons in a test train, the
reference wagons shall be distributed throughout the
train.
(3) In motion tests with liquid load
When using liquid load, tests shall be consistent
with the Intended use of the instrument.
Previous and next