Rule 23, Rajasthan Factories Rules,1951: Specifications of hygrometer
Rajasthan Factories Rules,1951 - Rule 23
Text of rule 23
Specifications of hygrometer
23. Specifications of hygrometer. — (1) Each hygrometer shall comprise two mercurial
thermometers of wet bulb and dry bulb of similar construction, and equal in dimensions,
scale and divisions of scale. They shall be mounted on a frame with a suitable reservoir
containing water.
(2) The wet bulb shall be closely covered with a single layer of muslin, kept wet by means of
wick attached to it and dropping into the water in the reservoir. The muslin, covering and the
wick shall be suitable for the purpose, clean and free from size or grease.
(3) No part of the wet bulb shall be within [75 mms.] from the dry bulb or less than @[25
mms.] from the surface of the water in the reservoir and the water reservoir shall be blow it,
on the side of it away from the dry bulb.
(4) The bulb shall be spherical and of suitable dimensions and shall be freely exposed on all
sides to the air of the room.
(5) The bores of the stems shall be such that the position of the top of the mercury column
shall be readily distinguishable at a distance of @[60 cms].
(6) Each thermometer shall be graduated so the accurate readings may be taken between
50 and 120 degrees.
(7) Every degree from 50 degrees up to 120 degrees shall be clearly marked by horizontal
lines on the stem, each fifth and tenth degree shall be marked by longer marks than the
intermediate degrees and the temperature marked opposite each tenth degree, i.e. 50, 60,
70, 80, 90, 100, 110 and 120.
(8) The markings as above shall be accurate, that is to say, at no temperature between 50
and 120 degrees shall be the indicated readings be in error by more than two tenths of a
degree.
(9) A distinctive number shall be indelibly marked upon the thermometer.
(10) The accuracy of each thermometer shall be certified by the National Physical
Laboratory, London, or some competent authority appointed by the Chief Inspector and
such certificate shall be attached to the Humidity Register.
----------------------------------------------------------------------------------------------------------------------
@. Substituted vide Notification No. F1(112)L&E/76 dated 24.3.1979; Rajasthan Gazette
Part IV-Ga(1). dated 26.4.1979